DEPENDENCE OF THE HELIUM II FILM TRANSFER RATE 0N PRESSURE HEAD, FILM HEIGHT, AND SUBSTRATE Thesis far '?:he Degree of Ph, D. MICHIGAN STATE UNIVERSITY CARL JAMES DUTHLER ' A 1970 r'rlfi”435! ‘IE‘Jljzallfit‘tél'zzl 1:! Michigan State University I“ This is to certify that the thesis entitled Dependence of the Helium II Film Transfer Rate on Pressure Head, Film Height, and Substrate presented by Carl James Duthler has been accepted towards fulfillment of the requirements for Eh.D. degree in_Eh¥S.i_QS Major professor Date August 28 , 1970 0-169 ’ 3" muomc av "TWA” IIIIAE & WW I ,, 590K BRIBERY INC. I , ‘ L'BRA B 1000 X, the thickness of the He film is equal to the thickp ness over a pure Ne substrate so that we have an effective Ne beakp er. One-half saturation, where the He film has the average of these thicknesses, occurs at E - 100 X. The thickness as a function of Na coverage, as shown in Figure l, is assumed to be qualitatively valid for a glass substrate instead of Xe substrate. That is, we assume that we need only put a 1000 X neon layer on a glass beaker to have an effective Ne beaker. Dzyaloshinskii at al.37 have objected to the use of the Hie-Lennard-Jones potential for a solid interacting with a liquid. They point out that the Mic-Lennard-Jones potential is strictly valid only for isolated atoms and that corrections must be made for the screening of neighboring atoms. This requires knowing the entire electromagnetic spectrum of both the helium and neon. This is beyond the scope of this thesis. However, the correction should be smallest for the case of neon substrate. The simple calculation above is for a static film interacting 360 340 0 He Film Thickness d(A) N N 01 (,4 (D (D O N O O O O N 43 O 220 0 Figure 1: Calculated thickness of the helium film, d with h - 1 cm, ssssss thickness of Ne coating, 5, on a Xe substrate. 18 l l i l .1 I I I l IO Ne Thickness {(A) no2 I03 oIo“ IOS 19 with an infinite wall. Kontorovich38 predicted that the thickness of a moving Film would be less than the thickness of a static F11m. However, Keller39 has found experimentally that the thicknesses are the same. Goodstein and Safr'ifmanl.0 have considered the theory of a moving film in more detail and find that both moving and static films have the same thickness as is determined by Equation 31 above. Various methods have been used to measure the thickness of the Film experimentally.‘ Bowers41 has weighed the Film adsorbed on a metal foil. Atkins42 has calculated the thickness of the film from measurements of oscillations of the film. Keller39 has also measured the thickness, using a capacitance technique which is sens sitive to changes in the thickness. The most direct and accurate measurements of the helium film thickness are by L. C. Jackson and co-workers.‘ They measure the ellipticity of light reflected from a surface that is covered by the He film. Ham and Jacksonl'3 k, is equal to 3.0 x 10.6 cm4/3. This is the basic measurement find that the thickness parameter, that is used for the thickness of the He film on a glass substrate in this thesis. C. Dynamics of the Film Transfer Dynamical limitations to measurements of the properties of critical velocity in beaker filling experiments are examined in this section. The flow is considered to take place in a channel h8V1n2 the vertical cross—section d - k/hl'l3 , as discussed before. In the experiment, the rate at which the beaker Fills is measured. The transfer rate per unit circumference, o, is related to the ve- locity of the inner level, dz/dt, for a beaker of radius r, by: 20 o - (r/2)dz/dt. (41) These measurements of o are related to the superfluid velocity by 0 ' (pa/p)v.d. (42) The superfluid obeys the two fluid equation, Equation 2: dv’s/dc - 4m)??? + air. (43) In the case of isothermal, irrotational flow, as we have for film transfer at subcritical velocities, this becomes”2 dvs/dt - ave/at + (383)38 - -(1/p)'v’1>. (44) For a co-ordinate 1 parallel to the surface of the beaker and par- allel to 3;, this becomes: 3 3 1 2 l 3? at V3 + a“? v3 - 73- !, (45) Integrating along the path 1 from the outside level of the beaker to the inside level, we get: inside [Give/add). + 6- v3) =- -(1/p)AP - -gz, (46) outside where z is the level difference. Because of the nearly zero velocity inside and outside of the beaker, the second term on the left hand side of Equation 46 drops out. Using the continuity equation, vhd - constant along path, (47) and substituting for v8 in terms of dz/dt from Equation 41, we get a M dzz/dt2 - -pgz, (48) a where M - (pz/Zps)f(r/d)dl. This relates the acceleration of the measured level difference to the driving pressure, age, for sub- critical flows To account for the critical velocity, we add a frictional 21 pressure, Ps’ to the right hand side of Equation 48:20 t n dzz/dtz - p. - pgz. (49) We are primarily interested in the frictional pressure as a function of superfluid velocity, v8, or transfer rate, a. This can be measured directly for low acceleration and large a. In this case M? dzz/dt2 is small compared to pg: so that Equation 49 says that the frictional pressure is equal to the observed driving pres- sure, i.e., P. - pgz. For small 2 and large acceleration, the acceleration term he- comes important so that the frictional pressure cannot be measured directly. In this case Pa is small and the levels oscillate with small clamping.1l"l'2 The apparatus used in this thesis, however, is not suitable for studying these oscillations. III. EXPERIMENTAL A. Clean Glass Substrate Apparatus. Atkins44 in 1948, and van den Berg and de Haasl'5 independently in 1949, reported transfer rates as much as twenty times larger than those originally reported by Daunt and Mendelssohn.13 Bowers and Mendelssohnl‘6 showed that these enhanced transfer rates were a result of impurities, such as air, condensed on the beaker sur- face. These enhanced transfer rates for contaminated beakers can arise either from an increased film thickness or an increased micro- perimeter of the beaker. In the case of a granular impurity, the microperimeter is larger than the apparent macroperimeter and, in addition, a thicker film can result from liquid helium.he1d between the grains by surface tension. McCrunb and Eisenstein“7 pointed out that polar impurities, such as water, can also result in a thick- er film. The thick film, in this case, is a result of a stronger attractive force, of the polar substrate to the helium, than the usual van der Weals forces. Smith and Boorse17 investigated these problems in more detail. They found that, by being careful to prevent impurities, the enhanced transfer rates could be eliminated but a, typically 10%, background variation in the transfer rate remained. These variations prevented them from seeing a dependence of the transfer rate on substrate 22 23 material for the meterials they investigated. The apparatus used in this thesis was designed to provide a very clean, and vibration free, environment for the beaker in order to minimize these problems. To do this the beaker was enclosed in an experimental chamber rather than exposing it to the possibly dirty environment of the liquid helium bath. In addition, rather than condensing possibly impure helium gas into the experimental chamber, the chamber was filled by admitting helium in liquid form from the bath. Another advantage of the chamber was that the helium level of the chamber remained constant, while the beaker was filling, whereas the helium level of the bath decreased slowly with time. The experimental chamber and beaker are shown in Figure 2. A 2 in. Kovar to glass seal was used to make the chamber. The champ her was demountable at the top by means of brass flanges that were sealed with a lead O-ring. Five ampere fuse wire was used to make the O-ring whose ends were soldered together to form a closed loop. Liquid helium was admitted to the chamber through the needle valve at the top of the apparatus. This valve was made from a Hoke #3242 M 43 needle valve which was modified to reduce its mass. Connections from outside the dewar to the experimental chamber were made with a 3/8 in. o.d. stainless steel tube. Two other 3/8 in. stainless steel tubes served as supports for the chamber. Con- nections were made from the experimental chamber to mercury and oil nanometers, a mercury bubbler, and a diffusion pump system. The nanometers were used to measure the vapor pressure of the liq- uid helium in the chamber in order to determine its temperature. The bubbler served as a vacuum tight, safety valve which would allow 24 To Monomeier ”a W He Inlet Bath Valve ’ ////// \ Pb O-ring ___ .——_-—_4 __ -—_._ .__. —4 ___.___. __ h——— —— -< .—————4 _— ~——— _——-— .1 v—————--q ——— l———— _— p—____‘ ___ _._—_4 _.__. _‘ ,__._ ._ .. __.__ ___ Figure 2: Schematic drawing of clean glass substrate apparatus. 25 helium gas to escape from the chamber in case of pump failure. The diffusion pump system consisted of a CVC model 21, air cooled diffusion pump which was backed by a Welch model 1402 B, 140 liter per minute mechanical pump. An, approximately 2 liter capacity, liquid nitrogen cold trap placed between the pumps and apparatus, served as e cryopump and in addition prevented pump oil from diffusing to the beaker. Before each run the diffusion pump system was used to degas the beaker. After each run the mechani- cal pump, without the diffusion pump, was used to remove the liquid helium from the experimental chamber. 3 Pressures above 10- Torr, in the diffusion pump system, were treasured with a CVC model GP-140 pirani gauge. Pressures below 10"3 Torr were measured with an H. S. Martin cold cathode ioniza- tion gauge. The highest vacuum attainable with this system was a- bout 10-5 Torr. Beakers used in these experiments, shown in Figure 2, were made of 3 mm i.d., 5mm o.d., and 10 cm height Corning #7740 Pyrex tube. The beaker was mounted on a 1 1/2 in. Pyrex base which had small Pyrex hooks for removing the beaker from the demounted cham- ber. Care was taken to specially select glass tube, from the stock of glass at the M. S. U. Glass Fabrication Laboratory, that was clean and scratch free. The rim of the beaker was formed by care- fully cutting the Pyrex tube with a Carborundum wheel. The rim was then lightly fire-polished to remove scratches and to smooth the edges. The exact radius of the beaker was determined volumetrically. The mass of room temperature, distilled water in the beaker was 26 measured as a function of height with a mettler analytical balance. Using the density of water, the resulting volume versus height curve was plotted and was fit by a least squares straight line. The slope of this curve gave the cross-sectional area of the inside of the beaker from which the inner radius was determined. Twa different, but otherwise identical, beakers were used in these experiments. Beaker 1 has an inside radius of 0.162 cm and beaker 2 had an in- side radius of 0.156 cm. When the Pyrex beaker'was removed from the demounted chamber, it could be cleaned in solvents that were not compatible with the metal parts of the apparatus. The following cleaning procedure gave reproducible transfer rates and was used before each run during the clean glass substrate study: The beaker was first rinsed with distilled water and then with methanol. This was followed by cleaning with detergent in an ultra- sonic bath and rinsing ultrasonically with distilled water. The beaker was then rinsed with nitric acid and rerinsed with distilled water. It was then heated to 200 °C in air with a heat gun to re- move adsorbed water. Finally, it was mounted in the experimental chamber and heated to roughly 75 °C with an infrared heat lamp while pumping to about 10'6 Torr with the diffusion pump system. The pumping*was continued for one or two days prior to cooldown. The entire cleaning procedure above may not be necessary be- fore each run. Allen and Armitagel'8 have used a similar cleaning procedure and report that, after the initial thorough cleaning, only heating and pumping between runs is necessary. With the neon sub- strate apparatus, described in the next section, it was also found 27 that it was only necessary to prevent impurities from entering the chamber and to diffusion pump between runs after the initial thor- ough cleaning. Impurity problems were encountered in the preliminary runs be- fore the actual thesis data were taken. It appeared that the beaker in the original non-demountable apparatus got dirtier each time it was cleaned, £.e., the transfer rate increased after each cleaning. This was most likely a result of water, used in the cleaning pro- cedure, being adsorbed on the beaker. These problems led to the design of the demountable apparatus so that the beaker could be re- moved to be cleaned in stronger solvents and to be heated to higher temperatures. These observations with the nonedemountable, demount- able, and neon substrate apparatuses indicate that the crucial step in the cleaning procedure is heating to 200 'C to remove adsorbed water. Although these observations indicate that the beaker may have to be cleaned thoroughly only once, the cleaning procedure, never- theless, was used with slight variations before each run as a test of the reproducibility of the cleaning procedure. This also ensured that the maximum initial cleanliness was reached. The experimental chamber'was immersed in a liquid helium.bath in an H. S. Martin, glass, liquid helium dewar with a removable liquid nitrogen dewar. Both the nitrogen and helium dewars were strip silvered with 3/4 in. viewing slits. Liquid helium bath temp- eratures were lowered by evacuating the helium dewar with a Heraeus-Engelhard #8225, air-cooled, 147 cfm vacuum.pump. The temp- erature of the helium bath was regulated to better than *0.01 K 28 for temperatures above 1.5 K.with a Cryonetics Mark II mechanical pressure regulator. Comparable temperature regulation was achieved for temperatures near 1.3 K‘by regulating the pumping speed with a 2 in. gate valve in the pumping line. The lowest temperature attainable with this apparatus was about 1.2 K. Vibrations of the apparatus were minimized by placing flexible bellows along the pumping line. A 4 in. x 24 in. copper bellows was located in the main 4 in. pumping line near the pump which was about 15 ft. from the apparatus. Also an asymmetric arrangement of five automobile radiator hoses connecting the helium dewar to the 4 in. pumping line further reduced vibration. Three 2 in. x 15 in. hoses, one 1 1/2 in. x 12 in., and one 1 1/2 in. x S in. hose were used in this arrangement. Two different means of illuminating the beakers during these experiments gave identical results. This agrees with the results 49 who found that the trans- of Bowers and Mendelssohn46 and Pious, fer rate was independent of the intensity of the illuminating radia- tion, as long as it was not too intense. For this experiment, radia- tion, other than that used for illumination, into the dewar was mini, mized by placing radiation shields, made of aluminum foil, in the helium bath. These shields were placed above the apparatus and around it, except for vertical viewing slits. Additional shielding was provided by covering the slits on the outside of the nitrogen dewar with aluminum.foil except for 6 in. near the bottom which was left uncovered for viewing. The first means of illumination made use of an 8 watt floures— cent light. This light was shielded by a plastic diffuser which, except for a narrow slit, was covered with aluminum foil. In 29 addition, a green glass filter was placed in front of this slit. Infrared radiation from this light source was filtered by a water cell placed in front the dewar slit. using this illumination the experiment was done with the laboratory completely dark except for other low intensity lights which illuminated the laboratory note- book and instruments. A second, more convenient method of illumination used no spe- cial light source. In this case illumination was provided by room light with the laboratory slightly darkened. Radiation shields on the nitrogen dewar and in the helium bath were again used in this arrangement. B. Neon Substrate Apparatus Neon beakers were made by coating clean glass beakers with neon. To have an effective neon beaker, as was discussed in section II—B, a 1000 Z or greater thickness neon coating on the glass beaker is required to saturate the forces between the helium and substrate material. However, the coating must be smooth and uniform to avoid the rough substrate problems that were discussed in the previous section. The neon substrate apparatus is similar to the clean glass substrate apparatus except that, in order to raise the temperature of the chamber above 6 K, it was provided with a vacuum space and heating coil. If neon is condensed directly onto a surface having a temperature of A K, enhanced transfer rates are seen presumably due to rough substrate conditions. This was observed in preliminary experiments. The rough substrate in this case was most likely a result of the fact that neon atoms stick, as soon as they strike, 30 a 4 K surface.50 At higher surface teuperatures, near 20 K for neon, the adsorbed neon atoms are distributed more uniformly because they have a lower probability of sticking when they strike the surface. Also because of their thermal energy, the adsorbed neon layers, at these temperatures, are somewhat mobile and as a result are able to smooth themselves.51 Therefore the apparatus was designed so that the neon could be condensed with the temperature of the glass beaker in the range 20K to 25 K. The neon substrate apparatus is shown in Figure 3. The inner chamber, containing the beaker, was made from a 1 l/2 in. Kovar to glass seal. The outer chamber, used to form a vacuum space a- round the inner chamber, was made from another Kovar to glass seal of 2 in. diameter. The tops of both chambers were soldered to brass flanges. Two 1/2 in. stainless steel tubes supported the apparatus. One of these tubes served both as an.inlet to the vacuum, or exchange gas, space and as an outlet for electrical leads from the heater and thermometer. The other 1/2 in. tube formed a vacuum space around a 1/4 in. stainless steel tube which served as an outlet from the inner chamber. The tube from the inner chamber, as in the clean glass substrate apparatus, was connected to mercury and oil mano- meters, the mercury bubbler, and the diffusion pump system. Liquid heliumnwas admitted to the inner chamber from the bath through a 1/8 in. copper tube from the needle valve at the top of the outer chamber. The needle valve used in this apparatus was constructed by the ML 8. 0. Physics Department Machine Shop. These chambers were not easily demountable so that the beaker, 31 Outer tube from inner chamber Exchange gas and wires Needle valve Pt thermometer Vacuum space Outer chamber Inner chamber Beaker Heater 32 TII'TIUOCD ' __.[[fl]]] Figure 3: Schematic drawing of Ne substrate apparatus. 33 which was fastened to the inner chainer, could not be removed for cleaning before each run. The beaker was initially cleaned and heated using the procedure given in the previous section. The inner cham- ber was then assembled taking care not to get the beaker dirty. Transfer rates obtained with the asseabled apparatus were the same as those obtained during the clean glass study. This indicated that the beaker was not contaminated during the assembly of the inner chamber. As was discussed in the previous section, the beaker was kept clean between runs by pumping. Occasional runs were made without a neon coating during the course of the neon substrate study. The transfer rates observed in these runs were again the same as the transfer rates obtained during the clean glass study. This demon- strated that the glass beaker in the neon substrate apparatus remained clean during the neon substrate study. The exact inner radius of the beaker was again determined volu- metrically. Because the beaker was fastened to the massive inner chamber it could not be calibrated directly. Another beaker, made from the same section of glass tube that was used in the construc- tion of the experimental beaker, was used for calibration. The inner radius of this beaker was found to be 0.160 cm. Manganin resistance wire having a resistance of 10 ohms per foot was used for the heating coil. Approximately 100 feet of this wire was wound longitudinally around the inner charter, and the base of inlet tubes, leaving vertical slits for viewing the beaker. Nylon thread and masking tape were used to fasten the heater wire to the inner chafler. The ends of the heater wire were connected 34 to a small terminal strip at the top of the inner chamber in the vacuum space. he leads from the terminal strip were routed through the exchange gas tube and were connected to 2 pins of a vactnm: tight, 9 pin connector at room temperature. A miniature platinum resistance thermometer was used to meas- ure the temperature of the inner chamber. The thermometer used in this apparatus was an Artronix #PS-l which had the serial nunber X-159. The thermometer had a square cross-section, 0.2 in. on a side, and was 0.05 in. thick. Thermal contact between the brass top of the inner chadaer and the thermometer was made using Apiezon N vacuum grease. The bottom of the terminal strip, used for connect- ing electrical leads, held the thermometer in place. The resistance of the platinum resistor as a function of tempera- ture was calibrated from basic measurements of the resistance at 4.2 K and 273 K. These twa measurements of the resistance were made with the apparatus immersed in a liquid helium bath, at atmos- pheric pressure and. an ice and water bath‘ respectively. The re- sistance, R, as a function of tenperature, T, is then given by Rm " ”1.2 x "' 201001273 K ' Rmz x ' (50) Here MT) is a tabulated function given by Whites;z for Pt. The miniature Pt resistance thermometer had only two short leads. In order to eliminate lead resistance, it was converted to a four-lead arrangement by connecting it to four inlet leads at the terminal strip. The of these leads supplied a 1 ma current to the resistor and the other two leads served as zero current, potential leads. has four leads also were routed up the exchange gas tube and were connected to four other pins of the nine pin connector. 35 The resistance of the thermometer was measured by connecting the potential leads from.the nine pin connector to a Leeds and Northrup model n+3 potentiometer. The auxiliary emfi termmnals of this potentiometer were used to monitor the l .A current through the platinum resistance thermometer. This was done by measuring the potential difference across a one ohm standard resistor which was in series with the resistance thermometer. The heater leads from the nine pin connector were connected to a Variac variable ac voltage source. A.1000 ohm resistor was placed in series with the heater and Viriac to reduce the heat imp put. Temperature regulation. for temperatures above 6 K, was achieved by manually regulating the voltage output of the Variac. This was done by adjusting the voltage while variations of the resistance of the thermometer, from the value set on the potentiometer, were monitored on the dc potentiometer null detector. . While the neon was being condeued, the chasbsr tenperature was above 4 K and a vacuum was maintained in the vacuum, or exchange gas, space. After the neon had been condensed,the inner chamber 'was brought to a temperature of 4 K. At this time helium gas was admitted to the vacuum.space before admitting liquid helium to the inner chamber. The helim exchange gas was stored at atmospheric pressure in a one-half liter, glass flask outside the dewar. A glass, vacuum stopcock‘was opened to admit the exchange gas to the evacuated vacuum space. This procedure ensured that the exchange gas pressure‘would never exceed atmospheric pressure even when the apparatus was warmed to room teqerature. In addition, the stopcock served as a safety 36 valve in case liquid helium should leak into the vacuum space from the bath. The amount of neon needed to coat the beaker was determined by assuming that the neon uniformly covered all the surfaces of the beaker and the inside surfaces of the inner chadaer. Using the area of these surfaces, the amount of, room temperature, neon gas required to form a desired thickness was easily calculated. The neon gas required to form a desired thickness neon coating was stored in a one liter glass flask outside the dewar. Typically a 1000 Z coating was used which required a room temperature, neon pressure of 5.3 Torr in the storage flask. The gas handling system for the neon is shown schematically in Figure lo. Mercury and oil nanometers were used to measure the pressure of the neon. After the desired partial pressure of neon was admitted to the flask, helium gas was added filling the flask to a pressure of one atmos— phere. Matheson Research grade neon having a purity of 99.9991 was used in these experiments. The helium gas having an estimated puri- ty of at least 99.99% was obtained from the M. S. 0. Physics Depart- ment. The purity of the helium gas was tested by admitting only helium gas, without neon, to the experimental chamber. It was then observed that the transfer rate was the same as for clean glass. This demonstrated that no impurities from the helium gas that was mixed with the neon gas were condensed on the beaker surface. During a run, the helium and neon gas mixture was admitted to the experimental chatter with the initial chamber temperature at 25 K. The helium gas served bath to distribute the neon and to 37 [“°‘”?_.L 1* H L Pumps] 1- ; IL Monometer l Expt. Chamber Figure 4: Gas handling system for He. 38 maintain thermal equilibrium within the chamber. After the gases were admitted to the chamber, the chamber temperature was than low- ered at a rate of 0.2 K/min from 25 K.to A K condensing the neon on the beaker. Transfer rate data indicate that this procedure re- sulted in a smooth surface for coverages of about 1000 A, although no independent test of the surface‘was made. For the condensation of the neon,the liquid helium.level in the dewar‘was set about half way up the outside of the experimental chamber. The tube from.the inner chamber to the needle valve pro- vided thermal contact between the inner chamber and the helium bath. The heat from the inner chamber was dissipated primarily in the cold helium vapor surrounding the top of the outer chamber. Heat was also dissipated in the liquid helium.by heat conduction down the glass walls of the outer chamber. The liquid helium transfer tube from the storage dewar to the experimental dewar remained connected during the neon condensation. A slight overpressure on the storage dewar maintained a slight flow of cold helium gas through the transfer tube to the experimental dewar. This cold gas served two purposes. It, first of all, main? tained a stream of cold gas on the top of the apparatus to provide cooling to the inner chamber. In addition, this cold gas kept the transfer tube cold so that more liquid helium.could be transferred after the condensation of the neon without warming the apparatus. 0. Experimental Procedure The measurements made during an experimental run to determine the He film.transfer rate are described in this section. As was described in the previous two sections, the surface of the beaker 39 was prepared for running before liquid helium was admitted to the experimental chamber. This preparation involved, in the clean glass substrate study, thoroughly cleaning the glass beaker one or two days prior to cooldown. In addition, during the neon substrate study, a layer of neon was carefully condensed on the surface of the beaker before admitting liquid helium to the experimental chamber. During an experimental run, approximately four liters of liquid helium were transferred into the experimental dewar and cooled by pumping to the desired temperature. Liquid helium was then admitted to the experimental chamber from the bath, by means of the needle valve, filling the chamber outside the beaker to a depth of about 2 cm. 'me position of helium level inside the beaker was then amas— ured as a function of time until the inner and outer beaker levels reached equilibrium. Usually the level difference, 2, was monitored over a distance of 1 1/2 on which took a period of time of about 30 minutes. While the beaker was filling, the outer level remained at a nearly constant distance h below the beaker rim. The position of the inner and outer levels with respect to the beaker along with the distances h and s are shown in Figure 2. The positions of the helium levels were measured with an ele- gant Wild-Hesrbrug model 104326 cathetometer. During a beaker fill- ing, the position of the inner beaker level was measured and recorded as a function of time with an accuracy of 0.001 cm taking height versus time readings usually every 30 sec. The time during the beaker filling was measured with a stopwatch. The stopwatch was started at the time of the first height reading and subsequent 60 readings were made with an accuracy of *1 see without stopping the stopwatch. Figure 5 shows the level difference, 2, as a function of time for three of five beaker fillings at different h's during the same run. The transfer rate is proportional to the derivative of the x(t) curve. we notice that these curves are nearly straight and have only slightly different slopes. This agrees with the original observations of Daunt and Mendelssohn;3 that o is nearly independent of pressure head, 2, and height, h, of the beaker rim.above the reservoir. However this thesis is concerned with the slight departures of o from this simple behavior. These curves in Figure 5 are slightly concave downwards reflecting the dependence of a on 2. Also the height dependence of o is manifested in the different slopes of the curves taken at different h's. 41 9» .Su homes; may on .m .9530: nonhuman sauna me 35:: use—see sauna new can» we nowuonmu e as .m .soneueuuum Hobs ow mm on mm ON m. 3.5 25... 0. an enough I ' [Eoomdu . _ xmm._u._. J. O. 0 lies. es Ill l'llIllll'Illll’lll-I‘lllll"ll'llnll'llllll .23 £8 .28 _ ..m..om_.wu.... _ _ _ _ 3.0223"; _ _ _ I. 0.. N._ md v.0 (Lu 0) z souslayga |aA91 42 D. Data Reduction and Experimental Accuracy The transfer rate per unit circumference, o, is proportional to the derivative of the 2(t) curves shown in Figure 5 of the pre- vious section, £.e., o - (r/2) dz/dt. (51) Here r is the inside radius of the beaker. In this section the computer method that is used to evaluate these derivatives or trans- fer rates is discussed. In addition, the random error in o from the measurements and the differentiation procedure is estimated. The data reduction was done using the computer terminal in the Physics - Astronomy Building. This terminal was connected to the Michigan State University, Control Data Corporation 6500 com- puter which was located in the Computer Center. The following procedure is used in the computer program to 53 calculate the derivatives. A quadratic equation, z(t) - A + Bt + Ctz, (52) is fit to the first eleven data points of a beaker filling. The derivative, dz/dt - B + 2Ct, (53) is then evaluated at datum.point number 6, the midpoint of the eleven data point segment. Next, data points 2-12 are fit in the same way and the derivative is evaluated at the midpoint of this segment. By taking overlapping segments of the z(t) curve in this manner, the derivative of the curve can be determined either as a function of z or t. Transfer rates, a, are then obtained by multiplying the deriva- tives by c - (r/Z) x 10’s. This gives oin units of 10'5 cm3/sec-cm 63 so that o is numerically between 1 and 10. Figure 6 shows a as a function of x for a typical run. Each point on this curve was obtained by evaluating the derivative, which was then multiplied by G, at the midpoint of one of the overlapping eleven data point segments. The solid curve is a fitted curve which shows the trend of data. This curve will be discussed in the next section. Details of the computer program, Sigma-s, are given in Appendix A. A Fortran listing of the program Sigma-s is given in Table A1. Table A3 gives the computer output for these data. Although the data for only one beaker filling are shown in Appendix A, the pro- gram was designed so that data from several beaker fillings could be submitted and analyzed at the same time. The first computer card of the input data in Table A2 gives an identification number for the data set. The second computer card of the input gives the height of the heliua level outside the beaker, no, and the geometrical factor G. Subsequent cards give the position of the inner level as a function of time. Six height versus tin data points are recorded on each computer card. The remaining points, which do not fill one card, are then recorded one datum point per counter card. The first line of the computer output .presented in Table A3 gives the identification number of the run; the liquid helium level outside the beaker, Ho; and the geostrical factor, G. TVo other numbers given on this line of the output are the parameters A and B which describe the 0(a) curve. These parameters are discussed in the next section. The remaining computer output is presented in eight columns Mo 'E 3 5.8 - .3."- cu .. m . I0\ .0 a.s/ ”g 5.6 h .. {4.4. e "‘ Q arr/.5. b 5 4 — «M. ”‘5’". .3 .2 f at; 5.2 — 35/. — 5'2 3' g 5.0 - 1 _ 2 s '— 4.8 l l l l O 0.5 LO I5 2.0 2.5 Level Difference 2 (cm) Figure 6: Transfer rates, 0(8), from the x(t) curve of a typical run. 45 in Table A3. The second and third columns of the Table list the input data, 11.5, the inside beaker level height, E, versus tin. These data points are numbered in column 1. The fourth column gives the nearly constant difference in the inside beaker level height between adjacent data points for proofreading purposes. The trans- fer rate, a, as a function of s is presented in tabular form in columns 5 and 6. In the next section the functional form of the 0(a) curve is considered. In that section use is made of the curve l/o versus in: which is tabulated in colums 7 and 8. The random errors in the 0(2) points, arising from the measure- mmlts and the differentiation procedure. are now estimated for the 0(2) points of a particular beaker filling such as these shown in Figure 6. To do this, the root mean square deviation of the z(t) data points from the fitted quadratic equation was calculated for the overlapping segments of a few typical sets of beaker filling data. Since the z(t) curve is nearly a straight line, the error in the derivative of the eleven data point segment was taken to be the difference in the slopes of two straight lines whose end points differed in s by twice the rms deviation. Using the differ- ence in time of 300 sec between the first and eleventh data points, the error in the derivative is estimated to be *2A/300 sec, where A is the rms deviation. The average percent error for the overlap- ping segamnts of these s(t) curves was found to be about 0.62. This agrees with the error that would be estimated for the derivative using the accuracy *0. 001 cm for the cathetometer readings. The percent error of o is the sum of the percent errors of the derivative and the beaker radius. From the volume versus height 46 curve used in calibrating the beaker, random deviations in the radi- us of the beaker along its length are estimated to be *0.22. This error has, however, already been taken into account in the deriva- tive error because the derivative error was estimated from experi- mental data. Therefore the random error in the 0(2) points for one beaker filling is estimated to be 0.62. There are also larger run-to-run differences in the transfer rate‘which arise from differing runrto-run substrate conditions. These and other errors will be discussed in the following sections as the experimental results are presented. IV. RESULTS AND DISCUSSION A. Pressure Head Dependence In this section experimental results of the dependence of the transfer rate on level difference, or pressure head, are presented and discussed. This dependence has already been mentioned in the previous sections regarding the curvature of the 2(t) curves in Figure 5. The critical velocity is observed to respond to pressure, causing the transfer rate to decrease by about 102 over a change in head of 1 1/2 on in these experiments. As was discussed in Section II—C in reference to Equation 69, the dependence of the frictional pressure, P., on 0 or vi can be measured directly when the driving pressure, pgs, is larger than the acceleration tern,‘M* d22/dt2. Figure 7 shows the dependence of 0 on 2 for three of five'beaker fillings during the same experi- mental run. These curves were obtained by differentiating the same 2(t) curves that were shown in Figure 5. Measurements of the accel- eration, dzz/dt2 , made from experimental 0(2) curves such as those shown in Figure 7 indicate that for 2 > 0.01 cm the acceleration is small enough that the frictional pressure, P., can be set equal to the driving pressure, 032. For 2 > 0.1 cm, the 0(2) data may be described by 54 0(2) - l/[A - Bm(cm)]. (54) Curves of this functional formnwith A and B as adjustable parameters 47 a;(l0'5cm3/sec-cm) Transfer Rate 48 8 5 I I I I I I I I I l I l I I 80— zfl'm'm h=0.2lcm— 7.5 _ _. 7.0 -— ,.- . ,.———'o°"'000'o.".. 0. ”000...... 9’ a: I 6.0 '4’- .4 5.5 -— ”4.4—”; /" h=8.20cm 5.0 —- .9/ / T= l.65K l l l ‘}55 I l, 1 1 l .I 1 1 1 ee..e' h=2.l2cm L l ' O 0.5 LO Level Difference 2 (cm) Figure 7: Transfer rate as a function of level difference for the three beaker fillings of Figure 5. L5 49 are shown ae eolid linee on Figure 7. Thin functional form ie a good fit to all the 0(2) dete obeerved in 80 beaker fillinge during the clean gleee eubetrate etudy end aleo for 35 beaker fillinge during the neon eubetrate etudy. Ihe curvee in Figure 7 can be interpreted ae giving the I-V characterietic of the film. Alternatively, Equation 54 cen be in- verted to give the dependence of the frictional preaeure on a or v.: P. " pg: - C WEI/Ba). (55) where C in a conetant equel to pg exp(A/B). The frictional preeeure will have the eene functional form ueing v. instead of a. mu 1e obtained by eubetituting for o in Equetion 55 ueing the relation: 0 - (p./p)v.d. (56) If 1/0 ie plotted vereue inz(cn) we get a etraight line of elope -B and intercept A et inr(cn) - 0. me parametere A end B were obtained by fitting the linearized date with e leaet equered atraight line in the computer program Sigma-z. Valuee of A end B are given in the firet line of the computer output euch ea in the output ehown in 'reble A3. Graphe were drawn after each run a vereua r end l/o vereue inz(cm) along with the fitted curvee to vieuelly check the parametere end goodneee of the fit. 'Ihe data pointe on theee graphe were plot- ted end the fitted curvee were drorn from the computer output ueing a Hewlett-Packard 9100A calculator end 9125A Calculetor Plotter. Valuea of the parametere A end B were obtained at two temere- turee during the clean gleee eubetrate etudy. At 1.65 K. the values of theee parametere ere: A - (1.80 * 0.09) x 104 nee/cm2 end 50 n - (8.9 . 1.9) x 1.02 eec/cn2 with h - 7 on. At T - 1.28 x, the valuee of theee paranetere are: A - (1.35 t 0.02) x 104 eec/cm2 2 eec/cm:z with h - 7 cm. The errore given and B - (8.3 t 1.4) x 10 here are the root mean equate deviatione of the neaeured A'e and B'e for h in the neighborhood of 7 on. Then errore will be die- cueaed in more detail later in thie eection. Theee and other pere- eetere neeeured during the clean glaee and neon eubetrate etudiee are preeented in tabular form in Appendix B. For 2. < 0.1 cm, the acceleration ten: in the equation of motion beconee inortant no that the frictional preeeure cannot be meeaured directly. If thie tern wee not included and the fitted curve for large 1: wee extended to enell r, we would have a + 0 ea 2 + 0. However, the acceleration tam caueee the inner level to overehoot the outer level with a 1‘ 0 at r - 0. AI wee diecueeed in Section 11-0, the inner level then oecillatee about the outer level. he critical velocity. or critical trenefer rate, ie defined to occur at the velocity where frictional diaeipation firet occure. In thin experiment the velocity could not be increaeed from zero until thie velocity ie reached, no one would like to do. Inetead the critical velocity wee approached from above with the reeulting dynamical limitation. To eetinate the zero preeeure head or criti- cal tranafer rate. a linear extrapolation is made from the fitted curve for large 2. lhe daehed linee in Figure 7 are euch lineer extrapolatione. Extrepolating from x - 0.1 on. using do/dz evalu- ated at x - 0.1 cm. we find the following eatimete of the zero pree- eure heed tranefer rate: ac - a(0.1 cu) - B[o(0.l ”>12. (57) 51 Here ac is taken to be the critical transfer rate of the beaker filling at that h. The parameter B is observed to be independent of h for h > 4 cm. For h < 4 cm, B decreases so that o is nearly independent of 2. as can be seen on the top curve of Figure 7. This decrease of B at small h may not be a property of the helium but may be a result of a rapidly changing film thickness when the level difference, 2, is comparable to the height. h. to the beaker rim. However for h=7 cm, the thickness changes sufficiently slowly with h that pure pressure head dependence can be seen. ‘Most of the height dependence of a. which is the subject of the next section, is contained in the parameter A. Abrupt changes in the transfer rate from one constant value to another have been reported by Harris-Lowe et a1.55 and Allen and Armitagef’8 In this thesis,behavior of the transfer rate has also been seen which is suggestive of these abrupt changes. As the level difference decays, it appears that the date shown in Figures 6 and 7 have a stepped structure as reflected in the departure of the data from the fitted curves. These deviations of the 0(2) data from the fitted curves are larger than would be expected from the measurement errors of the individual 0(a) points. As was calculated in Section III-D, the average random error in one 0(2) datum point is *O.6Z. The average rms deviation of the 0(2) points from.the fitted curves is. however, found to be 1.22. This is twice the deviation that is expected from the measurements. However. more quantitative information about these steps cannot be given because of the low signel—to-noise ratio . 52 The errors given earlier in this section with the values of the parameters A and B are the rms deviations of the measured A's and B's from their average value for data in the neighborhood of h - 7 cm. These random errors are primarily a result of the devia- tions on the 0(2) curves. However, the actual errors in the observed A's and B's are slightly greater than would be expected from the 2(t) data because of additional run-to-run deviations arising from differing substrate conditions. The magnitude of o is primarily determined by the parameter A. The approximately 42 error in A in this experiment is lower than the 101 run-to-run background devia- tion in a which was reported by Smith and Boorse. 1'7 Thou 0(2) results are in qualitative agreement with Atkins'su' original results and with recent results of Martin and Mendelssohn.S6 157 for flow Similar results were also obtained by Keller and flame through a narrow slit. The functional form of the pressure head dependence agreeswith Notarys's28 observations of superfluid flowing through narrow pores and with his extension of the hanger—slasher26 thermal activation theory. However. when this theory is extended to the experiments of this thesis on He film transfer, there are several disagreements with the dependence on other quantities. The most serious disagreement of these experiments with the Langer-Fisher26 theory is in the temperature dependence of the para- meters A and B. An extension of the Lenger— Fisher theory gives A(T), B(T) c T/(o./p)2. The temperature dependences of the experi- mental parameters are: A(l.65 K)IA(1.28 K) - 1.33 and B(l.65 K) /B(l.28 K) . 1.1, while the Lsnger-Fisher theory predicts 53 a ratio 1.84 for both A and B. This gives, for the narrow tempera- ture range investigated, a nearly temperature independent B and A(T) c 9/990 Hence, within experimental uncertainties we have the usually accepted temperature dependence of 06:1-7 ac ¢ pulp. These observed temperature dependences of A and B give an in- creasing do/dz with decreasing temperature in qualitative agreement with the results of Martin and Mendelssohn.56 As suggested by Martin and Mendelssohn, the increased curvature at lower temperatures may explain the anomalously increasing o's with decreasing temperature for T' < l X that were reported by some experimenters.1m7 The functional form of the dependence of the frictional pressure on superfluid velocity observed in these experiments, and also by Notarys, 28may be true for lower velocities, and pressures, than the limited range investigated in these experiments. If this is true for very small velocities, this would mean that there is no "critical" velocity. That is, for every superfluid velocity, no matter how small, there would be a sull frictional pressure. For dynamical reasons, discussed earlier, this could not be tested in this thesis. Although the hanger-Fisher26 thermal activation theory does not appear to apply to this experiment. it may be valid for tempera- tures near Tl’ As was discussed in Section II-C, there may be two competing processes for nucleating vortices. It is possible that a mechanism such as the vortex mill model proposed by Glaberson and Donnelly23 may dominate at low temperatures while thermal acti- vation dominates at higher temerstures. In this case, the transi- tion temperature from one region to the other would depend on the number of pieces of vortex lines present in the liquid helium. This could explain why Notarys28 observed the thermal activation 54 temperature dependence for temperatures down to l I while it was not observed in these experiments. B. Height Dependence The critical velocity occurs at the region near the rim of the beaker where the film has its minimum cross sectional area and the velocity is a maximum.58 Using the dependence of the film thickness on height, the dependence of the critical velocity on film thickness can be determined from measurements of the dependence of a on the height, h, of the beaker rim above the reservoir. To do this, we seems that the thickness, d, of the film at the beaker rim is given by d - k/h1/3, (58) as was calculated in Section II—B for liquid helium adsorbed on a semi-infinite wall. The critical velocity is expected to have the empirical depend- c d1“, as was proposed by Van Alphen at al..21 Using Equation 58, this gives cc 8 5-1/4. However there ence on film thickness, vc is a question concerning which height to use in this relation because the distance to the beaker rim from the bulk liquid is different for the inner and outer levels. Atkins“ originally suggested that the proper height to use was the height to the beaker rim from the source liquid. This is the distance, h which is shown in Figure 2, from the beaker rim to the outer level for this experiment. This height was used as a first approximation for the preliminary data analysis. The dependence of the critical, or zero pressure heed, trans- fer rate, oc,on h is determined from a log-log plot of ac versus h. 55 Figure 8 shows a log-log plot of ac versus h for the clean glass substrate data at 1.65 K. For h > 2 cm, the data are described by a straight line. A least squares fit to the data for h in the 0.19. 59 For h < 2 cm, the trans- range 2 cm to 8 cm gives ac c h- fer rate is nearly independent of h as well as 2. The observations of Figure 8 indicate that the film thickness at the beaker rim is not exactly determined by the height, h, from the outside,or source, level and that further corrections mat be made. Another possible distance that could be used to determine the film thickness is the instantaneous distance to the inner level from.the beaker rim. In this case, the film thickness would increase with decreasing r as the beaker is filling. As a result, when h is comparable to z, a should increase with decreasing s as the beaker fills. As can be seen on the top curve of 0(a) in Figure 7, this is not observed. The data suggest a third possibility. It appears that the data in Figure 8 are displaced by a constant factor in h. This leads to the plausible assumption that the proper distance to use for determining the film thickness is a distance: ‘1 - h + h. > be (59) corr This correction appears to be a result of the nonzero initial level difference, 2 , as suggested by the departure of the data from init a straight line in Figure 8 at h - 2 cm, £.s., when a is com- init parable to h. This might be similar to metastable behavior that has been seen for beakers that were filled by submersion, i.e., a metastable state of thickness,60 or vorticity, is formed appropri- ate to the distance to the beaker rim from.the initial inner level. 56 .awu uexeen on» on an ensues: use»: .M no; I H 53 oo .39.. wow-new» Heouuauo one «o no? uoanmoa .m super.— ..5: 5E 2 22.: 0. wk w m ¢ m N _ 0.0.10 md NO ._ _ n _ _ l _ I: _ _ _ _ e . 1 n. D l u s - w. x8...» new L m. .... .w - ._ m. TI 0 10 g o 1 O I 1 we. I . I V I 9 I L AW .l o a PhD I 1 M. 1: _ _ _ _ _ _ _. 2m 57 Other metastable behavior of the film.is also suggested by the abrupt changes in the transfer rate, which were mentioned in the previous section. The above suggestion wee tested by performing two experimental runs with a ”init of 0.3 cm instead of the usual 1 1/2 cm to 2 cm. If the correction is due to 'init’ we would expect that with a smaller zinit the data‘would both depart from a straight line at a smaller h and also have a larger magnitude slope in the straight line region. With zinit - 0.3 cm, the departure of the data from a straight line on the log-log plot appeared to occur at 0.6 cm instead of 2 cm as in the original data. However, the slope of the straight line wee —0.20, essentially the same as the slope for a larger 2 There- init' fore this experimental test was inconclusive and, at worst, may not support the assumption that the correction term h', was equal to zinit' The following considerations of the quantum mechanical phase, ¢. make the concept that the thickness at the inside rim of the beaker is determined by the distance to the outside beaker level, plus a possible constant displacement, more palatable.31 As was discussed in Section IIqA, the phase acts as a potential for flow so that on the outside surface of the beaker, where vi is subcriti- cal, we have a gradient in ¢,but ¢ is well defined everywhere in this region. This may again be true along the lower inside surface of the beaker. The dissipative region, where the critical velocity occurs, is near the inside rim of the beaker. In the dissipative region 0 is not well defined but instead slips at the Josephson frequency. Hence we have a well defined quantum.mechenical phase 58 up the outside surface of the beaker and over the rim up to the dis- sipative region. This could cause the thickness in the dissipative region to be coupled to the height from the rim to the outside level. Ewever, this does not explain the origin of the displacement, h', unless it is possibly due to the finite thickness of the beaker rim. Nevertheless, if we make the correction, hcorr - h + 1 1/2 cm, we not only get a straight line on the log-log plot but also agree with the empirical relation: ac ¢ hi”. This is shown in Figure 9. A least squares fit to the corrected 1.65 K data gives: arr -5 cc (1. 65 z) - (8. 3 : 0.2)hco1o cm3/sec-cm. (60) The error given here for the magnitude of (1c is the m scatter of the data from the fitted curve. The error of the exponent was estimated by taking the difference in the slopes of “o lines on the log-log plot whose end points differed by twice the rms scatter. Although we have less data at 1.28 X, if these data are treated in the same way we get: -5 a(1.28 x) - (10):? 1’ ‘ cm3/sec-cm. (61) corr‘ :10 These two results are now combined by removing the temperature de- pendence of a : -1/4 -5 a (h, 1:) - (10) (9 Mb" :10 cm3/sec-cm. (52) corrx Using the dependence of the film thickness on height given in Equation 58, we get the functional dependence of v c on d: vc c d-IM. This agrees with the empirical relation of Van Alphen at al..21 This empirical relation was proposed from the examination of several different experiments spanning several decades of chan- nel width of which data the helium film critical velocity formed 59 “N00 .5 a: a + a u e .233 332:8 3:3 .e no.“ .. a u not. o .3! nauseous 133.8 one no some nonsmom 2.. our»: 2.3 :3; 2.2.: 882.8 m e m a _ l IIIIIITII _ — am e. _ 1 _ _ o I u) l -D ”03 JO;8I.IDJJ_ v2.3.7... 1 / ‘0 (W0 -338/:u10 9-0” (0 N 9 .C U \ ,/ I. e e lLLllllllLl [s 60 only one point. In the present experiment, this empirical relation was verified over the narrow range of thicknesses covered by the film. This verifies that the proposed dependence is valid for this narrow range. This indicates that the mechanism responsible for the critical velocity in the film is probably the same as for the larger channels. The dependence, vcd c ind, predicted by many theories that use Feynman- Onsager vortices is not observed.1"7 This does not necessarily disprove these theories but may be a result of a lack of knowledge of how the vortices interact with the channel walls and how they are nucleated. This functional form of the dependence of a on h agrees with the experimental results of Allen and Armitagel'8 for the film. Their measurements were mde primarily for beaker emptying experi- ments and were reported at about the same time as these these experi- ments were begun. The magnitude of ac with h - 2 cm agrees with 1—7 many other experiments although it is larger than the results of Allen and Armitage."8 Van Alphen at al. predict l/ 3 -6 ve 8 ld-J’M (c.g.s.).21 If we use the value d - 3.0h- x 10 cm 16 -l/4 «as...» for the film thickness, Equation 62 gives vc - 1.6d c. Substrate Dependence Results are reported and discussed in this section of measure- ments of the transfer rate on a neon substrate. As was discussed in Section III-B, an effective neon substrate was prepared by coating a clean glass beaker with neon taking care to provide smooth and uniform surface conditions. Using this method, a decrease in a 61 was seen going from a clean glass substrate to an effective neon substrate in the same apparatus. Using the results of the previous two sections, the dependen- ces of a on h and 2 could be separated and the substrate dependence isolated. During an experimental run with a neon substrate, meas- urements were made of 2(t) at three h's: typically at h - 5 cm, 7 cm and 9 cm. The 0(a) curves were then determined and the a de- pendence of the data removed by using the zero pressure head trans- fer rate, °c° The dependence of a c on b was then removed from ac by normalizing the data to h II 7 cm using the dependence : °c ‘ -1/4. During the neon substrate study, three runs were made using only the clean glass beaker without a neon coating. The first clean glass run was made after the final assembly of the apparatus for comparison with the results of the previous sections. TVo other clean glass runs were made with the neon substrate apparatus dur- ing the course of the neon substrate study as a test of the clean- liness of the beaker. The transfer rates obtained for these runs agree with the previous clean glass substrate results demonstrating that the beaker was initially clean, and remained clean during the course of the substrate study. All of the data during the neon substrate study were taken at a temperature of 1.65 K, as were most of the clean glass substrate data of the previous sections. Transfer rates for a neon substrate were determined as a function of the thickness of the neon coating, 5. Figure 10 shows ac with h - 7 cm and T - 1.65 K as a function of E. The three points at the left of this figure are the clean 62 5-4 "V I 1 I I A 5.2 - ° — s ‘3 § 5.0 - - "E O 0 '9 4.8 - 9 b0 0 4.6 o -— E a; 4.4 — “.3 S s: 4.2 ° e 40 4 I l l J 0 IO :02 I03 . :04 lo5 Ne Thickness of (A) Figure 10: Semi-log plot of the critical transfer rate, ac with h-7cm and T - 1.65 K, versus thickness of He coating, 5. 63 glass substrate transfer rates obtained with this apparatus. The 0(5) results shown on Figure 10 are more meaningful than seems upon initial inspection. For a neon coating of about 1000 A, it was possible to see an on-off effect; 11.0., the substrate could be changed relatively reproducibly from clean glass to neon at will. Because of the logarithmic neon thickness scale, the larger than expected transfer rates at very high and very low coverages mks the 0(5) data look worse than they really are. Although some of the transfer rates are larger than expected, all of the neon sub- strate transfer rates for E > 500 A are, however, less than the clean glass transfer rates. The lowest transfer rates were seen over the decade of neon thickness from 500 A to 5000 A. Four experimental runs having a neon coating in this thickness range gave a critical transfer rate of oc(h-7 cm) - 4.1 x 10'5 cm3/sec-cm. Two other runs at 1000 A 5 thickness neon coating gave ac - 4.14 x 10. cm3/sec-cm. These larger transfer rates are most likely a result of accidentally rough neon coatings. For comparison, the average clean glass transfer rate 5 in this apparatus was: ac - 4.86 x 10- cm3/sec-cm. 0 At 100 A thickness neon coating, we expect to have a helium film thickness equal to the average of the film thicknesses on clean glass and pure neon substrates. Two measurements were made of ac at this neon coverage. One of these experiments gave 5 cm3 lsec-cm,approximately as expected. The other Cc - As’ 3 10 experiment, probably due to rough substrate conditions, gave the larger transfer rate: ac - 5.2 x 10'"5 cm3/sec-cm. 0 At 10 000 A thickness neon coverage, larger transfer rates 64 than the minimum are again seen. From the calculated helium film thickness as a function of 5, this is not expected. Two measure- ments were‘made of ac at this coverage. These transfer rates had 5 cmglsec-cm. .At this large thick- an average value of ac - 4.7 x 10- ness. it may not be possible to get a smooth neon coating using the method described in Section III-B. Because of the long annealing times used in this procedure to smcoth the neon coating, it may be that a granular layer is formed as a result of crystallization of the thick neon coating. The functional form of the expected dependence of ac on film thickness can be calculated using the results of the previous sec- 1/4. 3/4 tions: 0 ¢ vcd and vc c d. These results give oc c d . c Using this dependence, the expected dependence of ac on the thick- nose of the neon costing, E, can be determined from the plot of the helium film thickness versus E shown in Figure 1. This expected functional dependence is shown as the solid curve on Figure 10. This curve has been normalized to go through the experimental clean glass substrate points and the lowest 1000 A thickness neon coating 5 points‘which have average transfer rates of 4.86 x 10' cm3/sec-cm and 4.1 x 10-5 cm3/sec-cm.respectively. The ratio of the thickness of the helium film on the effective neon substrate to the thickness of the helium film on the clean glass substrate is new calculated from the experimental results. using the lowest neon substrate transfer rate and the clean glass substrate transfer rate, given above. we get )4/3 dN's/dglaes - (cue/aglass - 0'79' (63) 65 1/3 x -6 Assuming the film thickness, d 10 cm, for a clean glass substratef3ths thickness of the helium film on a neon 81". - 3e0h substrate can be calculated ming Equation 63. Solving this equa- tion, '0 86': d“. - 2.4h'1/ 3 x 10'6 cm. This experimental value of the film thickness agrees remarkably well with the result: d“. - 2.3h":l'/3 x 10.6 previously ‘calculated in Section II-B. V. SWY AM) COICLUSIONS Measurements of helium film transfer rates for filling clean glass and neon coated beakers have been reported in the previous sections. The apparatus used in this study was designed to provide a clean environment for the beaker, to avoid enhanced transfer rates and unnecessary background variations. With this apparatus, quan- titative measurements could be made of the slight dependences of the transfer rate, a. on pressure head, film height, and substrate anteriel. The pressure head dependence of a was measured over a change in head, a, from 2 cm to 0.1 cm. The data in this range are described by the functional dependence: 0(2) " l/[A - Bin(s)]. (64) This gives a dependence on superfluid velocity, v.,of the frictional presets-91", opposing the flow of P. c exp[f(h,T)/v.]. These results agree with previous, qualitative measIn-smsnts by other experimenters of the pressure head dependence for both the film and narrow channels. The dependence of the critical, or sero pressure head, trans- fer rate, ac, on the height, h, of the beaker rim above the outer helium level was measured for h ranging from a few millimeters to '10 cm. After applying corrections, which were discussed in Section IV-B, these data are described by: acmcorr) - Golgi: ’ (65) 66 67 where Go is a content for a given temperature and hoorr is the -1/4 a corrected height. This results in the dependence, vc c d of the critical velocity, vc, on channel width, d. This dependence (‘8 for the 21 agrees with recent measurements by Allen and Armitags film and the empirical relation proposed by Van Alphen at al.. A 16: decrease was seen in the transfer rate going from a clean glass substrate to an effective neon substrate in the same appara- tus. The helium film thickness on a neon substrate, derived from this change in o, is in reasonable agreement with the thicknesses calculated for rare-gas solids in Section II-B. Although the physical mechanism for the critical velocity is believed to be vortex production, basic theories using this mecha- nism are not sufficiently well developed to predict these measured dependences on pressure head and film height. We hope these results can be used as a guide in the solution of these problems. LIST 0? REFERENCES 1. 2. 3. 4. 5. 7. 9. 10. 12. 13. 14. 15. l6. 17. LIST OF REFEREMES K. R. Atkins, Liquid Helium (Cambridge University Press, Cambridge, England, 1959). R. J. Donnelly, Empsrtmntal Supsrfluidity (University of Chicago Press, Chicago, 1967). J. c. Daunt and n. 8. Smith, Rev. Mod. mm. .2_6_, 172 (1954). Le Cs JCChOn “d L. Gs 6:1... Phil. HIS. Suppl. L 635 (1958)e w. E. Keller, Haitian-3 and Helium-4 (Plenum Press, New York, 1969). C. T. Lane, Supsrfluid Physics (licGrawbHill, New York, 1962). J. Wilks, Liquid and Solid Helium (Clarendon Press, Oxford, w. 1!. Reason, Helium (Elssvier, Amsterdam, 1942). L. 11.2.. Nature £4, 913 (1938); Compt. Rand. 39;, 1035. 1186 (1938), J. Phys. Radium 1 165,350 (1941), Phys. Rev. _7__,2 838 (1947), and Phys. Rev. _7__,5 885 (1949). F.‘ London, Supsrfluids, Vol. II (Wiley, New York, 1954). L. Landau, Phys. Rev.6 60 357 (1941); J. Phys. (0.8. 8.1!.) 5, 71 (1941); J. Phys. (6318.11,) 8 1 (1944); J. Phys. (ms. s .R.) 11 91 (1947); and Phys. Rev. 7_5_, 884 (1949). B. V. Rollin and F. Simon, Physica6 219 (1939). J. G. Daunt and K. Mendelssohn, Proc. Roy. Soc. (London) _A_l__70, 423 (1939): Al__?_,0 439 (1939). X. R. Atkins, Proc. Roy. Soc. (London) A203, 240 (1950). no no Esel'son and Be as hum. me “no T'Ore F130 22. 552 (1952). P. w. 1". Gribbon and 1.. C. Jackson, Can. J. Phys. Ll 1047 (1963). B. Smith and E. A. horse, Phys. Rev. _9__,8 328 (1955); 99 346 (1955): _9_,9 358 (1955), 9__9_, 367 (1955). 68 18. 19. 20. 21. 22. 23. 24. 25. 26. 27. 28. 29. 30. 31. 32. 33. 34. 35. 69 L. Onsager, Nuovo Cimento Suppl. 2., 2249 (1949). R. P. Feynman, in Progress in Lad Tempsmturs Physics, Vol.I, edited by C. J. Gorter (North Holland, Auterdam, 1955), p. 17. Ref. 5, Chapter 8. w. H. Van Alphen, G. J. Van Eaesteren, R. DeBruyn Ouboter, and K. W. Taconis, Phys. Letters 22. 474 (1966). w. F. Vines, in Progress in Low WW Physics, Vol.III, edited by C. J. Gorter (North Holland, Auterdam, 1955), p. 1; in Diquid Helium, edited by G. Careri (Academic Press, New York, 1963), p. 336; and in When Fluids, edited by D. F. Brewer (North Holland, Amsterdam, 1966), p. 74. W. I. Glaberson and R. J. Donnelly, Phys. Rev. 141, 208 (1966). V. P. Peshkov, in Proceedings of the VII International Conference on Low Wmturs Physics, edited by G. M. Graham and A. C. H. Ballett (University of Toronto Press, Toronto, 1961), p. 555; in Progress in Lao Rnpsmturc Physics, "Vol. IV, edited by C. J. Gorter (North Holland, Amsterdam, 1966), p. 1. P. P. Craig, Phys. Letters 3;, 385 (1966). J. S. Langer and M. E. Fisher, Phys. Rev. Letters $2, 560 (1967); M. E. Fisher, in Proceedings of the International Conference on Fluctuations in Superconductors, edited by W. S. Cores and F. Chilton (Stanford Research Institute, Henlo Park, Calif., 1968). De 357s J. R. Clow and J. D. Reppy, Phys. Rev. Letters 3:2» 91 (1967). He As Hoary.. my.e We “cut. ‘12-. 12"0 (1969)e S. T. Beliaev, Soviet Phys.—UETP _7_, 289 (1969). P. W. Anderson, Rev. Mod. Phys. 38 298 (1966): also in Quantwn Fluids, edited by D. F. Brewer (North Holland, Amsterdam, 1966), Po 146s R. J. Donnelly, Phys. Rev. letters 3.! 939 (1965). Le Is Schiff, PhYIs ”Va 22.. 839 (1941)s T. L. 8111, Advances in Catalysts and Related Subjects, Vol. IV (Academic Press, New York, 1952), p. 211. c. 1.. Pollack, 1m. Mod. Phys. _2_._6_.: 748 (1964). Go Ks Horton, Aha Jo PhYI.‘2§, 93 (1968)e 36. 37. 39. 40. 41. 42. 43. 41.. 45. 46. 47. 48. 49. 50. 51. 52. 53. 54. 55. 56. 57. 58. 70 J. W. Rowlinson and J. R. Tovnly, Trans. Faraday Soc. 32, 20 (1953). I. E. Dzyaloshinskii, E. M. Lifshitr, and L. P. Pitaevskii, v. M. Kentorovich, Soviet Phys.-~JETP ;, 770 (1956). W. Es “llCr. Phfls h's “tar. &. 569 (1970)e De Le GOOdItQm .nd Ps Gs SCfm. W's RCVs uttCTU 2.8-. 1402 (1970). R. Bowers, Phil. Mag. 33. 465 (1953): 8.4.0 485 (1953). K. R. Atkins, Proc. Roy. Soc. (London) 203A, 119 (1950). A. C. Han and L. C. Jackson, Proc. Roy. Soc. (London) A240, 243 (1957). x. n. Atkins, uncut. 161, 925 (1948). We J4 d. um and Ge Js m d.“ B.r8. ”Vs "Ode Phns E. 609 (1949). Rs Bower. and Ks W.1I.&n. hoes Phy.s 803s (London) 63A. 1318 (1950). N. G. 11ch and J. C. Eisenstein, Phys. Rev. 32, 1326 (1955). J. P. Allen and J. G. M. Arnitage, Phys. Letters 22, 121 (1966). c. s. Pious, Phys. Rev. 23, 1459 (1954). J. N. Schubb and I. E. Pollard, Le Vide 22, 64 (1969). J. G. Dash, Phys. Rev. .A_1_, 7 (1970). G. R. White, Experimental Techniques in Low-Tauperature Physics (Clarendon Press, Oxford, England, 1959), p. 115. C. Hack, Essentials of Statistics for Scientists and Technologists (P1en1nn Press, New York, 1967). C. J. Duthler and G. L. Po11sck, Phys. Letters 314, 390 (1970). R. 1". Harris-Love, C. 1". Hate, K. L. McCloud, and J. C. Daunt, PhYIs Letters 2.2, 126 (1966)s De Jo “tin Q“ Ks mud“..°hn. fins L.tt.r. 30A, 107 (1969)s W. Es nHCr ‘nd Es Fe W1. “8”,.“ A 221 (1966). w. 8. Keller and E. P. flannel, Phys.;Rev. Letters _1_7_, 998 (1966). 59. 60. 71 C. J. Duthler and G. L. Pollack, Bull. An. Phys. Sec. 3.3, 912 (1968):.13, 96 (1969). J. P. Allen and C. C. Mathsson, Proc. Roy. Soc. (London) A290, 1 (1966). APPENDICES 72 Z.Nu~ mo 00 A—vztoru.~.Nm agvzpso.ocu.~.mh a o.cu.~.<»4moa mlmuz uwm Zn wt—h 024 .N .thwo whm Nshoohmod u .b.N .U—h410430 Ibu) whzwt¢wm .th .ha Am bum Anvrloxuaqu .uvths00oou.~.mh Adi—.IIA—vzn.~.