-1——, :~':‘~1‘"'11111M,W 11 '5!” ”if“ , “'yf'" ”Youhlr, 9.11 W ' ,U . “my“;fi', V‘ “16“. r‘b’hzv J k1 1,3,; 111171,” I" :1}. 11195113.?” 11' :1 ' ""vui‘“ 11. 1.3.1111 ‘ f1.”1'.1.'!'1 “1‘11 11:11 11131131; 1 13.111.11.11, $11171 11111111: ,1‘ {£11 117:? “111.» f2}, ”13.3111” ' .. 11‘11‘31111',1""!“"9 1I'1'11H1.’11.~'I’11111.1 11111211111111 "7'11'1 11 11, TEE/f!“ .o(.q ‘1‘}: “J1 "1’5‘.|:I1:'- ‘N’T‘ I),”'{l "'1‘ .\ 1' 3111'." 1.. 11-111." ‘1'1~I.[1r1"'zr111'{'1‘J371-.1,21 '1- i 111911-11”! - o 1-1,] ‘ ‘ €212: “’11 x , 9:91;? '1 “fig! .111 mfim¢MW 1 1M 1 1" . ‘ 1 1111,. 1111I 11.1.11“: [11" 11.1 [11,7 151%; 1111]“ Eggh‘r 3‘51 5112'1'1—11‘1‘11?’ “300' W 161}; ,‘,.‘1‘1.,,'1.1',‘..1,111f:,1.111,111'l1:~1,,,1 111 1111111111111. 11511,,W,Em“, '1'111'131 111,3,11131‘7IU1P1 11:}. 1,21% - $.11 . , 1,? ., -1 \ 1 , |1 ' " 1 ' L1: 111"" 11.": 1"‘l‘.""1 1"‘111 111111'11’111151"?’11' {“1611}!111111131'117112111’5117' '11P: ‘1 ”1,111. $1.11, ,"‘1,1,'Ul$1’11111}1 (1,11! ’3'. ,1‘11.I 1,!” . - 11:1! . .1 ”1,1,,“ “'1‘ 1"1'1’111'L1Jqui'!” . 10%,)?! I . 1‘ _. ‘ "iwwwgq! '1‘) .1)’ I fit 1 1.11.131“ 111’” 1“,“, 11111, J}? . , ’11, I, 1 1;? 3‘ 1’1" fij‘rfiflm . 1 .1 11-11. . 1 -~ 1. .1 =11 '11 11H 1 11:11.,11'15 "‘ I: 11. 1,1116}; . , .1 ,1 .11 .1 5,, V 1 ., . “1,1,, 11“ "1"..0‘11‘1“%1fa 1.! *1 1 1 . ', I I . ‘ 111, ..11 "3.1111 111111,? . ' 15 I ' '1 1 1111 11.11111" ”‘33? 1% I ‘1'):1'1-711111'25 “5‘”: $11., 1'”, ‘9 | h 1’]; 11'1'$J_1111$#'£i,11 1 'l‘ .. 111 ~~~~~ 11;!" 11,711 1.. (1,325,; {lifiiffi}. 1: .1!" ' :1. . 1‘1'.’ 1 '1' ‘ 11": 1111’ 111111111 1 1 '1 ' i ‘11! 1211' 11.11.: 1,,1 . 1 .11 ', ', , .,,-.,.1 1' 111111 . 1.1111111.111111111171511 "-1.1 1'. ' ‘ ‘ “ ‘ ‘-' ‘ :l', .‘ Pitt 1:": £..‘ ;.'.. , ,’.. . ’ :1" | . .. ‘ , 131111111 ,1'11.,1,,,,'1 1111,121'1111‘1” 1.,.,1...,..,1 .111 .1 ,. ,, 1 ffilflw 1.1, ",1;,',J,'1;]'1‘1,1,’ 11,511,115 .' '. MH- .1311] ‘11} ["1113 11111311” 1‘18}: .1, £11 «11 1 1 1. ' ‘1 111355;; {1" ‘ ' ,1 . 11,1)“, 1,.1r1{.'11111:1 1 ,. 2,1 1. 7.1.1111, \,,(;YJ,fp:.'1::‘;£ 111117.]?1-31! r. 1'; :: . '21.:‘1', ; ’ 1. 1 ,Ll‘l ’13:, "I,“ Fig; 11):! :1: fi:;|.1:,:1§:t;f.~rr:. . :5, ' 1.1 . h 1") I'r‘;:}‘13"11‘"l" .‘.g -. "1‘ .1' 1311119111.:1’ 11 ' ,1 131,, "'jgrg- -~ 1. 1,. . ,(f' .- 3-4— ' "'5: ‘ ~ —‘ ‘ 5%.: 1 ““2, 2-: .5 .4, l 1 n . 11,1515 {-1.1 4,31,. ‘11 1.1%.» 1, 3 41‘ 81 1131,11}: ‘ ,,,J, “111.,“ ’151111313’3‘ by}, 1111131"! 1,41,11,111‘2111‘1111“. ’1 1W1; 1511.111 1111 w{1—31521131111,3 1111 ”1,1,1 111»; 131’ .. 311 . .11., 1:. i“ ”1,, [1,1113% .11 1111111311‘11111"11Ji x, “'1' 1111‘“ '11 (21.1.1 1 1'11 ".1‘ " ';1~13fi! 31?. 1’11".)1 #INW'JM“! #151 111 111 11.,{.?.1’1",,, 1137‘"? :11, ,I’ V‘- r‘ 1 . 1 ~11 ,h’Iufi’IL g pt ’1‘“ x {.fi’: ., .3112; I; Sikh‘ : 9* ‘Q 1111““ 1111' - ~ 1 1111’. * I . :3- 1 o L 111111,,1 '. 1111‘ '1 . 1" ..... 1111111911111111111111111111111111111 00669 4958 LIERARY Michigan State University This is to certify that the dissertation entitled PRACTICAL CONSIDERATIONS FOR THE USE OF IMMOBILIZED ENZYMES IN FLW INJECTION ANALYSIS presented by Cheryl Lynn Mattson Stults has been accepted towards fulfillment of the requirements for 31.12. degree mm /£Z/éwc Mafor professor Date—WBL— MS U i: an Affirmative Action/Equal Opportunity Institution 0-12771 MSU RETURNING MATERIALS: Place in book drop to LJBRARJES remove this checkout from —:-—. your record. FINES will be charged if book is returned after the date stamped below. «rt-fr PRACTICAL CONSIDERATIONS FOR THE USE OF IMMODILIZBD ENZYMES IN FLOW INJECTION ANALYSIS By Cheryl Lynn Mattaon Stults A DISSERTATION Submitted to Michigan State University in partial fulfillment of the requirements for the degree of DOCTOR OF PHILOSOPHY Department of Chemistry 1987 Copyright by CHERYL LYNN MATTSON STULTS 1987 ABSTRACT PRACTICAL CONSIDERATIONS FOR THE USE OF MBILIZED ENZYMES ' IN FLOW INJECTION ANALYSIS By Cheryl Lynn Mattson Stults Numerous practical issues concerning the use of inobilized enzymes in flow injection analysis (FIA) have been investigated. Inobilized glucose oxidase (GO) was used predominantly. The enzyme reaction was coupled with a modified Trinder reaction in an FIA system. The basic components of the apparatus were: a peristaltic pump, a pneumatically actuated sample injection valve, single head string reactors (SBSRs), and a miniaturized filter calorimeter. Data acquisition, sample injection and pump speed were controlled by a microcomputer . The procedure for inobilizing GO on nonporous glass beads was studied and improved. Simplex optimization was used to find an optimum set of operating conditions for the FIA system containing a G0 SBSR. Seven variables were specified: carrier pH, reagent pH, temperature, flow rate, aminoantipyrine concentration, dichlorophenolsulfonate concentration, and peroxidase concentration. The results were similar to those obtained by univariate experiments. Several extended applications of the GO SBSR were investigated. The enzyme coated beads were used to make a microconduit SBSR. The response obtained was equivalent to that of the conventional SBSR. A Go SBSR was used to study the mutarotation of D-glucose in the presence of phosphate. The mutarotation coefficients obtained were in good agreement with the literature values. Galactose oxidase (GA) was i-obilized by a procedure similar to that for glucose oxidase. A GA SBSR was combined with a,G0 SBSR to make a parallel FIA system. The results obtained from the analysis of real samples matched known values. Two other issues that are related to the use of i-obilized enzymes were also studied. First, the effect of temperature on dispersion was examined for four classes of reactions—none, fast, medium, and slow. A dye was used to establish the effect of temperature. The reactions were: p—nitrophenol with sodium hydroxide, nickel with 4-(2-pyridy1azo)resorcinol, and the GO/Trinder reaction. The results indicate that it may be possible to calibrate FIA systems for temperature changes. Second, a partially automated system on which flow reversal/merging zones experiments may be carried out was designed and characterized. The results illustrated the advantage of using a second detector and two dimensional mapping of reaction surfaces fram a single injection. ACKNOWLEDGEMENTS I’m glad for the opportunity to officially thank those who have contributed to this achievement in my life. First and for-0st, I am thankful for Professor Crouch’s continuous, positive encouragement and patience. Special thanks are extended to Professor Watson fer being my Second Reader, and to the other members of my Guidance Committee (Professors Dye and Nocera) for their helpful suggestions and insight. I am indebted to Dr. Adrian Wade for his friendship, gentle pushing, and essentially "taking me under his wing". Without the help of Paul Kraus (AKA: my cohort in crime) I might still be taking data on a chart recorder. A big thank-you is in order for letting me use his version of the "Bruce Bus", and for the camaraderie we’ve shared. The friendship and support of the other members of the Crouch group is also much appreciated. It was a special privelege to teach for Professor Timnick, Professor Allison, and Dr. Funkhouser. A special note of thanks to Marty Rabb and the other members of the technical support groups in the Chemistry Department for going out of their way to be helpful. I also grateful for the graphics assistance provided by Bill Draper. I owe special thanks to Dr. Peggy Dine and other teachers who have encouraged me throughout my life to pursue goals that I mistakenly thought were unattainable. I must also thank my mom and dad for always believing in me and lending a hand when things got a little crazy in this last year. Last, but certainly not least, I am deeply grateful to my husband, John, for his constant support and taking care of the details of our personal life so that I could complete my work; and to my son, Michael, for providing happy moments of diversion during my writing binges . vi TABLE OF CONTENTS CHAPTER . PAGE LIST OF TABLES ................ . .............................. ....x LIST OF FIGURES ................................................. xi WHAT’S IN THE BOOK ............................................... 1 BACKGROUND INFORMATION.... ....................................... 3 Introduction to I-obilized Enzymes. . ........................... .3 Methods of Enzyme Immobilization ........................... 4 Analytical Applications of Immobilized Enzymes ............. 5 Introduction to Continuous Flow Analysis. . . ...................... 7 Principles of Flow Injection Analysis ............ . ......... 7 Types of FIA Manifolds ..................................... 9 Measurement Tools in FIA .................................. 10 Use of I-obilized Enzymes in PIA ............................... 12 THE GLUCOSE OXIDASE SBSR. ....... .......... ...................... l4 limobilizntion Procedure far Glucose axidhse .................... 16 Cleaning Procedure ........................................ l7 Whisker Formation. ........................................ 19 Silanization .............................................. 20 Glutaraldehyde Attachment ................................. 21 Enzyme Attachment and SBSR Construction ............... . . . .21 Alternate Method of I-obiJization .............................. 22 vii CHAPTER PAGE Determination of Amount [mobilized ............................. 24 Ca talase Acti Vi ty ............................................... 27 Specificity of [mobilized Glucose Oxidase ..... . ................ 27 4 OPTIMIZATION OF THE FIA SYSTEM .................................. 31 Instrmentation.................... ....... . ..... . ..... . ......... 33 Choice of Fixed Parameters ................................ 33 Temperature Calibration .............. . ......... . . . . . ...... 39 Microcomputer Interface to FIA Syst- ..................... 39 Data Acquisition. . . . ...................................... 45 . Improvement of System Performance ............................... 45 Univariate Experiments .................................... 47 Delay Flow Experiments .................................... 58 Simplex Optimization ..... ................ ..60 5 EXTENSIONS OF THE GLUCOSE OXIDASE FIA SYSTEM ...... ..............75 Design and Tea ting of the Microcondui t Glucose Oxidase SBSR. . . . . 7 5 Ascorbic Acid Interference Study ............. . .................. 78 Kinetic Study of D-Glucose Mutarotation. . . . . . . ............... . . . 82 Determination of Mutarotation Coefficients .............. ..87 Effect of Phosphate Concentration ....................... . .88 Effect of Galactose ....................................... 91 Parallel Determination of Sugars ...... . ...... . .......... . ..... . . 92 The Galactose Oxidase SBSR ................................ 92 Inobilized Galactose Oxidase Specificity ............ . . . . . 93 Design of the Parallel FIA Syst- ...................... . . .95 Analysis of Real Samples .................................. 96 viii CHAPTER PAGE 6 THE EFFECT OF TEMPERATURE ON DISPERSION ........................ 101 The Apparatus for Cases 1 through 3 ............. _ .............. . 104 Case 1: No Reaction ........................................... 107 Case 2: Fast Reaction ......................................... 111 Case 3: Moderate Reaction ..................................... 113 Case4.‘ SlowReaction................ ......................... 119 DEVELOPMENT OF A FLOW REVERSALS/MERGING ZONES FIA SYSTEM ....... 124 me FIA Flow Reversals System .................................. 125 Calibration with Dye ........................................... 127 Extension to a Chemical System. ................................ 134 WHERE DO WE GO FROM HERE? ...................................... 136 LIST OF REFERENCES ............................................. 139 ix TABLE 10 11 LIST OF TABLES PAGE Examples of immobilized enzymes. ................................ 13 Initial immobilization procedure ................................ l8 Glucose oxidase activity toward other sugars.... ............. ...29 Coefficients of the Steinhart-Rart equation for different flow rates ........ . ............ . ......... . ........ 42 Range and precision of variables for simplex optimization ....... 65 Initial and optimal experimental conditions for the FIA determination of glucose .............. . ............. 67 Effect of ascorbic acid in experiments without Brij-35. ......... 81 Mutarotation coefficients for D-glucose at 21°C ................. 90 Galactose oxidase activity toward other sugars. ................. 94 Temperature correction equations for various flow rates........106 Temperature effect on glucose oxidase/Trinder reaction ......... 122 LIST OF FIGURES FIGURE PAGE 1 10 11 12 13 14 15 16 17 Comparison of FIA profiles obtained with an SBSR and an open tube. Signal traces are only in the region of interest (see text for further explanation)....... ............ ..11 Single bead string reactor....... ............................ ...15 Final immobilization procedure .................................. 23 Loss of glucose oxidase activity for routinely used heads (a) and stored beads (A) .................................. 26 Actual (- -) and expected (--) effect of catalase in a 20-cm SBSR (O) and a 30-cm SBSR (o). .. .. ................... 28 Glucose oxidase/Trinder reaction.............. ...... . ........ ...32 FIA system for glucose determination ............................ 34 Effect of varied glucose sample size.... ...... ........ .......... 37 Effect of flow ratio (Buffer : Carrier) on FIA profile .......... 38 T-piece containing 30-kn thermistor ............................. 40 Thermistor response curve ....................................... 41 Effect of varied flow rate ...................................... 48 Effect of AAP/DCPS ratio ............. . .......................... 49 Three dimensional surface of the FIA peak for the Trinder reaction........ ................................ 51 Effect of peroxidase concentration .............................. 53 pH dependence of GO/Trinder reaction with Universal buffer ...... 54 pH dependence of GO/Trinder reaction with phosphate (- -) and borax (—) buffers .......................... 55 xi FIGURE PAGE 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 . Temperature response of a lO-cm GO SBSR ......................... 57 Results from delay flow experiment .............................. 59 Two dimensional simplex example ................................. 61 History of simplex optimization .............................. . . .63 Progress of the simplex ......................................... 69 Effect of experimental variables on response function: (a) carrier pH; (b) reagent pR; (c) flow rate, ml min“; (d) SBSR temperature, “C; (e) DCPS concentration, i4; (f) AAP concentration, In; (g) PO concentration, mg ml‘1 ........ 71 Calibration curves for optimum operating conditions arrived at by simplex optimization (0) and by univariate experiments (0). . .73 Side (a) and top (b) views of a microconduit SBSR ............... 77 Mutarotat ion process for D-glucose. . . . . ....... . . . . .............. 84 Effect of phosphate concentration on rate of D-glucose mutarotation. Lines represent curves of best fit. Upper curves were obtained with only B—D—glucose initially present. Lower curves were obtained with only c-D-glucose initially present. Phosphate buffer concentrations were: (Q) 0.00 M, (O) 0.05 M, (A) 0.10 M, (I) 0.20 M .................. 89 FIA system for parallel determination of glucose and galactose. .97 FIA profiles for : (a) Pepsi, (b) Sunglo fruit punch, (c) diluted serum, (d) deproteinized serum ...... . ............. ..98 FIA system for temperature studies ............................. 105 Effect of temperature on dispersion without reaction. Flow rates were: (0) 0.56 ml min‘1 , (A) 1.20 ml min’1 , and (I) 1.88 ml min‘1 ........................... .. ............. 108 Effect of temperature on peak area without reaction (a), peak area with fast reaction (b), peak height without reaction (c), peak height with fast reaction (d), m without reaction (e), and mm with fast reaction (f). Flow rates were: (0) 0.56 ml min'1 , (A) 1.20 ml min"1 , and (I) 1.88 ml min“1 .......................................... 110 FIA profiles obtained for the experiment with p—nitrophenol under the following conditions: flow rate 1.20 ml min”; coil length 3.0 m; and temperature range 20-70‘C ............... 114 xii FIGURE 34 35 36 37 38 39 40 41 42 Reaction progress curve for nickel (2x10'5 M) with PAR (10" M) at pH 6.85 ................................... 116 Effect of temperature on peak area (a) and peak height (b) for medium reaction with l-m (O) and 3-m (A) coils ............. 118 Effect of temperature on peak area(a) and peak height (h) for slow reaction. Concentrations of standards were: (0) 1.65 IN and (A) 3.3 nM glucose; (0) 0.14 m and (A) 0.28 IN H202 ............................................... 120 FIA system for flow reversal/merging zones experiments ......... 126 FIA profiles for two reversals and one detector. The three injections were: (a) 81 only; (b) 82 only; and (c) 81 plus $2 ............................................. 129 Contour plot for two reversals and one detector ................ 130 FIA profiles for three reversals and two detectors. The three injections were: (a) 81 only through the first detector; (b) 81 only through the second detector; (c) 82 only through the first detector; (d) 82 only through the second detector; (e) 81 plus 82 through the first detector; (f) 81 plus 82 through the second detector ......... 9 ............ 132 Contour plot for three reversals through the first detector (0) and the second detector (A). . . . . .................. 133 Six channel parallel FIA system for simultaneous determination of glucose, galactose, maltose, lactose, sucrose, and fructose ................................. 137 xiii CHAPTERl WHAT’S IN THE BOOK The i-obilization of enzymes is no longer a novel concept. Application of these reusable reagents has gained notoriety in the fields of food processing, clinical chemistry, and analytical chemistry. Research in the area of i-obilized enzymes has focused primarily on i-obilization procedures, reactor designs, and analytical applications. The research reported in this dissertation focuses on the practical considerations that must be made in order to utilize i-obilized enzyme technoloa to its fullest. This work centered around the use of glucose oxidase (GO) i-obilized on nonporous glass beads in the form of a single head string reactor (SBSR). This enzyme is inexpensive, has good activity, and can be bonded to a variety of insoluble substrates. Primary consideration was given to the use of the SBSR in a flow injection analysis (FIA) system. In this system the GO reactor is coupled with a reaction that utilizes the hydrogen peroxide formed from the enzyme reaction to produce a colored dye. Fundamental background information is provided in Chapter 2. A review of current i-obilized enzyme methodology is provided along with a more specific orientation to the analytical applications of these reagents in FIA. A brief review of the principles of flow injection analysis is presented; the SBSR is introduced. Chapter 3 describes the characteristics of the GO SBSR. A more efficient i-obilization procedure is enumerated. The activity, useful lifetime, catalase activity, and specificity of the reactor are discussed in detail. Another matter of practical interest, optimin operating conditions, is treated in Chapter 4. Three methods by which the response of the FIA system can be increased are presented. Several extended applications of the system are given in Chapter 5. The potential for using the i-obilized enzyme beads in a microconduit is explored. A preliminary investigation of ascorbate interference is described. A novel use of the system, to study a batch reaction, is presented. Finally, the knowledge gained from the work on glucose oxidase was applied to galactose oxidase and resulted in a parallel analyzer. The design, construction, and use of the system for the simultaneous determination of two sugars in real samples are discussed. Two topics of practical interest are presented following the enzyme work. In Chapter 6 the effect of temperature on the FIA profile is exuined. Four different reaction conditions are presented. The design and characterization of an FIA system which codines the flow reversal technique with merging zones is discussed in Chapter 7.. Ideas for future projects are given in Chapter 8. The practical information obtained in the course of this research should prove beneficial to those who continue to work with i-obilized enzymes in coming years. It should be possible to extend the procedures developed here to other i-obilized enzyme syst-s. CHAPTER 2 BACKGROUND INFORMATION This chapter provides a condensed review of the topics with which familiarity is assumed in the subsequent chapters. There are two major areas of focus: i-obilized enzymes and flow injection analysis (FIA). The first section gives the reader a perspective on the general methodology employed with i-obilized enzymes in analytical chemistry. More complete information on the subject of i-obilized enzyme technology can be found in recent books (1-11) and general review articles (12-18). Specific procedures for enzyme i-obilization can be found in Reference 8. The second section deals with some of the fundamental aspects of FIA which are co-on to all of the systcs in the following chapters. Discussion of subtopics which relate only to the material in one chapter are delayed until then. The authoritative book on FIA by Ruzicka and Hansen was published in 1981 (19). A recent monograph by Valcareel and Luque de Castro (20) provides a review of current FIA technology. An excellent bibliography containing 811 references to the work accomplished in this field was published in a recent review written by the sue authors (21). A complete bibliography has also been published by Tecator which covers the period from 1974 through February 1985 (22). Introduction to I-obilized Enzymes An enzyme is a protein which acts as a biological catalyst. It differs from an ordinary catalyst in that it has a high turnover rate at ordinary tmeratures, is selective with respect to a particular reaction pathway, and may be specific for one or for only a few compounds. Enzyme solutions are useful as analytical reagents, but they have a few major disadvantages. Some enzymes are unstable in solution, most cannot be used in organic solvents, and all are very sensitive to elevated temperatures. Enzymes are expensive reagents. (hoe an enzyme is used in solution it is not easily recovered, therefore making the use of these catalysts rather uneconomical. The i-obilization of these biocatalysts has been a mjor factor in the popularity which they have achieved in the past few decades. By rendering the enzyme water-insoluble many of the disadvantages of the soluble form are eliminated and some advantages gained. The enzyme retains its activity over a broader range of pH and temperature. The enzyme is reusable in this form and so provides an economical alternative to other methods. An additional advantage of this feature is that the enzyme is easily manipulated. It has been suggested that there are fewer interferences with i-obilized enzymes (11); thus, comlex samples are relatively easy to analyze. Methods of Enzyme I-obilization The methods for enzyme i-obilization fall into four categories: adsorption, cross-linking, entrapment, and covalent binding. The initial discovery by Nelson and Griffin of enzyme activity retention upon i-obilization involved the adsorption of invertase on charcoal (23). This type of i-obilization occurs any time an enzyme is adsorbed on some type of insoluble support. The chemical conditions required for this process are very mild. However, as easily as the enzyme adsorbs to a surface it also desorbs. This is a significant problem if the i-obilized enzyme is to be used in a system where it is exposed to a continuous flow of solution. Entrapment provides a sturdier environment than adsorption. A co-on method of entrapment involves creation of a polymer in the presence of the enzyme. The enzyme becomes enclosed in a lattice. Polyacrylnide is often used to i-obilize enzymes in this way. Gels of this type are co-only used in enzyme electrode applications. An alternative type of entrapment is accomlished by encapsulation of the enzyme in a semipermeable polymeric medarane. This usually involves a tedious and time consuming i-obilization procedure. A somewhat more harsh chemical method is cross-linking. The enzyme is covalently bonded with a bi- or multifunctional reagent to create an intermolecular cross linkage. Glutaraldehyde is co-only used as a linking agent. 0f the four methods of i-obilization covalent binding of the enzyme to a water-insoluble support provides the sturdiest environment. This is the most co-only used method of i-obilization. Some type of chemically modified glass provides an ideal inert support. The methods of enzyme attachment to the glass are as varied as the nui>er of applications. Analytical Applications of I-obilized Enzymes For a complete review of the many ways in which i-obilized enzymes have been used in analytical chemistry the reader is referred to the book by Guilbault (11). There are two main areas of application. I-obilized enzymes have been coupled with amperomstric or potentiometric electrodes in a nwer of ways. The enzyme may be bound directly to the surface of the sensor, but is more co-only attached in the form of a membrane. These probes have gained great popularity in biological and clinical applications. I-obilized enzyme reactors (IMERs) constitute the other major area of application. Many types of such reactors have been developed. They take the form of packed beds, open tubes, and single strings of beads. Packed bed reactors typically contain the enzyme i-obilized on very small particles (37-74 in) which are packed into a column. They have been used in batch as well as continuous flow methods. The great advantage of these reactors is the large surface area over which the enzyme is spread. However, when used with flowing streu, these reactors may develop channels and high dispersion may result. Open tubular reactors contain the enzyme i-obilized on the inner surface of the tube (usually about 1 - i.d.). These are co-only used in air-segmented continuous flow analysis. The small surface area available per length of tubing results in very low enzyme activity. There have been various attempts to increase the surface area of the inner wall of the tube. Onuska et a1 (24) demonstrated the use of mania hydrogen difluoride to grow whiskers on the inside of glass tubes. A recent report indicated that controlled pore glass particles had been edaedded on the walls of plastic tubing (25). The single head string reactor (SBSR) is used in flow injection analysis applications. Such reactors are made by i-obilizing the enzyme on porous or nonporous beads. These are packed into a tube which has an i.d. that is 1.3—1.7 times the diameter of the bead. The dispersion in such reactors is low while the enzyme activity is high. More will be presented about this type of reactor in the following section. Introduction to Continuous Flow Analysis Continuous flow analysis is the technique by which a sample undergoes treatment as it is carried by a flowing stream through a manifold made of interconnected tubing. It can be divided into two major areas: aimegmented and non-segmented continuous flow analysis. Skeggs is accredited with the development of the former technique (26), which is most often referred to simply as continuous flow analysis (CFA). In this, air segments are introduced into the analytical stream at regular intervals so that the stream is divided into a train of identical segments. The sample is introduced onto the manifold by aspiration for a defined period of time. In each segment complete mixing takes place so that the signal obtained at the output has a rectangular shape similar to what would be expected in the ideal case of a plug-shaped sample. This technique has been used extensively in clinical analysis applications; the best known co-ercial CFA instrument is the Technicon Autoanalyzer. Principles of Flow Injection Analysis The other branch of continuous flow analysis was developed by two separate groups, but has since emerged as a singular technique known as flow injection analysis (FIA). The approach of Stewart, Beecher, and Rare (27) was a conceptual outgrowth of liquid chromatography, whereas that of Ruzicka and Hansen (28) was founded as an alternative to CFA. In any case, the driving force for the development of this technique was the ability to manipulate and reproduce dispersion of the sample. This technique involves the injection of a snple into a nonsegmented carrier stream. Since the conditions are usually such that laminar flow is predominant, the development of a parabolic velocity profile is responsible for the dispersion of the sample along the axis of the tube. This dispersion, although much greater than that found in CFA, can be controlled by appropriate choices of tubing length and inner diameter, flow rate, samle size, and other components such as valves and flowcells which determine the overall volume of the reactor in the FIA system. Ruzicka and Hansen have proposed an empirical method by which dispersion can be measured (19). The ‘sample has an initial concentration of C“ as it enters the carrier stream. As the plug travels through the manifold, axial and radial mixing take place. This results in a predominantly Gaussian-shaped signal profile. The maxim concentration sensed by the detector, 0'“, is only a fraction of 0". The formula for the dispersion is: The height of the peak obtained with the undiluted sample is H’. After the sample has traversed the manifold a lower peak height, H, will be obtained due to dispersion. If the two constants are equal, as in the case of photometric detection for a system that adheres to Beer’s law, the peak heights of the signals can be used to determine the dispersion of the FIA cyst... This empirical value is classified as limited (D = 1-3), medium (D = 3—10), or large (D > 10). The amount of dispersion that can be tolerated in an FIA system depends on the application for which it is intended. For more transportation of a sample, limited dispersion is ideal. On the other hand, when a chemical reaction requiring reagent addition must take place, medium or large dispersion is desirable. Types of FIA Manifolds One way of manipulating dispersion is by selection of the appropriate type of manifold. A comparison between various types bf FIA manifolds was carried out by Ruzicka and Hansen (29). Straight open tubes yield relatively large amounts of dispersion. Coiled tubes show less dispersion due to the presence of secondary flow. This type of flow is a result of the centrifugal forces which affect the flow perpendicular to the axis of the tube. The sue phenomenon is observed in 3-D coiled tubes (tubes with regularly spaced knots tied in them). A relatively small amount of dispersion has been found in the use of packed tubes. This is due to the fact that the parabolic profile is broken up as the sample passes through the packed material. The high pressure drop associated with tubes that have been packed with very small diameter particles makes them difficult to use with the peristaltic pumps normally present in an FIA system. 10 The single head string reactor (30) has gained acceptance as a viable alternative to open tubes and those packed with small particles. The SBSR consists of a tube packed with beads that have a diameter 603 to 803 of the tube inner diameter. This type of reactor has the advantage of decreased dispersion due to the break up of the velocity profile. The pressure drop is small. Therefore the SBSR can be used to provide longer residence times without an increase in dispersion. An example of typical signal profiles obtained with an SBSR and an open tube of the same length and volume is shown in Figure 1. (Each length of open tubing had the same internal diameter). Although the arrival time of the peak at the detector did increase with reactor length, the point of injection is not given for the sake of simplicity. For use with i-obilized enzymes an added advantage of the SBSR is the additional surface area available compared to that of an open tube. Measuraent Tools in FIA Following the development of an FIA system with the appropriate amount of dispersion, the analyst must choose a measurement tool from a variety of peak characteristics. The complete signal is usually recorded either on a chart recorder or by a computer data acquisition system. The most co-only used measurement is that of peak height. In this case the FIA syst- may be simplified by the use of a peak detector. In some cases the height at some other place on the peak than the maximum is used (e.g., in gradient techniques). The full width of the peak at half the maximum height (FWHA) can be used as a measure of sample dispersion. Peak area may also be used as an analytical ll Assessed—Xe bonanza you 38» 005 000.395 mo sewn»: one 5 3.5 one soon: 133m .04.... some as was mama as :33 vocwsano 33.3.3 <2 .«o canton—loo 8mm. m2; n&_og om.aen§ o~ o ELL-PE mm? zumo mo...uoo cocoosuuosuh mmmm swoon macaw usouomeo: cu . 1;: me ousvwxo ouoosao ecu-acouus aco~s>oo Uwauulouogeosuuomm mmnm unsusaso oususo>cu uesaossdsz sauna chem vouaohaeoo an we ousvwxo onoosuc asalaosuus u=o~s>oo concussed-Sawmogo to: museum ousuuo>cu :::s oesaousasz sedan ouch vauaouusoo on ~v oaavflxo oeooaao ass-nosaas aeo~s>oo Omens-ouozmosuoomm can voxonm sedan ouch vaunouusoo 0» cc ouuaououvazov onoosdo ass-nosuus acmas>oo cwuuoloflu:0uom pom voxoam unsuso>cu onsuohsusl madam when vouuouuaoo c» an ousvwxo snousac ass-Joanne aco~s>oo oflsuQIUhQAI< pom vozosm newsflas nsosom 0» mm onsvwxo mucosao acolgusuus acoas>oo awuuolouomld mom moment sow noses» as onscaxo ouoosuo «nuisances aco~s>oo com voxosm salsams mason oew>on d an vasomxosom new: voxcwalnuouo cauaoIGLOQI< ovosuoouu ussvwxouvm \ousvwxo sewnassx onsvfixouom Amos sea»: 0» ovohuooHo mm \onsvwxo unoosuo acclaosuus aco~s>oo oflsuoleuoml< cahwcolwm Wwdwguuhom gfimm venue: nowunNMEAQ-IH v0.30! newuoouo: ms scuosmm .uomswso confluwaomla mo seamlsxm a u4m<8 CHAPTER 3 THE GLUCOSE OXIDASE SBSR As was reported previously (Chapter R) there is a minimal amount of dispersion associated with an SBSR made of nonporous glass beads; this allows longer residence times which. are desirable in enzyme applications. Glucose oxidase was chosen as a model enzyme since it is inexpensive, highly active, and easily obtained. The development of the procedure used to i-obilize this enzyme on nonporous glass heads is described. The dimensions of the SBSR assedaled with these beads are given in Figure 2. In order to use an immobilized enzyme single bead string reactor effectively one must first know its characteristics. In some cases it is desirable to know the amount of enzyme actually immobilized. A method fer such a determination is presented. Since it is desirable to work with a reactor that retains its activity for a long period of time, the useful lifetime of the immobilized glucose oxidase SBSR is discussed. One problem with many enzyme preparations is that they sometimes contain small amounts of impurities which can be immobilized along with the enzyme. These impurities sometimes affect the apparent activity of the enzyme. The presence of catalase in glucose oxidase preparations can be problematic in this respect. Results of the investigation of catalase activity in the reactor are presented. The specificity of the reactor is also examined. 14 15 9000000900 / \ 0.86 mm i.d. 0.6 mm diam. Teflon Tubing Enzyme Coated Glass Bead FIGURE 2 Single bead string reactor. 16 mobilisation Procedure for Glucose mjdase Covalent attachment of enzymes to an inert matrix provides a sturdy environment and yet allows the enzyme to be readily accessible for reaction in continous flow reactors. Glass provides an ideal inert support because of its low reactivity toward most compounds and mechanical stability. Nonporous glass beads provide added surface area without increased dispersion. For these reasons, covalent bonding of glucose oxidase to nonporous glass beads was the method of choice for the reactor to be incorporated into a flow injection system. Glucose oxidase is ca-only attached to glass by glutaraldehyde linkage. The specific conditions for this type of i-obilization as enumerated by Thompson (43) were used initially. There are four steps in the procedure: cleaning, whisker formation, silanization, and enzyme attachment. The aim of this work was to minimize the amount of time required to perform the i-obilization and to establish a procedure that gave reproducible results. The conditions for a given step were varied while those for the other three steps were held constant. In each case all four steps were carried out on a 0.5 g sample of nonporous glass beads. After the i-obilization of the glucose oxidase the beads were put into Teflon tubing (10 cm in length) and incorporated into the FIA system described in Chapter 4. A calibration curve was obtained with glucose standards in the 0 to 800 ppm range. The slope of the calibration curve was taken as a measure of the activity of the enzyme on the beads. l 7 Cleaning Procedure In order to i-obilize an enzyme efficiently it is necessary‘to eliminate any contaminants from the surface of the glass before the procedure is begun. There are a variety of ways by which glass surfaces can be cleaned. The most co-on chemical methods are by chromic acid, hydrochloric or nitric acids, or alcoholic potassium hydroxide. Chromic acid cleaning is a very harsh treatment that ensures the oxidation of any carbonaceous material. Other acids are used to leach any heavy metals from the surface of the glass. As shown in Table 2 the initial procedure specified treatment of the glass beads with both chromic and hydrochloric acids. It is known that in order for an acid treatment to be effective it must be carried out at elevated temperatures (48). The use of warm concentrated acids can be hazardous and is therefore not a highly desirable method of treatment. Also, when the beads were cleaned with chromic acid they darkened during the whisker formation process. These two factors suggested that an alternative cleaning process would be advantageous. Since the glass beads are not extremely dirty it is not mandatory that chromic acid be used. As an alternative cleaning solution, alcoholic XOR was chosen. The beads obtained by cleaning with only the alcoholic XOR did not darken and were found to have a level of activity similar to those treated with chromic and hydrochloric acids. This would suggest that the beads do not have an excess of heavy metals at the surface; these would require leaching by RC1. $.92 Cleaning Rinse Dry Whisker Formation (Dry Heat Silanize Decent Heat Glutaral- dehyde Enzyme 18 TABLE 2 Initial immobilization procedure. Regent Time Temperature ('0) Chromic Acid 15 min. 23 6 M 801 10 min. " Distilled 820 " Acetone " "2 n 53 (w/v) 1 hr. 23 Nflsf RF/MeOR Ne " 3 hr. 450 2% (v/v) 30 min. 23 3-aminopropyl- triethoxysilane/ acetone 24 hr. 90 13 solution in 30 min. 23 pH 8.0 buffer 5 mg/ml glucose 24 hr. 4 oxidase in pH 6.85 buffer l 9 Ihisker Pornt ion The best conditions for whisker formation were found by Onuska (24) and were not altered. This step involves soaking the beads in methanol saturated with a-onium hydrogen difluoride for an hour, decanting the solution and drying the beads, followed by beating them in a sealed glass tube at 450°C for three hours. The effectiveness of the treatment with a-onium hydrogen difluoride was investigated. Three experiments were conducted. The first involved treating the beads in the specified manner. In the second the heating of the beads in a closed tube was eliminated. The third experiment involved no treatment with a-onium hydrogen difluoride. It was expected that the beads from the first experiment would have the highest activity. Since the second experiment was carried out at room temperature it was expected at most that only a limited amount of etching would take place; this was expected to result in only a small increase in surface area and thus a small increase in enzyme activity over that obtained from the beads in the third experiment. The activity of the beads was found to vary as expected. The beads from the second experiment exhibited only 588 of the activity of those from the first; the beads from the third experiment had an activity that was only 49% of that associated with those from the first. This indicates that indeed the surface area is greatly increased by this treatment. It should be noted, however, that the enhancement is not due to the formation of real whiskers. Mottola has shown that the surface looks more like a series of "volcanic crater erosions" (45). The same author points out that 20 there is not a great advantage to be gained by roughening the surface of the tube and i-obilizing the enzyme on its surface as well. Silanization The silanization of the beads proved to be the most problematic part of the immobilization procedure. It was found that heating the beads with the 3—aminopropyltriethoxysilane in acetone caused the coverage to be quite non-uniform. It was suspected that the rapid evaporation of the solution was the reason for this. Therefore, it was thought that an aqueous solution might give better results. The silanization procedure specified by Mosbach (8) was tried. The resulting beads had low activity. A "trial and error" approach was used to arrive at a silanization procedure which gave reproducible results. The concentration of the solution, the temperature, and reaction time were varied. It was found that a 1" (v/v) solution gave sufficient coverage without leaving an excessive residue on the bottom of the beaker. By allowing the solution to evaporate from the heads at room temperature in the fume hood more uniform silane coverage was obtained. The temperature and reaction time were decreased in order to prevent darkening of the heads (a sign of silane decomposition). The time required for this part of the procedure was significantly reduced from 24 hours to 15 hours. Since the solution used to silanize the beads was made with acetone it was assumed that washing the beads with acetone after the treatment would be sufficient. However, the beads stuck to the bottom of the beaker. One solution to this problem would be to use Teflon 21 beakers instead of glass. However, the more economical solution proved to be the dropwise addition of distilled water (DH) to the acetone until the beads were loosened. If too much water is added the beads float. Thus, it can be concluded that this part of the procedure seals to be much more of an "art" than a ”science". Glutaraldehyde Attachent The glutaraldehyde was attached to the silane group as specified in Table 2. A 1% (v/v) solution of glutaraldehyde in pH 8.0 phosphate buffer was allowed to react with the silanized beads. This step was not modified. However, it was observed that the beads become salmon colored when this part of the procedure is successful. It was noted that the time required for the color change to take place was in the range of 0.5-2 hours. Enzyme Attachment and SBSR Construction The covalent binding of the enzyme was carried out under the same conditions as in the initial procedure (Table 2). A 5 mg nul‘1 solution of glucose oxidase (E.C.l.l.3.4, Sigma Type II from Aspergillus niger, specific activity approximately 17,800 U g'1 at 35°C) in phosphate buffer was allowed to react with the glutaraldehyde treated beads for 24 hr. It is desirable to immobilize the enzyme in its most active state. The use of a 0.05 M phosphate buffer, pH 6.85, facilitates this. After this step was complete, a single head string reactor was made by placing a suitable portion of the beads in a length of Teflon tubing. This task was most easily accomplished by crimping one end of the tubing 22 and inserting it into the end of a syringe. The tube was then filled by applying auction. The beads remaining were transferred to a glass vial. In any case the beads were stored in 0.05 M phosphate buffer at 4°C. The final inobilization procedure is sum-arized in Figure 3. A1 temate Method of [mobilization While periodically retesting the SBSRs, a rapid decrease in activity was observed during the first few days of use. This was attributed to enzyme being adsorbed on the surface. One remedy for the situation is to flush the SBSR with copious amounts of buffer prior to use. An alternative is to prevent adsorption from occurring in the first place. Therefore, the possibility of immobilizing glucose oxidase by continuously circulating the enzyme solution was investigated. The immobilization procedure was carried out in the usual manner up to the point of the glutaraldehyde attachment. The beads were then put into a suitable length of Teflon tubing which was connected to a pump tube. A small two-channel peristaltic pump (Ismatec) circulated the enzyme solution at the rate of 0.8 ml min'l. The resulting beads exhibited very low activity compared to that of beads obtained by the usual method. The trade—off of providing the beads with a continuous stream of "fresh” solution versus the time needed for attachment at the surface proved to be in favor of the latter. Perhaps a very low flow rate (e.g., 0.05 ml min”) would provide a suitable environment for the immobilization . .835& 330333 use: a an 3:3 22.9%.. s. no.0 s 822. 8.. 5... 8.8: 3 832.8 . o . o . . Esau- z.u2~:u.:u.zn.«:ozmo A 25.26. z~x+ozun.~:o.:o.zn.«:u: ...mo . 9v 9 23 . I m. ... .56522. 3:35.. .m 3 , 8:: 8:3 suing... 28.0 3 332.2 2 . . ... .5 ... ozomuzuezuémazuzmé A - 9 an ”15. + ~zz»..zu:m.-o JV 0. 010 mo 8:63 02522235 .N m 3.... of 28:3 3 2.32.2 . «:zn.~zo..m-o :o o l ”15.50%” u a + «zznmxuzw o 92. 9 an} . :8 :z :o m o.- 0 and a . cozoucgm _ 24 Determination of Amount I-obilized The are two viewpoints concerning the amount of enzyme i-obilized. The first is the literal interpretation-what quantity of enzyme is on each bead. The second is more pragmatic in nature-what equivalent amount of enzyme must be used in solution in order to do the same analysis. There are a under of ways to arrive at an answer for the first case. One might choose to do a nitrogen analysis. This would not guarantee an exact result unless the aminosilane content were previously known (or determined). An alternative would be quantitation of the protein present. However, the small amounts of other enzymes present as impurities would result in a positive error. In either case, the enzyme is destroyed in the process. Professor Mottola’s group at Oklahoma State University has been working on a nondestructive solution to this problem, but has only published. a method for the determination of aldehyde groups on glass surfaces (49). It therefore seemed appropriate to pursue another method of enzyme determination. The method chosen was that of an assay by difference. A glucose oxidase solution was prepared, half of which was used for the covalent binding step in the immobilization procedure and the other half as a control. After the enzyme was attached to the heads the reacted GO solution was poured into a volumetric flask. The washings were cowined with this and the mixture was diluted to the mark with distilled water. The final solution was used as the carrier solution in the FIA apparatus with a plain SBSR in place of a GO SBSR. Dilutions of the control solution were used in a similar manner to construct a calibration curve. For each GO solution a 400 ppm glucose standard was injected and the 25 absorbance measured. The absorbence of the reacted solution was located on the calibration curve. After correcting for the dilution factors, the amount of glucose oxidase i-obilized per gram of beads was determined to be 7.3 mg (approximately 1.4 ug -'2). From a more practical standpoint (the second viewpoint mentioned), the activity of the G0 on a 10-cm SBSR was found to be equivalent to a solution containing 0.5 mg of enzyme per m1. Cofiining the results from the two determinations, it would take less than two hours under continous operation (a flow rate of 0.5 ml min”) to use the entire amount of glucose oxidase needed for the i-obilization. These results were based on the initial activity of the GO SBSR. In order to make a more realistic comparison one must take into account the drop in activity that occurs. A typical scenario of the loss in activity is shown in Figure 4. It can be seen that after about 150 days the activity of the stored beads would match that of the beads which had been used for periodic testing. The activity in this particular SBSR (one of the first assdbled for this research) dropped to approximately 568 of the original. The G0 SBSR used in the determination of the amount of enzyme i-obilized was found to retain 66X of its original activity after one year (the i-obilization method was improved). Thus, it can be concluded that the glucose oxidase i-obilized by the modified method presented in Figure 3 has a very long useful lifetime. This evidence substantiates the cost advantage of enzyme imobilization. cmammou CURVE SLOPE mo“) 1.“ 26 FIGURE 4 UUUrrritf'UllI'IfirIrrIIT'UUFV' so 100 150 TEST DAY Loss of glucose oxidase activity for routinely used beads (C) and stored beads (A). 2 7 Ca talase Acti vi ty In all ihobilizations there is a risk of i-obilizing impurities which will interfere with the reaction of interest. Catalase is a major probl- in the case of glucose oxidase since it destroys the peroxide formed. In order to get some idea of the extent of the catalase activity in the GO SBSR, an experiment was carried out in which three lO-cm reactors were tested individually and then connected together to make reactors 20 and 30-cm long. The expected absorbance was obtained by adding the appropriate absorbances for i the two- or three-unit reactors used. As can be seen in Figure 5 there is a significant difference between the expected values and those actually obtained. In an effort to eliminate some of this effect the color-forming reagent stream was placed before the glucose oxidase SBSR. A wider range of linearity was found, but there was still a significant deviation from the expected results. Specifici ty of I-obi 1 i zed G] ucose Oxidase Since the ultimate intended use for the GO SBSR is in the analysis of real samples it was necessary to ensure that the reactor is specific for glucose compared to other sugars. Nine nutritionally—significant sugars were tested by preparing 3.3 #4 standards of each and using them in the place of B-D-glucose. The responses were then taken as a percentage of that obtained for fl-D-glucose. The results are compared in Table 3 to those found in the literature (50). The experimental n n values designated by a - indicate that the value obtained could not be differentiated from the baseline. All values except that obtained for ABSORBANCE S o ‘ g g E llllllllLlllllllllllllljllllljllllllllllllllllll 28 om UIVTr'I'UrriIUIIIUIIrTUUIUUU'I’T‘W‘IUrI—Ul FIGURE 5 200 400 600 [GLUCOSE] (pg/ml) Actual (- -) and expected (-—) effect of catalase in a 20-cm SBSR (O) and a 30-cm SBSR (O). 29 TABLE 3 Glucose oxidase activity toward other sugars. Response Ratio §uga£ §§§§ Literature p—D-glucose l . 00 l . 00 (“D-glucose 0.3% 0.0064 galactose - 0.0014 fructose - na mannose 0.011 0.0098 2-deoxyglucose 0.18 0.25 maltose 0.015 0.0019 xylose 0.0062 0.0098 glucosamine - (0.0005 sucrose - na "-" denotes no distinguishable response "as" means not available 30 a-D-glucose are as good as or better than those indicated in the literature. The unusually high value for c—D—glucose was a result of phosphate catalysis of the mutarotation of this species. A. more detailed discussion of this is undertaken in Chapter 5. In am, an inobilization procedure for glucose oxidase has been improved and characteristics of single bead string reactors made from nonporous glass beads containing the enzyme have been established. In the next chapter some practical considerations concerning the use of this reactor are discussed. CHAPTER 4 OPTIMIZATION OF THE FIA SYSTm The glucose oxidase/Trinder reaction (51) is one of the most cmonly used methods for the colorimetric determination of glucose. The reaction scheme originally proposed involves the reaction of glucose and molecular oxygen in the presence of glucose oxidase (GO) to produce hydrogen peroxide. The H202 then reacts with 4-aminoantipyrine (AAP) (also known as 4-aminophenazone) and phenol in the presence of peroxidase (PO) to produce a colored dye with an absorbance maximum at 505 nm. Trinder later found that when the phenol was replaced by 3,5-dichloro-Z-hydroxyphenyl sulfonic acid (DCPS) the reaction was four times more sensitive and the absorbance maximum shifted to 510 nm (52). In this work the latter Trinder reaction was coupled with the GO reaction as shown in Figure 6. Other workers have immobilized glucose oxidase and used it in flow injection analysis systems in conjunction with the Trinder reaction (45, 53, 54). However, this work is the first to be published (55) regarding the optimization of such systems. In this chapter the computer controlled FIA system incorporating the GO SBSR with the Trinder reaction is described. The salient features of the instrumentation are enumerated. Three different approaches toward improving system performance are discussed. 31 32 603000.. uevsuuh\ees3xo 0.836 U! n I n: z u: a J n 10.2, come: o~zm+©gzu0 +||oa z o o o 6 9532 $683.05 N0.4.: + o n 101 IIO I O xouzu O: I m 953.. n34 mauo u ZN: Inom + I9©6 + «ofm @ . wmouaoéum :o x 302.8 3820 NO + r“ I rm H... 0 ref... 33 Instrumentation A diagram of the apparatus is shown in Figure 7. The temperature of the solutions and the GO SBSR was controlled by a thermostatted circulating water bath. The temperature of the reagents was controlled by placing each solution in a jacketted beaker or flask. A jacket was made for the GO SBSR. This consisted of a piece of glass tubing (that fit snugly around the lO-cm length of Teflon tubing) fused into the open ends of a small condenser. The sample was injected via a pneumatically actuated 6-port sample injection valve with a 30-u1 sample loop (Rheodyne) into the carrier stream which was pumped by a lZ—channel peristaltic pump (Ismatec). Flow rated pump tubing (Technicon) was used for delivery of the reagents to the manifold. The sample first passed through the GO SBSR. It was then joined at a tee by a solution containing PO, AAP, and DCPS. The sample/reagent mixture then passed through an SBSR made with unmodified glass beads (also referred to as a "plain" SBSR) in order to promote mixing without added dispersion. After reaction the sample/reagent mixture entered a miniaturized flow-through filter colorimeter which was designed by Patton et a1 (56) and assembled in our laboratory. Data acquisition, sample injection, and pump speed were controlled by a microcomputer as discussed below. Choice of Fixed Parueters Initially there were a number of parameters that, for convenience, were fixed. These were: SBSR lengths, sample size, flow ratio of the carrier (buffer) and reagent streams. As mentioned previously 34 eczema-heal. 0803' no.“ use». 4: b 953.: “Sign"; ego:- _H ..eaoeaeml_ lull-.1, “I IIIII u . V A A Jaeemoemu :3... .....-..-..H...-.H-.i . u . 35 (Chapter 3) the effect of the catalase, as the GO SBSR is made longer, invokes the ”law of diminishing returns”. Therefore, a lO-cm length was chosen since its activity gave acceptable absorbance values. This also minimized the back pressure which results when a packed reactor is used in a flowing stream. A 40-cm SBSR consisting of unmodified glass beads was used as the reaction manifold for the Trinder part of the reaction. This length was necessary to allow sufficient time for reaction between reagent addition and detection. The placement of the PO reagent solution was also of interest. Theoretically, it would be better to add this solution prior to the glucose reaction so that any effect of catalase would be minimized. The hydrogen peroxide would be i-ediately consumed by the color-forming reaction and have less chance of coming into contact with catalase as the reacting plug travels through the manifold. A comparison of the calibration curves obtained with the reagent addition before and after the GO SBSR revealed that there was no significant increase in slope or linearity for a lO-cm length. The 20 and 30-cm lengths did exhibit a significant increase in linearity. Thus the choice of a lO-cm length was confirmed as the best one. The reagent solution could be added after the glucose reaction without loss in sensitivity and without risking contamination of the immobilized glucose oxidase. A sample volume of 30 ul was chosen to match the internal volume of the glucose oxidase single head string reactor. In order to investigate the relationship of absorbance to sample size a 6-port sample injection valve was connected in such a way that the sample size was regulated by the amount of time the valve spent in the "inject" 36 position. The injection time multiplied by the flow rate of the sample stream was used to determine the sample volume. The sample injection time was varied from 1 s to 25 s in l 3 increments with a flow rate of 0.44 ml min‘l. The peak heights obtained are plotted in Figure 8 versus sample volume. It is interesting to note that the relationship between absorbance and peak height is linear from 7 pl to 81 ul. The greatest sensitivity is attainable in this region and, thus, makes the 30-311 sample volume desirable. Beyond that region the response increases by successively smaller amounts and finally approaches the maximum amount of glucose that can be converted--approximately 74 ug (185 pl of a 400 ppm standard). In previous experiments related to this project (43), double peaks were obtained with volumes greater than 65 ul for a 20-cm GO SBSR through which the reagent stream was passed concurrently with the sample. This resulted in an optimum sample volume of 45 pl. No incidence of double peaks was observed for this system which indicates that complete mixing occurs in the manifold. The ratio of the flow rate of the carrier stream to that of the reagent stream was adjusted by appropriate choices of pump tubing. A range of ratios from 4:1 to 20:1 was investigated. The overall flow rate was kept at 0.5 ml min‘l. The resulting peaks are shown in Figure 9. There was no sampling time advantage to be gained since the baseline widths of all peaks are similar. The peaks obtained with the two highest flow ratios showed some distortion. Therefore, a 10:1 ratio was selected. This also helped minimize both the consumption of PO and dilution of the sample. PEAK HEIGHT (ABSORBANCE) 0.400 0.2” IIILIJJLJIIlllllllelllllllljl 37 r T I l I r I I ' r U r r l t r r r j 0 50 100 150 200 SAMPLE SIZE (pl) FIGURE 8 Bffect of varied glucose sample size. ABSORBANCE 0300 0300 0200 OJOO 38 FLOW RATIO (BUFFER:REAGENT) 3 5 7 9 11 13 15 17 19 21 23 r I I U fir ' l V r t l I l V r r i r I' I 4 4 n , fl 1 1 d J rrvwfirrrvwwqrrrrw1arrrv1 0 100 200 TIME (sec) FIGURE 9 Effect of flow ratio (Buffer : Carrier) on FIA profile. 39 TQerature Calibration Since it was necessary to control. the temperature of the FIA manifold, a method had to be devised to monitor the temperature. The temperature of the solution flowing through the GO SBSR was not the same as that of the water bath. A high precision 30-k0. thermistor was placed in a thin piece of protective Teflon tubing which was subsequently inserted into a glass tee (see Figure 10). The junction was sealed with Teflon tape to minimize the dead volume and prevent leakage. This temperature probe was calibrated by immersing it in a water bath at the same level as a thermometer. The temperature of the bath was increased in 2" or 3° increments and the resistance recorded after a stable value was reached. The calibration curve is shown in Figure 11. The relationship of resistance to temperature was fitted for 10° segments to the Steinhart-Hart equation ('1"1 = A + 3*(1n R) + C*(ln R)3, taken from Reference 57). The resulting coefficients are presented in Table 4. The probe was then connected to the terminal end of the GO SBSR. Once again the temperature of the bath was increased in small increments and the resistance measured. After converting the resistance to temperature, the following linear relationship between the temperature at the end of the reactor and that of the water bath was found: T = 0.505 It Ta + 11.1, where T is the temperature at the GO SBSR and Ta is that of the water bath. Microcmputer Interface to FIA Systa Original 1y an in-house designed (58) Intel 8088-based microcomputer (also known as the ”Bruce Bus") controlled the pump speed, 40 £35885 978 335380 83?... 2 952.. ll \1 an..." @\ =00 603* 35200 2 +1 x f ‘ 5562.22 .265 A RESISTANCE (RD) 41 8 LLllllLlLllllJlllllllelllLll] 20 10 O frrrtTIUIfI'rTTrrTITIUIITrUTlrUTrI 0 ' 10 20 30 40 50 00 70 TEMPERATURE (°c) FIGURE 11 Thermistor response curve. 42 TABLE 4 Coefficients of the Steinhart-Hart equation for different flow rates. Equation: T*1 = A + B (In R) + C (In R)3 T = temperature Temperature Range (°C) 21-30 31-40 41-50 51-60 0.133 0.0535 0.0358 0.0100 R = resistance B -0.0595 -0.0226 -0.0145 0.00112 C 0.00279 0.00159’ 0.00137 0.000527 43 injection, and data acquisition. The software was written in FORTH (Forth, Inc., Hermosa Beach, CA). Most of the "words" used in this work were strings of "core words” which had been written or modified by Paul Kraus. The final word, "PEAKS", combined injection timing, data acquisition of 100 points over a specified period of time, automatic storage of the data on a floppy, and repetition of the process for a specified number of times. More extensive information on the configuration of the microcomputer can be found in Reference 59. In the latter phase of the project computer control was relinquished to an IBM PC compatible microcomputer (Bentley-model T). The machine was also Intel 8088—based. It was equipped with two floppy drives, 512 Kb of memory, a color graphics board (IBM), and an RTI-815 interface board (Analog Devices). All routines were written with QuickBASIC (Microsoft). The timing routines were written in assembly and made use of the AM9513A counter/timer that was part of the interface board. Control of the pneumatic actuator on the injection valve was accomplished by connection to a digital output port on the interface board. The data acquisition routines utilized direct memory access (DMA). Plotting routines were developed for three different graphics modes-~EGA, CGA, and HERCULES (the HERCULES drivers were provided by Mark Victor, a graduate assistant in the Chemistry department). The initial DMA and plotting routines written by Pete Wentzell, a graduate assistant in the Chemistry department, were modified for this work. The main program "FIADATA" is a menu-driven program. A number of features were built into the program to make it flexible enough that almost any flow injection analysis experiment could be carried out on 44 the equipment presently available in the laboratory. One important aspect in this regard is that either the sample output (1 10 V) or the ”log-ratio” output (i l V) from the calorimeter can be monitored. The gain selection option on the interface board is controlled by the software. The general characteristics of the program are: user- selected experimental parameters, a flow reversals experiment option, data acquisition from one or two channels at a rate selected by the user, immediate data storage on a floppy, screen plots of the data, and printouts of the file containing the data. This program was modified to allow variation of the injection time; "FIAINJ" was the result. The need also arose for the capability to plot (on the screen) or print data from files previously stored on a floppy. ”FIAPLOT" incorporates a routine that will read the data regardless of its format. The only stipulation is that the comments be prefaced with ”;" and confined to the beginning lines of the file. By calculating the number of data points and the number of lines of data, the number of columns of data is empirically determined. The user selects which columns of data are to be used for the plot-one x and one or two y values. There is also an option in the program that allows the user to print the data file. Eventually, the three programs may be merged into one large generic program that will do everything but turn the computer on for you. An interface box was constructed in order to facilitate easy connection to the interface board. The mounts for 50-pin and 34-pin connectors were soldered to the back side of a solderless breadboard. The breadboard was then mounted into an aluminum box along with 6 ENG 45 and 6 banana plug connectors. The appropriate size ribbon cables were used to attach the box to the interface board. This arrangement provides a versatile way of linking the computer to the analytical system. Data Acquisition In either case each experiment involved injection followed by monitoring the output voltage of the detector. Each data point shown in the figures represents the average of the values obtained from 3-5 sequential injections. The baseline voltage was maintained at 5.0 V. The detector voltage was converted to a digital signal by the microcomputer’s lZ-bit analog to digital converter (ADC). For the in- house built microcomputer the ADC counts were stored in a file on a floppy disk. The files were later transferred to a DEC LSI 11/23 minicomputer, and the ADC values converted to absorbance units after first calculating transmittance values. The software for the PC compatible computer was written such that the absorbance was calculated immediately following the data acquisition period and stored on the floppy. In both cases the resultant data file contained time, absorbance pairs. Further data manipulation was usually carried out on the LSI ll/23. Improvement of Sys ten Performance The complete FIA system, when first assembled, was functional but it was felt that better sensitivity and sample throughput could be achieved. As mentioned earlier three approaches were used. The first 46 two approaches were focused mainly toward improving the sensitivity. In essence the goal was to achieve the largest peak absorbance possible. A third approach incorporated the achievement of higher sample throughput as well. In the first and third cases an optimum set of operating conditions was sought for six experimental variables; these were the flow rate, temperature, pH, and the concentrations of PO, AAP,and DCPS. All solutions were prepared using distilled water (DH) and filtered as necessary. A peroxidase solution was prepared by dissolving 20.0 mg of Horseradish peroxidase (E.C.l.1l.1.7, Sigma Type II, specific activity approximately 200 U mg'1 at 20°C) in 2.000 ml DH i-ediately before use to avoid decomposition. The reagent solution consisted of the desired volumes of PO solution, 10 i4 4-aminoantipyrine (Sigma), and 10 IN 3,5-dichloro-2-hydroxyphenyl sulfonic acid (Sigma) mixed together and then diluted with buffer to 10.00 ml in a volmetric flask. The reagent solution was prepared i-ediately before use. Glucose (Sigma) standards were made by diluting with buffer the appropriate quantities of a 2 g (c—D—glucose) 1‘1 stock solution that contained 0.5 g 1‘1 benzoic acid as a preservative (the stock solution was stored at 4°C). This stock solution was prepared at least 24 hours in advance so that mutarotation equilibrium was reached. Unless otherwise noted, a 400 ppm (2.2 #4) standard was used. The concentrations of PO, AAP, and DCPS were 0.8 mg ml”, 1 i4, and 1 14, respectively, unless otherwise stated. Phosphate buffer (0.05 M), pH 6.85, was used in the univariate and delay flow experiments. Universal buffer (60) was used in the simplex experiments. Alternative buffers tested in the univariate experiments 47 were Tris and Borax. The recipes for all but the Universal buffer can be found in the Ch-ical Rubber Co. Handbook. All experiments were carried out at room temperature (22°C) except, of course, those in which the temperature was varied. Univariate Experiments The most canon method of optimizing analytical systems is the univariate approach. In this, one parameter is varied while all others are held constant. The six variables mentioned previously were studied by this method. The flow rate was the first parameter investigated. As can be seen in Figure 12 there is a marked decrease as the flow rate is increased. The flow rate (FR) is related to the pump speed (PS) by the following equation: FR = 0.0141*PS + 0.00167. A rate of 0.5 ml min"1 was chosen for subsequent experiments. This gave an absorbance that was large enough to give very reproducible results yet small enough that values obtained with more concentrated samples would not be beyond the range of detection. This also allowed an acceptable sampling rate of one per minute. The effect of the concentrations of the two color-forming reagents was explored. In the initial study the DCPS concentration was fixed at 1 i4 while the concentration of the AAP was varied from 0.1 IN to 2.0 id. The AAP/DCPS ratio that gave the best results was 1:2. This was found to be the case for pH values between 6.0 and 8.6 as evidenced in Figure 13. From the reaction stoichiametry one would assume that the ratio ought to be 1:1. This caused suspicion as to whether or not the ABSORBANCE 0.200 0.150 00‘” 48 0.0 0.5 1.0 1.5 FLOW RATE (ml/min) FIGURE 12 Effect of varied flow rate. ABSORBANCE 0.500 0.400 0.200 0.100 49 I .pH 7.99 ‘ .pH 0.74 < .pH 5.00 I .pH 3.55 " A A 3 ¢ ¢ ; pH 9.03 I I I I I I I T r l l I l I l I I l I r I l Ifi—l 0.00 0 50 1.00 1.50 2.00 2.50 AAP/DCPS RATIO FIGURE 13 Effect of AAP/DCPS ratio. 50 dye was really formed by the binding of one AAP molecule to one DCPS molecule. The possibility was entertained that the initial compound formed might be 1:1 with later addition of another DCPS. Inspection of the proposed dye compomd structure would suggest that this is not highly likely since the addition of another DCPS would be sterically hindered. To rule out this possibility the diode array detection ysstem developed by Mark Victor (61) was connected in place of the calorimeter. Spectra were acquired every second as the reacted sample plug passed through the flow cell. It was assumed that a change in the structure of the complex (formation of more double bonds) would cause a shift in the absorbance maximum. The ends of the plug, where the greatest amount of reaction occurs, seemed the most likely place for this effect to present itself. However, as shown in Figure 14, there was no detectable shift in the absorbance maximum over the entire plug. Next, a continuous variations experiment seemed in order. The coinined concentration of the two reagents was kept at 2.0 IN. Ratios of 2:1 through 1:6 (AAP:DCPS) were investigated. The highest absorbance was obtained for the 1:4 ratio. This ratio is very different from the one obtained earlier, but the AAP concentration is similar (i.e., 0.4 ‘4 compared to 0.5 dd). Since the aminoantipyrine is considered the two- electran donor for the peroxidase (62), its concentration should be the critical factor. This is corroborated in these studies. The results also imply that for a reagent solution 0.4 lid in AAP the DCPS concentration must be at least 1.0 id. From these experiments it is 51 .8388 3.3: I: ..8 .12. 5.. I: so 88....- 5.8383: 8...: 3 use: OOIIO E l 2 a 1W”- n”, s . —~ g R an <54 3 ‘9' “fix. \. ..O 7- . \. s, a , - ,..“m...»... A. - a .... - “mmHMMNN’Illll/ . 1. v.4“ N\\J \ww - H ONOI/llll’ . QDWL..MW.O I. ,N~NNQ. .' ‘\ \‘h‘ “““W‘\- t \“\\\.H“‘~‘ 1- o 0 -1 . - I l 1 1 7 7 0 5 0 0.0110 “0 (3’9 “‘0 .VO \\'0 ”'0 W0 esuaqsosqw 52 also apparent that there would be no great advantage in increasing the total concentration of the reagents. The peroxidase concentration was varied from 0.05 mg ml"1 to 1.0 mg ml”. An 800 ppm standard was used so that the results would reflect the maximum amount of PO necessary for the range of standards normally used for calibration. A plateau region emerged as the concentration approached 0.8 mg ml"1 (see Figure 15). At this point a small variation in the PO concentration would not drastically affect the absorbance obtained. Since such small amounts of peroxidase are used on a daily basis it is desirable to operate in that region. The investigation of the effect of pH was twofold. First, the type of buffer had to be determined. The four types of buffer listed above were each made to be pH 8.00, since that is in the working range of each. The experiment was conducted in the usual . fashion. The phosphate and Universal buffers gave equal absorbazmm values while the Tris and Borax buffers proved to be less than 1. Because the Universal buffer has such a wide working range (pH 5.00-10.00) it was used for the second part of the study. The pH was varied from 5.00 to 10.00 in 0.50 increments. The results shown in Figure 16 indicate a pH optimum at 7.50. For the purpose of comparing this optimum to that obtained with other buffers, another set of experiments was carried out in which phosphate and Borax buffers were used to cover the range of pH 6 to pH 10. The maximum response was obtained at pH 7 (see Figure 17). It can be concluded from this study that the glucose oxidase/Trinder reaction is quite sensitive not only to the pH of the buffer but to the type of buffer used as well. The optimum values ABSORBANCE 0.400 0.200 0.100 53 q q i .1 q - I r j— r I I I— r I r f 1 I I I ' I I I I I I I 1 0.” 0.20 0.40 0.00 0.00 1 .00 1 .20 [P00] (mg/ml) FIGURE 15 Effect of peroxidase concentration. Absarbonce 0.150 54 q «A q 1 .1 d d .4 I I I I I r I T r l’ I r I I I r I I I I I I I I I I I I I l 4.50 0.50 0.50 10.50 pH FIGURE 16 pH dependence of GO/Trinder reaction with Universal buffer. ABSORBANCE 0.300 55 - .4 -’ \ I I \ .4 t \ \ T b q 1 q IIrIlrIIIIIrIIIIrrI'rTII'IIII[IITI'IIIr1 4.00 0.00 5.00 1 0.00 12.00 pH FIGURE 17 pH dependence of GOVTrinder reaction with phosphate (- -) and borax (-) buffers. 56 obtained here are slightly higher than the pH optimum for glucose oxidase in solution (63). This will be discussed below. The final variable examined was the temperature of the GO SBSR. It should be noted here that the temperature of the colorforming reaction was varied over the same range as that used for the glucose oxidase reaction and no significant improvement was attained. This demonstrates that the reaction is fast comared to the time required for the reacted sample to traverse the manifold. Since it is well known that enzymes are prone to denaturation above 37°C the operational temperature range was from room temperature (21°-22°C) to a few degrees beyond this upper limit. The bath temperature was adjusted and the components were allowed to come to thermal equilibrium before the sample was injected. The response obtained increased with temperature as illustrated by the plot in Figure 18. Between 31°C and 38°C a plateau region exists, where the response is at its maximum value. Beyond 38°C the response decreases rather rapidly because of enzyme denaturation. A complete set of univariate experiments would have involved going back to each variable systematically in order to determine a true set of optimum conditions. If such an approach were used to evaluate the optimum to within 20% in each of seven dimensions (an additional variable is created when the pH of the carrier stream is separated from that of the reagent stream), for this FIA system a_ total of 175 experiments would have been required. Since a great amount of time was expended in this small set of experiments, an alternative to this method was sought . ABSORBANCE 57 002m - 1 ' s q .. O a. ‘ m - q - q as” I r I f I I I I I r I I r I 1 20 30 40 50 TEMPERATURE (°C) FIGURE 18 Tuperature response of a 10 — cm GO SBSR. 58 Delay Flaw Experiments This second approach focused solely on increasing the absorbance obtained with a 400 ppm standard. Based on the results of the univariate experiments a standard set of conditions was defined: 0.5 ml min”, 22°C, pH 6.85, 1 fl AAP, l m DCPS, and 0.8 mg ml"1 PO. By integrating the FIA peak it was found that the total absorbance obtained corresponded to conversion of 7.3% of the total amount of glucose in the sample. It was therefore suspected that significant increases could be attained by stopping the flow after the entire sample plug occupied the GO SBSR. It was thought that a delay time would allow as much reaction to occur as possible before the flow was resumed. A preliminary study was used to determine the time at which the pump was to be stopped. The sample was injected and the pump was stopped at various times, while the delay time was kept constant at 10 s. Then the pump was restarted. The greatest response was obtained when the pump was stopped after 10 s. The stop time was then set at 10 s and the delay time varied. Data acquisition times were adjusted to ensure complete recording of the signal. Figure 19 shows that the maximum response can be obtained when the flow is delayed for 10 s. The decrease in absorbance for longer delay times is probably due to the destruction of hydrogen peroxide; as the sample is in the reactor for a longer period of time there is a greater probability that the peroxide will come into contact with catalase and, thus, be reduced. This method only served to increase the sensitivity of the system and actually decreased the possibility of Emmi-.m- ABSORBANCE 0.100 59 T U r f ' T T I T T r r I U r 1 O 20 30 TIME (sec) FIGURE 19 Results from delay flow experiment. 60 greater sample throughput. Therefore, an alternate method of optimization was sought. Simlex Optimization The underlying assumption in a set of univariate experiments is that there is no interaction between variables. If there is indeed some interaction a false set of optimum conditions may be found. Alternatively, a pattern search could have been used to determine the response surface for the seven variables investigated here. Under these conditions 78,125 experiments would have been required. A better alternative than these methods has been shown to be a simplex optimization (64), which allows simultaneous variation of all parameters. The simplex method is widely applied and accepted in analytical chemistry (65). A simple two dimensional surface, as pictured in Figure 20, can be used to illustrate the principles employed. The x and y axes represent the two parameters to be varied and each concentric circle represents combinations of those two which give the same response. This surface can be thought of as a topographical map; as the circles get smaller the response increases in magnitude. The simplex is generated initially by choosing a set of experimental conditions which are known to be suboptimal. After the response from that experiment is obtained another set of conditions is specified. This process is repeated until the geometric shape made of n+1 vertices is obtained, where n is the number of parameters. In the two-parameter case, three experiments are required and the simplex takes the shape of a triangle. 62 The response at each vertex is ranked based on its magnitude—the largest is taken as best, the next largest as next best, etc. In the triangular case, the point that gives the worst response (point 1 in the figure) is reflected an equal distance through the line between the other two points (points 2 & 3). A new set of conditions is specified, the response obtained, and the responses are again ranked. In this example, the response surface is shown so that the movement of the simplex can be understood. Normally, the response surface is unknown and becomes defined by the movement of the simplex. As the optimum set of conditions is reached the simplex may begin to oscillate. Such behavior can be thwarted by reflecting the next best point rather than the worst. The history of the simplex methodology (adapted from Reference 66) is su-arized in Figure 21. Since the pioneering work of Morgan and D-ing (67), many modifications have been suggested which attempt to improve upon the speed and reliability of the method. These have been reviewed elsewhere (68) . For this work, the method chosen was the Composite Modified Simplex (69). This method has additional features such as expansion when the optimum is far away and smaller movements when the optimum is near. The composite modified simplex program (OPTIMA—-written in BASICA) was provided by Betteridge and Wade (70). Simplex optimization was previously attempted on the original Trinder reaction in a static system (71), however, no optimum was found. It has been shown that a judicious choice of a response function can expedite the optimization process (72). Such a response function allows a number of FIA parameters to be optimized simultaneously, and 63 T... ..I mv. rem s maven-m m t m (REFLECTION. next 10 m1 mun-omens. amount n'um hem: "“ names a. nun max «new EXPANSION. mill-“Yuan. no to? "Rut!!!“ or 10 WORST names as m mu. mass was mummy | 7m ERNST m m common I m IMHO.“ I T... I m an a! nouns I mums m mam; am; “7‘ museum ‘—————— menus cum 1072 I—MELILAU (Y'EAY‘N' G museums» magnum ol- moms: cowsmeo me: mm mm» snncm an pun-no s ammo ' (LARGE mun. mm 1.7. KING, mm 6 WM ’0!“ 6 scam (REINTRODUCE NEXT 70 (Ilium K01 M m REJECTIONI ' 1 1070 ”3 mmmm a. neuron MRWPED m. mm m, rum Fm 1000 J RYAN, m & tom um on WIEL (CONTROLLED WEIGHT!!! (ESYINATED l) ammo macs: «muss»: Fm (WW A)" (WONTED “HMO! WEIGHTED cam Mason: 10.: has 3 commas ammo: a m m a wow I» «emu: m mx utmost 1”: wanna. 1.“ “In! t 70“! (MAIN! m FIGURE 21 History of simlex optimization. I. 64 takes the form of a mathematical equation that codaines two or more system parueters (peak height, cost, throughput, etc.) in an appropriately weighted fashion. The response function chosen was based on three considerations. First, the activity of the enzymes involved changes over long periods of time; second, the achievement of a sampling rate of at least 60 hr‘1 was desired; finally, the sampling rate factor should contribute no more than 30% to the overall response. The function chosen was: 80 3 Ann» Time Corrected Absorbance Ratio = x -- 60 8 + tp Absss where Asxp = Absorbance at peak maximum for a set of experimental conditions, Abase = Absorbance at peak maximum for the baseline set of conditions, and tp = time (sec) after injection at which the peak maximum occured. The initial simplex was obtained by entering the information listed in Table 5 into the composite modified simplex program. The simplex program was run on an IBM PC compatible microcomputer (PC Designs) while data were acquired with the Bruce Bus microcomputer. From the univariate experiments that were done prior to this optimization, acceptable ranges for each of the parameters were identified. The settings listed under "Reverse" are those which gave low absorbance values and those under "Forward" gave high absorbance values. Some other considerations played a role in this decision making process as well. The minimum pump setting (flow rate) indicates that a minimum sampling rate of 24 per hour was required. The bath temperature 65 TABLE 5 Range and precision of variables for simplex optimization. Experimental Foward Reverse Variable Boundary Boundary Precision Pump Setting 99 20 1 Bath Temperature (°C) 21.0 51.4 0.2 [POD] (mg ml‘l) 0.05 1.0 0.02 [AAP] (mM) 0.1 2.5 0.01 [DCPS] (mM) 0.1 2.5 0.01 pH Carrier, Sample 10.00 5.00 0.02 pH Reagent 10.00 5.00 0.02 66 was bounded at the low end by room temperature and at the high end by 37 ‘C-the maximum temperature shown to be advisable for use with this enzyme. The high end of the PO concentration was set at 0.8 mg ml‘1 since it was desirable to limit the cost of the experiments and this concentration was in the plateau region described previously. The upper limit for the AAP and DCPS concentrations was such that if both were used in their highest concentrations the reagent solution composition would be no more than 25% (v/v) in each. The pH of the carrier and sample was made independent of that of the reagent since it was unknown whether the two reactions would have different optimal pH values. The range for each was selected on the basis that the working range of the Universal buffer is between pH 5.00 and pH 10.00. The initial (baseline) conditions listed in Table 6 were known to be suboptimal, but exhibited acceptable reproducibility (18 RSD). It was found that a 400 ppm glucose standard, when used with a new GO SBSR, gave a maximum absorbance value that was in the middle of the detectable range and was, therefore, suitable for use in the optimization. Since there were seven variables, eight experiments were needed to form the initial simplex. These were performed and the results entered. The first point of the simplex was the response obtained from the baseline experiment with the initial conditions. Thereafter, each time the simplex program specified a set of experimental conditions, this experiment was performed and the time-corrected absorbance ratio was calculated and entered. In reality two experiments were done in order to obtain the response for one set of conditions: one at the specified conditions and the other at the baseline conditions. 67 ¢.N N.HN ~m.c m¢.m Nc.m Nm.m mm.m mm.m N¢.m o~.m oo.m acumen: mm ov.o 5% CD QQVQN Heme-:08 OD bfiOl‘bCD‘D 8. sewswmo mm m.N m.~m b.~m m.mN m.mm m.mN m.~m m.Nm N.NN Gov .osoosau mo cameosflILanv oa .vam m may mom 5.: ado oumao>< ob mm mm mm m mos A ammum Caz .mxm 68 The simplex rapidly improved the response at first. This was followed by a more gradual increase and then a plateau region as diplayed in Figure 22. This same pattern has been observed and explained elsewhere (73). After twenty—seven experiments, the data suggested that a "false maximum" had been found in experiment nine. To check this, that particular experiment was repeated and was found to be in error. The simplex was restarted with the original values for the first eight experiments and the new value for the ninth. After the thirtieth experiment with the new simplex it was intuitively felt that the rate of improvement might be enhanced by use of a simplex that was larger in the dimensions of temperature and PO concentration. Further optimization was carried out using the current simplex points, but replacing two of those experiments with two experiments at higher temperatures and/or POD concentrations. The optimization process was terminated when an optimum set of conditions was deemed to have been reached. Since the composite modified simplex method is a search procedure in which contraction around a maximum will take place, it is likely that there will exist several sets of conditions in the final data set that are near a true surface maximum. By taking the average of the values for each of the parameters for the experimental conditions that gave the six largest time corrected absorbance ratios (i.e., those yielding a response that was 758 or more of the largest value found), a rough measure of the width of the optimum in each dimension was possible. These results are given in Table 6. 69 qfiqfidJ—qqddqulddddddedddddlfiqdd nw w o o m. m. 02.5. uoz