\ W W i: M W) H HI fl N W W lell i t w m \l—3 } CDN .THS_ mummy a? A was mmmrocsmmm $EPAEAT30N ma DETECTEON rem-swam: FOR WATER vma Tmmwon mm mamas: The”: {as 95m Dogma «:5 M. S. MICHIGAN STATE UNIVERSITY James CL leaking 1965 ABSTRACT EVALUATION OF A GAS CHROMATOGRAPHIC SEPARATION AND DETECTION TECHNIQUE FOR WATER VAPOR TRANSMISSION RATE STUDIES A chromatographic method for the separation and detec- tion with quantitative measurement of water vapor was devel- oped and evaluated. The applicatinns of this specific method to packaging research are described. The applications of gas chromatography to packaging in general are reviewed. Evaluation of a porous polymer chromatographic column packing material was carried out utilizing thermal conduc— tivity and ionization detector gas chromatographs. The re- spective sensitivities to water and water vapor of the gas chromatographs were investigated and tabulated. Several instrumental anomalies were encountered. ex- plained. and effectively controlled. The major finding of this research was that separa- tions. detections, and measurements of water vapor in air by chromatography is a successful method. It is indicated that further deve10pment utilizing this method should be pursued actively. Also, many other separatinns may be uncovered by further investigation of the porous polymer beads, and packaging research should benefit greatly from these separations. EVALUATION OF A GAS CHROMATOGRAPHIC SEPARATION AND DETECTION TECHNIQUE FOR WATER VAPOR TRANSMISSION RATE STUDIES By James G. Jenkins A Thesis Submitted to Michigan State University in partial fulfillment of the requirements for the degree of MASTER OF SCIENCE School of Packaging 1966 TABLE OF CONTENTS Acknowledgements .............................. List of Tables ................................ List of Figures ............................... Introduction Purpose of this research ................ Present WVTR detection method ........... Theory of Gas Chromatography General theory of gas-solid and gas- liquid Chromatography 00.000000000000000. Columns: Their efficiency and resolution limitations 00.0.0.000...OOOOOOOOOOOOOOOO Instrumentation: Thermal conductivity detectors and ionization detectors ...... Experimental Procedure Effect of Water on Ionization Detector response 0.0.0.000...OOOOOOOOCOOOOOCIOOOO Sensitivity determinations of chromato- graphy with reSpeot to water ............ Conclusions and Future Work Results and conclusions ................. Future worli .0.00......OOOOOOOOOOOOOOO... Page ii iii U1 Kn #1 +4 10 16 16 23 “5 45 1+9 ACKNOWLEDGEMENTS The writer is deeply indebted to H. E. Lockhart for much helpful criticism plus much encouragement during the course of this research, to L. B. Westover, O. L. Hollis, D. Pegaz, and J. L. Mautz who gave freely of their time and efforts, and to my wife Mary Carolyn who worked long and hard to make this thesis possible. I shall never forget. I. II. III. IV. V. VI. VII. LIST OF TABLES Filament Stripping Experiment 0000000000.. conditioning EXperiment 00.000000000000000 Burrell K-7 Minimum H20 Detection Limit Experiment Using a 6.5 ft. x 3/16" Column. Burrell K-7 Minimum MeOH Detection Limit Experiment Using a 6.5 ft. x 3/16" Column. Burrell K-7 Minimum H20 Detection Limit Experiment Using a 2.5 meter x 2.0 mm. Stainless Steel Column ................... Thermal Conductivities of Various Gases at 0°C O00......00.0.0000...00.00.0000... Compilation of Data from the Fisher Gas Partitioner coo000000000000000000000000000 ii Page 27 29 30 31 37 38 1+2 II. III. IV. V. VI. VII. VIII. IX. X. XI. XII. LIST OF FIGURES Weight of Water Vapor in 1.0 cc. Air vs. Percent Relative Humidity OOOOOOOOOOOOOOOOO Illustrations of Normal. Langmuir, and Anti-Langmuir curves OOOOOOOOOOOOOOOOOOOOOO Composition Profiles of Tungsten Filaments. Conductance of Various TUngsten-Carbon Compositions Measured at Room Temperature.. Structure of Porapak R and Porapak Q ...... Performance Data on a New Ionization Detector. Burrell Corporation ............. Measured Area vs. Weight Loaded for Water Sensitivity Determination on the Burrell K-7 Instrument Using a 6.5 ft. x 3/16" column OOOOOOOOOOOOOOOOOOOOOOOOOOOOOOOOOOOO Measured Height vs. Weight for Burrell K-7 Water Sensitivity Determination on 6.5 ft. X 3/16" COlumn OOOOOOOOOOOOOOOOODOO00...... A Typical Chromatogram using Porapak Q .... Present Chromatographic syStem 000000000000 Fisher Chromatogram of Water in Air ....... PrOposed Chromatographic System ........... iii Page 13 14 17 2A 32 33 35 no 4# 5O INTRODUCTION EEEPPSB of this research The purpose of this research was to investigate and evaluate the practicality of using a chromatographic tech- nique. preferably compatible with existing equipment at the School of Packaging, so that a statistically reliable method of measuring water vapor transmission rates would be avail- able to the packaging engineer. A second goal of this re- search was to find a method by which a single measurement of the components. 02. C02 and water vapor. in atmospheric air would be possible. The final area of interest was to investigate and describe applications of gas chromatog- raphy to packaging research in general. Any gas chromatographic separation is dependent on the efficiency of the columns. The following character- istics must be inherent to the column material. 1) The column material must be easily prepared or commercially available. 2) The column material must be stable at temperatures above 200°C. 3) The separation agent cannot bleed or react with the instrument or components to be separated. U) The column material must be easily packed into small diameter columns and it must allow necessary carrier-flow characteristics. 5) The material must separate a wide range of com- pounds such as N2. 02. 002. water vapor, and non-polar organic vapors. 6) The material must not only separate water but be able to yield smooth symmetrical peaks. Briefly. the column must minimize tailing @dsorption-desorption) of water. At present. there is no permanent record made of water vapor measurements; this is not the case with the gas per- meants. The gas chromatograph will enable absolute measure- ments to be performed. Simultaneous comparison of water vapor content and gas permeant content are but one reason for attempting this research project. If a gas chromatographic technique may be developed for both gas permeants and water vapor measurements a source of instrumental error would be eliminated. This source of error is in using electric hygrometry for water vapor measurements. The many advantages of using one de- tection and measuring instrument for any given set of measurements has been demonstrated by Lockhart (13) in the utilization of gas chromatography for measuring permeability rates through barrier materials in the Davis cell (A). Present WVER detection methods A degree of laboratory SOphistication rarely found in technicians is involved in classical methods of measur- ing water vapor. Among these methods are gravimetric, chem- ical. electric hygrometric, dessicant-cup. and pressure in- crease methods (8.12.lh.15.16). Gas chromatography has been demonstrated to be a simple straightforward measurement device. It is used for routine testing as well as research. The School Of Packaging is presently using a testing system developed by H. E. Lockhart that measures gas per- meants by gas chromatography (13). The water vapor is measured by electric hygrometry (29). The level of detec- tion is indicated by Figure I (20.26); a sensitivity of at 6 least 1.0 x 10' gm./cc. for water must be obtained for gas chromatography to be a practical method of measuring water vapor in air. The collection and tabulation of permeability data is essential for future design of packages as the main cause of product degradation in many cases is due to water vapor transmission and gas permeability. FT SUP? I ‘- L4 Height of Water Vapor in 1.0 cc. Air vs. Percent Relative Humidity o 0 . 23 C. (73.4 F.) 20.23 gms. 190 vapor cubic meter of air 100a Percent 601 Relative Humidity 04’ 2 Q 2 A? 10 12 I3 6 18 20 Gms. x 10" of 0 Vapor/cc. of Air ‘ I I b-A—d \. {‘0 THEORY OF GAS CHROMATOGRAPHY General Theory of GaseLiguid and Gas-Solid_Chromatoggaphy. Small diameter columns of various solid supports, which may or may not have been previously coated with a thin layer of some organic compound have been used for chemical separations for years. If the solid support is coated, this coating is usually a high boiling liquid or solid. This coating is commonly called the liquid phase. A sample of various components may be inserted into the front of the column by some suitable method. The sam- ple is usually transported through the column by the use of an inert gas called the carrier gas. Separations are effected by differencesin vapor pressure, solubility of the sample components in the high boiling liquid coating the column solid support. or in the case of using a solid sup- port alone the elution and separations are dependent upon surface adsorption. molecular interaction between the sample and this support. The eluting components are detected by some means. the more pOpular being thermal conductivity cells and ioni- zation detectors. A series of signals are given by these detectors; when these signals are diSplayed as signal in- tensity versus time symmetrical peaks approximating Gaussian distribution curves are given. and the areas under these curves may then be normalized to obtain the percentages of the sample's components. Columns: Thei£_efficiency and resolution limitations. Sufficient column length must be present to separate the desired components completely and quickly. Means of describing a column's efficiency are approached by many workers in various ways (11.23). The method des- cribed here is based on the theoretical plate treatment of James and Martin in which N. the number of theoretical plates. is found by using peak dimensions. (A) N = 16(X/I)2 N may be calculated using Equation (A) where X is the re- tention time and Y is the baseline width intercepted by tangents to the points of inflection of the peak.. (See Figure II-A). In general. the separation requiring N plates by normal distillation is given by N2 theoretical plates. One other comparative method of measuring column efficiency is commonly used (11.23). This measurement H is defined as the height equivalent to a theoretical plate. It is obtained from N and the length of the column. L. (B) E = L/N In Equation (B) a measure of band broadening is essentially given as width of peak base. I. during time of elution. X. See Figure II-A. A means of comparing column resolution is FIGURE II Illustrctions of Normal, Langmuir, signal s i gnal Curves II-B L_____P time and Anti-Langmuir elutinr ——-———————-& sample given by (c) a = (II - Uta/x1 - 1) This expression for R, the distance between peak maxima, is expressed as a multiple of the standard deviation of the first peak. It is clear from Equation (B) and Equation (0) that if the resolution, R, is to be doubled, the cOlumn length must be increased by a factor of 4. Also the effective resolution may be increased by increasing the ratio of liquid phase to solid support within practical limits. Along with the resolution the symmetry of the peak for the eluted component is dependent on adsorption-desorp- tion effects taking place on the surface of the solid support and on the instrument walls. The instrument must be non— reactive, clean, and it must possess small hold up volume in order to minimize tailing and spreading effects (2.10.11). The shape of the curve obtained in gas chromatography may be varied in three ways (11,23,2h). The first varia- tion is shown in Figure II-A. The nearly Gaussian peaks are obtained when the concentration of the solute in the stationary phase is directly prOportional to the concentration of the solute in the gas phase. It is seen that the retention time is independent of concentration. The type called Langmuir or Classical Isotherm as shown in Figure II-B is caused by the main portion of the solute band being eluted more rapidly than the leading front edge due to a limited area of adsorption available. Langmuir curves are given by all columns to some degree. Peaks of the Langmuir type are usually Observed in gas-solid chromatography. In this type of column the retention time is a function of sample size and normal narrow peaks are Obtained with micro amounts. A third type is shown in Figure II-C. Its shape is due to the main portion eluting more slowly than the extremities. Again, retentinn time is a function of sample size which is, of course, dictated by sample volume available and the sen- sitivity of the detector. The column is the heart of a gas chromatograph be- cause the actual separations are Obtained in the column. In gas—liquid chromatography high thermal stability and low vapor pressure at the Operating temperature of the instrument must be possessed by the liquid phase. The separation of homologous mixtures is easy, and, of most importance, azeotropes are not formed in homologous series. The injection of samples must be done in such a manner as to approximate a "plug" introduction of vapor. For optimum results the smallest possible sample consistent with detection sensitivity is desirable, but this efficiency is reduced if the sample vapor is excessively diluted with carrier gas. The rate of elution as well as the degree of separation 10 is affected by the temperature of the column and the flow rate of the carrier gas. It is necessary to Optimize these variables in such a manner that a rapid, precise, and yet simple technique may be available to the packaging engineer. Throughout the remainder of this paper the term "gas chromatography" is used instead of the limiting terms "gas- solid" or "gas-liquid". Since the development of solid supports which will separate both polar and non-polar substi- tuents the scope of the field is widened in gas chromatog- raphy--both gas-solid and gas-liquid. Instrumentation Thermal conductivity detectors. A thermal conductivity cell is basically a balanced differential bridge by which the unbalance caused by various characteristic thermal conductivities of materials flowing over one arm of the electrical bridge is measured. These are of two types--the hot wire filament and the thermistor thermal conductivity cell. Samples as low as 10'9 to 10"10 grams can be detected using small volume thermal conductivity cells (6,11,23,24). Ionization detectors. There are two pOpular types used: the direct ioni- zation detector and the hydrogen flame detector. The hydro— gen flame is completely insensitive to water (23) and will 11 not be discussed here. The argon detector is a type of ionization detector which Operates on the principle of forming metastable argon ions (6.23) as a source of energy by which free electrons similar to those described by Ryce and Bryce (22) are produced. The separated components are ionized by the free electrons not much differently from the way a sample is ionized by a mass spectrometer (3). Ion- ization detectors are characterized by their fast response in the range of milliseconds (23,2#) and high sensitivity; 10"12 gram determinations are not uncommon fOr non-reactive organic compounds (23). Any material capable of being heated to incandescence and emitting free electrons will constitute a filament for an ionization source. These free electrons are accel- erated by a positive potential which gives them enough kinetic energy to ionize molecules. The molecules are ionized by re- moving electrons from the electron orbits of the molecules. The ions formed migrate to the collectorgiate by virtue of their positive charge. Upon striking the negative collector the ions discharge by accepting electrons from the collec- tor. This electron flow amounts to a minute current and this current, when amplified by an electrometer, is a sta- tistical representation of the sample introduced into the device. 12 In the case of a filament manufactured from pure tungsten it is necessary to precondition it with a hydro- carbon. The basic reaction is to form two distinct tung- sten carbides. 1W + ZHZC=CH2 .9 WZC + WC + w + 2CH4 A diagramatic presentation of this may be seen in Figures III-A, III-B, and III-C. Figure III-B is the ideal filament which has been described by Sharkey and many others (3,7,23,27). It is possible to overcondition a filament and this filament structure is represented by Figure III—C. The reaction involved is Similar to the above but it is driven further to the right causing a larger percentage of WC to be formed at the surface. Supposedly the carbon permeates (diffuses) through the tungsten metal and WC forming a large build-up of W2C in the center of the filament (3). A graph of conductance (reciprocal of resistance) is shown in Figure IV. The minimum conductance, which is the maximum resistance, is found at the point where the filament has the greatest amount of W2C present at the emitting sur- face. Briefly, this is the point where the electron emission is Optimum with reSpeot to a constant filament heating current. As previously mentioned, the filament is the source of free electrons. These free electrons when given 13 1,. _ -... BEACH f‘om YeYdOn. 1““s ~n~3tro~otnr nwa th " ‘1 . n4“ am 1 A 4— , ”1’ «L .‘ q (LP: -1" I") ‘ r) LIA-4") 1n dii L‘lejo P. 1000 III-A Pure tungsten before any conditioning VJ 2 C NW III-B A properly conditioned filament (III-A + hydrocarbons) % 'IT M III-C An overconditioned filament (III-B + excess hydrocarbons) A \\\\\\\\\‘1 \\X\\\\\\\ \\ N 14 *3. ’n v rlpure If Conductance of Various Turgeten-Ca.bon Compositions Ieasurcd at Room Temperature 7 O r— {‘0 b) 4: Um O\ on \O \Q \0 C3 \0 -\) *0 C\ \ C) Kn \ 100 7 [4‘3 .711 - o T) , v. r: 4- 7.-4- - —~., A -~ ° '- ragea LTOW neynon. Lass cpectr01etrg anfl Its npgllcatlooe 4- -...~'. n1...‘..,- 4. , H- a n LKD‘DTPfihwtg ulfljnlSLJéK. IAiQC?.LOy. 15 enough kinetic energy will ionize sample molecules. As part of the description of the ionization source it must be stated that the ionization of materials present in the source will give rise to one ionization signal un- less the ions are separated by some electrical or magnetic means. A chromatograph will. in most cases, allow Just one single component to be eluted at a time; in turn, this sin- gle component is ionized and the ions detected are representa- tive of that particular Specie plus any residual components present in the ionization source at the time of ionization. There are two gas chromatographs available at the School of Packaging. The Fisher Gas Partitioner is a ther- mal conductivity instrument which Operates at ambient temperatures (stabilized by a constant differential heater). The other instrument available utilizes an ionization detec- tor and is equipped with a programmed column heating element. It is the Burrell Kromo-tog K-7. Most of the evaluation was performed using the Burrell instrument in order to utilize its high sensitivity and column heating equipment. A tungsten filament is utilized in the Burrell Kromo- tog K-7. This tungsten filament is heated by a potential of 18 volts. and an emission of 0.8 milliampere is obtained when 5.0 to 6.0 amperes current is passed through the tung- sten filament causing it to become an electron emitter (Edison effect). 16 EXPERIMENTAL PROCEDURE Effect of water on ionization detector_response The first step in this procedure involved the selec- tion and preparation of a suitable chromatographic column. Initially. 5% Carbowax coated on Teflon powder was con- sidered to be the best water separation column. It has been reported by Kirkland (10) that this column packing separated liquid water with minor tailing. Kirkland also describes cooling the Teflon powder (after coating) to 0°C. and vibration packing 3/16” and l/h" one meter columns success- fully. Attempts to pack columns at room temperature invar- iably result in aggregated Teflon and undesirable plugging. In the past few years at The Dow Chemical Company gas - chromatograph methods of measuring water were tried and evaluated using the common technique of coating a fluorine containing polymer with a suitable liquid phase. That much of the tailing (Langmuir effect) might be eliminated by the use of a uniformly sized support (microspheres) was confirmed in a personal interview with L. B. Westover (30). A solid porous support had been developed by 0. L. Hollis (9) and was being released by Waters Associates, Inc. of Framingham. Massachusetts under the name of Porapak was also made known in this interview. Two of the more common Porapak structures are graphically demonstrated in Figures V-A and V—B (9. 17. 17 Frans: v Structure of Porapak R and Porapak Q V-A Porapak B----a copolymer of divinylbenzene and styrene. The styrene molecules are those with solid rings. V-B Porapak Q-—-—a polyethyl- vinylbenzene, 18 18.19.30). Both the liquid phase and solid support functions in the column are actually served by the porous polymer beads (17) for chromatographic separations. Due to the availability of this chromatographic column packing. Porapak, which packs easily and needs no liquid phase coating. Teflon powder support was not investigated further. The Porapak column had been recommended by West- over (30) and Hollis (9) for the separation Of water vapor from air. Hollis suggested using a small diameter column at least 6 feet long and Operating the column above 100°C. Past experience with coated Teflon indicated that bleeding Of the liquid phase would be another disadvantage with the Teflon column. The Porapak does not bleed even at temper- atures of 250° C for it has no liquid phase tO bleed. Pora- pak exhibits the unique prOperty Of separating polar and non-polar compounds with virtually no tailing of either type. A column 6.5 ft. I 3/16 in. 0.D. aluminum was packed with 18.3 m1. of Porapak Q by vibration. This column was conditioned at 230° C. for 4 hours while being continuously purged with helium. This column was bent in the hairpin shape common to the Burrell instrument and attached to it. While Optimum column conditions were being sought. the emis- sion current was noticed to drOp by a factor Of 25% (0.8 to 0.6) during detector response. Due to this loss Of emission current. the chromatOgrams were in nO way reproducible. 19 The reason for this was that the tungsten filament was being stripped by the oxygen-containing samples. This tungsten stripping effect is well known in mass Spectro- metry and is referred to frequently in the literature (1,3,11,27,28). Also, insulating deposits will be released by accelerated decomposition in the ionization chamber due to the presence Of oxygen-containing compounds. Erratic Operatinn will be caused by the release Of these deposits, especially if these deposits are subsequently exposed to electron bombardment from the filament, and the back- ground will be high. The filament itself will be damaged by some of these materials in this process Of decomposition. As an example, water vapor is decomposed when contact is made with the filament; the oxygen is chemically combined withthe filament and hydrogen is evolved as atomic hydrogen. H20 4- WZC —-—-> we + WC + 2H H O + W --‘§ W0 + 2H 2 230+wc —-—-> wo+co+uH 2 At the Operating temperature of a tungsten filament the tungsten oxide is distilled to the walls Of the ioniza- tion chamber. There it is reacted with the atomic hydrogen to reform tungsten metal and water so that the process is repeated. The process is not reversible because the tung- sten metal is now deposited on the waIlsurface. 20 NO + 2H -—e> w + HZO This effect is readily noticed in a mass Spectrometer after many water samples are run, and the effect is shown as a loss Of total emission and then the filament must be reconditioned by purging with a hydrocarbon. The loss of tungsten is irreversible, and in Order to Obtain the same total emission level the filament current must be contin- uously increased. This type of reaction proceeds readily at pressures Of 10'12 mm. Hg. and as the pressure Of the ioniza- tion chamber is increased the loss Of emission becomes notice- ably greater. If carbon-containing compounds alone were present, the surface Of the filament would be chemically changed to main- ly WC, the other carbide of tungsten, and the carbon would permeate to the center Of the filament forming WZC. This type Of filament is shown in Figure III-C. An entirely different work function Of electron emission is possessed by this filament, and this emission will be much less than that Of the case shown in Figure III-B for the same amount Of input current. All these reactions take place in mass spectrometers Operating at pressures Of 10'12 mm. Hg. In the case Of the Burrell K-7 this deleterious effect will be accelerated due to higher ionization pressures Of 0.1 and 1.0 mm Hg (6). 21 The effect would be Observed for small amounts Of an oxygen-containing sample. When a large amount is suddenly introduced the emission current decreases sharply, and erratic non-reproducible chromatograms are caused by this decrease. TO Obtain reproducible ionization of samples some source Of hydrocarbon must be introduced to maintain the filament as shOwn in Figure III-B. The chemical composition Of a properly conditioned filament has been determined by Sharkey (13.27.28) as previously mentioned (See Figure III). The Optimum emission will take place when the carbon content is 3.16%. This may be seen in Figure IV. ‘One is not concerned with the exact chemical composition of the tungsten filament being exactly 3.16% carbon, but rather that the filament easily reaches some stable reproducible level of carbonization. It may be readily demonstrated that when a slightly conditioned tungsten filament is exposed to an excess Of any hydrocarbon a further production Of W20 will take place and the resistance Of the filament will increase (1.27). This resistance increase will cause the filament to be maintained at a satisfactory emission level by succes- sively lower quantities Of filament heating current needed. Conversely, if oxygen is introduced, the filament heating current must be increased to maintain this same level Of electron emission. The exact reverse behavior would be ob- served with an over-conditioned filament (3,28). The 22 Burrell instrument has no provisions for automatic electronic control of the emission level (6) and drastic decreases and increases in the total emission current are noted during the Operation Of the instrument. In this research, water vapor decarbonized and methanol vapor carbonized the tungsten filament in the Burrell K-7. Of interest is the section entitled "Analytical He- sults" found in Performance_pata on a New_;pnization Detectgg, a Burrell Corporation publication (6). The analysis of hydrogen, oxygen, nitrogen, methane, carbon monoxide, argon, carbon dioxide,ethy1ene, and acetylene is referred tO as the analysis Of a "fixed gas." These gasses were adsorbed on molecular sieve and silica gel giving reproducible quanti- tative chromatograms. The reproducibility is caused by the presence Of hydrocarbons in this "fixed gas" sample. In general usage, the constituents N2, 02, C02, and water vapor are found in atmospheric samples, and are denoted by the term "fixed gas." Atmospheric samples dO not generally contain methane, ethylene, or acetylene in any great amount. A fil- ament having a coating Of tungsten carbides wzc and wc coated on pure w is Obtained in the presence Of those hydrocarbons. Also shown in the Burrell publication (6) is a chromato- gram which contained 90% water and the remaining 10% made up of the hydrocarbons methanol and ethanol. The ionization 23 energy was reduced after the methanol and ethanol were eluted; Figure VI is an exact scale drawing of reference (6) with the dotted line drawn in to indicate water. If this exper- iment were to be repeated with ionization energy great enough to ionize the water, erratic chromatograms should be obtained. It is not stated in the Burrell publication anywhere that water-containing compounds may be quantitatively analyzed in the absence Of hydrocarbons. The non-reproducible Situa- tion noted in this research may be overcome by prOperly con- ditioning the filament by the use Of hydrocarbons during the analysis Of small amounts Of oxygen-containing compounds. It is seen that, unless a prOportional amount Of hydrocar- bon is present when oxygen-containing samples are ionized, the electron emission will not be linear; thus, the signal measured will be non-linear also.' Sensitivitydetermination Of chrOmatograph gith respect to water Even though the Burrell ionization instrument gave less than completely successful results, the anomalous conditions are instrumental in nature.and would be eliminated by care- ful conditioning or instrumental deSign changes. These changes are covered in detail in the section on further work. ' Excessive high background was Observed at maximum sensitivity and minimum attenuation Settings. The background 24 flT.flTTH ‘1 T I d .- ‘ . l —L J iv1‘AAJ \] a -n.m ~ 1“ L ‘ T - c . r'r Performance Data on a let Ionics Turrcll Corporation 100 vita—nirv- 1'1 1. L LLJ. / \O O CO 0 [1’0 AIR I ZING “vs "V 1.41. ".11. K1 .1. PD til U C: O is C) H O n y - 4 A tion Detector ALCOHOL G—-ETHYL ALCOHOL 0 1+ 8 MTYJW H 1T Tzfi'fpa Lilla: "' liIiQL‘ldtJ Fl 0\ 12 25 was in the same order Of magnitude as that of water vapor in air. Due to high background levels the water vapor de- tection experiments were performed at lower sensitivity and higher attenuation settings. The values Obtained are relative values,and, in fact, will be 50 times lower than could be Obtained by Operating at the maximum sensitivity of the Burrell instrument. Various solutions Of water in methanol were prepared to determine the minimum practical limit of sensitivity Of the Burrell K-7 for water. The use Of methanol was dictated by the Observed deleterious effect Of water alone upon the tungsten filament in the Burrell KromO-tog instrument. There was still a noticable drOp in emission current, but this was due to the cooling of the filament below its most efficient emission temperature, and, as there would always be a hydro- carbon (methanol) present, the normal behavior of a properly conditioned filament will be experimentally created.‘ A series Of chromatographic experiments were per- formed on the Burrell K-7 using a 3/16 in. 0. D. aluminum column 6% ft. long packed with 18.3 ml. Of Porapak Q resin. The column was tested between 75° C. and 200° C. with varying helium flow rates in order tO Obtain relatively "tail-free" curves. All tables have the exact instrument Operating conditions Specified on them. The first experi- ments (EXP BK-7-1A and EXP BK-7-1B) were performed to 26 Observe the variation Of the peak area for pure water when no organic compounds were present. The 1.0/.1. sample was injected by a microliter syringe through the Burrell syringe port. TO eliminate any question of loading error, the syringe was loaded, weighed to the nearest 0.1 of a milli- gram, and injected; the syringe was immediately weighed again. The 1.0,al. load Showed no detectable variation to 0.1 milligram when the loads were weighed and injected carefully. 'Also the sample was drawn up through the syringe by a Squeeze bulb. This was to insure that no air bubbles were trapped in the syringe needle. Due care was taken in injecting the sample through the syringe port, as this ~ port was Operated at temperatures in excess Of 100° C. for all Of this project. The method used was to insert the syringe fully into the port, immediately to depress the plunger, and to remove it in one swift motion. If this were not rapidly done. the needle would become hot and preferentially volatilize lower boiling constituents. The results Of these two eXperimentS are Shown in Table I. A variation Of -h9.h% to +31.3% is demonstrated by these experiments, and this variation is due to the.water stripping problem mentioned previously. A second experiment was designed to demonstrate the effect an organic compound has upon stabilizing the emission characteristics of the filament. The results of this .4. .e u a ego on u m n.2_ one eta mo cnoanaeeoo oe.m ”mfi-s-mm mum .pgomfiaem.. clear:cc .p,omoag candoao on Q\. I a. u .. 3.1 .343 '3‘ A 0.1_\|O\lrs cm I a or m wg r oH_t:mc r H Hfi5\. 00 u c \u :1... :1 I)... o)“ . e I..- 4.4 ...1... CUP NH paw. CEDHOO Eigw rnrfidw :wfi\m 1r. .CCQCW ..Crlhwlwttr \Li. ssfiaoz once Beam ..0 27 Cl C 1 C x) l \C If x \r‘) + Q 0‘ O 1. '\ 0 € \ \f \ U\ 0 O O H :9- J a \"1‘ C 0’ + C 1 C I \Q O 0 CK) H C- L'\ 0 ;) r-I O: C’ \ (\3 C \ O C \ (\1 O O O H O O H OH 0.. C’ ) C‘\ + W V—I kW .2 O ( N H C? 11W O O .D ( J O O 0 r1 C) O H (‘1 \.O\1II ,J\o.1)\l \\ o .1 \o\.\. 0 0|. I ..i c; c ec+ woo Hr 0L Cc eve o o e r -..n e .d1. .-1. (If AL kknF L: C \3 H C J "t‘ if C \ I ~) {\ \3 Q as C‘ \ H O C) r1 C) O H L; ee.m . em.mfi+ \. em.m cm.ofi 02.0 o.a m FULL Smear e.Ce- mfi.e + om.“ eo.wfi me.o o.H m ((\ r4 0 \ + CD 0 O H I (\I o D- (\1 O ("x H C) H o H O o T‘I v—I 4a-nnwm an ..no 3. .r .....3. 3.5 ram .M a cone .o: .am Wong oamecane .pa .nm ogmaom seem nsmao: Hodge r... 3.0.40. ,3.. EQHP CO H3 ma 3an . ..1. >0 .mo $0.0m. mtflm @055... Heurim Him 9b mHQuH Gm... 28 experiment are given in Table II, and it is interesting to note that between Trial #1 and #2 no organic was introduced to the previously conditioned system, and a drOp of 15.00 cm.2 in the water peak area is noted. This is calculated to be a variation of 45.5%. The organic (methanol) 1.0/#1. load was injected between all of the other trials and after four injections of methanol (each followed by a 1.0/“l. load of water) steady state conditions were again attained by the instrument. The importance of prOperly condition- ing this instrument before attempting analysis work cannot be overstated. An experiment designed to determine the minimum water sensitivity of the Burrell was carried out by using a blend of methanol and water. A conditioned filament was created by the introduction of a source of organic material (methanol) and progressively smaller amounts of water. These injections were performed without sacrificing the previously demon- strated precision obtained by using 1.0/“l. sample loads. The results of these experiments are given in Tables III and IV. The practical minimum sensitivity of the Burrell K-7 instrument with electrometer set at a sensitivity of 10. and the attenuation set at 5 was no greater than 3.000 x IO'u gm./5 cm.2 for water in the presence of an ionizable hydro- carbon. A graphic diSplay of this data from Tables III and IV is shown in Figure VII. It is shown in Figure VIII where 29 Conditiofiir“ ‘"f0?‘u“flt F - ’ o v --_v o ‘T’ZC‘I (‘1 “I h ‘ ..‘I '4... 1 ‘1 ‘1' 7’) ,f! T" .,.1.. . ‘ ”ML: 4...» x ..a 1.... --1 .oignt Fa-“ a r--- “71‘.” Triccrl. 1 1.0 n:. 2.;0 cm. 23.30 on. 33. 9 cm. - ('3 2 1.0 ug. 1.75 on. 21.00 cm. 3.38 cm. 3 1.0 Mg. 2.7. cm. 16.30 on. 27.00 cm. I . 4 1.0 V”. 3.00 on. 19.20 on. 28.00 cm. / . ,. 5 1.0 T‘f . 3.05 0:71. I‘LJQEO CH1. 33.76 0171. ‘\‘I0\ H O O O k.) I I \ 1 O [...L CO 0 \. n O '3 '30 ‘ZQJIEJ T11. \ .3) O C\ O O a I—A CO KR O 9 ...3 In F“ ‘U. U? rT1-j. -. ‘&J_J. TVP P" IJJ‘A. JiL introduced a introduced until instrument \ ,3 J. .1. rar- .1r _ _ .)t\.d.uu. ESL/8.00. ‘ 1 a.proac:,o ~7—2: Speed 2, 3/16” aluminum column at 2000 C., Holiur flow 68.5 3 =10, .51 =2 10. oc./min., (\3 0.) NM (\D PO 30 .mUmz Sp psmasmfiu .m u a .ofi u m ..o own ..sflg\.oo m. .fi wmmmm .6 xmmmpom Spas QESHCO ESCHEfiHm .pm C) ofi.ofin mm.m- mow.m OH.: om.fi mmfim.o : Hm.m - mm.m- mam.m om.: m.H mmfim.o m :fl.m + mo.n+ :wm.m moa.m m:.: 00.: mm.fi mmHN.o m oo.flfi+ mm.@+ omm.m om.d o¢.H mmfim.o H om.fi . oo.m+ omo.m om.m om.fi mumm.o m \ mnfi.m mm.m . \ CU.H + 00.3- omm.m om.u mq.a vmmu.o a oq.u + Hfim.o+ omm.mfl mfi.wfi om.fi mam:.o m cm.qfi- mmo.o- amo.mfi mam.ofi mo.mH oa.mfi om.fi mum:.o m (u.m + Hum.o+ omm.mfi ofi.wfi ou.fi mam:.u H EL .50 .80 .FC .50 .bh mmu< .pm .xm mmg< mammmfiue .pm .xm psmfimm ozmm unmamz Hmfipa mmmgmbm anm SOprabmmkfi .®>< .mo ampm mmmhm>< mmmw Hafim maaidm msfil. sazfloo :ofi\m x .pm m.@ m pgmaflhamwm pfiaflq Sofiuompmm o m ESEHSHE mam Hammpsm HHH mum.v 31 .m M Q .OH H m ..o emu ..:H_\ 00 m.mQ C H5HHo: .H woman .w Mammhom 55H: SeSHoo EDSHHSrd .pm moo \ M :mfi\m Hfiumumm NH 55.5 - :H.o- oom.mm OH.mH om.m oQNQ.o : HN.H - mu.m- - oQo.mm 0H.wH cm.m onQ.o m mmr.or mm.mH --- :H.:+ oun.om 0Q.0H oH.m o CQNQ. o m B.HH+ mm.m+ omm.:m om.om oq.m :5 Q. o H 035.0 + --- QQQ. m OH.mH om.m onwm.o m Hmw.om H.mH - omm.o - --- mwo.o« 0H.mH VH.m oumu.o H :Q.N - mc.m- ouo.mm OH :H 03.x gmmm.o m NQ.m - Mo.m- mHm.Hm cum.mm m..HH 0H m5.m wmmm.o m mm.Q + mo.m+ 0mm.Qm om.QH oQ.m arm .0 H .,:g m.n:u .-o :50 .50 .mwu m¢5< .5m .55 Wm5< QHQQQHHB .55 .mm pgmHmm Qsam pgmHm: HmHHm mfldhwbfl thh HHOHWZ. >63 1 .O>< .mo deS vmfih®>d Mdom HHSm vdmfidm x? casHom =QH\m x .55 Q.Q m mgHmp pCmEHMHw 9H1 pHEHA :chompcm Mom: has; H: mum Hamppfim L >H Hand? I';('\f\ ."1 A 42'.“ \. natnr in milligram “- 7‘ ‘1' V“ ‘PLJ ~.'.’v -,l. o 6 L) ,_J FIC ' T ,. A ”,1... .d' t" |_ . 7- 1L ff. 32 T 4.. .L/ 7: f3 "1 l. 0;-HJ Loadcfl v-7 coljmui 1 7" - A UK} .0: L",1.‘-J“"‘= J '1 L I.— 3/1 VTI 6n A) Q? 4— .r 3“ .4..3.LIO-L rument uifiitlviby Tfii‘r 15.).L; \/ 15 ("‘3 M 1 33 10.0 lm.o mamamflaaflg CH no.0 nSchmaoo \m N .um m.m :o COHQGQHEpmme m Hamupsm 50% psmfim: .m> psmflmm wmuzmdmm HHH> mmerm 34 sample weight in milligrams is plotted against peak height in cm. that 0.05/11. (5 x 10"5 gm.) of water must be present before water detection (any signal) is observed. There must be enough signal to measure area before the water per- centage may be calculated in any case. A graph of grams of water per cubic centimeter versus relative humidity at 23° C. (73.40 F.) was given in Figure I; it is indicated that at least a sensitivity of 1.0 x 10'5 gm. of water at 50% relative humidity must be present in any instrument for detection at the 50% B. H. level. In order to compare the resolution efficiency of the Porapak column Figure IX is included. This is a reduced scale drawing of Trial #1 (EXP BK-7-3). The elution time for water is 2.58 minutes at 750 C. column temperature with a flow of c468.5 cc./m1n. of helium in the Burrell K-7 instrument using the 3/16 in. x 6% ft. aluminum column. A second column of stainless steel. 2.0 mm. I. D. x 2.5 meters long was packed with 10.2 cc. of Porapak Q and conditioned at 230° C. for 0 hours constantly being purged with helium. On this column an experiment was run to determine the relative sensitivity of this column versus the 3/16 in. column. The chart Speed was set at i. the sample load was 2.0/41. of distilled water. and the column flow was 40.0 co./min. with a column temperature of 125° C. The average of three determinations was 160.1 cm.26al. of 35 mafia msflmmopcsfi Hmcmflw we huHmSmpCm [J d xmamaom mQHmS Emamopmsoaso HmOHmmg d oom C C a. \Ja.4 O 7.7-1 N.)- .1 U00 t...- t...... 0 r firmHQ amass-HchSuem .0 0mm .9 on .Qflfi\.oo .SHE\.Qfl w e. at» CSSHOO mqu seHm toom padso WHH... < OH H m 36 water loaded. These figures may be found in Table V. One last experimental procedure was attempted with the Burrell instrument. The instrument was stabilized at 125° C. column temperature and a flow rate of 00.0 cc./ min. He was established. The sensitivity was set at 10 and the attenuation at 1. These settings gave very high noise to signal ratios. A sample of 1.0 cc. of room air was injected. The instrument did not give a response for water. The emission current drOpped to such a low level that it was not possible to increase the filament heater current high enough to obtain 0.8 milliampere emission current. This experiment was attempted 5 times and no meaningful water separation chromatograms were obtained. At this point it was necessary to evaluate the thermal conductivity detector using the same column packing material in order to complete column and instrumental evaluations. A survey of the literature indicated that thermal conduc- tivity is not as sensitive as ionization detection for organic vapors. A study of thermal conductivities of var- ious gases and vapors indicated that the detection of water vapor appeared possible by thermal conductivity provided a large enough sample is used. See Table VI. Table VI is a compilation of selected thermal conduc- tivities taken from International_Critical TablesI Vol. 5 "Thermal Cnnductivity: Gases and Vapors." pages 213 - 216. 37 H n m ..e emmH ..eHe .ee N CESHCO HCCPm mam-fi-SH?P C ‘ UJ H! \I ‘14. o —l t», 4 s \1 \l _) o 4 4 4.)! A44 0‘1! .l. u.. ' I -l . n1. f- (..IU 0r Hep B KFr-r r.|rFP .\. .‘IJ CH eon mm.m- m H @H.mH+ em.m+ Hs.QH mm.:H m HQ.OH+ mH.e+ s.eH s.om ms.H o.m H 0.20 d.:He .ro .To .so AW; mess .pm .mm moms mfiwsm a .pm .xm psmficm Usmm psmHmm adage mnsaebs moan Sempmemu . .mbd .so swam enmaobm ysem HHSh mflamdn . L :uufloo Hompm amoflsflmpm .EL Ohm lacmflucmum. .vfih Hmoflpoopofl -cww w momma m.u ( s5sHsHs a-a H‘H inermal Conductivities of Various 12’ ..° .. n = Lila-Hr "1.31" 1" Fe 13.9 Air rxz-.23 21:2 ~15.9 1120 ”209.0 : i T‘s.) [\J . (\) CO 39 The values are given at 0° C. which is a common reference point for which all could be compared meaningfully. That thermal conductivity is very sensitive to water vapor has been verified by much previous experimentation (2,9,10,11,23). Due to favorable thermal conductivity of water vapor and the fact that thermal conductivity does not destroy the sample being analyzed as does ionization detection, an experiment to determine the Fisher Gas Partitioner (Model 25V) water sensitivity was performed. An aluminum column 3/16 in. 0. D. x 3 ft. was packed with 10.0 cc. of Porapak Q and it was conditioned at 230° C. for 0 hours. This cOlumn was installed in the Fisher Gas Partitioner. This instrument was used as a one column in- strument by removing the drying tube and column #1 (see Figure x). A short 7% in. copper tube was substituted for column #1 and the Porapak Q column substituted for column #2. The instrument was run at room temperature with the temperature held constant by a thermal stabilizer Oper- ating at 50° C. The cOlumn flow (helium) was adjusted be- tween flow rates of 80 cc./ min. and 125 cc./min. for optimum separation and minimum elution times. Three distinct component areas on the chromatogram were given at the 100 percent sensitivity range with the introduction of 1.0 cc. of laboratory room air. The first area is due to the composite peak measured by thermistor Sl, 00 *3 a"??? 'f I'I‘ka ‘ ' .‘...'. -1. - ‘4 “1,. ‘fl -\ . r‘v r V\1‘- . Present chromatographic He sample inlet f drying tube column 'I co umn exhaust ”10 -f a Column f 1: 30 in. x i in. 30K hexanethylphosphoroamide l 6. wi th F~ F) N O O COlUJI‘L‘ 7 detectir: the filled with ’3 ‘f 3 ’1L rmistors. ft. X 3/16 in. 0. D. filled Molecular Sieve 1 01 the second component is air measured by S and the third 2. small component is water also measured by thermistor 82. The quantitative confirmation of the third component was done by using three separate measurement techniques. The first technique was to carefully set the helium carrier flow at 125 cc. per minute and inject multiple air samples, carefully measuring the elution time of the third component which was observed to be 0.60 minute. The second measure- ment technique was to introduce the same sample size of air from the humidity chest. The peak for the third component was observed to have the same elution time, 0.60 minute, but increasingly larger area. The third confirmation was to introduce a small sample of pure water and air by a microsyringe and to observe the elution time of the third component. It was -0.60 minute. Great care mustbe used in this latter verification step due to the necessity of over-loading this instrument by the introduction of such a-large sample; damage to the measuring thermistors may be caused. The controlled environmental room was at 07% relative humidity and 75° F. when the air samples were taken; this correSponds to 1.0 x 10-5 gm. of water in 1.0 cc. of air. The results of this data are tabulated in Table VII and instrumental conditions are Specified in the table. A reduced scale drawing of the chromatogram obtained . "(lT'I'l 3‘.- ~. '- VLU 4 I... .1 0...". \ hrea n .H i a I T‘\H-~--'.( 1.: .. 1 . ‘.-‘ \. .LCHtL'lCIi' --_Y VII 7? L4 TABL' ta from the lit. cm. L/' 'eight Pk. .,... ii .1: 1 -11. 9 l Full Ti 3and “mple «JUL c9. Volume F‘I - Trial 42 C C"? C\ M’\ (:2 km 0 \ C-\.:, r»: Cri- . O O O O I O O UNLW 1“- t- H U3 0‘] 0') (\J 1"! (\3 V1 Y—“I +-t+-I II I I U-\ L“ L“ ”W L“ UN I C) C3 01‘! CC- C.‘ C7" Cl 3' | (\- o o o o o o I o 0\0\c\0\0\0\ It» I I I + I I + I + (7- 1r\ 0 \O \O C {\H (\2 L" \ Cm ("N \f\ U\ C \ C‘ (\- \C' U\\O :1 :t 42' :r O O O O O O O O COOC‘CC‘OO O O c O C 0 mm Mom: ("\ ms,- (W3 0 I O O O I O I HHHHHrit—IH OCDCTC>OCDU\U\ C>O\C\C~DMD~P»V3 O I O O O I O O r-ICTCOCC‘CZIO C‘OOCOCDOC‘J 0000.000 HV—{r‘lr-Iri'r'lr-iv'i v—4 N (“x—Ll" UN)- Ce 0:: G:- i'vi c..- O 0H C. C}- 0 or‘i H 0 E; +3 C) \ .C- \ . C" o 0 PL: :50 a in) c... VHH \fl (\1 -r'I I H '11: C. VI 0 O L) 1 H (H O (\I C) L) a L'\ . ("r‘ 71 0’3 .531 CS >3 U) {.1- ,CJ r’l 03 C5 $4 Pd Z O 0) CW:- N G) --—-1 I: r-I “4-3 .,—: o "—2 ff: [1* I; C." +3 0 5:: a L'W E— L\- "‘ 0..» r4 $4 a O 2: r} O '-5 3 L"; ""1 12$ S—i re :5 C," ""3 5:; M E u": {—3 .H E G) #3 La 4—) r4 1‘ 0..“- E3 'P O ----«i Z 0 PK.) $.- 05 H F‘I\4—) C; 0“» c3 3;: pt:e ‘39. H L\~ . :5 :J 4.5 +3 e-m . 5.1 .1! O O 'r' ' 9-1 5.5 m E Q) - c3 4—) O 053:: 0 2‘5 C‘ S C F—l r4 0 .r1 0 H :3 o J-f‘ ,.) I. b- ”’3’? I. 1 x. o. q ..— leu '1 H (\3 C"\\'l7 \C: \C’ \. (‘fl\ 0 o o o C\\./') (P\ O‘\ (\I ('1' + I I I C \ (\‘\-t' Cx’ o o o o H V-{ + I + l C13 -:1' V'I 0 7‘1 0. :3 O O \O N T-I L‘r\ :i C‘\ H O O (3 C C) \i’\ (“\(T \lz“‘ o o o o H H ‘4 r1 OOCJO r-I' T’I H r'l H No\;‘* {tit-“lit . U“ ‘x a DOVE .n the Wee -et' d run b tl'l ,— C 4- Li A '3 '1‘211’3 temperature of the MI *1 A4- I‘O' 1 taken 1 -no control over 4 .a syring- he t 43 by the Fisher instrument may be seen in Figure XI. It should be noted that the noise to signal ratio is high and some loss of precision may be inherent in the calculation of peak areas by using the normal chromato- graphic calculation method based on triangulation of the peak areas (11.23). I 4h TH flTT’fififi ‘8”: i l"- . L \JViéL—l «L “Hun. 4 r“- +- w, ..AV 0 7n - - -°- 1'- JlLHOf vpfOJJqurifl OL hater in air ...J- Air - water 3 (_ Air + mate“ measured by 32 meas‘red by 31 Hater measured by 32 \I ,\ Helium flow 125 oo./min. 1.0 00. load of air from atmosphere at u7; R. H. and 750 F 45 CONCLUSIONS AND FUTURE WORK Results and conclusions A support material Porapak has been evaluated and found to possess excellent stability in a wide temperature range (up to 250° C.). These porous polymer beads are physically stable and they vibration pack well in small diameter columns at room temperature. Porapak did not bleed and it yielded excellent high temperature sensitivities. This material can be conditioned completely in four hours at 230° C. with a helium purge. Porapak has demonstrated the ability to separate water both in the liquid and vapor phase from other polar and non-polar compounds; the packing has shown the ability to separate water cleanly with minimum tailing and rapid elution times (0.60 to 5.0 minutes) dependent on column dimensions, carrier flow, and temperature. Porapak may be coated as any other column packing material and by possessing geometrical uniformity excellent chromatographic separations result. The research has yielded water sensitivity data for both of the instruments presently used at the School of Packaging. The data from the evaluation of the Burrell instrument supports the statement that water in liquid form as well as water vapor may be detected. The Burrell was operated at a sensitivity level of 10 and an attenuation of #6 5. The data shows that a level of 3.00 x 10'“ gm./5 cm.2 (See Figure VII) would be obtained with the 6.5 ft. x 3/16 in. O. D. aluminum column. The data obtained using the 2.5 meter x 2.00 mm. I. D. stainless steel column further indicates separations using this column yield approximately 6.8 times as much signalas the 3/16 in. aluminum column does. From this data it appears that the 2.5 meter x 2.00 mm. I. D. stainless steel would have an easily attainable sensitivity of 2.1 x 10'5 grams per 5 cm.2 for water. This figure is relative in that an arbitrary area (5 cm.2) is defined as the necessary minimum area required for chromatographic cal- culations. Even with this requirement the increase of the Burrell sensitivity by 50 times (the electronic maximum sensi- tivity level) would put the level of detection at 1.05 x 10"7 gm./5cm.2, well below the minimum necessary to detect water between 1 - 100 % Relative humidity range in 1.0 cc. air at 73° F. This sensitivity indicates that indeed one may use the Burrell for water vapor measurements in air. It must be stressed that long and arduous conditioning and many instrumental corrections will be necessary before the Burrell instrument may be used quantitatively for either water or water vapor measurements. The Burrell instrument is extremely sensitive and reproducible to most organic compounds, and, with the non-bleeding characteristics of Porapak, this instrument appears better suited at present 47 for organic vapor separation and detection applications. The Fisher instrument uses a thermistor thermal con- ductivity cell for detection and from the experimentation and the reduced scale drawing (Figure XI) there is undeni- able evidence of the ability to measure water vapor in air. The obvious candidate for a chromatographic detection system of water vapor is a small cavity thermal conductivity cell fitted with nonreactive thermistors. A reason for obtaining different instrumentation is that the preferred level of detection isin the order of 5 - 10% relative humidity. The Fisher Gas Partitioner appears to be limited at higher levels of water vapor due to the overloading and damaging of the metallic oxide thermistors. Liquid water sample measurements are present- ly prohibited using the Fisher instrument as there are no provisions for heating the columns and many adsorption- desorption (tailing) effects would be noted with liquid water samples. Because the column could not be heated, a flow rate of 125 cc./min. helium was necessary to minimize tail- ing when water vapor was separated on the Fisher instrument. Another reason for using thermal conductivity that is Just as important as the sensitivity requirements is that ioniza- tion detectors destroy the sample whereas thermal conduc- tivity cannot. One other advantage in thermal conductivity 48 detection may be in the intrinsic simplicity of this detector which means that lower sensitivity but more stability and re- liability in operation are usually noted. I As a means of comparing the Fisher with the Burrell K-7 the minimum sensitivity of the Fisher was calculated from the 8 trials of room air and using the arbitrary re- quirement of 5 cm.2 area for the water peak. The Fisher 4 grams per 5 will yield a maximum sensitivity of 1.01 x 10‘ cm.2 compared with the maximum theoretical Burrell sensi- tivity of 1.05 x 10"7 gm./5 cm.2. From the summarized results four major conclusions may be stated. 1. Detection of water vapor in air has been accom- plished using the Fisher Gas Partitioner. 2. The Burrell experiments indicate that under ideal instrumental Operation the detection of water vapor in air would be easily accomplished. 3. The preliminary work With the Burrell indicates much develOpmental and instrumental conditioning will be necessary for use of this instrument for water measurement studies. 4. The reproduction of the Fisher chromatogram pro- vides the School of Packaging with "evidence of thermal conductivities sensitivity and ease of water vapor detection. 49 The best choice of detection for water vapor in air appears to be a small cavity thermal conductivity cell fitted to a temperature programmed two column gas chromatograph utiliz- ing a system such as shown in Figure XII. Future work This section will first deal with suggestions to im- prove existing equipment, taking into consideration that the suggestion of acquisition of a temperature programmed small cavity thermal conductivity gas chromatograph appears to be the best approach to obtaining the desired sensitivity and versatility. . (The suggestions for improving the Fisher apparatus are as follows: First, a system of columns such as shown in Figure XII would be successful if some external means of temperature programming the column were to be developed. Secondly. the sensitivity increase noted when column volume was reduced on the Burrell would also be noted in a thermal conductivity instrument such as the Fisher. The sensitivity 1‘ em./ of the Fisher Gas Partitioner was found to be 1.01 x 10' 5 cm.2 using a 3.0 ft. x 3/16" column. If the volume of both the column and the thermal conductivity cell were re- duced by a factor of 10. the sensitivity would be increased by 10. The cell volume may be changed easily and will not affect the separation efficiency; however, the column dimen- sions must be such that clean separations will be made. 50 FIGURE XII V Reference - -~ ..- ., .1“ r: .J- iroposed ChrodatOwlaphlc gyfibed A f\ U K.) sample inlet l temperature programmed Porapak column for H20, CO separatIon exhaust tube column a 2 I Ir —\ ' 0" ’ y o J_ _ ~ 31 and 82 are detecting thermist0°s. 51 Water vapor and carbon dioxide would be trapped after de- tection at the S detector by the drying tube and column #2; 1 other separations (02, N2) would be done on the #2 column utilizing the $2 -detector. If the Burrell instrument is to be used for measuring oxygen-containing compounds (such as water vapor using the standard tungsten filament equipment), it must have some means of electronically maintaining stable electron emission levels. Another method less preferable would be to incorporate a method of injecting a hydrocarbon to chemically create a suitable filament environment. Another way to eliminate this emission level insta- bility would be to substitute a non-reactive material for the reactive tungsten. The tungsten filament may be changed to a non-reactive .material such as rhenium (21,22) which does not form stable carbides; its nitrides are also un- stable; its oxides are conducting and it is not involved in a water cycle in the same way as tungsten. A rhenium filament is limited by one drawback. This is that its vapor pressure is 156 times that of tungsten at the same electron emission. but the electrical resistivity of the material is higher so a larger diameter wire may be used. The life of rhenium is limited by evaporation. but it may be used for long periods quite well if prOper electronic 52 controls are observed. The life of the tungsten filament is not normally limited by evaporation, but rather by the fragility of tungsten recrystallized by prolonged heating at high temperature. and tungsten is caused to distill as the oxide by the forementioned water cycle phenomenon. Acceleration of the normal (brittle) hardening is also caused by this water cycle. With this filament material change one may also want to increase the amount of sample introduced into the ionization detector which would increase detector sensitivity. One last instrumental change is recommended. The Burrell electronics. especially the electrometer circuit, should be stabilized. This research dealt primarily with the chromato- graphic separation and detection of water vapor, and second» 1y with the application of this chromatographic separation and detection method to packaging research. The results of this work may be directly applied to further the technique of water vapor transmission rate measurements. This specific . chromatographic technique may be also used in determining the water content of packaging materials. It may be used to analyze water absorbed by hygros00pic products as well as materials giving the package researcher another way of studying package performance. A third area of interest was to determine the other area in which gas chromatography may 53 be applied to packaging research. The time spent with the ionization detector instrument indicated that the Porapak material coupled with ioniza- tion detection will be an excellent way to study product de- gradation such as those found in food products (25). This area alone would yield a wealth of information and develop- mental data. The study of odor transmission rates as well as permeants such as 02, N2, C02, and water vapor would give design data to extend shelf life. The use of chroma- tography in studying residual solvents by S. G. Gilbert (5) illustrates but another area of application to packaging research. One more tool for the packaging engineer is found in gas chromatography. This research has enabled the School of Packaging to evaluate and plan for further use of this valuable tool in packaging research. Hopefully. this type of research will continue and the results utilized for solving packaging problems. 1. 2. 7. 9. 10. ll. 12. BIBLIOGRAPHY Books, Articles. and Periodicals Andrews, M. R. JournalZQf PhysicalTChemistry. Vol. 27 (1923) 270. Bennett, 0. F. Analytical Chemistry. Vol. 36 (1964) 684. Beynon, J. H. Mass Spectrometry and Its Applications to Organic Chemistry. New York: Elsevier Publishing Company. 1960. Davis, Donald w. "Gas Permeability, An Isostatic Test Method." Modern Packaging. (May 1946) 145. Gilbert, Seymour G. Modern Packaging. (February 1966) 129. Guild, L. V.. M. I. Lloyd. and F. Arrl. "Performance Data on A New Ionization Detector." Instrument Society of America. A paper presented at the 2nd International Symposium of Gas Chromatography." E. Lansing, Michigan, June 1959. Hall. L. D. Review of Scientific Instrumentation. Vol. 28 (l957)7653. "Heat Sealing Characteristics of Materials." Modern Packaging Encyclopedia 1966. 475. Hollis, o. L. Analytical Chemistry. Vol. 38 (1966) 309. ' Kirkland, J. J. Analyticavahemistry. Vol. 35 (1963) 6. Littlewood, A. B. Gas Chromato ra h . New York: Academic Press, 1962. . Lockhart. H. E. "A Review of the State of the Art and Summary of the Permeability Theory." Multi- sponsor Technical Report #1 (March 30, 1962) Michigan State University. School of Packaging, E. Lansing, Michigan. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. Lockhart, H. E. Design and Application of an Iso tactic Permeability Test Method for the Study of Cellulosic and Plastics Barrier Materials. Packages, and Closures." A thesis. (1965) Michigan State University. E. Lansing, Michigan. "Method of test for Water Vapor Rates Materials in Sheet Form, E-96-63T." American Society for Testing and Materials. "Method of Test for Gas Transmission Rate of Plastic Film and Sheeting, D-l434-63." American Society for Testing_and Materials. "Method of Test for Water Vapor Permeability of Pack- ages, D-895-62T." American Society for Testigg and Materials. Moore, J. C. Jou;na1_of PolymeygScience. Vol. 2 (1964) 835. Porapak. A commercial publication of-Waters Associates, Inc.. Framingham, Massachusetts. 1965. Pprapak and AerOpak. A commercial publication of the Varian Aerograph Company, 1966. "Properties of Saturated Steam." gandbook of Chem- istry and Physicsgg41§t Ed. Cleveland, Ohio: Chemical Rubber Publishing Co. (1959) 2448. Robinson. C. F. and A. G. Sharkey. Review of Scien- tific Instrumentation. Vol. 29 (1958) 250. Ryce, S. A. and w. A. Bryce. "An Ionization Gauge Detector for Gas Chromatogra hy." Canadian Journal of Chemistry. Vol. 35 (1957 1293. Saframski, Leo w. and Stephen Dal Nagare. C & E News: Vol. 39 (June 26. 1961) 102 - 114. o C & E News. Vol. 39 (July 3. 1961) 76 " 830 Schwecke, w. M. and J. H. Nelson. Analytical Chem- istry. Vol. 36 (1964) 689. 26. 27. 28. 29. 30. Semat, Henry. Fundamentals of Physics. 3rd Ed. New York: Holt et als, 1960. Sharkey, A. G. A report. 5th Annual Meeting American Society for Testing and Materials Committee E-14. New York, 1957. Sugimoto, A. and H. Hoshino. Reports to Scientific Research Institute, Tokyo. Vol. 27 (1951) 431. Wexler, Arnold. "Electric Hygrometers." N B S Circular 586, September 3. 1957. Other Sources . Personal interviews with L. B. Westover. The Dow Chemical Company. Midland. Michigan. ‘immw