1111mmnlmm‘mmmmn lL 32193 00671 4707 This is to certify that the dissertation entitled Design, Characterization, and Applications of a Miniature Continuous Flow Analysis System presented by Charles J. Patton has been accepted towards fulfillment of the requirements for Ph . D . Chemistry degree in WKW Major professor Date December 1, 1982 MS U L! an Affirmative Action/Equal Opportunity Institution 0-12771 PLACE IN RETURN BOX to roman m- ohockout from your record. TO AVOID FINES rota-n on or More on. we. DATE DUE DATE DUE DATE DUE || 9K W ’5‘? I If” MSU Is An Affirmdlvo Action/Equal Opportunity Initiation emu»: DESIGN. CHARACTERIZATION. AND APPLICATIONS OF A MINIATURE CONTINUOUS FLOW ANALYSIS SYSTEM by Chas. J. Patton A DISSERTATION Submitted to Michigan State University in partial fulfillment of the requirements for the degree of DOCTOR OF PHILOSOPHY Department of Chemistry 1982 Gzaoei’j ABSTRACT DESIGN, CHARACTERIZATION, AND APPLICATIONS OF A MINIATURE CONTINUOUS FLOW ANALYSIS SYSTEM by Chas. J. Patton A.modu1ar, single channel. continuous flow analysis instrument that can be configured for either air-segmented continuous flow analy- sis (CFA) or flow injection analysis (FIA) has been developed. System components include a modified, commercially available. peristaltic pump, a dual-beam, fiber optic. filter photometer, and an electronic bubble gate that removes the air segment artifact from the detector signal when.bubble-through flow cells are used for colorimetric CFA determinations. Instrumentation and hardware designed and built for use with this system are described in detail. Performance of the miniature CFA system (0.1 cm ID manirold com- ponents and bubble-through flow cells) is characterized for a number of equilibrium based colorimetric determinations. Novel open tubular cad- mium reactors (OTCRs) were develOped and characterized that were used to reduce nitrate to nitrite in conjunction with automated colorimetric determinations of nitrate in water and seawater. Peaks with flats and less than 1% interaction were achieved in conjunction with OTCRs at a l sampling rate of 120 hr- . Comparable performance in conjunction with packed bed cadmium reactors generally used for automated nitrate determinations could only be achieved at half this sampling rate (60 hr'l). Results from an experimental comparison of the performance (sampling rates, sample and reagent consumption, and precision of analytical results) of CFA and FIA are presented. Reagent consumption and sample dispersion in the miniature CFA system were less than in FIA systems equipped with either coiled Open tubular reactors “L05 cm ID) or single head string reactors. 'The advantages of using single bead string reactors for merging zones FIA determinations are demonstrated. CFA and FIA are shown to be complementary techniques, and the relative merits of each for various applications are discussed. Fedor Mikhailovich Dostoevski. the Russian novelist. said one time that. 'One sacred memory from childhood is perhaps the best education.’ I can think of another quickie education for a child. which, in its way. is almost as salutary: Meeting a human being who is tremendously respected by the adult world. and realizing that that person is actually a malicious lunatic. Slapstick -- Kurt Vonnegut ii ACKNOWLEDGMENTS I am extremely indebted to Professor Stanley Crouch for his guidance, enthusiasm, and friendship throughout the course of my research effort. Many thanks also to Professor Chris Enke who served as my second reader. His advice was consistently helpful and incisive. Who else would have known that BALD. stands for Bradlee, Vorhees. and Day? This information is now pinned discretely to my B.V.D.'s. The efforts of the other members of my Guidance Committee. Professors Tom Pinnavaia and George Leroi. are also appreciated. It was also a pleasure to teach for Professor Andrew Timnick. who allowed me to incorporate a continuous flow analysis experiment into his Chemistry 333 course. A great deal of assistance (32; necessarily constructive) was pro- vided by my buddy and fellow sufferer. Eugene Ratzlaff - AKA: AZURE FLEETFANG. Together we managed to advance the 'Boy's Shop' approach to analytical instrumentation to new levels of perversity. Thanks. too, to Robert Thompson who shock-tested early versions of my instrument. You will be happy to know, Rob. that the leering grin of 'Bow tie Man' still haunts my troubled dreams. Sincere thanks are also extended to Keith.Trischan.who wrote several computer programs that were extremely useful in my work. The friendship and help of other Crouch Group members will not be forgotten. iii The help. good humor (ice cream?), programming expertise. and down home philosophy of Dr.‘Tom Atkinson were invaluable. Thanks. too, to Marty Babb for help with the design of electronic components of the miniature continuous flow analysis system. This work would have been impossible without the craftmanship of Russ Geyer, Ben Stutsman, Dick lenke. and Deak Watters in the Chemistry Department Machine ShOp. Andy Seer and Manfred Langer in the Chemistry Department Glass Shop. and Ron Haas and Scott Sanderson in the Chemistry Department Electronics Shop. Special thanks to JO Iotarski who prepared the figures and drawings in this dissertation and to Margy Lynch for technical typing. iv FOREWORD This dissertation is more like a collection of short stories with a common theme, than a fully deveIOped novel. With the exception of Chapter 1 where the origins and principles of air-segmented continuous flow analysis (CFA) and flow injection analysis (FIA) are discussed. the chapters are arranged in the approximate order that research reported was performed. As work progressed new lines of research began to suggest themselves. I have thus included more FIA experiments than I had planned at the onset of this project. In Chapter 2 the miniature continuous flow analyzer that evolved during the course of this research is described in detail. Here I have attempted to provide sufficient information to allow others to use the system as it now exists and to modify it for new applications. Also included are procedures for data reduction and analysis. A description of the electronic bubble gate that removes the air segment artifact from the detector signal when.bubble-through flow cells are used with air segmented continuous flow analyzers can be found in Chapter 3. Details of individual circuit elements in the bubble gate are included. ‘The results of experiments in which this bubble gate was used in conjunction with a commercial CFA system and the miniature CFA system described in Chapter 2 are also presented. and the performance of the latter is compared with performance reported for flow injection analysis. The developmental work on open tubular cadmium reactors for rou- tine nitrate determinations in water and seawater reported in Chapter 4 is the culmination of an idea that occurred to me in 1976 after I had returned from an ACS summer symposium on immobilized enzymes. The major obstacle to beginning this work was locating a source of small diameter cadmium tubing. Apparently it is used in the nuclear reactor industry. I had worked on this project sporadically for several years, but until about six months ago, the successful outcome of this project was still very much in doubt. The experimental comparison of PIA and CPA presented in Chapter 5 resulted from some lively, late-night discussions with several FIA enthusiasts in attendance at the ACS summer symposium on flow injection analysis held in New York in 1981. Upon returning from this symposium. I realized that the miniature continuous flow analysis system which can be easily configured for either FIA or CFA determintions was ideally suited for an unbiased comparison of these two techniques. Please note that the results of this comparison pertain only to equilibrium based colorimetric determinations. and they should not be extrapolated to include determinations in which the formation of concentration gradients can be exploited (titrations. viscosity measurements, refrac- tive index measurements). for which I believe FIA has some clear advan- tages over CPA. Finally, in Chapter 6. some future applications for the miniature continuous flow analysis system are discussed. vi CHAPTER TABLE OF CONTENTS FOREWORDO O O O O O O O O O O O O O O O O O O O O O O O O O O v TBLE 0F FIGURESO O O O O O O O O O O O O O O O O O O O O O O x LIST OF Tux-IE8 O O O O O O O O O O O O O O O O O O O O O O O O xv HISTORICAL AND THEORETICAL INTRODUCTION TO CONTINUOUS FLOW ANALYSIS.. . .. . .. . .. . .. . .. . .. . .. A. Overview. . . . . . . . . . . . . . . . . . . . . . . . B. Historical. . . . . . . . . . . . . . . . . . . . . . l. CFA. . . . . . . . 2. FIA. . . . . . . . . . . . . . . . . . . . . . . . C. Principles. . . . . . . . . . . . . . . . . . . . . . . l. CFA. . . . . . . . . . . . . . . . . . . . . . . . 2. FIA. . . . . . . . . . . . . . . . . . . . . . . . O O He 0 ‘OQQUINNHH INSTRUMENTATION, DATA ACQUISITION AND PROCESSING AND OPERATIONAL PROCEDURES. . . . . . . . . . . . . . . . . . . .25 . General Considerations - CFA. . . . . . . . . . . . . . .25 . Pump. .. . .. . .. . .. . .. . .. . .. . .. .29 . Air Segment Phasing. . . . . . . . . . . . . . . . . . . 31 . Detector. . . . . . . . . . . . . . . . . . . . . . . . .34 Bubble Gate. . . . . . . . . . . . . . . . . . . . . . . 37 Data Acquisition. . . . . . . . . . . . . . . . . . . . .41 Data Processing Techniques. . . . . . . . . . . . . . . .41 1. Data Reduction. . . . . . . . . . . . . . . . . . . .41 2. Graphical Procedures for Estimation of b and ot. . . 42 H. Manifold Components. . . . . . . . . . . . . . . . . . . 49 1. CFA. . . . . . . . . . . . . . . . . . . . . . . . . 49 2. FIA Manifold Components. . . . . . . . . . . . . . . 52 I. Plumbing. . . . . . . . . . . . . . . . . . . . . . . . .53 PHMUOG> DESIGN AND CHARACTERIZATION OF AN ELECTRONIC BUBBLE GATE FOR AIRrSEGMENTED CONTINUOUS FLOW ANALYSIS. . . . . . . 60 A. Overview. . . . . . . . . . . . . . . . . . . . . . . . .60 B. Experimental. . . . . . . . . . . . . . . . . . . . . . .62 1. CF Systems. . . . . . . . . . . . . . . . . . . . . .62 2. Determination of Nitrite and Silicate. . . . . . . . 63 a. Nitrite Reagents. . . . . . . . . . . . . . . . .63 b. Nitrite Standards. . . . . . . . . . . . . . . . 63 vii CHAPTER D. E. PAGE c. Silicate Reagents. . . . . . . . . . . . . . . . 63 d. Silicate Standards. . . . . . . . . . . . . . . .64 3. Bubble Gate. . . . . . . . . . . . . . . . . . . . . 65 Results and Discussion. . . . . . . . . . . . . . . . . .70 1. Introduction. . . . . . . . . . . . . . . . . . . . .70 2. Steady State Experiments. . . . . . . . . . . . . . .71 3. Sample Interaction. . . . . . . . . . . . . . . . . .77 4. Mixing Effects. . . . . . . . . . . . . . . . . . . .79 Analytical Results. . . . . . . . . . . . . . . . . . . .79 1. Nitrite Determinations. . . . . . . . . . . . . . . .79 2. Silicate Determinations. . . . . . . . . . . . . . . 80 Conclusions. . . . . . . . . . . . . . . . . . . . . . . 84 NOVEL CADMIUM REACTORS FOR DETERMINATION OF NITRATE IN WATER AND SEAWATER BY SEGMENTED. CONTINUOUS FLOW COLORIflmY O O O O O O O O O O O O O O O O O O O O O O O O O 88 A. B. C. A. B C. Overview. . . . . . . . . . . . . . . . . . . . . . . . .88 General Considerations. . . . . . . . . . . . . . . . . .89 Summary of Nydahl's Work. . . . . . . . . . . . . . . . .92 1. pH Effects. . . . . . . . . . . . . . . . . . . . . .92 2. Dissolved Oxygen Effects. . . . . . . . . . . . . . .93 3. Effect of Chloride. . . . . . . . . . . . . . . . . .94 Experimental. . . . . . . . . . . . . . . . . . . . . . .94 1. Reagents. . . . . . . . . . . . . . . . . . . . . . .94 2. Standards. . . . . . . . . . . . . . . . . . . . . . 95 3. Cadmium Reactors. . . . . . . . . . . . . . . . . . .96 4. Manifolds. . . . . . . . . . . . . . . . . . . . . . 98 5. pH Measurements. . . . . . . . . . . . . . . . . . . 98 Some Definitions. . . . . . . . . . . . . . . . . . . . .98 Preliminary Experiments. . . . . . . . . . . . . . . . .101 1. Open Tubular Cadmium Reactors. . . . . . . . . . . .101 2. Cadmium Foil Reactors. . . . . . . . . . . . . . . .110 Final Experiments. . . . . . . . . . . . . . . . . . . .112 Discussion. . . . . . . . . . . . . . . . . . . . . . . 120 EXPERIMENTAL COMPARISON OF CFA.AND FIA.. .. .. .. ..122 Introduction. . . . . . . . . . . . . . . . . . . . . . 122 Experimental. . . . . . . . . . . . . . . . . . . . . 124 1. Dye Dispersion Experiments. .. . .. . .. . .. . 124 2. Chloride Determinations. . . . . . . . . . . . 131 Discussion. . . . . . . . . . . . . . . . . . . . . . . 138 FUTURE APPLICATIONS O O O O O O O O O O O O O O O O O O O O O 14 1 A. B. C. D E. Overview. . . . . . . . . . . . . . . . . . . . . . . . 141 Automated Sample Pretreatment and Pre- and Post-Column Reactors for Liquid Chromatography. . . . . . . . . . . 141 Continuous Flow Kinetics Determinations. . . . . . . . .143 Active Open Tubular Reactors. . . . . . . . . . . . . . 144 Extended Data Acquisition and Processing Systems. . . . 145 viii CHAPTER LIST OF REFERENCES. ix FIGURE 1-1 1-2 1-5 TABLE OF FIGURES PAGE Diagrams of generalized CFA and FIA instruments. 8 I sampler, SV I sampling valve. L I sample loop. DB I debubbler. D I detector. PC I flow cell. 36 I bubble gate, CR I chart recorder. A I air, RIreagent. A)CFA. B)FIA................8 Signals recorded for the same flow stream with bubble gated and debubbled flow cells. . . . . . . . . . . . . . . 10 Expanded scale plot of smoothed and regenerated interaction test patterns shown in Figure 1-2. . . . . . .14 Variation of a as a function of n. d . andeort=300 2s. A)FI0.0083 mLs 1: dt:0.2cm. 2: ddtIOJ cm. 3: dltI 0.05cm. _?)d I01 cm. 1: Fl I0.0167mLs . 2: FIO.OO83 mLs . 3: F I 0.004 mL s 1. Minor variables: 8. 9 x 10- poise, 15 I c33 dyne m . D' 25 I 5 x 101.. . .. . .. . .. . .. 17 0.3 Dispersion in FIA as a function of injected sample volume (8 ) and reactor length.(L). Carrier stream I pH 9.5 borate buffer. sample I 10 a! phenol red in pH 9.5 borate buffer. flow rate I 2.0 mL min-1. coiled Teflon reactor diameter -=.O5 cm. A) L I 50 cm. 8': 1 I 25 uL. 2 I 50 uL. 3 I 100 uL. 4 I 200 uL. 5I 400uL. 6- 800 uL. B)Sv I50uL. L: 1 I 10cm. 2 I 50 cm. 3 I 100 cm. 4 I 200 cm. .. . . 21 Methods for phasing air injection with pump pulsations. A) Dual pump tube method. B) Solinoid valve method. C) Schematic diagram of solinoid valve controller. ICl I LM311 comparator. IC2 I 96LS02 dual monostable multivibrator, D1 I germanium diode. 01 I 2N3904 NPN transistor. All resistances in ohms. . . . . . . . . . . . 33 FIGURE 2-2 2-4 2-6 2-7 2-8 PAGE Schematic representation Of the dual-beam, fiber Optic, filter photometer designed for the mCFA system. S I source. F0 I randomized. bifurcated fiber optic bundle. H I fiber optic/flow cell holder. D I dove- tailed slide, PC I flow cell. M I flow cell/interference filter mount. F I interference filter, PD I photodiode. I/V I operational amplifier current-to-voltage converter. A I second stage amplitier.. .. .. .. .. ..35 Schematic diagrams of detector circuits. A) Photodiode (PD) and current-to-voltage converter. B) Second stage variable-gain inverting amplifier. ICl and IC2 I LF351 FET input operational amplifiers. A11 resistances in ohms. Power supply bypass capacitors omitted for clarity. .. . .. . .. . .. . .. . .. . .. .. .. ..38 Detailed schematic diagram of the bubble gate designed for the mCFA system. ....................39 Schematic diagram Of a modified circuit to trigger and clear the monostable multivibrator of the bubble gate. ICl I LF351 Operational amplifier. IC2,IC3 I LM311 voltage comparators. IC4.IC5 I SN74OO quad 2-input NAND gates. .All resistances in ohms.. . .. . .. . .. . 4O Extraction of b from rise curves of CFA peaks. Bubble gated flow cell: VI smoothed At points. VI smoothed (Ass-At) points. Debubbled flow cell: 4;: smoothed At points. AI smoothed (Au-At) points. . . . . . . . . . .46 Graphical estimation of “t by means of a cumulative probability plot.. . .. . .. . .. . .. . .. . .. . . 48 Manifold components designed for the mCFA system. A I air inlet, R I reagent inlet. S I sample inlet. E I epoxy cement. G I glass. A) sample inlet blocks: left. used for plastic coils. Right. used for glass coils. B) Reagent addition tees: left. used for plastic coils. Middle and right used for glass coils. C) Flow stream debubblers. .. . . .. . . .. . .. . .. .51 Manifold components for FIA. C I carrier stream inlet. M I to manifold. W I to waste, S I sample inlet. L I sample loop. A) Six-port rotary valve. B) Dual four-way slider valve. C) Single four-way slider valve. D) Reagent addition tee.. . . .. . . .. . . .. . . .. .54 xi FIGURE 2-10 2-11 3-1 3-3 PAGE Methods for interconnecting various manifold components used for continuous flow analysis. T I teflon tubing. P I pump (not shown). LS. I inner plastic sleeve. 0.8. I outer plastic sleeve. MC I mixing coil. R I reagent inlet. G I gripper fitting. C I standard HPLC connector. 8 I plastic sleeve.. . .. . .. . .. .. 56 A) Preferred method of sample introduction for FIA systems. C I carrier stream. R I reagent. S I sample. W I to waste. B) Pecked sample introduction used for CFA systems. S I sample. W I wash solution. IAS I intersample air segment. (3 Effect of intersample air segments on dispersion that Occurs in the sample line of CFA systems. . . . . . . . . . . . . . . . . . . . .57 Fluctuations in the detector signal caused by the successive passage of liquid and air segments through the flow cell: Position 1. air segment completely within cell. Postion 2. air segment exiting cell. Position 3. cell completely filled by liquid segment. Position 4, air segment entering cell. . . . . . . . . . . .66 Generalized schematic diagrams of bubble gate circuitry. All resistances in ohms. all capacitances in microfa- rads. A) Basic bubble gate. 1C1 I LF311 FET input comparator. IC2. 74LS123 dual monostable multivibrator, IC3 I LF398 sample-and-hold amplifier. IC4 I AD755 logarithmic amplifier. IC5 I TL084 quad FET input Operational amplifier. B) Differential edge sensor. IC5 I TL084 quad FET input operational amplifier. 1C6 I LF311 FET input comparator. C) Sample-and-hold amplifier update logic. 1C7 I 7410 triple 3-input NAND gate. D) Automatic threshold adjustment for comparator. ICl. IC8 I TL084 quad FET input operational amplifier. . . 67 Composite diagram of various CF manifolds used for nitrite determinations. ‘Values outside or within parentheses refer to AAII or mCF manifolds. respectively. . . . . . . . . . . . . . . . . . . . . . . . 72 Results of percent steady state 0588) experiments: 1. standard AAII CF system. 2. bubble gated AAII CF system. 3. bubble gated mCF system without peeked sampling. 4-6. bubble gated mCF system with peeked sampling and n equal to 1.5. 2.0. and 3.0 s-l. respectively. . . . . . . . . . . . . . . . . . . . . . . . 74 xii FIGURE 3-5 4-5 PAGE Effect of segmentation frequency and liquid flow rate on SSS with the residence time of the sample slug in the mCF system held constant at 110 s: l. n.I=1.5 s-l. F .. 0.0067 mL .‘1. 2, n - 3.0 .‘1. F .. 0.013 mL .‘1. . . ..76 Recording of 2. 6. 10. 14 and 18 “M nitrite standards run in ascending and descending order. followed by an interaction test pattern and replicate determinations of thel6 a! standard. A) 120 samples hr-l. B) 360 samples hr O O O O O O O O O O O O O O O O O O O O O O O O O O O O 82 Diagram of the manifold used for silicate determinations 'ith the “CF 'y‘temO O O O O O O O O O O O O O O O O O O O O 83 Recording of 5. 15. 25. 35 and 45 UM silicate standards run in ascending and descending order followed by an interaction test pattern and replicate determinations of the 15 u! :fandard. A) 120 samples hr-l. B) 180 samples hr . . . . . . . . . . . . . . . . . . . . . . . . 86 Diagrams of manifolds used to determine nitrate and nitrite with the mCFA instrument. A) System for Open tubular and foil cadmium reactors. B) System for packed bed cadmium reactors. . . . . . . . . . . . . . . . . . . . 99 Percent recoveries of nitrate and nitrite. A) Open tubular cadmium reactor. B) Packed bed cadmium reactor. . 106 Effect of chloride concentration on percent recoveries of nitrate and nitrite with Open tubular cadmium re‘ctor‘O O O O O O O O O O O O O O O O O O O O O O O O O .109 Peaks recorded for seawater spikes with a 60 cm open tubular cadmium reactor. A) Nitrite spikes. Nominal concentrations from left to right: blank. 0.5. 1.5. and 2.5 9!. B) Nitrate plus nitrite spikes. Nominal concentrations of N02- + N03- from left to right: 0.5 + 5.0. 1.5 + 15.0. 2.5 + 25 ufl. Sample and wash times were 23 s and 7 s. respectively. . . . . . . . . . . 117 Peaks recorded for seawater spikes with a packed bed cadmium reactor. A) Nitrite spikes. Nominal concentrations from left to right as in Figure 4-4. B) Nitrate spikes. ‘Nominal concentrations from left to right: 5.0. 15. and 25 pg. Sample and wash times were 40 s and 20 s. respectively. . . . . . . . . . . . . . . . 118 xiii FIGURE 4-6 5-2 PAGE Correlation of data obtained with open tubular and packed bed cadmium reactors. O I seawater spikes. II== distilled water standards. . . . . . . . . . . . . . .119 Composite diagrams of manifolds used to compare FIA and CFA. A I air. S I sample. C I carrier. R I reagent. A) Manifold used for dye dispersion experiments. B) FIA manifold for chloride determinations. C) mCFA manifold for chloride determinations. . . . . . . . . . . . . . . . 125 Curve tracings of peaks from dye dispersion experiments with Open tubular reactors. single head string reactors, and air segmented reactors. . . . . . . . . . . . . . . . .129 Peaks recorded for chloride determinations with the single line FIA manifold. . . . . . . . . . . . . . . . . .133 Peaks recorded for chloride determinations with a merging zones FIA manifold. .A) 50 cm open tubular reactor. B) 50 on single head string reactor. . . . . . . 134 Peaks recorded_ior chloride determinations at a sampling rate of 360 hr . A) FIA with merging zones and a 25 cm single bead string reactor. B) mCFA. . . . . . . . . . . .136 xiv TABLE 1-1 2-2 2-3 3-1 3-2 3-3 3-4 LIST OF TABLES PAGE Evolution of Techniconfs air-segmented continuous flow ‘n‘lyzersO O O O O O O O O O O O O O O O O O O O O O O O O O 4 Theoretical values of o as a function of F and (1t calculated with Snyder's model for t I 300 s. . . . . . . . 26 Minimum values for at as a function of F and d ‘when limitations in n imposed by flow cell volume are taken into ‘ccountO O O O O O O O O O O O O O O O O O O O O O O O 28 Segmentation frequencies and delivery factors as a function of speed control settings on the modified Brinkmann IP-12 peristaltic pump. . . . . . . . . . . . . . 30 Flow rates for standard pump tubes at various speed control settings of the IP-12 pump. . . . . . . . . . . . . 32 Data for the first 10 s of the rise curve for the steady state peak shown in Figure 1-2. This data set pertains to the bubble gated flow cell. . . . . . . . . . . . . . . .44 Data for the first 10 s of the rise curve for the steady state peak shown in Figure 2-1. This data set pertains to the debubbled flow cell. All other notation is shown in Table 2-5 O O O O O O O O O O O O O O O O O O O O O O O O 45 Major variables for SSS experiments. . . . . . . . . . . . .75 Percent intOIOCtione O O O O O O O O O O O O O O O O O O O O78 Comparison'between mCFA and FIA for colorimetric nitrite detemin.tion‘O O O O O O O O O O O O O O O O O O O O O O O 81 Performance of mCFA system for colorimetric silicate detemin.tion‘O O O O O O O O O O O O O O O O O O O O O O O 85 XV TABLE 5-3 5-4 PAGE Percent recovery of nitrite and nitrate standards for copperized OTCRs of various lengths. . . . . . . . . . . . 102 Linear regression parameters and percent recoveries calculated for nitrate and nitrite standards as a function of pH. . . . . . . . . . . . . . . . . . . . . . .104 Linear regression parameters and percent recoveries calculated for nitrate and nitrite standards with imidazole buffers and a 30 cm copperized OTCR. . . . . . . 108 Effect of segmentation gas on the pH measured for the effluent from a 60 cm OTCR. . . . . . . . . . . . . . . . .111 Data from determinations of nitrate and nitrite in distilled water and seawater with OTCRs and PBCRs. . . . . 113 Linear Regression parameters for data listed in Table 4-5O O O O O O O O O O O O O O O O O O O O O O O O O O O O 114 Sample volume as a function of sample time for dye dispersion experiments. . . . . . . . . . . . . . . . . . .126 Reactor lengths used for dye dispersion experiments. . . . 126 Precision of dye dispersion experiments. . . . . . . . . . 127 Comparison of mCFA and FIA: Colorimetric chloride deterMInatiOn-SO O O O O O O O O O O O O O O O O O O O O O O13? xvi CHAPTERl HISTORICAL AND THEORETICAL INTRODUCTIW TO CONTINUOUS FLOW ANALYSIS AL Overview Until about 1975 the term continuous flow analysis was used almost exclusively to describe a technique now known as air-segmented con- tinuous flow analysis (CFA) that was invented by Leonard Skeggs [1] and developed commercially by the Technicon Corporation under the trade name. AutoAnalyzer. For a number of years an air segmented analytical stream was considered a prerequisite for continuous flow analysis. but in the mid-1970's two promising experimental approaches toward prac- tical nonsegmented continuous flow analysis were reported almost simul- taneously by Kent Stewart and coworkers in the United States [2] and E10 Hansen and Jaromir Ruzicka in Denmark [3]. Despite pronounced differences in hardware. the basic concept for nonsegmented continuous flow analysis was essentially the same for both groups. Ruzicka and Hansen named this technique flow injection analysis (FIA) by which it is commonly known today. There have been a number of fairly thorOUgh reviews of CFA [4-6] and FIA [7-10] which I shall not attempt to duplicate here. Instead I shall focus on the strengths and weaknesses of the two techniques in relation to each other. To do this. it is necessary to place the de- veIOpment and theoretical foundations of CFA and FIA into historical perspective. since there is little question that both these elements have had a profound effect on the manner in which these two techniques are currently perceived. g; Historical 1... 93 Details concerning Skeggs invention of CFA and its ensuing early history are somewhat aprocryphal and inexorably linked to the Technicon Corporation. It appears that both the increased demand for clinical laboratory services and the technological boom that followed World War II provided Skeggs with both motivation and inspiration to automate routine colorimetric determinations. He dismissed the possibility of building a machine to manipulate the test tubes and pipets used for conventional batch analysis as being overly complex and impractical [1]. Instead. he began to explore the feasibility of performing assays sequentially in a continuously flowing stream contained within a net- work of small diameter open tubes (now called a manifold). At some point in the late 1940s or early 1950s. Skeggs began to build prototype CFA systems and devised novel solutions to problems posed by the CFA approach. He found. for example. that dialysis was particularly adap- table to CFA. and he used it to separate low molecular weight analytes from blood and serum proteins that interfere in many assays. His most significant innovation. however. was unquestionably 'the bubble'. In- termixing of samples as they passed through the manifold in boxcar fashion severely limited the performance of Skeggs' early systems. He soon discovered. however. that insertion of air bubbles into the analytical stream at regular intervals reduced this mixing dramatically. Thus the bubble became and remains the 3125 pg; 323 of CFA. In the mid 1950s. the Technicon Corporation bought and patented Skeggs' CFA concept and prototype instrumentation. and in 1957 they marketed a single channel CFA system. the AutoAnalyzer. which was the first in a series of highly successful and profitable CFA instruments. It must be appreciated that virtually none of the hardware that CFA practicioners presently take for granted was available when Skeggs began his developmental work. Reliable multichannel peristaltic pumps. and more importantly. durable and inert pump tubes that could deliver relatively small volumes with good precision had to be developed and manufactured in quantity. The same is true for the multitude of specialized manifold components such as dialysis blocks. mixing coils. and fittings that now can be obtained from a number of suppliers. Technicon must be credited with developing and constantly improving the hardware and technology that transformed CFA from a laboratory curiosi- ty into a highly practical and sOphisticated technique. While a typical present-day CFA instrument is much the same as the first AutoAnalyzer with respect to the basic components and their general arrangement in the system. the hardware itself and the way in which data are acquired and processed has now evolved through three major design changes as summarized in Table 1-1. Several comments are in order here. First. CFA was develOped by Technicon primarily for clinical determinations and in general. new technology has been aimed at this market. While first generation CFA instruments were general purpose systems that were readily adapted for non-clinical .ema-m_m .mfl .eeon Luz .Eo_em-:oumcmzv .eOcu .=.Lu .: .Lommmepmzuo: A.e.n .mmmoxme .mcwmmououn mam cemuwmwzcom mums pounceou .mucoconaoo myomficme mam .mssg .pofiasmm mo cufimomou ucmmz .mcomum0wgmmm ”newcwfluuco: you xfimumewum .Epom Hoccmco OHmCMm cw zuomoccuou <2m .m-ou 30am .mucocomsoo naomwcme .massa mo :uwmomop Home: .ucoespumcw oaxuouopm m.pOmmmuum;oo: mam mmuoxm mo ~cvoe newuoncOLQ .upmcu camcwm a co umpuouou mcofiumcfiau0uom fifim pom mufismo: .muo>:o o~nm>oe mam Houoouop o~mcmm .xuuoewpofiou >9 evapoficu mam ommxoMm :ospmu MILuOE0uoca osmfim xp ESwOOm vcm ssfimmmuoa mo m:0wum:flsuouom mzoocmunsewm .cowouuw: mug: voofin mce mmoozfiu mo mCOMHmCMEuouop msoocmufinswm .mofiasmm _m0mcw~u cw coucpuw: mop: coo—n no omoonfiw mo :ofiumcmsuouom pom mocuflmon mucceeou i III l:.9.l(ll..1 I'.‘ .1.-- (l1.l(.‘.( ..1.! -.1 A O 'I'.'I 'y' I In- lll'i'lfli omH DNA u<2m .I mum“ Eoumxm :oMumpocoo thank ce-o~ H CH emsxnm=m .H-~ ofinmh applications. a trend toward specialized multichannel systems that perform standard determinations at a single work station is apparent in the second generation SMA 12/60 (sequential. multichannel analyzer. 12 channels. 60 samples per hr) and in the third generation SMAC (sequen- tial multichannel analyzer with computer) system. Second. Technicon has tended to reserve its best technology for its large. clinically- oriented customers. Although SMA technology was made available to non- clinical customers in the form of the AAII single channel instrument. state-of-the-art SMAC technology in single channel form is still unavailable. .2... EIA The origins and early history of FIA have recently been reviewed by Kent Stewart [12] and Horacio Mottola [13]. Stewart traces the FIA concept-sample injection into a continuously flowing stream with con- tinuous downstream detection-~back to James and Martinis original work on gas chromatography in 1952 [14]. Mottola sets the clock somewhat later. at Blaedel and Hicks' [15] develOpment of a continuous flow kinetics instrument in 1962. FIA in its present form. however. did not appear until the mid 1970s and evolved through two distinct pathways which have now merged for the most part into a single technique con- taining elements of both original approaches. FIA as developed by Stewart. Beecher. and Hare [2] has very obvious connections to the non- segmented. continuous flow. post column LC reaction colorimeter designed by Spackman. Stein. and Moore [16] for amino acid analysis. The carrier (reagent) stream was propelled through 0.025 cm ID teflon tubing at relatively high pressure. and samples were introduced through a multiport sampling valve. It is unfortunate that much of this group's early work is buried in rather obscure literature sources. Ruzicka and Hansenfls first apparatus on the other hand seems much more related to CFA systems. A peristaltic pump was used to propel the carrier stream through much larger (0.15 cm ID) polyethylene tubing. Samples were literally injected. at first directly through the walls of the tubing and later through a septum. with a hypodermic needle and syringe. Clearly FIA. was not invented by a single group. but Ruzicka and Hansen deserve a tremendous amount of credit for their developmen- tal efforts. They coined the generic term for nonsegmented continouous flow analysis. recognized the great potential of the technique. developed many novel applications for its use. and continue to promote it with unflagging enthusiasm. At the time FIA was being develOped. the hardware required was readily available from a number of suppliers. By the early 1970s. peristaltic and reciprocating LC pumps were for the most part perfected and a wide variety Of small diameter polymeric tubings from which FIA reactors are usually fibricated were readily available. SO too were low dead volume connectors and fittings designed primarily for use with LC systems. In addition. the absence of air segments simplified inter- facing the flow stream to a variety of detectors. Part of the great appeal of FIA. especially in small industrial and academic labora- tories. is the ready availability of hardware and the ease with which the FIA technique can be adaptd to existing equipment. In addition. FIAfs lack of association with commercial concerns (until quite recently) has led to freer and faster dissemination of information and technological advances. The sampling rates achieved with relatively simple single channel FIA.systems often surpass those of third generation CFA systems. In general. high (>200 hr-l) sampling rates are possible only when the flow rate is high and the residence time of samples in the reactor is short. ‘This was particularly true for early systems that were often wasteful of reagents and limited to very fast chemical reactions. As sample dispersion in FIA systems became better understood. however. these shortcomings have been minimized. The trend in FIA has been toward lower flow rates. smaller diameter reactors. and smaller sample volumes. Multiport valves have gained almost universal acceptance as the sample introduction method. Innovations in FIA hardware and methodology continue to be reported at a rapid pace. 9; Principles LEE; The spatial arrangement of interconnectd modules for a generalized CFA instrument is shown in Fig. 1-1A. It is common practice to aspirate samples sequentially from containers positioned on the sampler into the manifold where chemical separations and reactions are ef- fected. Usually an analyte-free wash solution contained in a reservoir on the sampler is aspirated between each sample. The sampler probeis residence time in a sample or the wash solution. t‘ or t'. respective- ly. and the flow rate of the pump tube to which it is connected deter- mine the volume of sample drawn into the manifold. Failure to control these variables rigidly. therefore. can adversely affect the precision of analytical results. Sample and wash slugs are uniformly segmented with air bubbles as they enter the manifold and are thus divided into a A) pump WASTE 2:: <3— .m\ IQ)” i \..\' —~ : l I l __. .. B B) PUMP R WASTE L l} é [ _ WASTE ——3— FC D Figure 1-1. Diagrams of generalized CFA and FIA instruments. sampler, SV I sampling valve, L = sample loop, DB I debubbler, detector, PC = flow cell, 86 = bubble gate, CR = chart recorder, air, R I reagent. A) CFA. B) FIA. >60: uu number of nominally identical subunits. The segmented stream is pro- portioned with reagents at various points along the manifold as it is propelled toward the flow- through cuvet of a recording photometer. The highly reflective air segments pose special problems for photo- metric detection in CFA. In first and second generation systems air segments are removed from the analytical stream just prior to detec- tion; in third generation systems the air bubble artifact is removed electronically. a technique known as bubble gating. In the absence of dispersion. the trace observed at the recording photometer of a CFA instrument would resemble a train of square waves. Transitions to or from some time invariant signal level proportional to the analyte concentration in the wash solution or a sample (base line or sample gggggy ;;;;53) would be instantaneous. In the trace actually observed. however. these transitions are more gradual and steady states are approached along skewed rise and {all gurves shown in Fig.1-2. Data for this figure were obtained experimentally with the miniature CF system that was developed during the course of this research and they will be discussed in more detail in paragraphs that follow. and in Chapter 2. Deviation of these recorded signals from the idealized square wave output function is quite obvious. Data sets plotted to the right and left Of the dashed vertical line in this figure were obtained with bubble gated and debubbled flow cells. respectively. Note that steady state was fully attained in the broad center peak (t‘I30 s). but that steady states were not achieved at shorter sample and wash inter- vals (t'I8s. t'I2s) which were used to generate the peaks on either side of the center peak. The sequence of peaks to the right of the center peak is known as an interaction test pattern. Here samples were 10 .mHHoo 30am wofinnzsom was coumu o~£nzn saw: Ecouum zofiw 05mm ozu pom OOUHOOOH mfimcwmm .NIH whzwwm $va? :9% 03m Oxm me c—m cm? om? cm? ca 00 on AV _.- _ _ m _ _ _ _ _ _ _ _ 0.0 _ _:ll+¥ _ fi/J QJ _ k _ _ d _ 1.. No _ 11 Go V _ ”.350 m. : .jE I: to o _ J an _ NEDU J] 0.0 w _ mm... o _ 11 ed a _ J] 50 E g _ 94pm .i we zamCE 5m: >9qu I .. 20.54%? _ i no .38 38... _ .38 33... owkdu MJmmDm OBmQmeo 11 determined in the concentration sequence of low. high. low and it is apparent that the tail of the high-concentration peak contributes significantly to the height of the second low-concentration peak. Percent interaction (‘11) is defined as the difference in absorbance between the second and first low-concentration peaks divided by the absorbance of the high-concentration peak. This quantity multiplied by 100 is the percent interaction. The extent of sample interaction depends on the magnitude and nature of dispersion experienced by samples as they pass through the flow system. Dispersion in CFA arises from two distinct processes. The first is large scale convective mixing (longitudinal dispersion) that occurs in unsegmented zones of the system--e.g.. sample lines. flow cell debubblers. flow cells. Here flow is essentially laminar and. as a result. recorded signals are deformed exponentially [17] relative to the idealized square wave output function. An empirical linear rela- tionship between concentration and time for longitudinal dispersion is given by Equation l-l. lnm E :5 cow .1. e :3 8e .1. m :3 SN 1 e :3 2: u m :3 cm 1 N .4: mm H H ">m .50 cm a A< .Eelmc. u mouoemwm pouumou commoh cofiwou .~1:«E as o.~ n ouch zofiw .hommsn Campos m.m :9 CM mop fiocozm 2: o“ u ofiasmm .uomwzn Campos m.m :2 u seekum uofluumu .nmv caucofi heuueoa mam n>mv oss~o> o~c5mm m0u00wcfi mo comuocsm m we <~u cm :OMmuommMa .m1H opzwwu Amv NEE. 35.2.... mm Omvm m. N. w 0 Nvmm0m ¢N m. N. w 0 , u u q )r I i q d d d - 0.0 1 Qmu . v .._.o . 1_.o . m ow- N .. V O 8 3 l V m“ 1 . MW nwu . on NO% / .NOMW «J l 1 . nu wH ‘m.uN1 m” 1 8 m w — 1M.O 10.0% 1 0.1 Am -vo v .Io o . m 0 [[1111 U [[Ill'llmo o.—l[[OU[[[l [[[[[[ 0.0 22 being equal. dispersion in a tightly coiled tube is about 2 to 3 times less than in a straight tube [5.35]. The extent of secondary flow increases as the coil diameter approaches the inside diameter of the tube and as the mean velocity of the carrier system increases. Note that secondary flow actually stabilizes the laminar flow regime [36] so that there is no physical basis for the notion of‘incipient turbu- lance' in coiled Open tubes. It is also possible to minimize dispersion in FIA systems by using packed rather than open tubular reactors. More care must be taken in the preparation of packed reactors. especially if small diameter parti- cles are used. and the pressure drop may be too great to allow a peris- taltic pumping system to be used. Recently. however. FIA reactors have been reported in which a small diameter tube is packed with solid glass spheres with diameters only slightly less than that of the tube [37]. The pressure drop across these so-called single head string reactors (SBSR) or pearl string reactors are only slightly greater than that of an open coiled tube of comparable diameter. and dispersion in an SBSR is 2 to 3 times less than that of tightly coiled Open tube. Also the extent of dispersion in SBSRs is less affected by changes in the flow rate then are coiled Open tubular reactors [38]. This is an obvious advantage for FIA.because reagent consumption decreases with decreasing flow rates. Furthermore. the lower dispersion obtained with SBSRs should allow longer reaction times with less dispersion. SO far we have only considered dispersive effects in the reactor. The total dispersion of the system. however. is also dependent on dispersion that occurs in the injector and detector. which is seldom negligible. Chemical kinetics. too. greatly influence FIA peak heights 23 and widths. If the FIA system is used only to transport the sample to a detector. limited dispersion is desirable so that even though the two ends of the sample plug experience mixing with the carrier. there remains a central undispersed zone. This is clearly not acceptable when it is desired to measure the end product of some reaction between the sample and the carrier stream. Here the entire sample volume must mix with the reagent and remain in the reactor for a sufficient period of time for the reaction to proceed to an appreciable extent prior to detection. Medium dispersion systems are generally used in these cases and a compromise must be reached between adequate mixing and reaction time on one hand and excessive dispersion of the sample zone on the other. If the sampling rates generally associated with FIA (90-150 hr’l) are to be maintained. maximum residence times are on the order of 30 s. With the advent of SBSRs. however. the maximum residence time might be extended to 60-90 s. if several samples can reside in the reactor at a given time with minimal interaction. At present there is no general model for FIA that allows disper- sion to be calculated 5 2512;; from experimental variables such as tube ID. flow rate. and residence time. Betteridge provides a good summary of dispersion in FIA based on the tanks in series model [39] in Table II of reference 7. It must be noted. however. that at present only trends can be predicted. This is because subtle differences in the geometries of sample injectors and flow cells. as well as differences in wall roughness in reactors and connectors can profoundly influence the total observed dispersion [40]. Also. because there is a gradient both axially and radially within the dispersed sample zone. measurement of some property within this zone will always result in a composite 24 average of concentration distributed across or along the reactor. This is entirely different from CFA where each segment is essentially homo- geneous. Nonetheless. for a given FIA system. dispersion generally is highly reproducible and therefore the precision of analytical results is good. Furthermore for a number of applications such as kinetic assays. titrations. viscosity measurements. and other techniques where a gradient can be exploited. FIA offers possibilities that are not easily achieved by CFA. CHAPTER 2 INSTRUMENTATION. DATA ACQUISITION AND PROCESSING. AND OPERATIONAL PROCEDURES A. General Considerations - CFA In Chapter 1 the origins of dispersion in CFA systems were discussed largely from a theoretical point of view. It is now worthwhile to examine limitations to this theoretical treatment that are imposed by hardware used to implement real CFA systems. Returning to Snyder's model and assuming that the dwell time. t. is fixed at 300 s and that the values of minor variables [32] are assigned as follows: 7 I 32 dynes cm-1;n I 8.9 x 10"3 poise; Dw.25 I 5 x 10-5 cm2 s'l. it is a simple matter to calculate the magnitude of axial despersion. °t' as a function of liquid flow rate. F. segmentation frequency. n. and manifold inside diameter. dt' Results of such calculations. shown in Table 2-1. reveal that for each value of dt and F. there exists a value of n (nopt) where at is minimum. Two trends are observed in this table. First. increases as dt decreases. and second for any fixed nopt value of dt' as F decreases. nopt decreases continuously. Note. however. that “t at first decreases as F decreases. but at some point further reduction in F causes at to increase. The minimum Ot value (otmin) and the associated value of nopt for each value of dt are underlined in Table 2-1. Note the steady decrease in “t that occurs as dt and F decrease. This trend is without theoretical limit. In 25 26 Table 2-1. Theoretical values of (It as a function of F and dt calculated with Snyder's model‘2 for t = 300 s. d = 0.2 cm d = 0.1 cm d = 0.05 cm d = 0.025 cm t t t t I:1 (“L 5-1) nopt O1: r_‘_op_t_ at nopt Ct _n_op_§_ or 50.0 1.31 2.94 5.0 2.84 19.0 3.08 70 3.60 25.0 1.00 2.46 4.2 2.13 16.4 2.14 61 2.38 16.7 0.90 2.29 3.7 1.87 14.9 1.77 56 1.90 8.3 0.60 2.16 2.9 1.58 12.4 1.35 48 1.33 4.2 0.45 2.15 2.2 1.44 9.80 1.10 40 0.98 1.7 0.25 2.31 1.4 1.42 6.8 0.96 30 0.74 0.8 0.15 2.53 0.9 1.48 5.0 0.93 24 0.65 0 4 0 08 2 80 0 6 1.60 3 3 0 96 17 0 62 aValues for minor variables: Y = 32 dyne cm-1 2 ; n = 8.9><10-3 poise; Dw 25 = S><10'5cm1 5-1. Underlined values indicate minimum n and at values at each value for dt' 27 practice. however. the range over which n. dt' and F can be varied is quite limited due to restrictions imposed by the pumping system and flow cell volumes. In CFA the multichannel peristaltic pumps used to aspirate samples into the manifold. and to proportion samples with air and reagents. are far from pulseless. Pump pulsations can cause nonuniform proportioning that is manifested as noise on the steady state portions of recorded signals. This problem is minimized if air segments are added in phase with pump pulsations. that is. at the time each roller leaves the platten surface. Under these conditions, however. n is restricted to a fairly narrow range of values that is determined by the pump speed and the number of pump rollers. The range of allowed segmentation frequen- cies is further restricted if bubble-through flow cells are used. because in order to measure the absorbance of individual liquid seg- ments. the flow cell volume (Vc) must be considerably less than the liquid segment volume (V‘). Flowcells with volumes of about 2 pl. (path length 8 1.0 cm. ID - 0.05 cm) are commercially available and so the minimum value of V‘ can be conservatively set at about 4 pl. (twice Vc). At any flow rate. the maximum segmentation frequency (11 ) can be max determined by dividing F by V‘. If the values for at are now calcu- lated assuming these restrictions, results shown in Table 2-2 are obtained. In this table values in parentheses are for u when noptnonss onooomocoe “use Nommom u Non .poumumaeoo -m24 n fium .po-opu:oo o>~m> wfiocwfiom mo Empumwn odomeocom flu .vonuoe o>~m> wfiocwfiom mm .vozuoe oasuxrsxhfimso A< .m:0wuwmfisa can; cu“: cofluooncw haw ucwmmcm pow mwocuoz .Hum opswwu o 1 .o 8 o. 0.0 mutawwpwwu .. n I I I. r 1 1 #N w1 1 _ e . _ > .. _ _ _ _ r I II I .66 com >~$w 8 123a Q35.”— m m 10%|.— m 1PT.. m > m 7.4 5.4 4 Am 2 34 circuit diagram for the valve controller is shown in Fig. 2-1C. This method of air injection has two advantages. First. a source of compressed gas external to the pump can be used to segment the analyti- cal stream. Thus the number of pump channels required for a given manifold can be reduced. Second. if a variable-modulus counter is added to the circuit. air segments can be added in phase with with every second. third. etc.. roller lift-off. The latter feature would be particularly useful for teaching excercises. For routine work. however. I used the dual pump tube method of air segment phasing because it was simple and reliable. Q; Detector A schematic diagram of the dual beam. fiber optic photometer designed for the miniature continuous flow analysis system is shown in Figure 2-2. Models 03000 and 03100 miniature tungsten-halogen lamps (Welch-Allyn. Skanetelles Falls. NY) were used interchangably as the light source. Both lamps have identical voltage and current require- ments (3.5 V at 0.75 Amp). but the model 03100 has a flame formed lens that approximately doubles the radiant energy impinged on the fiber optic bundle. These lamps are rated for only 20 hours of operation at 3.5 Volts. but lifetimes in excess of 150 hours were obtained when the lamps were operated at about 3a3‘volts. as was customary. An EK-15 randomised. bifurcated fiber Optic bundle (Dolan-Jenner. 'oburn. NA) made from either glass or quartz fibers was used to transmit light from the source to the front windows of the flowcells. The fiber optic-flowcell interfaces are rectangular. black delrin blocks that slide along dovetail grooves machined in the base plate Of 35 03:35 oumum vacuum u < .pouco>:oo ommuHo>lounu:OHn=o nowmflfiasa Hacoflunnomo n >\H .owOMvOuosm n on .nouamw oocopomnoucw n u .ucsos noufifim ooconomnoucfi\fifiou 30am u 2 .fifioo zofiu n Um .Ouflfim vofiwmuo>ov n a .hovH0: HHoo zofim\omuao gondw n : .oficcsn owuao honfim umpmupswfln .vonwfioucmh n Om .OUHSOm n m .Eoumxm - 6 I“\ SIGNAL 3 £52 5 ~ .1 OUTPUT 4 g, t v O’BV Figure 2-3. Schematic diagrams of detector circuits. A) Photodiode (PD) and current-to-voltage converter. B) Second stage variable-gain inverting amplifier. ICl and IC2 = LF351 FET input operational amplifiers. All resistances in ohms. Power supply bypass capacitors omitted for clarity. 39 HTP ‘7 . lOk OFFSET LOW CLIP LEVEL LO - Chip 0 2. Circuits on us indicated: @ 1108‘: ® LF398N @(7) LFSIIN © SN74L5123N © snmon 3. All PS. Bypass Cups on .0471”: Figure 2-4. Detailed schematic diagram of the bubble gate designed for the mCFA system. 40 .mszo :« moozmumfimou HH< .moumm az Hfimzq u mu~.muH .howmflfimem fimCOWHNHQQO Hmmmg u ~UH .Oumw ofinnsn ecu mo houmhnfi>wumsa o_£mumo:OE ozu nwofio vcm youuflhu ou umsohmo voflmwpoe m we Eupumwu Umumsozom .mnm ohzwmm do 00. >n.o+ ..oo. . .: e + ‘ non o a m.) .3 .w . m n 3350202 0» c x: .m > —___‘-Ju- e >mr+ & >m+ O m 1. :89”: .1de0 m 0 .— 0 .lm m C E a only ml. Go 1 v. . >9- m. c. m.-. ole fl .1. E v. . . o >m ._ o F 1 + >m+ : . ) e NB —1 _ u. m >38 50ng . n ie m >o_+ 41 characterized fully at present. II” Data Acquisition Data were recorded with a Model SRrZSS 10 in. strip chart recorder (Heath Company. Benton Harbor. MI). In many experiments. data were logged simultaneously with a microcomputer. FORTH based software rou- tines written by Eugene Ratzlaff [41]. in this laboratory. acquired data points (voltage and time of acquisition) on command from the update pulse of the bubble gate. These data were temporarily stored in RAM and later shipped to an LSI 11/23 minicomputer (Digital Equipment Corp“, Maynard. NA) for further processing. I." Data Processing Technigues L. M Reduction When data are recorded on a strip chart recorder. data reduction is a simple matter of reading and recording the steady state absorbance value Of each peak. Before analytical results can be calculated. how- ever. the baseline absorbance . which is equivalent to the reagent blank in batch colorimetry. must be subtracted from the peak absor- bance. Under usual sampling conditions (t‘)t') the signal level does not return fully tO baseline between samples. and the baseline absor- bance for each peak is estimated by linear interpolation between the initial and final baseline absorbance values. Baseline drift due to instability of electronic circuits and the light source is less than 1% per hour when the mCF system is Operated in the dual beam mode. Larger drifts. however. can result from other factors such as build up of particulate matter on flow cell windows. instability of reagents. or partial blockage of pump tubes. If the refractive index of samples is 42 significantly different from that of the wash solution. a small frac- tion of the total absorbance measured for a sample is due to refractive index effects. The contribution Of refractive index to the absorbance measured for a sample can be quantitated if the normal reagent solution is substituted with one from which the reagent necessary for chromo- phore formation has been omitted. When samples are redetermined under these conditions. the baseline corrected absorbance of the resulting peaks is due only to refractive index changes. Corrections are made by subtracting these values from the peak heights measured under standard conditions. 2; Graphical Procedures £25 Estimation 23 b 52d at- As discussed in Chapter 1. two constants. b and at. provide mea- sures of longitudinal and axial dispersion. respectively. and therefore describe for the performance of a given CFA system. Both these constants can be extracted from rise or fall curves by simple. graphi- cal procedures. The data plotted in Figure 1-2 will be used to provide working examples of these procedures. Data for this figure were Obtained experimentally with a manifold similar to the one shown sche- matically in Figure l-lA. In this system values for n. dt' F. and t were 2 s-l. 0.1 cm. .014 ml. s-l. and 85 s. respectively. Samples were aqueous phenol red solutions and the reagent was borate buffer (pHle) containing Brij-35 surfactant. The analytical stream flowed sequen- tially through two identical flow cells (1.0 cm x 0.05 on ID). The first flow cell was bubble gated‘while the second was debubbled, and data (time and voltage) from both flow cells were acquired simul- taneously with a microprocessor; Data were then shipped to the L31 11/23 minicomputer where they were converted to absorbance and then 43 smoothed and derivatixed by means Of a moving. 7-point polynomial Savitxky-Golay algorithm [42]. The smoothed data points were also regenerated according to Equation 1-3. Processed data acquired for the rise curve of the high concentration sample (the broad center peaks in Figure 1-2) with the bubble gated and debubbled flow cells are listed in Tables 2-5 and 2-6. respectively. Values of b used to regenerate these data sets were determined as described below. Inspection of Equation l-l reveals that b is simply the inverse negative slope of the linear portion of a plot of ln(A‘s-At) versus time. In Figure 2-6 smoothed absorbance values and smoothed values for the difference between A,8 and At are plotted as a function of time for both data sets. The right y-axis in this figure is logarithmic and pertains to the (A‘s-At)‘values. while the left y-axis is linear and relates to the At values. See the figure caption for further details. The lepes of the linear portions of the logarithmic plots for data Obtained with the bubble gated and debubbled flow cells were ~0Jl36 and ~0.37l. respectively. which correspond to b values of 1.36 s and 2.70 s. Note that the flow cell debubbler approximately doubled the magni~ tude of longitudinal dispersion for this CF system. It is also possible to determine b emperically by regenerating the data sets with different values for b and plotting the results on a graphics terminal. If the value chosen for b is tOO small. regenerated peaks will still appear to be exponentially deformed. Larger than Optimunlvalues for'b on the other hand result in regenerated peaks that are excessively noisy and have large spikes at the leading edge of rise and fall curves that extend above and below the sample and base line steady state absorbance levels. Optimum b values determined in 44 Table 2-5. Data for the first 10 s of the rise curve for the steady state peak shown in Figure 1-2. This data set pertains to the bubble gated flow cell. . a .*b * .fc * + time (5) At At dAt/dt At lr1(ASS-At) At norm 63.49 .0096 .0043 .0429 .0472 -.2000 .9524 64.01 .0180 .0225 .0837 .1061 -.2225 .8801 64.54 .0590 0732 .1312 .2044 -.2879 .7594 65.05 .1600 .1593 .1737 .3330 -.4099 .6015 65.53 .2771 .2726 .1999 .4725 -.5971 .4303 66.01 .3938 .3931 .2045 .5977 -.8442 .2766 66.52 .5031 .5013 .1888 .6902 -1.1341 .1630 *67.03ci .5945 .5921 .1595 .7516 -1.4658 .0877 67.54 .6640 .6629 .1253 .7882 -1.8320 .0427 68.04 .7136 .7140 .0930 .8071 -2.2164 .0195 68.55 .7490 .7490 .0655 .8146 -2.6037 .0103 69.03 .7722 .7720 .0443 .8163 -2.9757 .0082 69.50 .7863 .7869 .0291 .8160 -3.3215 -— *70.00 .7955 .7960 .0190 .8150 -3.6119 - 70.50 .8021 .8022 .0126 .8149 -3.8728 —— 70.98 .8062 .8066 .0088 .8154 -4.1105 -— 71.46 .8103 .8097 .0067 .8164 -4.3200 -— 71.99 .8116 .8123 .0054 .8176 -4.5375 -— 72.51 .8144 .8147 .0043 .8190 -4.7915 - 73.01 .8172 .8169 .0033 .8202 -5.0995 - 73.50 .8186 .8186 .0022 .8208 -5.4262 -— aAt = raw absorbance. bA: = smoothed absorbance. 2A: = regenerated absorbance. * ..* inclusive data points used to calculate regeneration factor. See text for details. 4S Table 2-6. Data for the first 10 s of the rise curve for the steady state peak shown in Figure 2-1. This data set pertains to the debubbled flow cell. All other notation is shown in Table 2-5. . * * + * + time (5) At At dAt/dt At 111(ASS-At) At norm 68.55 .0118 .0100 .0248 .0670 -.2070 .9243 69.03 .0184 .0210 .0480 .1315 -.2206 .8392 69.50 .0449 .0492 .0765 .2250 -.2564 .7170 70.00 .0968 .0976 .1050 .3390 -.3210 .5695 70.50 .1643 .1629 .1280 .4574 -.4154 4179 70.98_ .2391 .2385 .1418 .5647 -.5370 .2824 71.46 .3157 .3169 .1451 .6506 -.6810 1757 *71.99 .3944 .3931 .1391 .7129 -.8442 1003 72.51 .4644 .4627 .1270 .7548 -1.0208 .0517 73.01 .5239 .5232 .1118 .7804 -l.2046 .0240 73.50 .5733 .5735 .0960 .7942 -1.3883 0108 73.99 .6158 .6167 .0810 .8030 -1.5784 .0037 74.50 .6524 .6526 .0675 .8080 -1.7696 .0009 75.01 .6850 .6832 .0559 .8117 -1.9675 - 75.53 .7060 .7077 .0459 .8133 -2.l602 -— 76.03 .7281 .7268 .0374 .8129 -2.3413 —— 76.54 .7420 .7426 .0304 .8125 -2.5195 —- 77.02 .7550 .7564 .0250 .8139 -2.7091 —— 77.50 .7685 .7669 .0207 .8145 -2.8806 —- 77.99 .7760 .7758 .0171 .8151 -3.0534 -— *78.47 .7824 .7836 .0140 .8159 -3.2340 —— 46 .mucmom nu~m> povwam xmzuuaom ofiwcwm mu .noo~ cumsmm u A .uoch Ouaswm n m .m-~ oosmna .oou :onofiooo ucomooz no .o>~w> xpmuop whonuxmm m< .o>~w> howwum xmzuusom “one am .> [0.4L 8 mw Fhuw54‘_n:un. MW >> mw >> 1 i i 1 ya 0 D; We... ..I .... I. WILL 0 ob “v ooa\ /~o call-36 55 the Teflon tube. The same techniques were used to connect the delivery end of the pump tubes to manifold fittings (see Figures 2-10A and 2- 108). A stock of worn pump tubes were kept on hand from which sleeves were made. Plastic sleeves were also used to connect glass mixing coils to fittings on CFA manifolds. Note that the coil and fitting ends must be butted securely within the sleeve. Otherwise the integri- ty of the air segmented stream was degraded (e.g. -- air bubbles were trapped in the void volume between the coil and fitting ends) and increased dispersion due to mixing effects was noticeable. The LC fittings used in conjunction with plastic mixing coils (for both CFA and FIA) insured zero dead volume connections (see Figures 2-10C and2- 10D). A few words about sample introduction methods used for FIA and CFA are in order here. In FIA it is best to withdraw samples from their containers into the sample loop as shown in Figure 2-llA. Note that in this configuration. the sample does not pass through the pump before it reaches the sample loop. Therefore the time required to fill the sample loop is short and cross contamination of successive samples is minimized. In CFA on the other hand. samples generally pass through the pump before they enter the manifold a shown in Figure 2-11B. Because the sample moves through the pump tube under essentially laminar flow conditions. considerable mixing occurs either between successive samples. or between a sample and the wash solution. The intersample air segment (IAS) that enters the sample line when it is cycled between a sample and the wash solution (or the reverse) provides a partial barrier to intersample mixing. If additional IASs are intro- duced at the start of each sample and wash interval by rapid. S6 [I’ll/1111111111110,” I \\\\\\\ D f os \§\ammmmmm T_____JEE@ \\\\\\\\\\\\ MC s ll...""'o"fi.a.I.-.A.l . l oilife'o'ohffl.‘ O Figure 2-10. Methods for interconnecting various manifold components used for continuous flow analysis. T = teflon tubing, P = pump (not shown), I.S. inner plastic sleeve, 0.8. = outer plastic sleeve, MC = mixing coil, = reagent inlet, G = gripper fitting, C = standard HPLC connector, 8 plastic sleeve. "50" 57 A) PUMP C or R 2,—‘__> :7— C) ‘c’ O .D 3 8 o q I 0 so 67) Time (5) Preferred method of sample introduction for FIA Figure 2-11. A) systems. C = carrier stream,R = reagent, S = sample, W = to waste. = sample, B) Pecked sample introduction used for CFA systems. S W = wash solution, IAS = intersample air segment. C) Effect of IAS on dispersion that occurs in the sample lines of CFA systems. 58 repetitive withdrawal and insertion of the sample line ('pecking'). mixing within the pump tube can be greatly reduced as shown in Figure 2-11C. This figure is a generalized composite of data collected from a number of individual experiments in which no manifold was used. Instead. one end of a pump tube was connected directly to the flow cell and the other end was fitted with a 20 cm length of Teflon tubing (.025 cm ID) that served as the sample withdrawal line as shown in Figure 2- 118. The sample and wash solutions used in these experiments were alkaline phenol red and distilled water. respectively. The absorbance at 540 nm (11“ for alkaline phenol red) was recorded while the sample x line was manually cycled between the two solutions at 60 s intervals. As previously mentioned. one IAS forms naturally when the sample line is transferred from one solution to the other. The IAS can be e1 imi- nated. however. if the pump is turned off during this Operation. Ad- ditional IASs can be introduced by pecking. In this way the amount of dispersion that occurred in the sample line and pump tube was monitored as a function of the number of IASs introduced at the leading edge of each sample and wash slug at several different flow rates. Results for one set Of experiments where the flow rate was 0.32 mL min"1 are shown in Figure 2-llC. With no IAS. dispersion in the sample line and pump tube was large. With one IAS. dispersion was reduced. but still significant. With three IASs. however. dispersion was almost eliminated. As expected. in the absence of an IAS. dispersion decreased as the flow rate increased. When three IASs were introduced at the leading edge of each sample and wash slug. however. dependence of dispersion on flow rate was minor. These experiments underscore the 59 importance of pecked sampling for minimum dispersion in CFA experi- ments . CHAPTER 3 DESIGN AND CHARACTERIZATION OF AN ELECTRONIC BUBBLE GATE FOR AIRrSEGMENTED CONTINUOUS FLOW ANALYSES A; Overview As already mentioned. photometric detection with CFA systems is complicated by the presence of air segments in the analytical stream. In commercially available first and second generation systems. the analytical stream is debubbled just prior to detection. Unfortunately. this expedient allows previously segregated liquid segments to mix in the debubbler and flow cell. and therefore the rate at which analyses can be performed is reduced. As shown in Chapter 2. mixing effects that occur in flow cell debubblers and flow cells can be removed (after the fact) by analog [45.46] or digital [47] curve regeneration tech- niques. or they can be eliminated by using bubble-through flow cells and a gated detector (bubble gating) as is done with the Technicon third generation SMAC clinical analyzer. This chapter deals exclusive- 1y with the latter approach. Habig and co-workers [27] appear to have develOped the first bubble gate. which was activated by conductance changes within a specially designed flow cell. They concluded. however. that their design was preliminary and was not suitable for routine use. Sub- sequently a more robust electronic bubble gate was described by Neeley. 6O 61 et a1. [28].'The same group later reported a slightly modified version of this circuit [48] which they incorporated into a high performance colorimeter for a miniaturized CF analyzer. This bubble gate repeti- tively sampled and stored the detector signal at a frequency determined by an adjustable internal time base. Then. for a fixed time interval. the stored signal level was compared with the real-time signal level by means of a window comparator circuit. When the stored and real-time signal levels were within the limits Of the window comparator. the real-time signal level was stored in a second sample-and-hold circuit connected to the readout device. The bubble gate reported here is much simpler. It uses the periodic fluctuations of the detector signal. caused by the successive passage of air and liquid segments through the flow cell. to syn- chronize the time Of data acquisition and temporary storage with the brief interval during which each liquid segment completely fills the flow cell. Data are presented to demonstrate that improved performance of both commercially available and custom built CF analyzers can be achieved using the bubble gate described. The analytical performance Of a miniature CF (mCF) system with 0.1 cm ID manifold components and a bubble gated detector used for the colorimetric determination of nitrite and silicate in water is presented. Results of nitrite deter- minations Obtained with mCF system are compared with those reported for flow injection analysis. 62 13... W L _C_F Systems The filter photometer used with the mCF system is described fully in Chapter 2. The 0.2 cm ID manifold CF (AAII CF) system consisted Of a Pump III (with air bar). and an Industrial S.C. colorimeter equipped with 1.5 cm x .15 cm ID debubbling flow cells and 540 nm interference filters (Technicon Instruments Inc.. Tarrytown. NY). In bubble gating experiments with the AAII CF system. the Technicon colorimeter was modified by removing capacitors C-201. C-203. and C-204 from the 'control module' circuit board. This modification reduced the rise time of the detector signal from about 0.3 s to about 10 ms. Also. the sample side debubbling flow cell was removed and replaced with a 1.5 cm x .1 cm ID bubble-through flow cell (Gamma Enterprises. Inc.. NF. Vernon. NY). The cell was mounted directly onto the phototube housing with an adaptor to occlude the phototube entrance and exclude stray light. The colorimeter was then adjusted in the usual manner. The 'telemetry output' (0-5 V) Of the colorimeter was connected to the bubble gate described below through a unity gain inverting amplifier with offset. This made the logarithmic output of the calorimeter compatible with the logic of the bubble gate which is described in detail below. Manifold components for the AAII CF system were obtained com- mercially from standard sources. while those for the mCF system were for the most part custom made as described in Chapter 2. Technicon SMA Flow-Rated Pump tubes were used for both systems. Data were recorded with a Model SR-255 10 inch strip chart recor- der (Heath CO.. Benton Harbor. MI). In some experiments data were simultaneously logged with a microcomputer (see Chapter 2). 2‘ Determination pf Nigrite ppg Silipate a. Nitrite Reagents The sulfanilamide (SAN) reagent was prepared by dissolving 10 g SAN in 500 mL of deionized. distilled water (DDW) and 100 mL of concen- trated HCl contained in a l L volumetric flask. This solution was diluted to the mark with DDW. mixed. and transferred tO an amber bottle. Then 0.5 mL Brij-35 wetting agent was added. The N-(l-Naphthyl) ethylenediamine dihydrochloride (NED) reagent was prepared by dissolving 1.0 g NED in l L of DDW. This solution was transferred to an amber bottle. and 0.5 ml Brij-35 was added. b. Nitrite Standards The primary nitrite standard was prepared by dissolving 0.345 g (5 mmod) of dried. analytical reagent grade sodium nitrite in.l L DDW. Working standards in two ranges were prepared by dilution Of the pri- mary standard using digital microliter pipets and volumetric flasks. c. Silicate Reagents The molybdate reagent was prepared by dissolving 10.8 g of am- monium molybdate and.2.8 mL of concentrated sulfuric acid in about 500 mL of DDW contained in a l L volumetric flask. This solution was diluted to the mark with DDW. mixed. and transferred to a plastic bottle. Then 2 ml.of an aqueous sodium lauryl sulfate solution was added. The tartaric acid reagent was prepared by dissolving 100 g of tartaric acid in 950 mL of DDW contained in a l L plastic bottle. The stock stannous chloride solution was prepared by dissolving 10 g of stannous chloride in 20 mL of 50% (V/V) hydrochloric acid. 64 Heating is sometimes required to effect complete dissolution. This solution was stored in a plastic bottle. The working stannous chloride reagent was prepared by adding 0.5 mL Of the stock reagent to 50 mL of 1.2 N HCl solution contained in a plastic bottle. This solution is not stable for more than about 8 hours. The wetting agent was prepared by dissolving 10 g Of sodium lauryl sulfate in 100 mL of DDW. d. Silicate Standards The primary silicate standard was prepared by fusing 0.3007 g (5 mmol) of dried silicon dioxide (Alfa-Ventron. 99.9%) with 0.7 g of sodium carbonate in a platinum crucible. Heating was effected with a compressed air-natural gas flame. After the melt formed a glass (:5 min). heating was discontinued and the crucible was allowed to cool. When cooling was complete. DDW was added to the crucible which was then covered and allowed to stand overnight. The next day the contents of the crucible were transferred quantitatively to a l L volumetric flask. This solution was diluted to the mark with DDW and transferred to a plastic bottle. Working standards in the range of 5 to 45 pH were prepared by dilution of the primary standard by means of digital microliter pipets and 100 mL volumetric flasks. Working standards were transferred to plastic bottles immediately following their preparation to avoid the possibility of silica in the glass volumetric flasks being leached into the standards. 65 3; Bubble Gate When bubble-through flow cells are used with colorimetric. air~ segmented CF analyzers. the detector signal closely resembles a square wave. With an air-segmented blank solution the detector output varies between the 100% T level (5.0 V in our case) and the 074 T level (0.0 V). as shown in Figure 3-1. As an air segment traverses the flow cell. it reflects most of the light away from the photodetector. and the signal approaches the 0* T level (0.1 to 054 V in our case). The actual magitude of the signal observed in the presence of an air segment is somewhat erratic. and it decreases as the transmittance of the two adjoining liquid segments decreases. When the air segment exits the flow cell. the detector rapidly rises to a level that cor- responds tO the transmittance of the liquid segment (see Figure 3-1). The bubble gate described here uses the information encoded in the periodic fluctuations of the detector signal to synchronize the time of data acquisition and temporary storage with the brief interval during which each liquid segment completely fills the flow cell. The schematic diagrams presented in Figure 3-2 depict the func- tional units of the bubble gate circuit. In the basic bubble gate circuit (Figure 3-2A) comparator ICl converts the detector signal to a TTL logic level signal that indicates the presence or absence Of an air segment in the flow cell. This is accomplished by setting the threshold level of the comparator at about 0.5 V. A HI logic level at the output of the comparator indicates the absence of an air segment in the flow cell. Note. however. that the logic level transitions of the comparator lag the entrance and exit Of air segments into and out of the flow cell by several ms because of the time required for the 66 V0 LTS () 4_:-- TIME Figure 3-1. Fluctuations in the detector signal caused by the successive passage of liquid and air segments through the flow cell: Position 1, air segment completely within cell. Position 2, air segment exiting cell. Position 3, cell completely filled by liquid segment. Position 4, air segment entering cell. I0 67 ' k I m ‘( I000 ' LEDZ 1:471: ——I_. 00 00 .. . .4 > =m- @09E900 . k ,. > IOk ”122.40 rcz-e © an +mv I C) 1’59 0 ..— Okuvsarro 11’" v@ LEDI To .I- ggocn: »_. ': PW. 2 C H Log new“; "I S/H Our S/H Out 4 39k ‘ 1C3 R24 1c: ’ S/H LOG A 3 From b© L A h S/H 00' Thread! LeveI Fnure 20 Frggc© FM 20 A) Detector Snnol OQI In K 19' IM . p v TO '1' - Figure 2A 0) Figure 3-2. Generalized schematic diagrams of bubble gate circuitry. All resistances in ohms, all capacitances in microfarads. A) Basic bubble gate, ICl = LF311 FET input comparator, IC2, 74L8123 dual monostable multivibrator, IC3 = LF398 sample-and-hold amplifier, IC4 = AD755 logarithmic amplifier, IC5 = TL084 quad FET input Operational amplifier. B) Differential edge sensor, IC5 = TL084 quad FET input operational amplifier, 1C6 = LF311 FET input comparator. C) Sample-and-hold amplifier update logic, 1C7 = 7410 triple 3-input NAND gate. D) Automatic threshold adjustment for comparator, ICl. IC8 = TL084 quad FET input operational amplifier. 68 detector signal level to cross the threshold level of the comparator. The rising edge of the comparator output triggers monostable multi- vibrator IC2-A. which generates a time delay pulse. The duration Of this pulse is adjusted with potentiometer P1 to terminate at the approximate midpoint of the interval during which a liquid segment completely fills the flow cell. Two light emitting diodes (LEDl and LED2) facilitate this adjustment by giving‘visual indication of the logic states of the comparator and monostable. respectively. The falling edge of the time delay pulse triggers monostable IC2-B which generates a pulse with a fixed width Of approximately 100 us. A gated version of this 100 pa pulse is used to update sample-and-hold ampli- fier IC3 which samples on a TTL HI and holds on a TTL LO. Irregularities in the segmentation pattern of the analytical stream. generated when the sample probe is cycled between samples and the wash solution. can produce two conditions which cause the basic bubble gate to malfunction. IFirst. irregular liquid segments that completely fill the flow cell for a time interval much shorter than the selected time delay interval cause the comparator and the monostable to get out of phase; several liquid segments Of normal length must pass through the flow cell before these two circuit elements are again synchronous. Erroneous sample-and-hold amplifier updates could occur in the interim. 'This problem was corrected by clearing monostable IC2- A with the falling edge of the comparator output so that the time delay pulse is aborted in the event that it is still in progress when an air segment enters the flow cell. Second. irregular liquid segments that completely fill the flow cell for a time interval approximately equal to the duration of the delay pulse can cause incorrect sample-and-hold 69 amplifier updates because of the lag between the initial entry of an air segment into the flow cell and the logic level transition of the comparator output as discussed previously. This problem was eliminated with the differential edge sensor circuit shown in Figure 3-2B. which generates a TTL HI at its output whenever the detector signal has a non-zero derivative. Because the output of this circuit responds to changes in the detector signal level almost instantaneously. it can be used to eliminate the possibility of sample-and—hold amplifier updates at the time of initial entrance or exit of an air segment into or out of the flow cell. Three conditions must be sastisfied in order to update the sample-and-hold amplifier: 1) The output of the monostable IC2-B must be HI. 2) The output of the differential edge sensor must be LO. 3) The ouput of comparator 101 must be HI. Figure 3-2C shows the logic that prevents sample-and-hold amplifier up- dates when the above conditions are not satisfied. This bubble-gate circuit will not operate when the signal level of a liquid segment is less than the threshold level of comparator 101. If the threshold level Of the comparator were to be fixed at 0.5 V. liquid segments with transmittance less than about 0.1 would be indistinguishable from air segments. and the bubble gate would be restricted to an operational range Of 100% T to 10% T. However. because an air segment's effective transmittance is not constant but decreases as the transmittance of the two liquid segments adjoining it decreases. it is possible to extend the lower Operational limit of the bubble gate considerably by using the detector signal level stored in the sample-and-hold amplifier for automatic adjustment of the 70 comparator threshold to an appropriate level. This function is per- formed by the circuit shown in Figure 3-2D. A buffered voltage divider applies one-half of the sample-and-hold amplifier output to the threshold level input of comparator 101 through a unity gain buffer (IC8-D). Two active diode clippers limit the maximum and minimum threshold levels to about 2 V (fixed) and 0.1 V to 0.5 V (adjustable). respectively. This circuit extends the bubble gate's lower limit of operation to about 3% T (1.5 A). and in addition improves performance in the high transmittance range because it provides greater separation between the comparator threshold level and the somewhat erratic detector signal produced by the passage of air segments through the flow cell. A logarithmic amplifier (Model AD755. Analog Devices. Norwood. MA) is also included in the bubble gate circuitry so that peak heights are linear functions of concentration. A detailed schematic diagram of the bubble gate can be found in Chapter 2. It should also be noted that bubble-gating can be accomplished with a microcomputer and apprOpriate software. Apparently this approach was used by the designers of the Technicon SMAC clinical CF system. Q; Resulgs ppd Dispuggion L Introduptipn The bubble gate was tested with the Technicon AutoAnalyzerII CF System (AAII CF System) and the mCF system described in Chapter 2. The colorimetric determinaion of nitrite via a modified Griess reaction 71 procedure [49] was chosen as the reference assay with which the performance of the two CF systems could be evaluated and compared. A composite manifold diagram for both CF systems is shown in Figure 3-3. Pecked sampling did not improve the performance of either the standard or the bubble-gated AAII CF system noticeably. As described in Chapter 2. however. it did improve the performance Of the mCF system. There- fore. pecked sampling was used routinely with the mCF system. For both systems the time required to reach steady state and the percent inter- action were determined as described below. g_,_ Steggy State ggperimgngs A set of five aqueous nitrite standards with nominal concentra- tions Of 5 11!. 15 (1!. 25 11!. 35 31!. and 45 II! were prepared. and the detector gain was adjusted such that these standards produced a response of approximately 10%. 30%. 50%. 70%. and 90% full scale. respectively. on the strip-chart recorder. With the AAII system each standard was sampled in triplicate for 5 s. 10 s. 20 s. 30 s. 40 s. 50 s. and 60 s with a 60 a wash between each sample to minimize carryover. With the mCF system each standard was sampled in triplicate for 5 s. 10 s. 15 s. 20 s. 25 s. 30 s. and 60 s with a 30 s wash between each sample. The average peak height for each sampling interval was divided by the steady-state peak height (60 s sampling interval) and multiplied by one hundred to obtain the per cent of steady state (%SS) as a function of sample time. .A grand average of %SS for all five standards at each sampling interval was then calculated. The results of this calculation showed that %SS was independent of concentration to within about 1% for the 5 s and 10 a sampling intervals and to within about (L5% for sampling intevals greater than 10 s. The increased scatter in 72 NOMINAL FLOW RATE [mL/min] AIR [Fl—(JO) 1 , WASTE 54C) 32(05) 4.? a _-_ flit. E31,? I5mmxlmmID SAMPLE .3206) 1 .4” a _0 0000 _»00mm05mmm) edgngfiiflfl} iflfiflfiw.. . a 9:41 :DILUEN_ l..80(.23 .4182 SAN 10(05) NED ,|o(,05) I5 mm x l.5mmlD "°° 9 WASTE PUMP Figure 3-3. Composite diagram of various CF manifolds used for nitrite determinations. Values outside or within parentheses refer to AAII or mCF manifolds, respectively. 73 the peak heights at shorter sampling intervals is probably a result Of the manual sampling employed. Results of the %SS experiments are presented graphically in Figure 3-44 ‘Values for major experimental variables can be found in Table 3-1 as well as the estimated dispersion. at. of the air segmented sample slug during its passage through the CF systean The latter quantity was calculated according to the model of Snyder and Adler [25.20]. Values of minor variables shown in Table 3-1 were in keeping with a similar set of calculations by Snyder [32]. As shown in Fig. 3-4. bubble gating reduced the sampling time required for the AAII CF system to reach 98 %SS by about 8 s. A similar reduction in the sampling time required for the mCF system to reach 98 %SS was achieved by using the peeked sampling techique. An apparent trend for the mCF system also shown in this figure is a decrease in the sampling time required to attain a given.%SS as the segmentation frequency. n. increases. Note. however. that with the mCF system. n was increased by increasing the pump speed (and therefore the liqiud flow rate. F) in order to keep air segmentation in phase with the roller lift off of the peristaltic pump. as discussed in Chapter 2. Thus the residence time. t. for the sample slug in the manifold decreased as n and F increased (see Table 3-1). To differentiate between the effects of n and F. the experiment was repeated for n . 1.5 .‘1. a - 0.0067 .1. .‘1 and n - 3.0 s". F - 0.014 mL .‘1 with t held constant at 110 s by altering the length and number of mixing coils in the manifold. As shown.in Figure 3-5. the same trend was observed. These results were were contrary to the predictions of Snyder's model [10] (see Table 3-1). This was not surprising. because 74 1C)O '-F' __ ._____ ___,_.__. - ‘.— .1- .__ 6 5 m 9‘) —-_t 4 o .__ on >\ BI) --» 3 '0 o _— 0 w 7 0 -- }; _._ o 2 5(3 ..L. 2 O) 0. 5() -u-' 1 I l l I I I I I 1 fl] 0 5 1 0 1'5 2(3 255 3() Sonprng Intervol (s) Figure 3-4. Results of percent steady state (%SS) experiments: 1, standard AAII CF system. 2, bubble gated AAII CF system. 3, bubble gated mCF system without pecked sampling. 4-6, bubble gated mCF system with pecked sampling and n equal to 1.5, 2.0, and 3.0 5'1, respectively. 7S .xHo>wuuommou .No.~ mam mm.~ one: mucwuuom Honucou conumpnwfimu wumvcmam .mEoumzm mu HH<< voumwumnnnan mam vumvcmum ecu mom .Hum «EU mIoHoAm n mmN.3oU acumowmmooo scamsmmflm o>wuoommo H IEU ocxv HoHoA~.m u nxv ”coflmcou oommHSm .Omfiom a IoHoAm.w u may xufimoomw> "moHnmflum> momma I. H m u no.H 96.6 Nw.o em.o wo.H wo.H mH.~ I. . Hog oHao oo oHH cHH mo ooH omH omH com com H83 0 mHo.o aooo.o mHo.o mwoo.o aooo.o aooo.o mmo.o -o.o HH-m Hey an H.o H.o H.o H.o H.o H.o ~.o N.o H563 no o.m m.H c.m o.N m.H m.H m.o m.o HH-mU = wcwuuom Houumoo 4w me am om N4 NV .I I. 6666. mesa . . . . . . . .. . .. ooeoonomoo mo mo we we mo mo oon mom. ooHaomHHE . . . . . . . . H563 HomeoH o H o H o H o H o H o H m H m H good HHoo 36H“ wcHHooa gon wcHHooa oooao oHoosm onaocoom oHoaHno> unoguwz Hmucoawuomxm zmem>m mos zmem>m mu HH<< .mucosfinonxo mww HomNomofinmwpm> Hana: .HIM canes 76 1C)O -——- 0> _._J -I- C) U) 955 -*- x ~— 13 C3 __ .3 __ 2 (f) 4, -_ C: 9(3 -- 0> 0 HI. k. 0) _- O. _ 1 I l l .gj 855 I I I I I I O 5 10 15 20 25 30 Sampling Intervol (s) Figure 3-5. Effect of segmentation frequency and liquid flow rate on %SS with the residence time of the sample slug in the mCF system held constant at 110 s: l, n = 1.5 s’ , F = 0.0067 mL 5’1. 2, n = 3.0 5‘1 F = 0.015 mL s-l. 77 Snyder’s model assumes that mixing effects in unsegmented zones of the CF system are negligible. As discussed in Chapter 2. this assumption is weak even for CF systems with bubble-through flow cells. Other factors being equal. mixing effects decreae as the flow rate increases. At standard pump speed. approximately 20 seconds elapse between.the time of sampling and the time at which the leading edge of the sample slug is segmented with air. Doubling the pump speed halves this inter- val. As demonstrated earlier. pecked sampling and bubble gating reduce mixing effects. but clearly do not eliminate them. 'This set of experi- ments suggests that the decreased dispersion of the sample slug observed when the pump speed was doubled resulted primarily from a decrease in mixing effects and that the concomitant doubling Of the segmentation frequency had little effect. 3‘ Sppplp Interaction Percent interaction (%I) was determined by the method of Thiers. _L _l. [50]. Nitrite standards in the concentration sequence Of 5 pg. 45 11!. 5 a! were used throughout the experiments. Sampling intervals for the AAII and mCF systems were fixed at 30 s and 25 s. respectively. and %I was determined as a function of wash interval. When the second low standard produced a shoulder peak. the peak height just prior to the fall was recorded. Results Of these experiments can be found in Table 3-2. On the AAII CF system with and without bubble gating. 1.0 %I was achieved using 10 s and 15 a wash times. respectively. On the mCF system a 5 a wash time resulted in 1.0 %I and 0.5 %I at standard and twice standard pump speeds. respectively. The decrease in %I at higher flow rates is again presumed to result primarily from decreased mixing effects as discussed previously. 78 an em ma OH m I I I H.o Nd H.ov H80 H8 :3 m5 :3 NH. NS 3. we Nd 4.6 33 as H..H .2. no oH .2 5H 3 gum H-m mHuc H838 6:55 a 5.5»... "as 55$ no :5 .cofiuomnoucw vacuums nmuHm>Poucm awe: .Nnm O~DMH 79 L Iixipg Efgegts It is apparent from the foregoing experiments that mixing effects contribute significantly to the Observed loss of wash in both of the CF systems used for this study. I was able to minimize mixing effects in the mCF system by using standard pump tubes with nominal flow rates approximately half that Of the desired flow rate. and then Operating the pump at about twice standard speed. Under these conditions analy- sis rates which approximate theoretical limits assuming Snyder's 8 0t criteria [10]. were achieved as is demonstrated in the final section of this chapter. A better approach would be to use a miniaturized pump fitted with shorter pump tubes. It is also possible to remove the effects of longitudinal dispersion by means of curve regeneration as described in Chapter 2. 1L Analytical Resulgs 1,, Nitrige Determinations Aqueous nitrite standards were determined with the mCF analyzer with the manifold shown in Figure 3-3. Note that the dilution pump tube and first mixing coil (shaded area of Figure 3-3) were omitted to increase sensitivity. A nitrite concentration range of 2 n! to 18 u! was chosen to allow direct comparison of the mCF procedure with a recent flow injection analysis (FIA) procedure for the determination of nitrite [51]. The FIA procedure used the merging zones technique to minimize reagent consumption and a novel 'intermittant flow' technique to increase the sampling rate. Performance of the mCF system at several sampling rates in terms Of linearity of standard curves. %I. and the precision Of five replicate determinations of the 6 .1! standard can be 80 found in Table 3-3. Recordings of data Obtained at analysis rates of 360 hr.1 and 120 hr'1 are shown in Figure 3-6. Also listed in Table 3- 3 are experimental conditions and performance criteria reported for the FIA.procedure. Inspection of Table 3-3 reveals that even at a sampling rate Of 360 hr’l. where precision was limited because of manual sampling. the performance of the mCF system is comparable to that Of the FIA system which had a maximum sampling rate Of about 70 hr’l. Note also that on a per assay basis. the sample and reagent requirements of the mCF system were less than those of the FIA system. These results conflict with the usual claims made about the performance of FIA relative to CFA. ‘The sampling rate of about 70 hr"1 reported.by ZagattO. et a1. [12] is somewhat superior to sampling rates generally achieved with second generation CFA systems. but is decidedly inferior to sampling rates achieved with the mCFA system. Furthermore. the FIA system required an elaborate system of valves to inject both samples and the reagent into the carrier stream. and to vary the flow rate of the carrier stream during the course Of each determination. In this respect the mCFA system is much simpler than the FIA system with the exception of the gated detector. ‘2‘ Silicate Determination; Aqueous silicate standards were determined with the manifold shown schematically in Figure 3-7. All mixing coils*were made from 041 cm ID Micro-line tubing. The lengths Of tubing used to make the first. second. and third mixing coils were approximately 250 cm. 100 cm. and 50 cm. respectively. These provided about 230 s. 70 s. and 30 a reaction times after the addition of the molybdate. tartaric acid. and 81 .Hm>nou:w 5mm: mam oHnEmm one now confiscop oEHu on» wcfinsm vomo>wHov mucoumou omz mam z vocwneoua .mucoswhomxo :m pom m 2: up can AaIm m.Nu5 on uvoonm 9556 some oz oo.H oo.H oo.H oo.H oo.H oo.H H.Hmooo .Honuoov zxo some oz wooo.o omoo.o Hooo.o wmoo.o mmoo.o «moo.o oaooooocH > some oz mmo.o mmo.o mmo.o mmo.o mmo.o mmo.o ooon H m~.o 4H.o am.o o~.o om.o mm.o H4 m.o m~.o 4N.o HH.o mH.o wm.o 44.o Hamzau :onHoooo omH mm mm om om N4 mm Han HoesHo> snowmom com wNH om om 4o 46 mm Han oasHo> oHosam on oNH ONH owH owH cam com HHInzv oooz mHmzHoa< II oH mH m CH m m HmVoeHP can: II om mH mH oH 0H m HmVoEHP oHosam 5... 6 <5... II] 111]] 1- 15 .m:0wpmcwsnouov Omaha“: OHPuoEwHoHoo pom 137 137 172 206 Precision (%RSD) 1.1 0.8 0.4 0.4 Percent Interaction 0.07 0.90 0.97 0.16 $10pe .010 .0105 .011 .011 Y Intercept 0.000 0.000 0.000 0.000 rxy (correl. coeff.) 1.00 1.00 1.00 1.00 aPump speed = 56 (n = 2.0 5.1) and t = 300 s for all experiments. Combined volume of molybdate, tartaric acid, and stannous chloride reagents delivered during the time required for one sample and wash interval. 86 Absorbonce j :I 04.: (1 .1 '1U[1 {1 ..LCIIII. .. 1,077qu I 1 1 I . I l 0 90 ISO 270 360 450 540 630 Time (s) 0.5 -P .._- II I q) .. U co.-- I I C .D .I. L 8 02-- I I ..D <( .. 0.1 4- " IIIII III I 0.0 , 4 4 E 4 I j 4 4 I 0: . i 4 4 , . 9 ; * s~1 0 60 120 180 240 300 360 420 Thne (s) Figure 3- 8. Recording of 5,15,25,35 and 45 uM silicate standards run in ascending and descending order followed by an interaction test pattern and replicate determinations of the 15 uM standard. A) 120 samples hr 1. B) 180 samples hr 1 87 assembled and Operated with moderate ease, and that such CFA systems need not be more elaborate or expensive than comparable FIA systems. In fact. the combined cost of electronic components for the bubble gate and a 2 9L bubble-through flow cell (:3500) required for the mCFA system is comparable to the cost of sampling valves and valve control- lers required.for FIA systems. vhile the cost of other major system components (samplers. pumps. detectors) is about the same for either technique. Hopefully the results reported here will encourage researchers in academic and small industrial laboratories to think seriously about custom designed mCFA systems. Such systems should be particularly advantageous for the automation of routine colorimetric determinations in vhich the analytical reaction requires more than about 30 s to reach an appreciable degree of completion, especially mhen lov dispersion of samples is required to achieve maximum sensiti- vity. For an experimental comparison of FIA and mCFA. see Chapter 5. W4 NOVEL CADNIUH REACTORS FOR DETERMINATIW OF NITRATE IN 'ATER AND SEAWATER BY SEGMENTED. CONTINUOUS FLOW COLORIMETRY A; Overview Nitrite in water and seawater is routinely determined colorimetri- cally with high specificity. sensitivity. and precision by diasotixa- tion with sulfanilamide and coupling with N—(l-Napthhyl)-ethylene- diamine as described in Chapter 3. An equally simple and sensitive colorimetric assay for nitrate has yet to be devised. and for this reason nitrate is generally determined as nitrite after it is reduced to that species with a suitable reagent. Although procedures for reduction of nitrate to nitrite with zinc [54]. hydrazine/Cu2+ [$5]. and immobilized nitrate reductase [56] have been reported. they have not been widely applied. This is because the reduction with zinc is too vigorous at ambient temperatures. reduction withhydraxine/Cu2+ is sluggish and difficult to control. and enzymatic reduction. although highly specific. requires reagents that are relatively expensive and difficult to prepare. Cadmium metal has proved to be a much more suitable reagent for reducing nitrate to nitrite. and is generally used in the form of packed bed reactors. although reactors fabricated from cadmium wire inserted into narrow plastic tubes [57.58] have also been reported. 88 89 unfortunately, packed bed and tubular wire cadmium rmactors are incompatible with segmented.Icontinuous flow analyzers because the analytical stream must be debubbled prior to entry into these types of reactors. Air segments cause channeling in the packed bed reactors and as a result, reduction efficiency decreases and sample dispersion increases. Tubular wire reactors will not support a stable segmenta- tion pattern and so generally the analytical stream is segmented on the downstream side of such reactors. By now it should be obvious that other factors being equal, dispersion will be greater for a system in which air segments must be removed and reintroduced downstream, than for a system where segmentation is uniform and continuous. This means that in determinations incorporating packed bed or wire reactors. the sample and wash intervals required to achieve acceptable levels of precision and interaction will increase. which of course decreases the rate at which determinations can be performed. Furthermore. the flow characteristics of packed bed reactors can change during the course of a run due to accumulation of particulate matter. which will change the back pressure. and therefore the flow rate through the reactor. Reactors fabricated from cadmium tubing should eliminate these problems. They would support a stable segmentation pattern in a similar fashion to other mixing coils that make up the manifold. the difference of course being that the inner walls of the cadmium tube are active rather than passive. 5‘ General Consideratiggg Reduction of nitrate to nitrite is generally 90 to 98% complete in packed bed cadmium reactors (PBCRs) under neutral'to mildly alkaline 90 conditions. The dimensions of the reactor, as veil as the size distri- bution of the cadmium particles with which it is packed. will have a major influence on reduction efficiency. Larger reactors and smaller cadmium particles lead to higher reducing power. It must be stressed. however, that the active surface area of cadmium particles with similar size distributions can be quite variable and may change during the course of the reduction. For example Davidson and 'oof [59] reported that different forms of cadmium -- e.g.. spongy. electrolytically precipitated. filings, powder -- had different reducing powers on a weight-to-weight basis. These conclusions are somewhat suspect because they did not consider differences in the surface areas of the different forms of cadmium. nor did they account for changes in pl! that are likely to occur during the course of the reduction due to reaction of dissolved oxygen with cadmium. Certainly one of the major difficulties in achieving good precision with cadmium reactors is the maintenance of a reproducibly active surface during the course of experiments. There are several ways to increase the active surface area of cad- mium particles. For example, if cadmium filings are reacted with nitric acid prior to packing. their reducing power is increased [60]. This is because nitric acid pits the cadmium and thus the surface area is increased. It is also well established [59,61] that amalgamated. silverised. or capperised cadmium is more reactive than pure cadmium. Amalgamated cadmium tends to form lumps [61.62] and is therefore not well suited for the preparation of PBCRs. Copperized cadmium on the other hand. does not lump and is highly reactive. For this reason it has been widely used for the preparation of PBCRs. Several workers 2+ [62,63] have suggested that Cu is a product of nitrate reduction in 91 such reactors. Recently. however. Sherwood and Johnson [64] presented convincing evidence that copper is not directly involved in the reduction process; instead it functions primarily as a surface activa- tor. Other variables such as pH and the amount of dissolved oxygen and chloride in samples and reagents also affect the reduction efficiency of a given cadmium reactor. These effects may go unnoticed in rela- tively large (25 cm x 0.8 cm ID) PBCRs used for batch reduction of nitrate to nitrite in natural water samples [62]. because of the great excess of cadmium relative to the amount of nitrate present in samples. As PBCRs are miniaturized to make them compatible with continuous flow analyzers [65.49]. however. control of reaction parameters becomes much more critical. and changes in the active surface of the cadmium that can occur during the course of the reduction become more apparent. Nydahl [61] reported a detailed study on the cptimum conditions for the reduction of nitrate to nitrite with PBCRs. This study was highly useful in my work with open tubular cadmium reactors (OTCRs). I will therefore summarize his results in the paragraphs that follow to provide some background information necessary to understand the results with OTCRs presented later. 92 _C_,_ Summary; mghahl's !Qrk L pl_i Effects Equations 4-1. 4-2. and 4-3 [61] suggest that no; + 211* + 2e'—>N02’ + 320 (4—1) we; + 611* + 4.’—>33Non* + 1120 (4-2) no; + 311* + 5.'—->Nn4+ + 2n20 (4-3) the pi! at which the reduction occurs has a major influence on reduction products. Nydahl [61] found that in the pH range of l to 3. hydroxyl- amine is the major product. Small amounts of ammonia are also formed. Reduction rates decrease sharply in the pH range of 3 to 5. and nitrite is the major reduction product. The rate of nitrate reduction con- tinues to decrease in the pH range of 7 to 9. and 95 to 994% of nitrate originally present in samples is recovered as nitrite. The fraction depends on the flow rate through the PBCR and its size. As the pi! increases in this range. either the size of the PBCR must be increased. or the flow rate of sample through it must be decreased to achieve the same degree of reduction. The rate of reduction of nitrite to hydroxylamine also decreases as the pl! increases. but even at pl! 9.5. slight reduction of nitrite still occurs. Nydahl reported the same trends for copperized cadmium but found that the rate of reduction observed for both nitrate and nitrite was a factor of 10 to 20 times greater for copperized cadmium than for pure cadmium. These results emphasize the importance of pH control during the reduction process for all types of cadmium reactors. The presence of dissolved oxygen in 93 samples and reagents makes pH control more difficult than might be expected. 2‘ Qissoived ngggg Effects Cadmium(II) ions are a product of the reduction of nitrate to nitrite as shown in Equation 4-4. - + 2+ - Cd + N03 + 2!! ->Cd + No2 + £20 (4—4) Because hydrogen ions are consumed in this reaction. the pH of unbuf— fered samples increases slightly during reduction. For samples con- taining micromolar concentrations of nitrate. however. this increase is negligibly small. Nydahl [66] pointed out. however. that a much more significant increase in pH occurred within reactors because dissolved oxygen in samples and reagents reacts quantitatively with cadmium as shown in Equation 4-5. ca + 1/202 + 2n*—>Cd2* + 1120 (4—5) Under mildly alkaline conditions this reaction proceeds at a rate about 30 times faster than the rate at which nitrate is reduced to nitrite. The concentration of oxygen in dilute aqueous solutions in equilibrium with air is on the order of'0.5 mg-At L"1 [67] which. in accordance with the reaction stoichiometry. is equivalent to 1 mmol of base per liter of solution passed through a PBCR. For this reason. the buffer capacity of the sample must be much greater than would be predicted on the basis of nitrate reduction. if constant pfl'is to be maintained throughout the reduction process. Note also that oxygen reduction 42* simultaneously increases the concentration of C to a value of 94 approximately 0.5 ml_l. Nydahl estimated that at this concentration. cadmium hydroxide would begin to precipitate at a pH of about 8.5. and was likely to adhere to the surface of the cadmium particles. As a result. the reducing power of the reactor would decrease. This can be prevented if compounds such as ammonium chloride or EDTA that form soluble complexes with Cd2+ are added to the sample prior to reduction. Nydahl [61] suggested the use of imidazole (pg-7.09) which can be used to buffer samples and complex Cd2+ (84:17.5) simultaneously. L M 21 Chlorigg Nydahl [61] found that chloride ions greatly retard the rate at which nitrate is reduced to nitrite. Thus the reduction of nitrate in seawater ”Cl-120.3 )1) proceeds more slowly than in fresh water. This is generally of little consequence when PBCRs are used because of their relatively large reducing power. In fact. because chloride ions also decrease the rate at which nitrite is reduced to hydroxylamine. yields of nitrite may actually be increased in the presence of chloride. especially when highly reactive capperized cadmium is used as the re duc ing agent. Q. Experimental L Reagents Sulfanilamide (SAN) and N-(l-Naphthyl)ethyenediamine (NED) reagents were prepared exactly as described in Chapter 3. The ammonium chloride reagent was prepared by dissolving 10 g (0.19 mol) of ammonium chloride in l l. of 00'. Buffer solutions (0.1 )1) with nominal pH values of 5.0. 6.0. 7.0. and 8.0 were prepared from acetic acid and sodium acetate. 95 2-(N-morpholino)ethanesulfonic acid (HES) and sodium hydroxide. imidazole and hydrochloric acid. and tris [hydroxymethyl] aminomethane (TRIS) and hydrochloric acid. respectively. These buffers were chosen because in each case. their pK.s or plbs are within 0.25 units of the desired pH. Brij-35 surfactant (1 ml. L-l) was added to each buffer. Copper sulfate (0.01 _l_) was prepared by dissolving 2.5 g of CuSO4'5H20 in l L of DDW. 2... W Primary (5 m!) and working nitrite standards were prepared exactly as described in Chapter 3. Primary nitrate standards (5.0 m!) were prepared by dissolving 0.5056 g of dried. reagent grade KN03 in 1 L of DDW. 'orking standards were prepared by dilution of the primary standard by means of digital microliter pipets and volumetric flasks. Seawater standards (spikes) were prepared as follows. Volumetric flasks (100 mL) were filled to the mark with low-nutrient filtered sea- water. Then l.00 mL of the seawater was withdrawn from each flask with a digital pipet and discarded. Next the required volume of primary nitrate or nitrite standard was added to the flasks by means of digital microliter pipets. The flasks were then diluted to the mark with DDW and mixed. This procedure ensured that the dilution of the seawater was small (150 and that the dilution was uniform regardless of the concentration of the standard prepared. The filtered. low nutrient seawater. collected in midsummer from Long Island Sound. was a gift of the Oceanographic Sciences Division of Brookhaven National Laboratory. 96 1,, ngmium Reactor; Packed bed cadmium reactors were prepared as follows. Cadmium filings (40-60 mesh) were stirred with a magnet to remove ferrous metal and then washed with methylene chloride to remove grease. The filings (:30 g) were air dried on filter paper in a hood. and then transferred to a 100 m1. beaker. The filings were etched with 25 mL of 8 g nitric acid for about 1 minute and then rinsed thoroughly with DDW. 1.2 ! hydrochloric acid. and again with DDW. Next the filings were treated with several portions of the copper sulfate solution until the blue color no longer faded from the solution. The capperized filings were then rinsed with ammonium chloride solution to remove colloidal copper and cadmium hydroxide. COpperized cadmium filings prepared in this way can be stored indefinitely under a solution of ammonium chloride. PBCRs were prepared in small glass tubes (7 cm x 0.2 cm ID) with a constriction at one end. A small plug of glass wool was pushed to the constricted end of the tube with a stiff wire. and a small funnel was attached to the unconstricted end of the tube by means of a plastic sleeve. The funnel and tube were filled with ammonium chloride solu- tion and copperized cadmium filings were added to the funnel. one spatula tipful at a time. Between additions. the walls of the tube were tapped with the spatula to ensure uniform packing. 'hen the filings were within 0.2 cm of the top of the tube. a second plug of glass wool was added to secure the filings. The funnel was then removed and the PBCR was ready to install in the manifold. Open tubular cadmium reactors were prepared from 0.127 cm ID x .229 cm 0D cadmium tubing (Reactor Experiments. San Carlos. CA). Tubing was cut into desired lengths and wrapped around a 1 cm diameter 97 glass rod to form a closely spaced helix. Next the ends of the coiled tube were connected by a short length of plastic tubing. and the exposed surfaces of the cadmium coil were sprayed with several coats of acrylic lacquer. This allowed the reactors to be handled freely without fear of cadmium poisoning. A syringe was used to wash the inner walls of the OTCRs with methylene chloride. which was subsequently removed by flushing the 0TCR with nitrogen for several minutes. The OTCRs were copperized as follows. A 50 mL beaker was filled with equal volumes of imidazole buffer (pH-7.0) and copper sulfate solution (0.1 l). A syringe was attached to one end of the OTCR with a plastic sleeve and a short piece of plastic tubing was attached to the other end of the 0TCR. The syringe was then used to draw the buffered cepper solution into the OTCR. The solution was forced back and forth through the reactor for several minutes. after which the reactor was ready to be installed in the manifold. 'hen the OTCR was not in use. it was removed from the manifold. filled with imidazole buffer. and sealed by joining the ends with a short length of plastic tubing. Cadmium foil reactors were prepared from a 0.0025 cm cadmium foil (Alfa Ventron. Danvers. NA). The foil was polished with steel wool. rinsed with methylene chloride. and dried between tissues. The foil was then mounted between standard 6 or 12 inch dialysis blocks (#178- 2538-07 or 178-2540-01. Gamma Enterprises. It. Vernon. NY). designed for use with the Technicon SMAC clincial CFA system. The foil was then capperized within the dialysis blocks by the same procedure used for OTCRs . 98 1‘ Nanifolgs Nanifolds used for nitrate and nitrite determinations with OTCRs and PBCRs during these experiments are shown schematically in Figures 4-lA and 4-1B. respectively. Note that the flow rates of the air delivery pump tubes are twice the normal values in the manifold used with OTCRs. Recall from Chapter 2 that the minimum air segment volume necessary to sufficiently occlude a tube of diameter dt is equal to 0.92 dt3. The only cadmium tubing commercially available had an inside diameter of 0.13 cm. so an air segment volume of about 2 9L was required. This is about twice the volume required when dt . 0.1 cm. as is the case for other components that make up the manifold. 5‘_ pg leasurements A11 pH measurements were made with an Orion 611 digital pH meter and a Ross combination pE/reference electrode (Orion Research. Inc.. Cambridge. MA.) _E_,_ Some Definition; As discussed in the previous section. both nitrate and nitrite are reduced to lower oxidation state species in cadmium reactors. and the extent of reduction depends primarily on the size of the reactor. the flow rate of solution through the reactor. and the pH and chloride con- centration of the solution in contact with the reactor. Since only nitrite reacts with the Griess reagents. ideal experimental conditions would be those where reduction of nitrate to nitrite is quantitative. while reduction of nitrite to lower oxidation state species is negli- gible. In this case determination of nitrite in the cadmium reactor effluent would yield the sum of nitrite and nitrate originally present 99 .muouomop sawsvmo won moxomn you soumxw Am .muouomou ezwsvmo awom mam magmas» mono uom.eoumxmmmm< .ucoempumcfi oEou «who one new: mcnmvcmum cameo“: you woummmos moocmanmnm Eonm voumfimofimo one: monogamoumm cw mo=Hm>o o.we w.om w.Hm o.ma m.~m e.wm H.om ~.oe sao>oooa w “mam. ”mamm.voooo.m mama. fleamm.voooo.m Rama. hemmm.vmmmm. emma. hemmm.vwmmm. axe ooo.- fiNoo.-Veoo.- moo.- ”moo.-vooo.- emc.- Avoo.vmmo.- ”Ho.- Amos.-V¢~c.- eaooaoeem-» mama. ”memo.voe~o. moms. AwVNo.UemNo. ono. Awmwo.vmmmo. «boo. NeaoVNo.Ummoo. odomm -moz -Noz -moz -moz -moz -moz -moz -Noz c~.w II oc.e cc.c me.v ":8 .mn mo cowuocmw a ma mvumvcmum momma“: can mummy“: pom woumfisofimo mowno>ooou ucooeom mam muouoemnma :oflmmouwom neon“; .Nuv oanmh 105 Percent recoveries of nitrite and nitrate were further investi- gated over the pH range of 6.5 to 8.5 using imidazole (pH-6.5. 7.0. 7.5) and Tris (pH-8.0. 8.25. and 8.5) buffers. A short (10 em) OTCR was used in these experiments so that changes in percent recoveries of nitrate would be more apparent. This experiment was also performed with a cepperized PBCR using the manifold shown in Figure 4-lB. Per- cent recoveries calculated from experimental results obtained with the OTCR and the PBCR are plotted as a function of pH in Figure 4-2A and 4- 28. respectively. As before. percent recoveries were calculated on the basis of the ratio of the slopes of nitrite and nitrate calibration curves (5. l5. and 25 pl concentrations) with and without a cadmium reactor on line. Nitrite recoveries increase slightly as the pH increases in experiments with both reactor types. as expected. Nitrate recoveries calculated from data obtained with the PBCR also increases as the pH increases in agreement with Nydahl's study. Nitrate re- coveries calculated from data collected with the OTCR. however. first increases slightly as the pH increases. and then decreases sharply with further increases in pH. Although this trend can be rationalized in terms of the decrease in the rate of nitrate (and nitrite) reduction that occurs as pR increases. it seems likely that the ability of d2+ is also involved. There was no complexing imidazole to complex C agent in the Tris buffers and so formation of Cd(OR)2 on the surface of the OTCR was a distinct possibility. This seems even more probable in light of some later experiments that are discussed shortly. Throughout these experiments nitrate recoveries were always highest when imidazole buffers were used. 106 A) '00 F W (O O l m 0 fl 0 I O) O U Percent Recovery T 50£ (51’ I 41 l l J 6.0 6.5 7.0 7.5 8.0 8.5 B) pH >~ §l00- E; 535). o: 98- C 97 b 0N0; £95, 0N0; 95- 0% L 4 1 1 1 6.0 6.5 20 75 8.0 8.5 pH Figure 4-2. Percent recoveries of nitrate and nitrite. A) Open tubular cadmium reactor. 8) Packed bed cadmium reactor. 107 In the next set of experiments percent recoveries were determined for nitrite and nitrate standards in the concentration range of 5 to 25 '1! using the three imidazole buffers from the previous experiment and a 30 cm copperized OTCR. Slopes. Y-intercepts. correlation coef- ficients. and percent recoveries calculated for data from these experiments can be found in Table 4-3. These results were encouraging. Recoveries for both nitrite and nitrate were in the range of 95 to 97‘. When low nutrient seawater that had been spiked with nitrate was deter- mined with this OTCR. however. nitrate recoveries were less than 80%. although the calibration curves remained linear. This observation was not unexpected because Nydahl [61] reported a sharp decrease in the rate of nitrate reduction when the chloride concentration of the solu- tion in contact with PBCRs increased. A plot of nitrite and nitrate recoveries as a function of chloride concentration is shown in Figure 4-3. Here calculations of percent recoveries are based on the absor- bance measured for 25 ll! nitrite and nitrate standards to which sodium chloride was added. Imidazole buffer (pH-7.5) and a 15 em copperized OTCR were used for this experiment. At this point it became apparent that longer OTCRs would be necessary to approach quantitative reduction of nitrate in a seawater “Cl-1:03!) matrix. When a 60 cm copperized OTCR was installed in the manifold. however. the segmentation pattern of the analytical stream emerging from the reactor was irregular. recorded baseline signals were noisy. and peak shapes were severely distorted. These problems were initially attributed to irregularities or par- tial blockages within the reactor. When several shorter OTCRs that individually supported stable segmentation patterns were connected 108 .vfiomwnme can seem uo>oaou 2090 on» saw: mvpmvcmum comma“: mom poemmmos moozmpuomnm scum voumfimomwo one: monogamonmm a“ momam>o 0.0m 2.5m 0.8m ~.cm o.ma o.mm muo>ouo¢ 3 oooc.~ Rocco.mvoooo.m oooo.~ floooo.mvmmmm. oooo.~ .flcooo.mvcoco.m axe Hoo.- flooo.-vmco.- voo.- flees.-vmoo.- moo.- ”woo.-vwoc.- udooaoeem-s OMNO. ameD.VHm~o. ammo. ”ammo.vm-c. ammo. ”VHGMNo.Ue~No. odomm Inez -moz -moz . -Noz -moz -Noz m.e . o.e m.o. ”mg .mupo vouwnonmoo Eu cm a can muommmb oHonvflaw new: mvuevcmum comma“: was museum: mom woumfisu~mo mofiuo>ooon acoouom new muouoammmm cowmmouwou among; .nue oHnmh 109 80 Percent Recovery V o O) O 50¢ °o.o 6| 0.2 0.3 6.4 6.5 6.6 6.7 6.8 69 Chloride Concentration (M) Figure 4-3. Effect of chloride concentration on percent recoveries of nitrate and nitrite with open tubular cadmium reactors. 110 together. however. the same symptoms were observed. ‘This was puzzling at first. but it finally became apparent that oxygen in the air segments was being reduced at the reactor walls. which caused the air segments"volume to decrease. In longer OTCRs. the air segment volume became too small to occlude the tube and so the integrity of the analytical stream was lost. When nitrogen rather than air was used as the segmentation gas. stable segmentation patterns were achieved for OTCRs as long as 150 cm. At this point. the p11 of the buffered analy- tical stream was measured before and after it had passed through the 60 cm OTCR. first with air. then with nitrogen as the segmentation gas. Results are shown in Table 4-4. The buffer capacity was clearly exceeded when air was used as the segmentation gas and a 60-cm OTCR was on-line. which confirmed the hypothesis that oxygen in the air segments was being reduced at the OTCR walls (see Equation 4-5) and also explained the distorted peak shapes and noisy base lines observed when the seawater spikes were passed through the 60 cm OTCR with air as the segmentation gas. As is demonstrated in the final section of this chapter. nitrate recoveries comparable to those measured with PBCRs can be achieved with a 60 cm OTCR when nitrogen is used to segment the analytical stream. 2‘ Cadmiumqui; Reactors As mentioned earlier. the ID of commercially available cadmium tubing is somewhat larger than the ID of other manifold components in the mCF system. Recall that. other factors being equal. dispersion in gas segmented reactors increases as the inside diameter of the reactor increases. It occurred to me that a cadmium reactor with improved wash characteristics could be fabricated by inserting a piece of cadmium 111 Table 4-4. Effect of segmentation gas on the pH measured for the effluent from a 60 cm OTCR. Measured pH Segmentation Gas: Air Nitrogen Nominal Distilled Distilled pH Water Seawater Water Seawater a 6.5 6.80(6.54) 6.96(6.63) 6.52 6.64 7.0 7.35(7.01) 6.38(7.11) 7.01 . 7.12 7.5 8.59(7.49) 8.38(7.64) 7.50 7.64 apH values in parentheses were measured before the buffered analytical stream entered the OTCR. 112 foil between the blocks of a dialyzer designed for use with the SNAC CF system. This approach was tried. but unfortunately it proved very difficult to produce a good seal between the dialysis blocks and the cadmium foil. Even when a seal was achieved temporarily. leaks began to develop within a few hours of operation time. This was not considered acceptable for routine use. In the few successful experi- ments with 12 inch dialyzer blocks. recoveries calculated for nitrite and nitrate standards were on the order of 977- and 75%. respectively. It did appear that sample interaction was lower in this reactor than in OTCRs. although insufficient data were collected to make a definitive comparison. In these experiments the analytical stream was buffered with imidazole (pH-7.0) and segmented with nitrogen. These results were encouraging. and if a solution to the problem of forming a good seal between the dialysis blocks and the cadmium foil could be solved. this system would warrant further study. Qp Einal Experiments The results of the final experiment in which the performance of a 60 cm cepperized OTCR and a 7 cm x .2 cm ID cOpperized PBCR were compared for both distilled water standards and seawater spikes are summarized in Tables 4-5 and 4-6. The concentration range of standards and spikes for both nitrate (0 to 25 p!) and nitrite (0 to 2.5 u!) was chosen to be representative of concentrations normally encountered in unpolluted seawater. Data listed in Table 4-5 resulted from five replicate determinations of each standard and spike. Three sets of standards and spikes were prepared for this experiment. One set con- tained only nitrate. another set contained only nitrite. and a third 113 Table 4-5. Data from determinations of nitrate and nitrite in distilled water and seawater with OTCRs and PBCRs. Baseline Corrected Absorbancea Nominal Concentration (uM) OTCR PBCR - Distilled Distilled N02 + NO3 Seawater Water Seawater Water 0.5 + 5.0 .1491.001 .1401.001 .l35t.001 .136i.002 1.5 + 15.0 .4121.001 .409i.001 .4071.001 .4071.001 2.5 + 25.0 .660r.001 .669:.001 .6681.001 .6681.003 NO3 5.0 .127.+..002 .127i.001 .124i.001 .1211.001 15.0 .373 :t .001 .374 i .003 .366 r .001 .368 i .002 25.0 .6111.001 .607i.001 .6111.001 .6011.002 NO2 0.5 .Ollt.001 .009:.001 .0111.001 .0121.001 1.5 .035:.001 .0321.001 .03Si.002 .037i.002 2.5 .065:.001 .059:.001 .0621.001 .0641.001 (Cadmium Reactors Off-Line) 0.5 + 5.0 .013i.001 .011:.001 .0101.001 .0121.001 1.5 + 15.0 .037i.001 .0381.001 .0371.001 .040i.001 2.5 + 25.0 .0621.001 .063:.001 .0601.001 .0651.001 N02 0.5 .013i.001 .011¢.001 .0111.001 .0121.001 1.5 .036r.001 .0381.001 .037:.001 .039:.001 2.5 .067i.001 .0631.001 .0641.001 .0671.001 aAverage of five replicate determinations. .0041 .002. Refractive index correction 114 .momceucou ore ofimfiw xx mama. mmmm.e mama. mama. e nmm. amm. mmm. mmm. .eea mo eoeeo .mem Sm. e 8m. 8m. a 8m. mmm. e mmm. mmm. e mmm. .m.m e mama—See; mmmmm. e mamam. mmmmm. e e28. mmmmm. e Seam. mmmmm. e. Seam. .m.m e emcee 8:358 mamm. mama. amam. mmmm. axe _mm. emm. mmm. mmm. .eaa ea eoeea .eem Nam. e So: So. e Nam: 8m. e «mm: So. an mm: 5.3 emaoeoemeIa ammm. e 88. ammm. e 3.8. ammm. e mmem. ammm. e maNm. .m.m e emcee -Noz mmmm. mmmm.e mmmm.e mama. axe «mm. _mm. mmm. mmm. .eae mo eoeeo .mem 8m. e 8m. 8m. e Nmm. 8m. e amm. Nmm. e m8. emooeoeeeIa emmm. e emam. ammmm.nammm~m. ammmm.nemme~m. emmm.ne emem. emcee Inez mama. mama. mmmm.e mama. axe mmm. mmm. mmm. mmm. .eea ea eoeeo .mem mmm. e mmm. 8m. x «mm. 8m. e mmm. 8m. e mam. .m.me emooeoemeIa ammm. e 2.8. mmmmm. a 23m. mmmmm. e mmxam. mmmmm. e mammm. .m.m e emcee Hope: ecumzmom noun: ecumzmom Inez + -Noz Imoemweaem maeemeemm mama ammo emoeaeoewem .mav ofinmh cm woummm meme eom meoaoememn cowmmouucm emcee; .ose smack 115 axe .emo mo eoeeo .mum uncoeouceua oaoem accessou axe .umo mo eoeeo .mum «acoeouceua omoem mama. Nmaa. mama. moaa. moo. moo. goo. moo. 8o. H So... So. H woof So. H go? So. H So... mooo. H 3.8. mooo. H ammo. mooo. H memo. mooo. H ammo. mamm. aaaa. mama. coma. com. com. moo. mac. 80..” woof 89H Non..- 39H Nocf 89H moo... aooo. H mama. Boo. H memo. Boo. H memo. aooo. H came. mama. Haaa. mama. amaa. Ham. moo. moo. coo. So. H Sci 8o. H woof 8o. H woof 8o. H 8o. aooo. H memo. Boo. H ammo. aooo. H memo. Boo. H memo. eoewz eoemzmom ecumz ecumzmom vofifiwumem wofifieemam mzmqaumo mm0ku4 v-U- I— I «o - o—— ._ .......Y —I .4 f - I I I : ' ' ‘ I I IL‘ -qt- _4I—-——T—e—. e—A‘L—--.-— -- ..PI) «>144» “—4-. .0. ..xI—I J -.l.._... -.. _é- . . - -- I h I I . I . . , I I 4 : qp—- — -T- I dy‘ a- - ,—-—.J ..I 1.---.. . I z - I g _ . . I I -—J .-.---'-—- «---J«-«I *“I‘ I- ~ J-I- I 9 - . : ! I ' F F I . ' 5“”‘I ”r “I“ ‘II ‘II‘rv- -~1~I Ir“ - I ‘ ' I I 5 I ' . I l I I e I . n 'I I _, -.. .' --ab. «I I-I ..n 1 [I1 .— vq I .1 —. i I I a ‘ I I . l I I . I ”1* - -‘ r 1 5 :1 - l x I I . \- I I '} TI 7 I -- 1' 1 I r C -i. , -_‘ _ r I - , _ E I ' . I _ ' ' D ' . ‘ 4 , _. _.._1-_...-l_-__. .. . ..-- _ -_ _ o , «'4- fl .- _ . -. - Figure 4-4. Peaks recorded for seawater spikes with a 60 cm open tubular cadmium reactor. A) Nitrite spikes. Nominal concentrations from left to right: blank, 0.5, 1.5, and 2.5 uM. B) Nitrate plus nitrite Spikes. Nominal concentrations of NOéVTNoj'from left to right: 0.5-+5.0, 1.54-15.0, 2.54-25 uM. Sample and wash times were 23 s and 7 s, respectively. '- 118 I s I -i-. I Figure 4-5. Peaks recorded for seawater spikes with a packed bed cadmium reactor. A) Nitrite spikes. Nominal concentrations from left to right as in Figure 4-4. B) Nitrate spikes. Nominal concen- trations from left to right: 5.0, 15, and 25 pg, Sample and wash times were 40 s and 20 s, respectively. 119 0.7 .0 m Q 0| Absorbonce (PBCR) Q o o: 45 Q N 9 1 OJ 0'2 d3 074 6.5 6.6 57 Absorbonce (OTCR) Figure 4-6. Correlation of data obtained with open tubular and packed bed cadmium reactors. O = seawater spikes; - = distilled water standards. 120 estimate (0.005). Y-intercept (0.002 t, 0.002). and correlation coef- ficient (0.9998) of this plot indicate that results obtained with both systems are equivalent within the limits of experimental error. L Dis u n The goal of this research was to develop a new type of cadmium reactor that would be more compatible with air segmented. continuous flow analysers than the PBCRs that are most frequently used for routine nitrate determinations in seawater samples. The OTCRs described here fulfill this objective. Inspection of Tables 4-5 and 4-6. and Figure 4-6 reveals that equivalent analytical results can be obtained with either reactor type. both for standards prepared in distilled water and for seawater spikes. Furthermore. dispersion that occurred in OTCRs where the analytical stream remained segmented during reduction was much less than that which occurred in PBCRs where the analytical stream was not segmented during reduction. This is shown clearly in Figures 4-4 and 4-5. Peaks with flats and less than 1% interaction were achieved in conjunction with the OTCR at a sampling rate of 120 hr-l. Comparable performance in conjunction with the PBCR could only be achieved at half this sampling rate (60 hr'l). OTCRs are much more convenient to use than PBCRs that require elaborate procedures both for preparing cadmium filings and for packing prepared filings into small glass columns. OTCRs are also much less prone to clogging and channelling than PBCRs, and their use eliminates the need to debubble and rebubble the analytical stream before and after reduction. Thus when an OTCR is used in place of a PBCR for nitrate determinations. fewer pump channels and manifold components are 121 required and therefore less can go wrong. The fact that the analytical stream must be segmented with nitrogen rather than air when OTCRs are used is somewhat inconvenient, and might be particularly cumbersome for shipboard work where compressed gas cylinders must be secured to the outside bulkheads for reasons of safety. Some work still remains to characterize OTCRs completely. No ex- periments were performed to determine the long-term stability of OTCRs. The optimum concentration of Cu2+ in the imidazole buffer should also be investigated more thoroughly. Addition of small amounts of copper to this reagent kept the activity of the OTCR high and uniform, but in larger amounts, reduction of nitrite to lower oxidation state species increased. Nonetheless. results presented here suggest that OTCRs are a practical alternative to PBCRs for routine nitrate determinations. Some field studies are now needed to determine how well OTCRs will per- form in extended. routine operation. CHAPTER 5 AN EXPERIMENTAL COMPARISON OF CFA AND FIA A; Introduction From the time of its inception. FIA has been promoted as a simpler, less costly. and more efficient alternative to CFA, although this assertion has been hotly contested in the recent literature [68- 73]. Snyder [5] recently compared CFA and FIA and demonstrated that from a theoretical standpoint. dispersion is generally much lower in air segmented flow reactors than in nonsegmented flow reactors. This study predated the first report of single head string reactors for FIA. Very recently Rocks and Riley [11] compared FIA and CPA from the standpoint of clinical applications and concluded that FIA would eventually replace CFA in the clinical laboratory; This review appears to have compared second, rather than third. generation CFA.systems with FIA and therefore the conclusions drawn seem overly optimistic. To date. however. there has been no direct experimental comparison of CFA and FIA, perhaps because ground rules to prevent bias are difficult to establish. and also because a single channel. third generation CFA instrument is not commercially available. The miniature continuous flow analysis system that I developed during the course of my research can.be configured for either CFA or FIA operation. and is therefore ideally suited to an experimental comparison of these two continuous flow analysis techniques. 122 123 The experimental comparison of CFA and FIA presented in this chapter is unique in several respects. lost importantly. instrumen- tstion bias is minimal because the same pump. flow cell. and detector were used for both FIA and CFA experiments. Furthermore. for the first time. dispersion in open tubular reactors (OTR). single bead string reactors (SBSR). and air segmented reactors (AS!) is compared under identical experimental conditions. In the first set of experiments. dye dispersion in the three reactor types was monitored as a function of the volume of dye injected into the reactors and the liquid flow rate through the reactors. Here chemical reaction was not required for detection. and therefore recorded peak profiles provided an unambiguous indication of the extent of dispersion that occurred in each reactor at each flow rate. The situation is more complex when chemical reaction is required for detection. as is often the case for continuous flow determinations. Here recorded peak profiles will depend not only on the dispersion characteristics of the reactor. but also on the kinetics of the chemical reactions involved. Analytical rections with rela- tively slow kinetics were expected to bias the comparison of CFA and FIA in favor of CFA. To avoid this type of bias in the second set of experiments. I chose the colorimetric determination of chloride by reaction with mercuric thiocyanate and ferric nitrate [74] as a reference assay by which the performance of CFA and FIA could be com- pared. This reaction reaches equilibrium in a few seconds of reaction time. 124 _B_,_ er men al L m Dispersion Experiments The simple flow system used for dye dispersion experiments is shown schematically in Figure S-lA. Here a single four-way slider valve was used to direct either the dye solution [phenol red (1.0 uncle) in 100 ml. of pi! 9.5 borate buffer] or the carrier solution (pH 9.5 borate buffer) into the reactors. Valve switching was controlled by a microprocessor as described in Chapter 2. The volume of dye injected as a function of pump speed and valve actuation time can be found in Table 5-1. Pump tubes with nominal flow rates of 1.0 ml. min"1 were used for all experiments with OTRs and SBSRs. For experiments with ASRs at the two lower flow rates. however. pump tubes with nominal flow rates of 0.42 ml. min"1 were used. This was necessary to achieve 1 at all flow rates (see segmentation frequencies of at least 1.5 s- Table S-l). Plastic OTRs, SBRs. and ASRs were fabricated in the lengths listed in Table 5-2. Fabrication details can be found in Chapter 2. Letters in parentheses in Table 5-2 are used to identify each reactor in the remainder of this discussion. Reactor lengths were chosen such that the dwell time of the dye slug in each reactor type would be about the same in each set of experiments. Signals that resulted when 25. 50. 100. 200. 400. and 800 pl. of dye were injected sequentially into each of the twelve reactors at four different flow rates were recorded with a strip chart recorder. In addition. five replicate determinations for one sample volume (usually 50 pl.) were obtained at each flow rate for each reactor. The average and the standard deviation for each set of replicate determinations can be found in Table 5-3. Relative standard deviations of 1% or less were 125 NOMINAL A) FLOW RATE Jul-mm)- : _ .019. 4. .1 1A"; I..._ SL952 #— —————— .1 SAMPLE I. O. 2) WASTE “EV '- O BUFFER "00.42” .—I [_J 540nm B) REAGENT __ 9.8_._ __ _ _ .. ._, WASTE REAGENT -8 H or DDW .' O. A |.O : 480nm [ w WASTE SAMPLE C) twp SPEED ~56 r‘ ""’ '" -' W n-2 I T_' a * I (31C) J VMAErrE: AIR L-._Q£§ 0.3m REAGENT 0 32 1 SAMPLE ° '5 -———J 480nm Figure 5-1. Composite diagrams of manifolds used to compare FIA and CPA. A = air, S = sample, C = carrier, R = reagent. B) FIA manifold for chloride determina- C) mCFA manifold for chloride determinations. for dye dispersion experients. tions. A) Manifold used 126 Table 5-1. Sample volume as a function of sample time for dye dispersion experiments. Sample Volume (uL) 25 50 100 200 400 800 Flow Rate Pump Speed (mL min'l) Samplipg Time 21“, 45b 0.5 3.0 6.0 12.0 24.0 48.0 96.0 42“, 90b 1.0 1.5 3.0 6.0 12.0 24.0 48.0 63a 1.5 1.0 2.0 4.0 8.0 16.0 32.0 84a 2.0 .75 1.5 3.0 6.0 12.0 24.0 aPump tubes with nominal flow rate of 1.0 mL/min at standard speed. Pump tubes with nominal flow rate of 0.42 mL/min at standard speed used with ASRs only. Table 5-2. Reactor lengths used for dye dispersion experiments. Reactor Type Lepgth (cm) OTR 10 (A) 50 (B) 100 (C) 200 (D) SBSR 10 (E) 25 (F) 50 (G) 100 (H) ASR 10 (I) 30 (J) 50 (K) 75 (L) 127 a Table 5-3. Precision of dye dispersion experiments. ------------------ Flow Rate (mL min’l) -—---——-------- Reactor 0.5 1.0 1.5 2.0 PA 0.217 1.003 .241 1.002 .258 1.007 .274 1.004 B_OTR 0.232b1.001 .280b1.000 .170 1.002 .164 1.002 c 0.153 1.001 .150 1.001 .160 1.001 .204 1.001 _0 0.124b1.001 .227c1.002 .257 1.001 .305’01.000 _ E 0.323 1.002 .334 1.002 .350 1.003 .364 1.003 1=__SBSR 0.247 1.002 .254 1.002 .273 1.003 .285 1.003 G 0.173 1.002 .183 1.002 .194 1.002 .203 1.001 _H 0.128 1.001 .138 1.001 .148 1.002 .146 1.002 ' 0.436 1.004 .430 1.000 .408 1.004 .392 1.015 J_ASR 0.424 1.002 .422 1.003 .371 1.007 .364 1.009 0.440 1.002 .435 1.001 .368 1.008 .352 1.014 LL 0.433 1.002 .431 1.002 .354 1.009 .309 1.017 “Sample volume = 50 uL unless otherwise indicated. b100 uL sample volume. 0200 uL sample volume. 128 calculated for all data sets except those obtained with ASRs at the two higher flow rates. The higher relative standard deviations calculated for these data (3-5fi) resulted from unfavorable experimental conditions that could not be avoided. At the two lower flow rates the dye slug was divided into about 10 segments (e.g.. 6 s/slug x 1.6 segments/s - 9.6 segments/slug; 3 s/slug x 3.2 segments/s - 9.6 segments/slug). while at the two higher flow rates. the dye slug was divided into only about 5 segments (e.g.. 2 s/slug x 2.25 segments/s - 4.5 segments/slug; 1.5 s/slug x 3.00 segments/s - 4.5 segments/slug). For this reason proportioning errors were much more pronounced at the higher flow rates. Under normal operating conditions used for CFA. this situation would not arise. In any case. the pre- cision of these two data sets is sufficient for qualitative compari- sons. Tracings of peaks recorded when 50 pl. of dye were injected into each reactor are shown in Figure 5-2. This figure provides a great deal of qualitative insight into the dispersion characteristics of the three reactor types. Examination of the peak profiles obtained with the four OTRs (A.B.C.D). reveals that peak heights decrease and peak widths broaden as the reactor length increases. Of course. reactor length and dwell time of the sample slug in the reactor are two sides of the same coin. It is generally accepted [75] that dispersion in OTRs increases with the square root of either the reactor length or the mean dwell time of the sample slug in the reactor. This relationship was found to hold to a fair approximation under all experimental condi- tions. Notice also that as reactor lengths were (increased. the peaks became more symmetrical as predicted by the 'tanks-in-series'model of 129 FLOW RATE 0. SmL Imin. l.0 mL/m‘n. l.5 mL/min. 2 0 mL lmtn. <13- FUXWDpentube g 0.24 A g a a: ‘1 0.l- F. I C , ‘ D r—fi f——T_I—fi T T T fl" r T T T T l T 0 IE) 0 90 0 45 0 36 0.4- FIAISBSR 0.3- E F 8 3 0.21 G a N “4° 0.:- OH H J I—_I—-1 7—1—4 fi—T—fi r T I T fifi—I 0 l20 0 60 0 45 0 30 (ls- rnCEA I .I 0.44 K t; L ggcis- a: é 2_O.2« 0.l- OJ L Msk J \L .1 LL}; L f—‘l—fi [_T_—I r—T'fi l T I T 1—1 060l20 03060 0|5304506l2l82430 TIME“) Figure 5-2. Curve tracings of peaks from dye dispersion experiments with open tubular reactors, single bead string reactors, and air segmented reactors. 130 dispersion in (Ills [76]. Another clearly visible trend is the increase in peak heights and decrease in peak widths that occurred as the flow rate increased. This trend is due primarily to the decreased residence time of the dye slug in the reactor. and to a lesser extent. an increase in secondary flow effects. These same trends hold for peaks recorded for the four SBSRs (E.F.G.H). Note. however. that at com- parable residence times. peak widths for SBSRs were about half those recorded for OTRs. Also tailing. which was particularly noticeable in peaks recorded for the shortest OT! (A). was much less pronounced in the SBSR (E) of comparable length. especially at lower flow rates. In general. peaks recorded for SBSRs were more symmetrical than those recorded for OTRS. This suggests that the mixing efficiency of SBSRs is better than in OTRs as reported by Reijn. et al. [37] who stressed that dispersion is not synonymous with adequate mixing. especially in OTRs. The peaks recorded for the four ASRs (I.J.K.L) exhibited very different characteristics. Notice first that reactor length had very little effect on either peak heights or peak widths. Note that subtle differences in peak heights and widths that might be expected were masked by mixing effects that occurred in the sampling valve before the dye slug was segmented with air. Under normal operating conditions for CFA where peeked sampling is employed. sharper peaks would have been recorded. Nonetheless the ability of air segments to minimize disper- sion is readily apparent. This study shows very clearly that under comparable experimental conditions the extent of dispersion in the three reactor types can be ranked as follows: ASR RgC12+28CN— (5-1) SCN'+Fo3*—> Fe (SCN) 2" (5—2) Max : 480 mm Although it has been suggested [77] that the major product of Reaction 5-2 is Fe(SCN)2+. and that even higher order complexes might also be formed. the 1:1 complex indicated is the mostly likely product [74]. The reagent solution was prepared by dissolving 0.626 g of mercury (II) thiocyanate. 30.3 g of iron (III) nitrate. 3.3 ml. of concentrated nitric acid. and 150 mL of methanol in about 500 mL of dionixed distilled water (DDW) contained in a 1 L volumetric flask. The solu- tion was diluted to the mark with DDW. mixed. and transferred to an amber colored bottle. Then 1 ml. Brij-35 surfactant was added. This reagent formulation is identical to the one recommended by Hansen and Ruzicka [77]. The primary chloride standard (1000 ppm) was prepared by dissolving 1.648 g of sodium chloride in l L of DDW. Working standards in the range of 5 to 45 ppm were prepared by dilution of the primary standard with digital microliter pipets and volumetric flasks. The FIA and CFA manifolds used for these experiments are shown schematically in Figures 5-lB and 5-1C. respectively. Samples were introduced into the FIA system by means of a six-port rotary valve equipped with a 60 pl. 132 sample loop. Pecked sampling was used to introduce samples into the CFA manifold. 'hen chloride standards in the concentration range of 5 to 35 ppm were injected into the single line FIA manifold (Figure 5-lB) at 30 s intervals. the peaks shown in Figure 5-3 were recorded. Note that there is a large blank signal (i.e.. injection of DD! resulted in a peak) and also that each positive peak was followed by a negative peak before the signal returned to base line. These effects are frequently observed when a colorless sample is injected into a colored reagent stream [78] or when there are large differences between the viscositics of the sample and the reagent stream [79]. The effects are due to temporal changes in refractive index and concentration along the dispersed sample slug. It is interesting that this behavior was not observed in an almost identical experiment described by Ruzicka and Hansen [77]. probably because the internal volume (18 pL) of the flow cell they used was nine times greater than that (2 p1.) of the flow cell used in these experiments. The above problem was eliminated by adding a second pump tube to the FIA system (dashed line in Figure 5-1B). Here samples were injected into a distilled water carrier stream that merged with the reagent stream in a tee mixer. Thus the extent to which the reagent stream was diluted remained relatively constant‘when samples were injected. Peaks obtained with the two line FIA system (50 em CTR) are shown in Fig. 5-4A. As expected. no response was observed when distilled water was injected into the system. but peaks and the base line signal were excessively noisy. This was due to poor mixing between the reagent and sample carrier streams. When the 50 cm OTR was 55 PPM 25””! 5P?“ " LI M «I u u 134 {15' . __ '35 25 I5 NM Figure 5-4. merging zones FIA manifold. km 8) 50 cm single bead string reactor. 15 3S K. k 45 K L Peaks recorded for chloride determinations with a A) 50 cm open tubular reactor. 135 replaced with a 50 cm SBSR. the peaks shown in Figures 5-4B were recorded. Note that base line noise was reduced to negligible levels and that peak heights increased and returned to baseline more rapidly. Here we see clearly the advantages of using SBSRs for routine FIA determinations. Dispersion is decreased so the sampling frequency can be increased. and the mixing efficiency is much improved relative to a comparable OTR. Further experiments showed that good results were still obtained when the length of the SBSR was reduced to 25 cm. At a flow rate of 1.6 ml. min—1 (carrier. 0.8 ml. min-1; reagent. 0.8 ml. miufl). sampling rates of 180 hr"1 were achieved using the 25 cm SBSR without significant interaction. 'hen the pump speed was doubled. it was possible to increase the sampling rate to 360 hr’l. Peaks recorded at this sampling rate shown in Figure 5-5A. Next the same chloride standards were determined with the CFA man- ifold shown schematically in Figure 5-lC. 'The pump speed and segments- 1 tion frequency were 56 and 2 s- . respectively. Peaks recorded with the CFA system at a sampling rate of 360 hr.1 (t' . 5 s. t - 5 s) are w shown in Figure 5-53. Data from chloride determination experiments for both FIA and CFA systems are compiled in Table 5-4. Note that in all cases. less reagent and sample were required for CFA determinations. Furthermore the linearity of the standard curves and the precision of analytical results is somewhat better for the CFA system than the FIA system. 136 ...... ! l 0 i 1 I Figure 5-5. rate of 360 hr‘l. string reactor. 8) mCFA. Peaks recorded for chloride determinations at a sampling A) FIA with merging zones and a 25 cm single bead 137 MO OESHO> OFF ”uou=w oswu on» wcwusu veno>wfiov psowmoa .Hm>uou:« cm“: can ofiasmm 6:0 wcwnsm v6u6>fifiom ucowmmn mo osmao> cab " 6H6. one. 156. Ame. “No. omo.o 66656 ..A HA N-Hnu 6.6 6.6 6.6 1603666661H 1 56.6 66.6 Na._ 6m.o m~.o m6.o “enema *1 662606666 mmm 666 new N62 Rom Ha Aefigav 6s=Ho> 6166661 :6 66 co 6m 6m 12 56:6 6e=~o> 6~asmm -11. .1. .111 m m m Amy 6206 :66: .111 .11 .111 mm as m Amy 61“» 6Haeam owm oem 66m owm cum 66m ”5-;V 6666 6261~61< 651 egos .m:0mumcfishouov opwpofizo ofipuoeflhofiou "