a‘A.. f i an” .Jv.» . . .(9‘ . ‘ V is.“ 3a...» :NQ‘ , Avwnafif . ‘ In ‘ .. ywwvdk» . 7.. . ls”) 2a. . . +3 . ’l lv -' ‘ 2.5L (fir... 4:. | 4. $th A at»; . ‘ 4! u$ ’l .5 .1 ,. 5.9 1 931D. . u. < 2. .§ . ,. 1. , v . .nflwtfimmmwcn . . on“; nixolxih .o. $~Iuka . ul :1 3! £45.. . . xf . . 17 V «"5. 1.3.? THESlS J memo»: STA‘i' u RABIES 2) lllllllllllllllllllllllll‘lllllllllllll 3 1293 01771 0892 LIBRARY Michigan State University This is to certify that the thesis entitled Experimental Measurement of the Slit Response Function and Corrected Infrared Thermographic Measurements presented by Paul Bryan Hoke has been accepted towards fulfillment of the requirements for MS Mechanical Engineering degree in j)? QC M4 6M Date August 21, 1998 0.7639 MS U is an Affirmative Action/Equal Opportunity Institution PLACE IN REI'URN BOX to remove this checkout from your record. TO AVOID FINES return on or before date due. MAY BE RECALLED with earlier due date if requested. DATE DUE DATE bus DATE DUE MAY 0 5m? 00kg; 2003 NPR—144M- 033039129" 1]” COMM EXPERIMENTAL MEASUREMENT OF THE SLIT RESPONSE FUNCTION AND CORRECTED INFRARED THERMOGRAPHIC MEASUREMENTS By Paul Bryan Hoke A THESIS Submitted to Michigan State University in partial fulfillment of the requirements for the degree of MASTER OF SCIENCE Department of Mechanical Engineering 1998 ABSTRACT EXPERIMENTAL MEASUREMENT OF THE SLIT RESPONSE FUNCTION AND CORRECTED INFRARED THERMOGRAPHIC MEASUREMENTS By Paul Bryan Hoke Infrared thermography utilizes emitted radiation to measure temperatures optically. An experimental facility and procedures were defined to measure emissivity and temperature of unresolved (target angle < 0.05 rad) targets. The facility consists of a flow system with instrumentation for measuring flow velocity, temperature, and humidity. The emissivity is determined with the infrared camera and the target temperature set by the known flow conditions. Once the specimen emissivity and size are known, its temperature can be measured in any environment. An Inframetn'cs 6OOL infrared camera and the associated data acquisition system were characterized to maximize thermal measurement sensitivity. The minimum specimen size that can be resolved thermally with the infrared camera was determined. The effect of specimen size on thermal signature was quantified for the infrared detector utilizing the slit response function (SRF). A standard procedure to measure emissivity was validated and a SRF measurement correction was verified. These two pieces of information increased the measurement capabilities of the infrared system and minimized experimental error when working with thermal targets smaller than the thermal measurement resolution of the system. ii To Carol, for always believing and of course, Madeleine iii ACKNOWLEDGMENTS I would like to thank my advisor Dr. John McGrath for having the patience to let me try things on my own and guiding me when I needed it. His advice, insight and assistance to this work have been invaluable. I would also like to thank Dr. Craig Somerton and Dr. Andre Benard for joining my committee and editing this work. Dr. Somerton has been an aid through out my stay at Michigan State University as an advisor, reference and an open door. Dr. Benard’s aid and insight are greatly appreciated. Dr. Foss for his advice, guidance, and equipment. Dr. Foss was a wealth of support and knowledge, especially on the finer points of fluid mechanics. Mike McClean for his mechanical assistance and precision work. The aide I have received from my colleagues has been immeasurable. Heidi Relyea, Jon Darrow, Al Aksan, Ramez AbdulNour, Mark Minor, Gloria Elliot, Bin Lian, Ingo John, Doug Neal and Scott Morris have all helped me out along the way. It ranged fi'om moving equipment, helping solve nasty mathematical problems, studying for class, or just listening, but it all helped me through and for that I say thanks! TABLE OF CONTENTS LIST OF TABLES. .................................................................. vii LIST OF FIGURES .................................................................. viii NOMENCLATURE .................................................................. xi CHAPTER 1. INTRODUCTION .................................................... 1 CHAPTER 2. WIND TUNNEL TEST FACILITY .................................. 4 2.1 DESIGN OBJECTIVES ................................................. 4 2.2 WIND TUNNEL BASE ................................................ 6 2.3 TEST MODULE ........................................................ 7 2.4 THROTTLE ASSEMBLY ............................................... 9 2.5 FLOW CONDITIONING ................................................ 10 2.6 DATA ACQUISITION .................................................. 12 2.7 PRESSURE MEASUREMENT .......................................... 13 2.8 HOT WIRE SENSORS .................................................. 13 2.9 CALIBRATION NOZZLE AIRFDOW HEATER ........................ 14 2.10 CHARACTERIZATION OF THE WIND TUNNEL FACILITY ......... 15 2.1 1 SUMMERY ........................................................... 19 CHAPTER 3. CALIBRATION 0F INFRAMETRICS IR CAMERA AND ASSOCIATED DATA ACQUISITION SYSTEM ...................................... 20 3.1 INTRODUCTION ......................................................... 20 3.2 STATEMENT OF PROBLEM ............................................. 20 3.3 EXPERIMENTAL PROCEDURE ......................................... 22 3.4 RESULTS ................................................................ 27 3.5 OPTIMAL CONFIGURATION TESTING ................................ 31 3.6 RESULTS OF OPTIMIZED EQUIPMENT CONFIGURATION ........... 31 CHAPTER 4. SLIT RESPONSE FUNCTION ....................................... 34 4.1 INTRODUCTION ........................................................ 34 4.2 IMAGE PROCESSING EFFECTS ON DATA ........................... 38 4.3 EXPERIMENTAL TEST PROCEDURE ................................. 39 4.4 RESULTS ............................................................... 41 4.5 PREDICTTVE CAPABILITIES ........................................... 44 4.6 CONCLUSIONS ......................................................... 47 CHAPTER 5. EXPERIMENTAL VERIFICATION 0F THERMOGRAPHIC TEMPERATURE MEASUREMENT UTILIZING SLIT RESPONSE CORRECTION ................................................... 49 5.1 INTRODUCTION ....................................................... 49 5.2 EXPERIMENTAL PROCEDURE ....................................... 49 5.3 EMISSIVITY MEASUREMENT ........................................ 51 5.4 EXPERIMENTAL VERIFICATION ...................................... 53 5.5 DISCUSSION ............................................................ 63 CHAPTER 6. CONCLUSIONS ...................................................... 69 CHAPTER 7. RECOMMENDATIONS FOR FUTURE WORK .................... 73 7.1 SLIT RESPONSE FUNCTION .......................................... 73 7.2 INFRARED CAMERA RESPONSE TO TEMPERATURE GRADIENTS. 73 7.3 HOT-WIRE SENSORS .................................................. 74 7.4 DATA ACQUISITION SYSTEM ........................................ 75 APPENDICES ........................................................................ 79 APPENDIX 1. Two DIMENSIONAL FLow CONTRAcnON .................. 80 APPENDIX 2. QUICKBASIC‘D PROGRAM To PERFORM DATA ACQUISITION AND STEPPER MOTOR CONTROL ............................. 84 APPENDIX 3. FLAT PLATE ISOTHERMAL VELOCITY PROFILES ........... 108 APPENDIX 4. HOT-WIRE CONSTRUCTION AND CALIBRATION ............. 111 APPENDIX 5. MAXIMIZING THE DYNAMIC RESPONSE OF THE INFRAMETRIX IR CAMERA AND DATA RECORDING SYSTEM ............. 121 APPENDIX 6. VELOCITY AND TEMPERATURE CALCULATION PROGRAM FOR Two HOT-WIRE METHOD .................................... 124 BIBLIOGRAPHY ................................................................... 129 LIST OF TABLES Table l. SRF summary for the Infiametrics 6OOL infrared camera in the HTRL ............................................................ 48 Table 2. Emissivity measurement results (e=0.88) ........................... 54 Table 3. Test results from test 1a (MAF sensor) (11 amps to flow heater, 42.1-62.10C IR temperature range) 56 Table 4. Test results fi'om test lb (MAF sensor) (4 amps to flow heater, 20.0-40.00C [R temperature range) 57 Table 5. Emissivity measurement results (e=0.95) ........................... 59 Table 6. Test results from test 2a (11 amps to flow heater, 32.9~52.9°C IR temperature range) 61 Table 7. Test results from test 2b (6 amps to flow heater, 23.3-43.30C IR temperature range) 62 Table 8. Null hypothesis statistical test results .............................. 68 Table 9. Two dimensional contraction machining coordinates ............ 81 Table 10. Post calibration data ................................................ 119 vii LIST OF FIGURES Figure 1. Wind tunnel test facility concept drawing ............................. 4 Figure 2. Wind tunnel test facility ................................................. 5 Figure 3. Modular concept utilized to guide flow facility design .............. 6 Figure 4. Structural drawing of sub-atmospheric test section ................. 8 Figure 5. Sluice throttle mechanism used to control airflow rates (top View) ................................................. 10 Figure 6. Airflow conditioner ..................................................... 11 Figure 7. Calibration nozzle airflow joule heater ............................... 15 Figure 8. Axi-symmetric calibration nozzle and associated coordinate system ...................................................... 17 Figure 9. Calibration nozzle exit velocity profile .............................. 17 Figure 10. Calibration nozzle exit temperature profile ....................... 18 Figure 1]. Ideal gray scale-temperature response for the Infiametrics infiared camera and associated data acquisition system ........... 21 Figure 12. IR camera tool for area temperature measurement ............... 22 Figure 13. Schematic of different equipment configurations tested .......... 23 Figure 14. Playback data image sequence from the equipment test experiment conducted on 2/11/97 ..................................... 25 Figure 15. Equipment configuration effects on the IR camera gray scale-temperature response ...................................... 28 Figure 16. Comparison of equipment tests with and without the For-A VTG-33 video timer ........................................... 29 Figure 17. IR camera gray scale output as a function of contrast setting. (Low end of Infiametrics camera temperature scale) .............. 3O viii Figure 18. Gray scale-temperature relationship with optimal contrast settings ............................................... 32 Figure 19. Optimal equipment configuration gray scale-temperature relationship ............................................................... 33 Figure 20. Schematic of IR camera optical configuration and definition of terms ....................................................... 35 Figure 21. Comparison of scan direction effect on IR camera response 36 Figure 22. Temperature profile of narrow dimension of the MAF sensor. Sensor parallel to scan direction ...................................... 37 Figure 23. Temperature profile of narrow dimension of the MAF sensor. Sensor perpendicular to scan direction .............................. 37 Figure 24. Schematic of the experimental configuration for SRF measurement .................................................. 39 Figure 25. Slit response function with 95% confidence region (n=5) for the Infi'ametrics 6OOL IR camera ................................ 41 Figure 26. SRF confidence region/mean value for Inframetrics 600L IR camera with external optics ...................................... 42 Figure 27. SRF of the Inframetrics 6OOL infrared camera with 3x Telescopic and 6” close lenses ................................... 43 Figure 28. Thermographic image of MAF sensor ............................ 44 Figure 29. Comparison of SRF between system configurations (no optics compared to external optics) .............................. 47 Figure 30. Emissivity variation observed in semi-conducting materials by Inagaki et.al., 1994 ................................................... 65 Figure 31. Correction factor (l/SRF) and % uncertainty in CF for Infi'ametrics 600L IR camera with close up optics in place as a function of target Size ................................................... 71 Figure 32. Cumulative average of the hot-wire data ........................... 77 Figure 33. Two-dimensional flow contraction .................................. 8O ix Figure 34. Compiled jet velocity profiles (w=2cm) ........................... 109 Figure 35. X/W=l .59 velocity profiles (w=2cm) .............................. 109 Figure 36. X/W=4.52 velocity profiles (w=2cm) .............................. 110 Figure 37. X/W==10.95 velocity profiles (w=2cm) ............................ 110 Figure 38. Schematic of hot-wire probe ....................................... 111 Figure 39. Tungsten wire copper plating station ............................. 112 Figure 40. Precision solder station ............................................. 113 Figure 41. TSFL hot-wire calibration flow contraction ..................... 114 Figure 42. “Coke Bottle” contraction utilized for hot-wire calibration 115 Figure 43. TSFL contraction hot-wire calibration curve .................... 1 17 Figure 44. HTRL facility hot-wire calibration curve ......................... 118 Figure 45. Optimal equipment configurations for data acquisition and processing ...................................... 122 Figure 46. Results of optimized equipment configuration ................... 123 ENGLISH: A A,B,n ATF Cd R0.) Re SRF NOMENCLATURE Area Hot-wire cah'bration coefficients fi'om Collis and Williams [1959] Aperiodic transfer fimction Discharge coefficient Diameter Voltage Blackbody emissive power (W/mz) Convective heat transfer coeflicient Infiared Radiosity Confidence region Mass air flow Pressure Prandtl number Thermal energy Radiated flux Wavelength dependent radiosity Reynolds Number (=Ux/v) Slit response function Temperature T Non-dimensionalized temperature [(T - Trange Milo/Tunge] Tm...“ True temperature value Tm Infi'ared measured temperature Tm; MAF sensor temperature T"Inge Temperature span on infrared camera Tum“... Lower limit of temperature span on infrared camera Tm: Infi'ared measured, SRF corrected temperature U Streamwise velocity V Velocity Vm Maximum velocity W Wall jet Slot width f Focal distance g Acceleration due to gravity k Thermal conductivity Id Working distance 1. Slit width 1.- Internal focal length n Number of data points t Statistical parameter from student T distribution x Streamwise direction y Direction orthogonal to streamlines and test plate 2 Direction aligned with gravity or direction orthogonal to xy plane - context willclarify xii GREEK: (I) Measurement error O Target angle .0 Electrical resistance a Absorbtivity e Emissivity A Wavelength p Density or Reflectance - Context will clarify 0' Standard deviation or Stephan-Boltzmann constant - context will clarify r Transmissivity v Kinematic viscosity xiii CHAPTER 1. INTRODUCTION 1.1 MOTIVATION The original motivation for this study was provided by a project for General Motors Corporation. The manufacturing process of mass air flow (MAF) sensors for an engine intake manifold requires that each sensor be individually calibrated. The cah’bration method utilized was expensive and time consuming. The method consisted of mounting the MAP sensor in a mold and filling the mold with an isothermal, electrically neutral and environmentally friendly liquid. This created an environment of known temperature in which the sensor resistance could be measured. The method provided a two point temperature-resistance calibration curve. The material cost for the fluid utilized was approximately 1.5 million dollars per year not incorporating the energy costs of heating it. Infrared thermography and an alternate heating method were proposed to obtain the necessary temperature/resistance information in order to replace the current cahbmtion method. The project was complicated by the size and material characteristics of the MAF sensor. The sensor is a 1500 um long and 500 pm diameter cylinder constructed from platinum wire surrounded by a glass body. The glass surface could not be coated with any materials and was expected to experience a large amount of manufacturing variability. Thus, the problem of measuring the surface temperature the small sensor Size was complicated by unknown and variable emissivity properties of the specimen. A wind tunnel facility was constructed to support and augment the Heat Transfer Research Laboratory (HTRL) and Specifically to investigate the capabilities of the infi'ared measurement equipment. This facility was designed to provide a flow of known velocity and temperature and the sensors necessary to monitor this flow. Chapter 2 details the wind tunnel facility and its construction, the data acquisition system, and the measurement transducers utilized. This facility development was essential as a test platform for the current project. The laboratory utilizes an Inframetrics 600L infiared camera and the associated data acquisition system to obtain Spatial temperature information. This system was characterized and optimized to obtain the full dynamic temperature measurement range. The effects of specimen Size on the measurement capabilities were determined experimentally utilizing the Slit response function (SRF) of the infrared sensor. The SRF is the attenuated response of the sensor to targets smaller than its measurement resolution. The combination of the flow facility and the characterization of the infrared system allow the goals of the original proposal to General Motors to be realized. 1.2 PREVIOUS WORK The capabilities and applications of infi'ared thermographic equipment have been explored by a previous investigators. The sensors response to an unresolved target (subtends an angle less than the measurement capabilities of the sensor) has been explored extensively by Holst [1993]. The target angle to produce a 50% response (the 50% Slit response fimction (SRF)) is generally provided by the manufacturer of the equipment as a measure of the performance (Inframetrics [1988]). These quantities are provided to define the limitations of the equipment so as to avoid inaccurate measurements. It is the goal of this study to quantify the Slit response function to allow meaningful measurements to be taken in these circumstances and thus to exploit the maximum capabilities of the infrared camera systems. Infiared thermography has also been applied in wind tunnel studies of fluid mechanics and heat transfer. One application has been the study of supersonic heating of structures (Lafferty and Collier [1991]). Infrared thermography has also been applied to study the surface temperature of structures in flows to measure the convective heat transfer coefficient (Willenborg, et al. [1996], Okamoto and Inagaki [1994]). Gartenburg and Roberts [1990] studied the temperature gradients on a heated wire in a flow field with an infrared imaging system. The study involved imaging an unresolved target (the wire) and compensating for the error in the temperature by using an ‘apparent’ emissivity. This method compensated for the targets emittance and Size but the study was fixed in one configuration because a change in the working distance caused a change in the apparent emittance. This study involved calibrating an infrared camera and the associated data acquisition system. An ideal linear system response to temperature was defined to maximize the dynamic range and the equipment was configured to obtain this result. The infrared camera response to unresolved targets was determined experimentally. A wind tunnel facility with well characterized flow conditions was developed to support the experimental work and a series of experiments to validate and explore the infrared data acquisition system were conducted. CHAPTER 2. WIND TUNNEL TEST FACILITY 2.1 DESIGN OBJECTIVES A wind tunnel testing facility was built to advance the laboratory testing capabilities and to generate constant temperature and velocity flow conditions. The facility was designed to allow cah'bration of various sensors, to achieve specific testing goals as well as to allow for future expansion. The entire system was designed to be modular so that any component could be replaced with minimal efl‘ort (Figure 1). System Inlet Steam Supply Armstrong Model 90 Steam Humidifier Condensate Return Duct Flow Conditioner (Honeycomb and Screens) 2-Dimensional Contraction Splitter Plate Thermally Active Test Section . Sub Atmospheric Test Chamber . Probe Traverse System . Calibration Nozzle . Wind Tunnel Base . Fluid Collection . Prime Mover . Exit Throttle Valve 1 .. i /,2 I .h 99°.“ 9‘1"? P’P’?‘ l \O au-II—lu—I NI—‘o 12 —a w 11 u— an 10 u—I—D 05M /16 13 . Figure 1. Wind tunnel test facility concept drawing. Several physical quantities are controlled or measured in this facility so that the target emissivity and temperature can be obtained with infi'ared thermographic techniques. These include: pressure (and thus mean flow velocities), velocity and temperature of the air flow. Humidity of the air flow was measured but not controlled in this study. Specific controls or measurement sensors are described and their application to the infrared testing procedure are defined in this chapter. Figure 2. Wind tunnel test facility. 2.2 WIND TUNNEL BASE The base of the testing apparatus was designed to fulfill several functions. The base design followed the modularity concept for the entire system. This concept allowed different flow configurations with respect to the test piece and the gravitational field to maximize testing flexibility. The design allows for the air flow to be oriented in the same or opposite direction as the gravitational field (Figure 3). Air flow orthogonal to gravitational field can be obtained by turning the system on its side. The base structure houses the prime mover for the system. A condenser and drain section may also be added to remove condensed water vapor for future experiments. The test section was supported physically by the wind tunnel base section. The entire structure was sealed to maintain a vacuum during operation. Figure 3. Modular concept utilized to guide flow facility design. The base was constructed with 2”x4” boards for the support structure and covered with sheets of 15/32” plywood. These materials were chosen to reduce costs and allow for easy and inexpensive modifications. The assembly was completed with 3/8” hex head bolts and wood screws to allow for easy disassembly if necessary. The joints were sealed with GE silicon sealant. One end of the base structure was designated as a door and made removable to allow access to the interior. The access regions were taped with closed cell foam tape to seal the unit while it was operational. The interior wood surfaces were treated with a wood preservative to protect against water damage resulting from high humidity experiments. 2.3 TEST MODULE One of the primary design requirements for the test section was to allow variable orientation of the test section with respect to gravity. The test section was thus designed as a cube so that the sides could be interchanged to accommodate difi‘erent testing configurations. The simplicity of this design (Figure 4) was its greatest strength. Two sides of the module have 0.375” thick plate glass to allow optical access to the interior environment. One side was designed with a removable door and the remaining side was solid. The top and bottom of the module have openings appropriate to the air passage through the respective sides. The flame of the test section was constructed of 2”x2” angle iron with a 15/32” skin of plywood. The plywood skin was bolted onto the iron fi'ame for easy modifications and reconstruction. The interior of the testing module was treated with shellac to prevent degradation of the wood during experiments with high humidity. Four extra 2”X2” angle iron support members were included in the original design. These members (Figure 4) can be utilized for additional support if necessary or to mount equipment in the testing environment. To p \71 0 w ' T i ' ’ j/ ("eiillbr'alion if M Nozzle i3 IQ)III'Ir..-IISI()ITIII CR)HIITICII(HI IITICI'IarC f Front View Side View L3 (5 X 2'3 10 ‘ ‘ , 'r‘ ““““““““““““““““ j ‘ Class [“h“ i 53 X 23 l i .. ‘ » - I I \l“111. (1) Assume Vle, and -g(Z1-Zz) zO results in: V2 = M (2) i ,0 These equations assume incompressible, inviscid, steady flow along a stream line through the nozzle). The pressure can be controlled by sluice position to increments of approximately 0.001 in-H20 which was within the resolution limits (0.0025 inHzO) of the pressure transducer. This corresponds to a velocity resolution of approximately 1 5 cm/s at typical experimental flow settings. Figure 5. Sluice throttle mechanism used to control airflow rates (top view). 2.5 FLOW CONDITIONING A structure was built for flow conditioning at the inlet of the system. This structure was designed to remove variability caused by changing ambient conditions and was necessary to insure experimental reproducibility and flow homogeneity. This system consists of a preconditioner and a flow contraction (Figure 6). The pre- conditioner consists of 3 wire screen stretchers and a honeycomb straw structure. The contraction is two-dimensional and constructed from 1/4” thick Lexano. Figure 6. Airflow conditioner. The design of the preconditioner was based on a design from the TSFL and the assistance of Professor Foss. The honeycomb structure removes large scale turbulent momentum eddies from the fluid. The wire screen used consisted of 30 x 30 wires per square inch mesh. This mesh removes smaller scale eddies from the fluid. The wire meshes were placed 3.5” apart. The disturbances created by the previous mesh are greatly dampened prior to reaching the next mesh (Blevins [1984]) at this distance. A two dimensional contraction was utilized to accelerate the flow and to provide a uniform velocity profile at the contraction exit plane. The wall curvature was the critical design criteria for the flow contraction. The design was obtained fiom the principles defined in “Design of Two-Dimensional Wind Tunnel Contractions” by Morrel [1977]. The mechanical details of the contraction are presented in Appendix 1. This curvature prevents thick boundary layers fi'om being produced at the exit plane of the contraction and it prevents flow separation in the contraction resulting from flow acceleration. The design accomplishes these ends by maintaining favorable pressure gradients and smooth transitions throughout the length of the contraction. 2.6 Data Acquisition System and Motion Control A Keithley MetraByte DAS-TC data acquisition board and a DAS-02 analog output board were utilized with sofiware written in QuickBasic° (Microsoft [1988]) (the QuickBasic program is presented in Appendix 2). These two boards can sample 16 channels of data and are capable of 2 channels of analog output. The DAS-TC board was utilized to obtain temperature and voltage measurements and the DAS-02 to output voltage control signals to a VCO-4050 stepper motor controller. The voltage measurements were from the hot wire anemometer, the pressure transducer, the humidity probe and the stepper motor control signal. The VCO-4050 stepper motor controller and a CMD-SO chopper driver from American Precision Inc. and the Eastern Air Devices Model #LA23ECK-81 step motor with the custom built traverse system allowed the hot wire sensors to be moved in increments of 6.6 pm. 12 2.7 Pressure Measurement A MKS Baratron Type 225A pressure transducer with a V2” H20 full scale reading was utilized to measure pressure differentials in the system. These pressure measurements were used to calculate velocities for hot wire calibrations and to meter the flow for the actual experimental work. The MKS transducer can resolve pressure measurements with an accuracy of i0.0025” H20. The relationship between velocity and pressure in the calibration nozzle is defined in Equation 1. The transducer was utilized to read the pressure drop across the contraction as well as the pressure difference between the atmosphere and the interior of the test section. Measurement of the pressure drop through the contraction allows for the inviscid core velocity in the test section to be calculated. This velocity was utilized to non-dimensionalize the experimental results. The pressure difference between the atmosphere and the sub- atrnospheric test section allows the inviscid velocity through the cahbration nozzle to be calculated. 2.8 Hot Wire Sensors The hot wire sensors utilized in this facility consist of 5 mm diameter tungsten wire controlled by a TSI constant temperature anemometer (CTA) system (model 1051-2). The wires are powered electrically to produce an elevated temperature to maintain the prescribed resistance value. The CTA system utilizes a bridge circuit to control the current through the hot wire sensor and thus its resistance. The CTA system outputs the voltage signal required to maintain the sensor resistance. This voltage Signal can be correlated with a known flow rates (through the calibration l3 nozzle) to calibrate the hot wire sensor so that it can be used to measure unknown flow velocities. The calibration results fi'om a balance between the electrical heating of the sensor and the convective heat transfer. The governing equation used during the calibration is E 2 = A —— B V ". (3) Essentially, the square of the voltage output (E) from the anemometer bridge was related to the velocity (V) by three calibration coefficients (A, B, and n). The hot wires are capable of measuring velocities within the range of the calibration to a resolution of 352%. 2.9 CALIBRATION NOZZLE AIRFLOW HEATER A stainless steel 30 x 30 wires per inch mesh screen was utilized to heat the flow through the calibration nozzle (Figure 7) to provide an thermal energy source to heat the thermographic targets. The mesh has a resistance of 2 Q and was heated electrically with a HP6274B DC power supply. The low resistance of the screen requires a power supply with a high amperage output to supply adequate energy to the flow. The power supply was capable of providing 15 amps (30 watts). l4 30 x 30 Stainless Steel Mesh Electrical Input Calibration Nozzle Figure 7. Calibration nozzle airflow joule heater. The screen was able to create a temperature rise of 30°C above ambient when powered by 8 amps with an output flow velocity of 10 m/s through the calibration nozzle. The core of the flow has a measured temperature stability of i l .5 °C. The design utilized presently was limited to 9 amps due to electrical arcing. 2.10 CHARACTERIZATION OF THE WIND TUNNEL FACILITY Velocity profiles of the flows through the calibration nozzle and the two- dimensional contraction exit were obtained. Only the calibration nozzle was used in the current study. However, the test section results were obtained to explore the capabilities of the facility and there are presented in Appendix 3 for completeness. Figure 8 Shows the calibration nozzle schematically and the coordinate system used. Figure 9 shows the velocity profile at X/D = 0.5 in the calibration nozzle. The majority of the exit area of the nozzle has a laminar uniform velocity profile. The flow was determined to be laminar from the hot wire voltage signal. This signal appeared smooth on an oscilloscope and had an RMS value of 2% of the mean flow value. It was important to note that the Reynold’s number, U-D ReD = U a (4) based on the flow velocity and the nozzle exit diameter was equal to approximately 13,000. This number seems high enough to indicate that the flow should be turbulent, but because the flow was still developing, the central region of the nozzle was still unsheared and laminar. l6 Flow Figure 8. Axi-symmetric calibration nozzle and associated coordinate system. Calibration nozzle velocity profile VNInvIscld Nozzle Exit ~~ Dia=21mm , l 5 10 15 20 Position (mm) Figure 9. Calibration nozzle exit velocity profile. 17 Temperature profiles of the air at X/D = 0.5 below the exit plane of the calibration nozzle were also obtained. The profile for 10 m/S inviscid flow velocity and ambient temperature equal to 22°C with 8 amps and 6 amps provided to the flow heater are shown in Figure 10. Temperature profile at calibration nozzle exit (10 mls flow velocity) ac +8Anperes m —-x— 6 Anperes ___ __ .- __- L. .._ . _.._.._.,—__.———-r___—._._4 Temperature (C) -15 Position (m rn) Figure 10. Calibration nozzle exit temperature profile. This figure shows that there was a core of relatively constant temperature approximately 1 centimeter in diameter for each current setting. This was much larger than the specimens that were tested and was considered adequate for this study. The stability of :1 .5 0C was larger than the uncertainty of the infrared camera (: 03°C) utilized but appeared to be the stability limit of the heating system used. The room 18 temperature above the nozzle inlet was monitored and it experienced a standard deviation of approximately 3: 1 0C over time period of 1 minute. This variation in the ambient conditions probably accounts for the majority of the fluctuation found in the nozzle temperatures. 2.11 SUMMARY A wind tunnel facility was developed with the capability to create and monitor a flow with acceptably uniform velocity and temperature profiles in core regions of the calibration nozzle. The velocity, temperature and humidity of the flow can be individually measured and recorded with the data acquisition system. Fine motion control to position the sensors was obtained with stepper motors and a translation stage. This facility was characterized specifically to obtain measurements with infi'ared thermography of temperature and emissivity of optically unresolved specimens and to support other research projects in the Heat Transfer Research Laboratory (HTRL). l9 CHAPTER 3. CALIBRATION OF INFRAMETRICS IR CAMERA AND ASSOCIATED DATA ACQUISITION SYSTEM 3.1 INTRODUCTION Infrared cameras produce temperature data as a function of spatial location and time. The appropriate data recording system typically includes a time marking video overlay, a video recording system (VCR), video monitors and image processing sofiware to analyze the data. The video overlay allows time data to be saved simultaneously with the temperature field information. The video recording system allows data storage and post processing of the information. The experiments can be viewed in real time on the video monitors. The sofiware allows digitizing of the video signal and provides necessary quantitative analysis tools. The following equipment were available to achieve these functions: an Infiametrics 600L infrared camera; a For-A VTG-33 video timer; a Sony PVM-l343MD Trinitron monitor; a Panasonic AG-2400 monitor and Image Pro Plus v2.1 (Media Cybernetics [1995]). The interfacing and optimization of this system to obtain optimal experimental analysis is explored in this chapter. An ideal system response is outlined and an equipment arrangement to achieve that response is defined. 3 .2 STATEMENT OF PROBLEM The Inframetrics infiared camera outputs an 8 bit digital signal in R8170 video format that includes the temperature field information. An ideal system output is defined 20 for this work as linear response that spans the entire dynamic range (Figure l 1). The linearity of the output maintains uniform sensitivity over the experimental range and Simplifies processing. Ideal Output of the Infi'ametrix IR Camera 250 200 I- Guy 150 4. Scale100 0 50 AL L—o— Ideal Response , I 0 .L r J. t i 0 0.2 0.4 0.6 0.8 1 Non Dimensional Temperature Figure 11. Ideal gray scale-temperature response for the Inframetrics infrared camera and associated data acquisition system. Different equipment or equipment arrangements are often utilized in the data recording circuit depending on the experiment, experiment location (the camera is portable), and equipment available . There was evidence that different equipment combinations could result in variations in the data when data obtained at the top of the lnfrarnetrics camera range would saturate at different levels for different users. It was determined that this was directly caused by the data recording and processing equipment. 21 3.3 EXPERIMENTAL PROCEDURE 3 .3 . 1 EXPERIMENTAL CONFIGURATION An PIP-6236B Triple output power supply powered a plate heater to a temperature between 5 and 10 degrees centigrade above ambient temperature. The plate temperature was monitored with a calibrated T-type thermocouple (i0.1 0C accuracy). The IR camera was then set with its temperature range lower limit just below the plate temperature. The system was allowed to stabilize for 15 minutes to insure it was at steady state. The IR camera was then used to define the plate temperature field with several different equipment arrangements. The area fiInction of the IR camera was used to define the region of interest (Figure 12). This insured that the same surface area was analyzed each time. This avoided variability due to non-uniform surface properties on the plate. Figure 12. IR camera tool for area temperature measurement. 22 Data were obtained with a variety of equipment configurations. These equipment arrangements were chosen to represent potential data acquisition arrangements or systems that are currently being utilized in the laboratory. The different equipment configurations that were tested are shown schematically in Figure 13. Soliciiinlic of equipment l(’~'.\‘t (-onfiglirritiOIiS [P Plus IR I: (fi‘mmnrd Computer IP Plus [R Sony l: (i‘amerzi \lonilor Computer 1P Plus IR Sony C (Mignon-T \-'(_‘R Monitor Computer IP Plus IR Video Sony ‘ ., E ("x-”nerd ”‘l Timer ‘ Monitor Computer Fr‘-l'ir*I'I'ie-ili~ E O 100 «» + Playback w/o ‘ Titer l 50 «» —I— Playback wl Trier i (750hm on) I 0 P t i 1r l A. t i 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 Nondimensional Tern perature Figure 16. Comparison of equipment tests with and without the For—A VTG-33 video timer. The Panasonic AG-2400 VCR did affect the RSI 70 signal and thus the gray scale to temperature relationship. Specifically, it created a negative offset on the RS-170 signal. This counteracted the positive offset fi'om the camera and was necessary to capture the entire dynamic capabilities of the IR camera. The best results were obtained with the VCR in the data acquisition circuit but not recording the data stream. This arrangement would be possible during steady state tests, but is impractical due to the time required to process the data from a typical experiment with the Inframetrics camera. Based on these results, the recommended equipment configuration for data acquisition is to connect the Infiametrics IR camera serially to the For-A VTG-33 Video Timer, to the Panasonic AG-2400 VCR, to the Sony PVM-l 343MD Trinitron monitor (in that order). 29 The results of the brightness and contrast tests are shown in Figure 17. The brightness setting moved the overall curve upward, maintaining a constant slope. The contrast setting altered the slope of the resulting line. These settings appeared to operate independently of one another. These tests showed that the best settings to maximize the sensitivity were Brightness 48, Contrast 66. This setting was selected to lower the y- intercept of the gray scale as a function of temperature response. Figure 17 may make a contrast setting of 65 look like the best way to lower the response. However, this setting limited the upper response to a gray scale value of approximately 150 (thus sacrificing a large portion of the dynamic range). Based on these results, the optimal data processing system to achieve the ideal response shown in Figure 11, is the Panasonic AG-2400 VCR connected to the Sony PVM-1343MD monitor to the Image Pro Plus software (Figure 13). Gray Scale as a function of Contrast Setting 250.00 1 5 re H—w—fi ". ‘ 1 200.00 x + Brightness=48 + Brightness-=50 a Bnghtness=55 150.00 T + Bnghtness=65 Gray ' Scale 100.00 R l 50.00» g l 0.00 r i t i J. 60.00 70.00 80.00 90.00 100.00 110.00 Contrast Setting Figure 17. IR camera gray scale output as a function of contrast setting. (Low end of Infrarnetrics camera temperature scale). 30 3.5 OPTIMAL CONFIGURATION TESTING An optimal data recording system and a data playback system were defined from the results discussed above. The data recording system consisted of the Infiametrics IR camera connected to the For-A VTG-33 Video Timer and the Panasonic AG-2400 VCR (in that order). The data playback system consisted of the Panasonic AG-2400 VCR connected to the Sony PVM-l343MD Trinitron monitor with the signal then fed to the Image Pro Plus software (Figure 13). This optimal equipment configuration was then tested. The test procedure was altered to reduce the time required. The plate was powered with the PIP-6236B Triple output Power Supply to approximately 10 degrees above ambient. The plate was allowed 20 minutes to stabilize thermally. During these experiments, the temperature range of the Inframetrics IR camera was adjusted instead of changing the plate temperature. This procedure had the same effect on the IR sensor as adjusting the plate thermally. Several verification tests were run to insure that the results were identical for each procedure. This modified test procedure allowed for many more data points to be taken throughout the temperature range. This procedure allowed verification of the linearity of the Infrarnetrics IR camera output. 3.6 RESULTS OF OPITMIZED EQUIPMENT CONFIGURATION Previous work (White [1996]) had Shown linear gray scale response to temperature but with a different Slope and concluded that the offset was a characteristic of the radiometer itself. This previous study did not incorporate the affects of the contrast and brightness settings of the Image Pro Plus software which allow the offset to be 31 adjusted. Figure 18 and Figure 19 Show the IR system response with optimal brightness and contrast settings and a comparison to previous calibrations obtained without these controls. The equations in the upper corner of Figure 18 Show the system insensitivity to contrast settings between 66 and 70. The correlation coefficient (R2) values Show the degree of linearity. The x-axis of these plots is presented in terms of non-dimensional temperature (T') which is defined as T. T- T(lower IR range limit) _ . 5 IR camera Range (i. e. 10° C) ( ) Optimal contrast tarts 250.0 ,_ B48066 Result /)T y = 246.36x + 7.6393 200.0 ,_ R2 = 0.9992 Gm y Scale “DDT 500" 0 OJ 02 03 0A 05 03 OJ OS 09 1 Non Dimensional Temperature Figure 18. Gray scale-temperature relationship with optimal contrast. 32 Optima] equrpment tests 20 degree IR range 250.0 .. Test I Result y = 240.07x + 5. I882 200° T R2 = 0.9996 9.) 3 150.0 _0_ Ideal R5311]! y = Z55x E E3 100.0 ~- —x-— B48C6’6‘ Tea 1 500 w ' - B48C6'6‘Tcst2 I '—i—- Heal Resrxmse 0.0 '// r I 4 + l 0 0.2 0.4 0.6 0.8 1 Non Dimensrbnal Tempera tare Figure 19. Optimal equipment configuration gray scale-temperature relationship. The system response, Gray Scale = 2455- T‘ + 7.7, (6) is almost identical to the optimal system response defined at the beginning of the investigation. The ideal response includes 100% of the dynamic range over the entire temperature scale. The optimized equipment configuration response is linear and provides 95% of the dynamic range for experimental work. The results presented here should also be considered typical for any range setting on the IR camera but the specific experimental configuration should be characterized prior to each experimental sequence with the relevant equipment. The procedure described can be run in a reasonably Short time flame and will help remove bias error created by the data acquisition system. 33 CHAPTER 4. SLIT RESPONSE FUNCTION - THEORY AND EXPERIMENTAL DERIVATION 4.1 INTRODUCTION The Slit Response Function (SRF) is a measure of the ability of an infrared system to sense radiation from targets that subtend an angle which is smaller than the measurement resolution. An observable target smaller than the measurement resolution will produce a data signal that is attenuated as a function of target size. The SRF must be known so that these Signals can be processed correctly and accurate measurements obtained. The lower thermal and optical measurement limits of the system can be defined if the SRF is known and corrections can be applied. The instantaneous field of view (IFOV) and the field of view (FOV) are important factors in determining system performance. Generally a large field of view is desired with a small instantaneous field of view. The field of view is the total area the camera can sense and the instantaneous field of view is the projected detector area (Figure 20). As an analogy to a digitized image, consider the field of view the number of pixels and the instantaneous field of view as the pixel Size. 34 lnstmrtaneous Field of View (projected detector size) IR Camera Field Target of View Figure 20. Schematic of IR camera optical configuration and definition of terms. Scanning infrared systems tend to have different SRFs in the scan and cross scan directions due to factors such as detector size and shape, line to line interpolation schemes, optics and scan rates. The SRF for a given system defines the limits on imaging and measurement resolution. The optical (imaging) resolution is typically defined as the target angle that will produce a 50% response (that is, it will measure 50% of the full scale temperature above or below background). The measurement resolution is defined as a target width that will produce a 99% response (Holst [1993]). The measurement resolution defines the smallest target angle that may be thermally measured accurately. The effects of the scanning direction and target alignment can be seen in Figure 21, Figure 22 and Figure 23. The scan direction is from left to right. The MAF sensor is at the same temperature for both images (T ~ 56 0C). The sensor aligned with its narrow dimension in the scan direction did not produce as large a response as the sensor aligned with the scan direction. 35 Figure 21. Comparison of scan direction effect on IR camera response. The MAF sensor alignment with respect to the scan direction affects the length of time that the infrared detector has to respond to the radiation emitted. Figure 22 shows a much flatter response to the target than Figure 23. This is because the SRF is created by step changes or steep gradients in the emitted radiation. When the sensor is aligned with the scan direction, its long dimension allows the infrared detector more time to equilibrate. Therefore, the scan direction defines the critical dimension related to accurately measuring temperature and was the focus of the study. 36 Non-dimensional temperature 0.7 Gray scale profile of narrow dimension of sensor Sensor parallel to scan direction 0.6 a+r+p=1. (15) The energy incident on the target (QR) must either be absorbed (a), transmitted (r) or reflected (p). This is further reduced to 8 = 1 — p (16) if the target is opaque (its transmissivity (r) is equal to zero) and the gray body assumption ofa=€ (Kirchofl’ S law) (Weast [1986]) is applied. 51 This method can be difficult to implement. Difficulty arises from three requirements of the method. The reflection of the two reference sources must be viewed on the test specimen. The reflection dimensions must be known if the slit response function correction is to be applied. The reflection can be very difficult to measure accurately. Finally, if the test Specimen has a high emittance (absorptivity) (>0.8) obtaining a reflection that can be imaged clearly is almost impossible unless very high temperature reference sources are utilized. The second method, which was employed in this study and included in step 8 of the procedure in section 5.2, was to place the target in an environment such that its temperature was known. The dimensions of the target must be known prior to the application of this method because the unresolved target requires the SRF correction to measure temperature. The isothermal environment was obtained by using the calibration nozzle and the electric flow heater (Chapter 2, section 2.9) to provide a jet flow with a known, uniform temperature profile (within : 1.5 0C). It is important to raise the target temperature above ambient to increase the ratio of emitted radiation to reflected radiation (effectively, the signal to noise ratio). The target temperature can then be measured directly with the infrared camera (with the slit response function correction applied). The emissivity setting on the camera can then be manipulated until the gray scale output matches the calculation from the appropriate transfer firnction. It is important to note that although the inflated imaging system was calibrated previously (Chapter 3), new software and a new VCR were utilized during the experimental verification. The addition of the new equipment allowed for the ideal response (Equation 12) to be realized and thus Equation 12 was utilized throughout this chapter. 52 It is critical to use the SRF correction to accurately measure the emissivity. An ‘apparent’ emissivity will be measured without the SRF correction. This restricts the experimental configuration to the original working distance (Gartenberg and Roberts [1990]). Any deviation from this working distance will cause the ‘apparent’ emissivity to change and introduce systematic errors to the measurements. 5.4 EXPERIMENTAL VERIFICATION Two experimental procedures were conducted. The first procedure utilized the MAF sensor shown in Figure 28. This target required that the external optics were utilized and produced an SRF value of 0.9. The second target was a metal cylinder (painted black) with a diameter of 1.75 mm. The IR camera was positioned 34 cm away fi'om the target and the external optics were not utilized. This produced a target with an SRF of 0.57. The cylinder could have been viewed from closer or with the external optics to obtain a well resolved target. However the test was conducted to investigate the SRF correction. In the first procedure, the MAP sensor was mounted below the calibration nozzle in the HTRL facility and the procedure in section 5.2 was followed through step 8. The IR camera emissivity setting was adjusted, an image was acquired and processed, and the target temperature was calculated from Equation 12. This procedure was followed until the target temperature matched (approximately) the flow temperature measured by the calibrated thermocouple. This required processing each data point prior to proceeding. Six data points were taken with the final emissivity setting. Each thermographic image was 53 the average of 16 times to reduce the effects of noise. The results of the emittance determination are Shown in Table 2. Table 2. Emissivity measurement results (e=0.88). SRF Corrected Measured Gray SRF Corrected Gray Temperature (0 C) Scale Scale (Equation 12) Test 1 225 250 49.6 Test2 238 264.4 50.8 Test 3 240 266.7 50.9 Test 4 225 250 49.6 Test 5 226 251.1 49.7 Test 6 223 247.8 49.5 Mean 229.5 255 50.0 Variability (95% 14.6 16.2 11.58 confidence level) Thermocouple 49.6 :2.8°C (95% measurement data confidence) The SRF corrected gray scale values averaged 255 and some values exceeded the 255 limit of the 8-bit scale. Section 4.5 warned about the possibility of introducing error by extrapolating temperatures beyond the IR temperature range setting (or gray scale values greater than 25 5). The gray scale - temperature relationship is assumed to be a linear relationship in this analysis. This relationship potentially becomes non-linear beyond the IR camera range setting. Preliminary experiments showed that the temperature calculation became less reliable if temperatures beyond the infrared camera temperature range were extrapolated. That affect may be present in these validation experiments as a decrease in the predictive accuracy and an increase in the standard deviation. The variability presented in all of the tables in this chapter are the experimental variance (L) defined at a 95% confidence level with 4:2 ‘2 L: ”a , (17) where n is the number of data points, a' is the sample standard deviation and t is the variable from the Student’s t statistic. The power to the flow heater above the cah'bration nozzle was then set to two different levels to simulate unknown operating conditions. A calibrated thermocouple was utilized to monitor the flow conditions. The results of tests 1a and 1b are shown in Table 3 and Table 4 respectively. 55 Table 3. Test results from test 1a (MAF sensor) (11 amps to flow heater, 42.1 - 62.10C IR temperature range). SRF Corrected Measured Gray SRF Corrected Gray Temperature (0 C) Scale Scale (Equation 12) Test 1 253 281.1 64.1 Test 2 211 234.4 60.4 Test 3 159 176.7 55.9 Test 4 197 218.9 59.2 Test 5 252 280 64.0 Test 6 215 238.9 60.8 Test 7 190 211.1 58.6 Test 8 209 232.2 60.2 Test 9 232 257.8 62.2 Test 10 219 243.3 61.1 Test 11 222 246.7 61.4 Test 12 250 277.8 63.8 Test 13 206 228.9 60.0 Mean 216.5 240.6 60.9 Variability (95% 52.4 58.2 15.1 confidence level) Thermocouple 61.0 i 3.4°c (95% measurement data confidence) 56 Table 4. Test results fiom test 1b (MAF Sensor) (4 amps to flow heater, 20.0 - 400°C IR temperature range). SRF Corrected Measured Gray SRF Corrected Gray Temperature (0 C) Scale Scale (Equation 12) Test 1 210 233.3 38.2 Test 2 191 212.2 36.6 Test 3 210 233.3 38.2 Test 4 229 254.4 39.9 Test 5 228 253.3 49.8 Test 6 181 201.1 35.7 Test 7 201 223.3 37.4 Test 8 225 250 39.5 Test 9 231 256.7 40.0 Test 10 203 225.6 37.6 Test 11 195 216.7 36.9 Test 12 223 247.8 39.4 Test 13 185 205.6 36.1 Test 14 229 254.4 38.1 Mean 208.6 231.8 38.1 Variability (95% 33.0 36.6 :3.15 confidence level) Thermocouple 36.6: 196°C (95% measurement data confidence) 57 The second experimental verification utilized the Inframetrics 600L infrared camera with no external optics. A metal cylinder with a 1.75 mm diameter was placed under the calibration nozzle in the HTRL facility and the procedure outlined in section 5.2 was executed. The cylinder was positioned 34 cm in front of the infrared camera. This created a target angle of 2.92 mrad and the resulting SRF was calculated from a curve fit fi'om Figure 25 as 0.57. The IR camera emissivity setting was adjusted, an image was acquired and processed, and the target temperature was calculated fi'om Equation 12. This procedure was followed until the target temperature matched (approximately) the flow temperature measured by the calibrated thermocouple. This required processing each data point prior to proceeding. Six data points were taken with the final emissivity setting. Each thermographic image was the average of 16 frames to reduce the effects of noise. The results of the emittance determination are Shown in Table 5. 58 Table 5. Emissivity measurement results (6:0.95). SRF Corrected Measured Gray SRF Corrected Gray Temperature (0 C) Scale Scale (Equation 12) Test 1 218 382.5 55.8 Test2 217 380.7 55.6 Test 3 216 379.0 55.5 Test 4 212 371.9 55.0 Test 5 216 379.0 55.5 Test 6 237 415.8 58.4 Mean 219.3 384.8 56.0 Variability (95% 17.43 30.58 12.9 confidence level) Thermocouple 55.4 :1.7°C (95% measurement data confidence) The power to the flow heater was then adjusted to two different settings. The system was allowed to equilibrate and images were taken of the metal cylinder. Sixteen images were taken and averaged at every data point to reduce electronic noise. The results of the two experiments are shown in Tables 6 and 7. Each of the experimental results of the temperature measurement (Tables 3,4,6 and 7) includes a different number of tests. This was due in part to images that were taken but were saturated and were discarded. The images were post processed and therefore saturation was not detected until after the experiment was over. Also, each test series ended when the IR camera displayed a “low liquid nitrogen” signal. This indicates that the 59 dewar required refilling and this effectively ended the test series because the IR camera was closed in the HTRL flow facility. 60 Table 6. Test results fi'om test 2a (11 amps to flow heater, 32.9 - 52.90C IR temperature range). SRF Corrected Measured Gray SRF Corrected Gray Temperature (0 C) Scale Scale (Equation 12) Test 1 224 393.0 63.6 Test 2 213 373.7 62.09 Test 3 216 379.0 62.5 Test 4 223 391.2 63.5 Test 5 237 415.8 65.4 Test 6 238 417.5 65.5 Test 7 232 407.0 64.7 Test 8 228 400.0 64.2 Test 9 239 419.3 65.7 Test 10 239 419.3 65.7 Test 11 245 429.8 66.5 Test 12 245 429.8 66.5 Test 13 246 431.6 66.6 Test 14 247 433.3 66.8 Test 15 245 429.8 66.5 Test 16 221 387.7 63.2 Test 17 207 363.2 61.3 Test 18 213 373.7 62.1 Test 19 224 393.0 63.6 Mean 230.6 404.6 64.5 Variability (95% 25.3 44.4 :3.7 confidence level) Thermocouple measurement data 63.5: 232°C (95% confidence) 61 Table 7. Test results from test 2b (6 amps to flow heater, 23.3 - 433°C IR temperature range). SRF Corrected Measured Gray SRF Corrected Gray Temperature (0 C) Scale Scale (Equation 12) Test 1 172 301.8 46.9 Test 2 165 289.5 45.9 Test 3 167 293.0 46.2 Test 4 172 301.8 46.9 Test 5 172 301.8 46.9 Test 6 173 303.5 47.0 Test 7 182 319.3 48.2 Test 8 166 291.2 46.0 Test 9 168 294.7 46.3 Test 10 170 298.2 46.6 Test 11 172 301.8 46.9 Test 12 174 305.3 47.2 Test 13 176 308.8 47.4 Test 14 177 310.5 47.6 Test 15 177 310.5 47.6 Test 16 166 291.2 46.0 Test 17 183 321.0 48.4 Test 18 182 319.3 48.3 Mean 173 303.5 47.0 Variability (95% 11.0 19.4 11.6 confidence level) Thermocouple 43.7: 16°C (95% measurement data confidence) 62 5.5 DISCUSSION The infrared measured temperatures fiom Table 2 and Table 5 agree with the temperature measured with the calibrated thermocouple within the combined uncertainty of the IR camera and the thermocouple. The best match attainable was with the emissivity set on the IR camera at 0.88 and 0.95 for each specimen respectively. The emissivity setting is adjustable in increments of 0.01 and the afl‘ect on the gray scale output is highly non-linear (Herring [1992]). A slight change in the emissivity was noted to have a large effect on the gray scale (and therefore temperature) output fiom the infi'ared imaging system during the emissivity determination portion of the experiment. Measuring the emissivity over a wide variety of temperatures could result in better measurement accuracy for future work. The SRF correction was critical to obtain accurate results in tests included in Table 3, Table 4, Table 6 and Table. 7. The effect of the SRF correction is more visible in test 2 (Tables 6 and 7) due to the smaller target angle (and smaller SRF value). If the gray scale outputs in test 2 would have been utilized directly the results would have been as follows: Test 2 (part a) Gray Scale = 230.6 Uncorrected Temperature = 509°C SRF Corrected Gray Scale = 404.6 Corrected Temperature = 645°C Flow Temperature = 635°C 63 Test 2 (part b) Gray Scale = 173 Uncorrected Temperature = 368°C SRF Corrected Gray Scale = 303.5 Corrected Temperature = 47.00C Flow Temperature = 437°C The infrared camera measurements fiom part a in both tests (Table 3 and Table 6) appear more accurate than the results from part b (Table 4 and Table 7). The infrared measurement results in Table 7 are the least accurate. The confidence regions of the temperature measured by the IR camera and the thermocouple do not overlap in Table 7. This could be the result of increased experimental uncertainty fi'om the lower temperature in test 2b decreasing the ratio of emitted radiation to reflected radiation (Signal to noise). The inaccuracy of the SRF prediction in Table 7 may also be compounded by the extrapolation beyond the temperature range setting on the Inframetrics 600L infrared camera. It is also interesting to note that the 95% confidence regions on the thermocouple and the infrared measurements are the smallest in this particular test. It may be that the narrow confidence regions are somewhat of an anomaly and that re-testing would Show that the confidence regions would overlap. A third possibility is a change in the emissivity of the component. Other researchers have noted that the apparent emissivity of semi-conductors can vary over the temperature ranges investigated here (Okamoto and Inagaki [1994], Inagaki et a1 [1994], Daryabeigi et a1. [1992]). Figure 30 presents results fiom Inakaki et.al [1994]. 1.2 "'- [10030-0345 "" [13553013292 W “’ 1.0 ' 3: 15 n ‘3 1.3 0.8 6 05 ‘ ‘ ‘ ‘ ‘ ‘ 4 ‘ L ° 280 300 320 340 360 330 Radiation Temperature Ts (K) Figure 30. Emissivity variation observed in semi-conducting materials by Inagaki et.al., 1994. The infrared measured temperature would appear to be insensitive to variance in the emittance if the radiosity - black body temperature relationship, T,T, —1— 1;, 5.01:“ ”8): ”E” . (18) is considered, where E, is the equivalent blackbody emission and J is the radiosity measured by the inflated sensor (Welch and Van Gemert [1995]). The radiosity, J, is proportional to the fourth power of absolute temperature (the Stephan-Boltzmann law). The blackbody emission - temperature relationship is also non-linear and for the 8-12pm wavelength band maybe curve fit with a fifth order polynomial (Welch and Van Gemert [1995]). The Infrarnetrics infiared camera utilizes an internal transfer fimction in the 65 hardware to manipulate this relationship to obtain temperature and electronically adjusts for the reflected radiation. At elevated target temperatures, the ratio of the emitted radiation to the reflected radiation (from the environment) is larger and thereby reduces error by reducing the signal to noise ratio. A variation in the measured radiosity will be approximately proportional to the fluctuation in the temperature, a variation in the emissivity will affect the temperature result differently at different temperatures with the sensitivity to emissivity variation increasing with the temperature. Valvano and Pearce [1995] state that “. . .all of the errors in thermal imaging contribute to an under-estimation of the true surface emitted flux, eEb, and thus an underestimate of surface temperature is obtained”. A Simple analytical model relating the black body emission and the camera radiosity measurement was created. A trend similar to the one presented was noted, the computed IR output was consistently below the “true” temperature and the error was increased at smaller temperature differences between the target and the ambient conditions. This could be corrected by decreasing the emissivity at lower temperatures indicating that the emissivity perceived by the IR camera decreases as the target temperature approaches the ambient condition. However, these results presented could not be replicated exactly. It is also unclear if this is a change in the actual emissivity or the emissivity as perceived by the infrared camera in specific wavelengths viewed. However, either effect will alter the measurements. The SRF correction does improve the accuracy of the infi'ared camera when imaging unresolved targets. The measurement error ((1)) as defined in 66 (ISRF — new) (T... — T...) ' was reduced by at least 50% (47.8% in Table 7) in the results presented (Tsar is the SRF (19) corrected temperature, Tm was the uncorrected IR temperature measurement). Typically the correction adjusted the infrared measurement so that the actual temperature was included within the measurement uncertainty range of the infrared camera. In the worse case, the thermocouple and IR camera confidence ranges do not overlap by a gap of 01°C. Another method to review the data is to perform a null hypothesis statistical test to determine if the difference between the corrected value and the SRF corrected value is statistically significant. A standard 2 test statistic (Miller and Freund [1965]) can be defined as z 2 3:11, (20) a, 0'2 _ + __ "I "2 where z is the test statistic that will be compared to a Student’s t statistic to determine the confidence level, 5? is the mean value, 0' is the standard deviation and n is the total number of data points. The results of applying the null hypothesis test are shown in Table 8 which also includes a summary of the true values (as measured with a calibrated thermocouple) and the SRF corrected values. Table 8 indicates that at the higher temperature values the SRF corrected value and the true value are statistically indistinguishable. Test 1b with the MAF sensor almost passed the statistical test (it failed by 0.1 and is marked No***). Test 2b with the cylinder 67 failed. It is important to note that the uncorrected value for test 2b was firrther from the true value than the SRF corrected estimate. Table 8. Null hypothesis statistical test results. MAF Sensor 0 = 0.007 radians SRF = 0.90 e = 0.88 SRF Corrected True Number of 95% Null Temperature Temperature Data Points Hypothesis Test 1a 60.91 5.1 61.01 3.4 n =13 Yes Test 1b 38.1132 36.6119 n= 14 No“ Cylinder 0 = 0.0029 radians SRF = 0.57 e = 0.95 SRF Corrected True Temperature Temperature Test 23 64.5 i 3.7 63.5 1 2.3 n = 19 Yes Test 2b 47.1116 43.7116 n =18 No In conclusion, the application of the SRF correction procedure outlined in section 5.2 can reduce thermal measurement errors by at least 50%. This improvement extends the capabilities of the HTRL and the applications of the Inframetrics 600L infrared camera. The error created by imaging unresolved thermal images was substantially reduced. The application of the SRF correction does increase the experimental uncertainty. The uncertainty in the experiments presented above was approximately :25 0C. This uncertainty can be derived from the variability of the SRF fimction or from the variability of the particular experiment as shown in Tables 3, 4, 6 and 7 by utilizing Equation 17. It is recommended that the specific experimental variability be applied if possible. 68 CHAPTER 6. CONCLUSIONS A facility was constructed to provide air flows of known velocity profiles and temperatures with the necessary sensors to monitor these parameters. The facility was characterized specifically to study small scale devices with the infrared imaging system in the HTRL. A procedure to calibrate the infrared imaging system and the associated data acquisition equipment was developed. The slit response function (SRF) for the infrared system was experimentally measured for the basic system and with the external optics (3x telescopic lens and the 6” close up lens). A procedure to use the SRF to correct experimental measurements was defined along with the increase in the uncertainty in the corrected temperature. Finally, an experimental verification of the procedure was conducted. The experiments contained in the investigation presented were all conducted under the calibration nozzle with the airflow heater in place. This provided an environment with known and controlled velocity and temperature profiles to compare the infrared radiometer measurements against. The control of these environmental conditions were utilized to simulate “unknown” operating conditions. A wind tunnel facility was designed to expand the experimental capabilities of the HTRL and to support the current research project. The facility contains a two- dirnensional flow contraction that is being characterized currently for a separate project. The facility also has the capability to inject steam into the flow to study the effects of humidity in experiments. The experimental facility contains a traverse system to take detailed spatial measurements and custom software to control the traverse system and acquire data. The details of the system are contained in Chapter 2. 69 The infrared imaging system and the associated data acquisition system were calibrated to obtain the maximum useful dynamic range. This procedure was outlined in Chapter 3. The result of this calibration was the inclusion of 95% of the dynamic range with a linear response over the entire measurement extent. These results agree very well with the ideal system response that was defined as the goal of the cahbmtion. The SRF function determined for the Infrarmetrics 600L infrared camera did not agree with the manufacture’s specifications. This was determined to be the result of delinquent calibrations and potentially contaminated lenses in the optical path. The SRF measured was repeatable and appeared to be steady over the time flame of the work presented here. Understanding the SRF is critical to obtaining valid thermal measurements on objects near the limits of the IR camera resolution. Figure 31 presents the SRF correction factor (CF) and associated uncertainty as a fimction of target size for the Inframetrics camera with the close up optics. The SRF correction (Figure 31) reduces the amount of error when imaging targets smaller than the thermal measurement resolution. There is an increase in the uncertainty associated with these measurements as the target angle is reduced. The mean temperature correction can be substantial (changing the gray scale value by up to and exceeding a factor of 10) and the uncertainty of this correction can be increased to as much as 25% of the correction for target Sizes below 0.25 mm. 70 Correction Factor (CF) and Uncertainty Specifications 12 f 30 1 A 10 ) —-CF 25 . o . . E 8 T A /o uncertainty In CF 1 20 E '2 e l ‘ . 15 E '| l II. o 4 r + 10 A A a! 2 l e 5 ~11 o l____._ _-~ I;— — — * +A‘ we. .__.__‘__. ‘———— LL 0 0 0.25 0.5 0.75 1 1.25 Target width (m m) Figure 3]. Correction factor (l/SRF) and % uncertainty in CF for Infiametrics 600L IR camera with close up optics in place as a fimction of target size. Target emissivity can be measured within each thermal measurement resolution (TMR). The TMR for the Infiametrics camera with the close up optics is 700 am. The experimental verification of the SRF correction (Chapter 5) shows that the method provides a valid measurement correction and reduces the error associated with measuring unresolved thermal targets. The goal of the original research proposal was to apply infrared thermography to the calibration of the MAF sensors. The MAF sensor resistance as a function of temperature can be determined utilizing an ohm-meter and the IR camera. Furthermore, the sensor can be heated electrically in a flow field and its temperature determined with the infrared system This flow configuration can be utilized with a standard Nusselt correlation (Incorpera and Dewitt [1990]) such as hD m ; Nuz-k—=CReDPr , (21) 71 with C and m as calibration coefficients, the Reynolds number (Re) defined in Equation 4, the Prandtl number for air (Pr), and the heat transfer coefficient (h) defined as I2 R h = , (22) A(TM4F _ Tflow) where 12R is the Joule heating occurring and A is the surface area of the sensor. All of these variables can be measured directly and the constants C and m can be obtained from an ordinary least squares curve fit procedure. The variable In is approximately 0.5 which will cause the velocity and the temperature measured by the infiared camera to be related as V at 1 2 . (23) (TA/14F - Tflow) Equation 23 can be utilized to reduce the error in the velocity calculation created by uncertainty in the MAF sensor temperature by maximizing the (TMAF-Tnow) quantity. This is advantageous with the results presented here because the SRF correction is more accurate at higher temperatures as discussed previously. These results indicate that the original goal of the project can be obtained. 72 CHAPTER 7. RECOMMENDATIONS FOR FUTURE WORK 7.1 SLIT RESPONSE FUNCTlON The SRF determined for this camera is not within specifications for the system as manufactured. This could be the result of optical misalignment, optical contamination or possibly scanner calibration. Inframetrics recommends that the 600L infrared camera be re-calibrated each year. The camera used for the experiments has not been re-calibrated for at least 3 years. The multi-point calibration consists of viewing a blackbody at known temperatures and adjusting the electronic processing to the correct value. The scanning optics are also realigned and cleaned during this procedure. Any combination of these issues can cause the SRF function to degrade. Never the less, the results presented here can still be considered valid for the scanner in question because it is repeatable. Studies in the future should measure the SRF function to see if it has drifted further. If the system is re-calibrated or if the optics are cleaned, it is expected that the SRF will improve. 7 .2 INFRARED CAMERA RESPONSE TO TEMPERATURE GRADIENTS The SRF determines the infrared system response to a step function in temperature. Typically a step function will only occur at the edges of the target. An extension of the SRF work would be to quantify the maximum temperature gradient that the infiared detector can accurately measure. This would be especially relevant in experiments involving heat transfer such as steady state conduction with a large heat flux (Q) through a material with a low thermal conductivity (k) as shown in 73 2. dx "1 Q _. 20 k < > Gartenberg and Roberts [1990] conducted a study in which electrically heated wires were suspended in an air flow and imaged with an infrared system. The target was unresolved Optically but this was corrected for with the use of an ‘apparent’ emissivity. They found that while the IR imaging system could accurately measure the wire center (where dT/dx =0), as the gradient increased, the measurement accuracy decreased. 7.3 HOT-WIRE SENSORS Preliminary work was performed with 2 hot wire sensors in the flow field to measure both temperature and velocity Simultaneously. Computer code was written to solve the two equations iteratively for the two variables and is included in Appendix 3. It appears that the maximum accuracy and sensitivity to both velocity and temperature can be obtained by setting one over-heat ratio as high as possible (approximately 1.7 for tungsten hot wires) and the other over-heat ratio reasonably low. Preliminary experiments have shown that an overheat value of 1.3 provides heighten sensitivity to the flow temperature. Five micron diameter platinum wire has been ordered. This wire is more stable in an oxygen environment and can be run at an over-heat of up to 2.2. This wire will be evaluated for future work. It is expected that the larger difference in the over heat values will reduce measurement errors in both temperature and velocity. The platinum wire will also be evaluated as a cold wire sensor. This is very similar to hot wire measurement 74 methods except that a minimum electrical sensing current is used to measure the resistance of the wire. The resistance is sensitive to temperature changes and therefore will allow the wire to be calibrated and used to measure temperature directly. The cold wire is insensitive to flow velocity and the output can be utilized to temperature compensate the hot-wire output. 7.4 DATA ACQUISITION Several improvements are scheduled for the data acquisition system. These include obtaining two 200MHz Pentium computer configured as follows: A. Pentium 200MHz computer with 32 Mbytes of RAM w/ a. A Keithley-MetraByte DAS-TC/b board b. A DAS-l802 333kHz data acquisition board B. Pentium 200MHz computer with 32 Mbytes of RAM w/ a. Image-Pro Plus v3.0.1 for Windows 95 software b. Data Translations DT3152 Frame Grabber board The Pentium computer with the DAS-TC and the DAS-l802 board will greatly enhance the data acquisition capabilities and the speed at which data can be obtained. A single DAS-TC board is being used for all thermocouple and hot wire data captures. This board limits the data acquisition rate to 25Hz on each channel. This is much too slow to capture any flow features aside from the mean velocity. At this rate, a data file of 1000 data points from a single hot wire and a single thermocouple requires 40 seconds. This is the equivalent to 20,000 slot widths of flow (10 m/s velocity with an exit width of 2 cm), 75 essentially, 1 data point every 20 Slot widths of flow. This is much too slow to capture transient events. It is also possible that periodic events in the flow (such as vortex shedding) could suffer from measurement aliasing and not be captured or understood from the data obtained. The DAS-TC/b board will be utilized for thermocouple and humidity sensor measurements. These sensors have relatively slow response times due to the nature or mass of the sensor (the bead Size of the thermocouple or the capacitor in the humidity probe). The DAS-1802 board will be used strictly to control the stepper motor motions and to read the data from the hot wires and the pressure transducer. The DAS-1802 with 3 channels active is capable of acquiring data at 100 kHz per channel. This will allow 200 data points per slot width of flow. The increase in data rate as well as the increase in memory will also help to insure data is convergence. The current system is limited to 2 data channels with 4000 points each. This results in a cumulative average shown in Figure 31. The cumulative average appears to be converged within 4000 data points. 76 Cumulative Data Averages 41.- _ -..- - _ .~ .__L%L__.._._._____.__-L,__.H-fi_ -— -- .—-_. w .._ ..-1_3 4.08» (>12 4.064» >11 4.04 l l H A 4.02 I E v . > 3.9 . A 0.5 0 500 1000 1 500 2000 2500 3012!) 3500 4000 Data Point: Figure 32. Cumulative average of the hot-wire data. A method for quantifying the degree of convergence is presented by Mandel [1964] as shown in (23) This equation uses the Student t statistic (t) and the estimated standard deviation (0') with the number of degrees of freedom (data points) (11) to estimate the uncertainty (L) in the mean value. The confidence level is obtained from the Student t statistic. The distribution of the data is assumed to be gaussian and n was always greater than 1000 data points. The data in Figure 31 has a mean velocity of 1.21 m/s with a standard deviation of 0.718 m/s. When Equation 23 is applied to the data utilized in Figure 31, the estimated uncertainty in the mean is 0.045 m/s at a 95% confidence level. If a 99% confidence level 77 is desired, the uncertainty becomes 0.058 m/s. Typically in this study, 2000 data points were utilized for hot-wire and thermocouple measurements. This results in an uncertainty of 0.063 m/s with a 95% confidence for the hot-wire data shown in Figure 31. This information also indicates the need for the new data acquisition boards. The equipment utilized for this study acquired data at 25 Hz, thus, it took 40 seconds to obtain 1000 data points. This is much too long for transient flow tests and made steady state tests take much longer than necessary. The longer each experiment was run, the greater the drift in the ambient temperature, the greater the drift in the hot-wire calibration and the greater the chance that a particle would ruin the hot-wire and the entire data set. 78 APPENDICES 79 APPENDIX 1. TWO-DMENSIONAL FLOW CONTRACI'ION The design for the two-dimensional contraction was taken from Morrel [1977]. The method designs the contraction by using two cubic functions and matching the slopes so that the flow experiences a favorable pressure gradient to prevent separation along the contraction surfaces and a gradual smooth change in the derivative of the profile shape. The design process outlined by Morrel is iterative. A program was written in Mathematicao to simplify the design process and to plot the resultant profile. The Mathematicao program is included at the end of this appendix. The contraction can be seen in Figure 33 and the manufacturing coordinates are detailed in Table 9. The contraction was manufactured from '/4” thick Lexan° by machining the necessary profile in a suction mold. The Lexan was heated until it became pliable and then was pulled into its final shape by the vacuum applied. The machining and forming was completed by Prodigy Machine and Gage, Lansing Michigan. Figure 33. Two-dimensional flow contraction. 80 Table 9. Two dimensional contraction machining coordinates. Mmeters) y (meters) x (meters) y (meters) 0 0.1905 0.115 0.12951 0.005 0.190495 0.12 0.121203 0.01 0.19046 0.125 0.112175 0.015 0.190365 0.13 0.102395 0.02 0.190179 ._ 0.1335 0.095086 0.025 0.189873 0.135 0.091924 0.03 0.189417 0.14 0.082231 0.035 0.188781 0.145 0.07378 0.04 0.187933 0.15 0.066484 0.045 0.186846 0.155 0.060259 0.05 0.185487 0.16 0.05502 0.055 0.183828 0.165 0.050682 0.06 0.181838 0.17 0.047159 0.065 0.179487 0.175 0.044368 0.07 0.176745 0.18 0.042222 0.075 0.173582 0.185 0.040637 0.08 0.169968 0.19 0.039529 0.085 0.165872 0.195 0.038811 0.09 0.161265 0.2 0.038399 0.095 0.156117 0.205 0.038208 0.1 0.150398 0.21 0.038153 0.105 0.144077 0.215 0.038149 0.11 0.137124 The next two pages contain the Mathematica program written to Show the profile and to aid the iterative process. 81 Clear-[81, HZ, 3:, Ba, Ba, EX, L] HI: .1905; 2 .2. £92, 2 L= 0.213; Ex: 0.627; Ha[x] 82 (El 82) 1 (if; = + - - — ,‘ " n2 (HI-112) (1- !)3 m3[x]=82+ ,-_ v—J' - (1-Ex)2 I blag] = Oxfialx] ; fiblxj = axmlxl ; Plan: Plot[Ha[x] , {3, O, .1336} , PM» "Cartractim Profile", Dim-t Identity] ; Plath: Plot-483m] , (x, .1336, 0.213} , W" Ichttitzy]; Plate: Plouhixl , {3, 0, .1336} , We 1&1:in ; P191311: Flanagan] , (x, .1336, 0.213}, W» MW] ; Show[Plota, Plath, OW» Wm, We {0, O} , Erase-b Tun, We Autumtic, Flow» {{0, .225}, {0, .225}} , W110» 1 i , W» (meters, matersfl 82 Output Graphics Contraction Profile l 0'21 l 0.15 m . Ll (l) .1.) 2 0.1 _____ \. 0.05 V . \ l O . a MA . 1 0.05 0.1 0.15 meters 83 APPENDIX 2. QUICKBASIC© PROGRAM TO PERFORM DATA ACQUISITION AND STEPPER MOTOR CONTROL The following is a quick basic program developed and written by Dr. John McGrath and Paul Hoke to operate the data acquisition and stepper motor boards. This program allows for the stepper motor to move to position in increments of 6.602 pm (this is determined by the motor and mechanical translation device). The stepper motor control consists of a CMD-50 chopper drive and a VCO-4050 controller board fiom American Precision Industries. The stepper motors are controlled fi‘om the data acquisition board inputs and outputs. The data acquisition card is a DAS-TC utilized in conjunction with a DAS-02 card from Keithly-MetraByte. These boards collect 16 channels of input and are capable of driving 2 channels of analog output. 84 'itttttttit*tt*titttttt********#*##3##titttttt*ttttttt******t***t*t**ttt ***** REM This is the program recalb.bas REM WEDNESDAY, Dec 17, 1997: 12:10 PM ' Include the supplied Q41FACE.BI file. This file ' contains all function DECLARAtion supported by the driver for the DASTC board. ' $INCLUDE: 'Q41FACE.BI' ' Define ALL local variables required by the program here. NOTE ' that you must avoid declaring and using QuickBASIC variables on the ' fly. DIM NumOfBoards AS INTEGER DIM DERR AS INTEGER DIM DEVHANDLE AS LONG DIM BoardNum AS INTEGER DIM CJC AS SINGLE DIM CHAN AS INTEGER DIM GGAIN AS INTEGER DIM ADDATA AS LONG DIM TEMPGBV AS LONG DIM TEMPGB AS SINGLE 85 DIM TEMPINV AS LONG DIM TEMPIN AS SINGLE DIM TEMPOUTV As LONG DIM TEMPOUT AS SINGLE DIM TEMPSURFV AS LONG DIM TEMPSURF As SINGLE DIM RELHUMV As LONG DIM RELHUM AS SINGLE DIM PDIFFVl AS LONG DIM PDIFFV AS SINGLE DIM HWVII AS LONG DIM va1 As SINGLE DIM va21 AS LONG DIM va2 AS SINGLE DIM de AS LONG DIM dp AS SINGLE DIM th AS LONG DIM J AS LONG DIM NUMBER AS LONG DIM SUMI As SINGLE DIM SUM2 As SINGLE DIM SUM3 AS SINGLE 86 DIM SUM4 AS SINGLE DIM HWlAVEL AS SINGLE DIM HWZAVEL AS SINGLE DIM dpavg AS SINGLE DIM HWV1(2000) DIM HWV2(2000) DIM dp(2000) DIM tcfl(2000) DIM VINVISCID AS SINGLE DIM DELP AS SINGLE 'OPEN FILES FOR WRITING HOTWIRE DATA 'OPEN "HW1.DAT" FOR OUTPUT AS #1 'OPEN "HW2.DAT" FOR OUTPUT AS #2 'OPEN "dp.dat" for OUTPUT as #3 CLS ' This step initializes the internal data tables according to the ' information contained in the configuration file DASTC.CFG. PRINT "Initializing the DASTC board - - - PLEASE wait" a8 = "DASTC.CFG" + CHR$(0) DERR = DASTCDEVOPEN%(SSEGADD(a$), NumOfBoards) IF DERR <> 0 THEN 87 BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING '..DEVOPEN'" STOP END IF BoardNum = 0 DERR = DASTCGETDEVHANDLE%(BoardNum, DEVHANDLE) IF (DERR <> 0) THEN BEEP: PRINT "ERROR, DEVICE HANDLE IS null. . .": STOP CLS REM This part of the program is using the DAC-02 board to control sensor motion REM Set base address at 784 for DAC-02 board to NOT conflict w/ DAS-TC board BaseAddr% = 784 REM D/A Output Channels are defined REM Channe1#0 will be used for MOTION: move (>25 v; OutData% = 4095) REM or stop (<2.5 v; OutData% = 0 ) REM Channe1#l will be used for DIRECTION: up/down or in/out for vertical and lead screw assemblies REM Set initial voltages on D/A channels to 0. This Means that ChannelO volts = 0 -> motor stopped REM Output 0 volts on Channel#1 (DIRECTION); Ch$l first because of prob detected Ch1% = 1 OutData% = 0 88 HByte% = INT(OutData%/ 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte lowportl = BaseAddr% + 2 * Ch1% OUT lowport, LByte% 'Write low byte hiportl = BaseAddr% + 2 * Ch1% + 1 OUT hiportl , HByte% 'Write high byte REM Output 0 volts on Channel#0 (MOTION) Ch1% = 0 OutData% = 0 HByte% = INT(OutData%/ 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte lowportO = BaseAddr% + 2 * Ch1% OUT lowportO, LByte% 'Write low byte hiportO == BaseAddr% + 2 * Ch1% + 1 OUT hiportO, HByte% 'Write high byte 'This segment inputs parameters needed for calculating velocity and local ' temperature in the boundary layer 'Clear screen CLS ' REQUEST AN INITIAL POSTION DEFINITION FROM USER IN CASE DIFERENT THAN ZERO LOCATE l, 3: INPUT "Define the initial location in the boundary layer (mm): ", position 89 THEBEGIN: 'REM Set initial position equal to zero 'position = 0 REM Set initial count equal to zero count = 0 DIRECTION: REM Decide which direction to move the motor and use Channel#l to effect decision REM Set channel number to zero LOCATE 2, 3: PRINT "Positive direction (+y) is AWAY from the wall" LOCATE 3, 3: INPUT "(D/A Chl) To move HOTWWIRE: TO WALL- Enter 5; AWAY FROM WALL- Enter 0 ", decide% IF decide% < 2.5 GOTO OUTDOWN ' LET SIGN : -1 FOR MOTION TO THE WALL SIGN = -1 REM Output 5 volts on Channel#1 to move IN or UP Chl% = 1 OutData% = 4095 HByte% = INT(OutData% / 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte OUT BaseAddr% + 2 * Chl%, LByte% 'Write low byte OUT BaseAddr% + 2 * Chl% + 1, HByte% 'Write high byte GOTO END.DIRECTION 90 OUTDOWN: REM Output 0 volts on Channel#1 to move OUT or DOWN ' LET SIGN = +1 FOR MOTION AWAY FROM THE WALL SIGN = 1 Chl% = 1 OutData% = 0 HByte% = lNT(OutData%/ 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte OUT BaseAddr% + 2 * Chl%, LByte% ‘Write low byte OUT BaseAddr° o + 2 * Chl% + 1, HByte% 'Write high byte END.DIRECTION: REM Define distance(s) to be moved (This will require motor-specific calibrations later) LOCATE 4, 3: INPUT "How far (in mm) should the motor move? ", DISTANCE count.desired = DISTANCE * 151.352 REM Decide whether or not to move the motor and use Channel#0 to effect decision REM Set channel number to zero LOCATE 5, 3: INPUT "Do you wish to move the motor (D/A Ch 0)? (5: move; 0 : stop) "; decide% IF decide% < 2.5 GOTO DONTMOVE MOVE: REM output 5 volts on Channel#0 to move motor Chl%=0 91 OutData% = 4095 HByte% = INT(OutData%/ 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte OUT BaseAddr% + 2 * Chl%, LByte% 'Write low byte OUT BaseAddr% + 2 * Chl% + 1, HByte% 'Write high byte GOTO PULSECOUNT DONTMOVE: REM Output 0 volts on C hannel#0 to stop motor Chl% = 0 OutData% = 0 HByte% = INT(OutData%/ 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte OUT BaseAddr% + 2 * Chl%, LByte% 'Write low byte OUT BaseAddr% + 2 * Chl% + 1, HByte% 'Write high byte GOTO IN F OOUT PULSECOUNT: REM This part of the program counts pulses from an MD channel REM The DAC-02 board can not be used for this, so the DASTC board will be used. REM Use Channel#8 for pulse counting REPEAT: 'get an initial voltage for edge detection in a loop 'The next step gets an initial voltage reading on the pulse counting channel 92 (HLMN=8 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, ADDATA) IFDERR<>0THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP ENDIF ' Calculate the effective voltage. CALCULATEVOLTAGEI : 'The next step calculates a voltage from the ADDATA value measured 'Call this first voltage V1 Vl = ADDATA / (1000000) CHECK: 'this is the beginning of a loop which checks whether successive voltages are different (>2.5 v) than initial 'voltage DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, ADDATA) IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRea '" STOP 93 END IF ' Calculate the effective voltage. CALCULATEVOLTAGEZ: “The next step calculates a voltage from the second ADDATA value measured 'Call this first voltage V2 V2 = ADDATA / (1000000) 'Calculate the difference between the an initial voltage and all successive voltages in loop DIFF = ABS(V1 - V2) 'If a 0 to 5 volt or a 5 to 0 volt change is not detected on channel 8, then continue to take measurements, 'Other wise increment a counter because an edge has been detected. IF DIFF < 2.5 GOTO CHECK count = count + 1 IF count > count.desired GOTO the.end GOTO REPEAT the.end: This stops the motor Chl% = 0 REM Output 0 volts on Channel#0 to stop motor OutData° o = 0 HByte% = INT(OutData%/ 16) 'Convert to high byte LByte% = 16 * (OutData% - HByte% * 16) 'Convert to low byte 94 OUT BaseAddr% + 2 * Chl%, LByte% 'Write low byte OUT BaseAddr% + 2 * Chl% + 1, HByte% 'Write high byte ' CLEAR SCREEN BEFORE WRITING MOTION INFORMATION TO THE SCREEN INFOOUT: CLS calibration = count / 151.352 position = position + SIGN * calibration LOCATE 10, 3: PRINT "Hotwire has moved "; calibration * SIGN; "mm" LOCATE 11, 3: PRINT "Hotwire is now at "; position; "mm" LOCATE 12, 3: PRINT "Where +x is away from the wall" DUMMY = 0 LOCATE 15, 3: PRINT "Move again OR set zero position?" LOCATE 16, 3: INPUT " -1 to quit; I to set current position; 0 to move again; 2 TO TAKE DATA: ", DUMMY IF DUMMY = -1 GOTO FINISH IF DUMMY = 1 GOTO ZERO IF DUMMY = 2 GOTO DATALOOP GOTO THE.BEGIN: ZERO: POSITION = 0 count = 0 GOTO THE.BND 95 FINISH: END CLS 'WRITE TO SCREEN THE CHOICES MADE FOR DIRECTION AND MOTION 'LOCATE 1, 1: PRINT "YOU CHOSE TO MOVE (UP/DOWN, ETC)" 'LOCATE 2, 1: PRINT "YOUR STARTING STREAMWISE POSITION (X/W) WAS:" 'LOCATE 3, 12 PRINT "YOUR STARTING BOUNDARY LAYER POSITION (Y IN MM) WAS:" 'LOCATE 4, 1: PRINT "INITIATE LOOPING TO UPDATE EXPERIMENTAL DATA" DATALOOP: CLS LOCATE 1, 3: PRINT "The Hotwire is now at y= "; position; "mm" LOCATE 2, 3: PRINT "Where +y is away from the wall" LOCATE 3, 3: PRINT "The Hotwire is now at X: "; position; "mm" LOCATE 4, 3: PRINT "Where +x is DOWN- in the F low Direction" LOCATE 6, 3: PRINT "Current data are being displayed for checkirng" DO CHECKS = INKEYS 'This section is for measuring experimental variables 'Measure ambient Cl C temperature BoardNum = O 96 DERR = DASTCGETCIC%(BoardNum, CJC) IF (DERR <> 0) THEN BEEP: PRINT "ERROR GETTING CJC TEMPERATURE...": STOP LOCATE 8, 5: PRINT USING "AMBIENT CJC TEMP(C)= ###.##"; CJC 'Measure TEMPERATURE IN GIELDA BOX, TEMPGB CHAN = 7 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, TEMPGBV) IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF TEMPGB = ((TEMPGBV/ 1000000) - 1) * 25 LOCATE 9, 5: PRINT USING "TEMPERATURE IN GIELDA BOX (C) = ###.##"; TEMPGB Measure PLATE INLET TEMPERATURE, TEMPIN CHAN = 5 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, TEMPINV) IF DERR <> 0 THEN BEEP 97 PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF TEMPIN = TEMPINV / 100 LOCATE 10, 5: PRINT USING "PLATE INLET TEMPERATURE (C)= ###.##"; TEIVIPIN 'Measure PLATE OUTLET TEMPERATURE, TEMPOUT CHAN = 6 GGAIN = l DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, TEMPOUTV) IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF TEMPOUT = TEMPOUTV / 100 LOCATE ll, 5: PRINT USING "PLATE OUTLET TEMPERATURE (C)= ###.##"; TEMPOUT 'Measure PLATE SURFACE TEMPERATURE, TEMPSURF CHAN = 2 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, TEMPSURFV) 98 IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF TEMPSURF = TEMPSURFV/ 100 LOCATE 12, 5: PRINT USING "PLATE SURFACE TEMPERATURE (C)= ###.##"; TEMPSURF 'Measure ambient relative humidity CHAN = 9 GGAIN = l DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, RELHUMV) IF DERR <> 0 THEN BEEP PRINT "ERROR"; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF RELHUM = ((RELHUMV / 1000000) - 1) / .04 LOCATE 13, 5: PRINT USING "RELATIVE HUMIDITY (%)= ##1##"; RELHUM 'Measure Pressure DIFFERENCE, PDIF F CHAN = 12 GGAIN = l 99 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, PDIFFVl) IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF PDIFFV = PDIFFVl / 1000000 LOCATE 14, 5: PRINT USING "PRESSURE TRANSDUCER VOLTAGE (V OLTS)= ##.##"; PDIFFV DELP = PDIFFV * .1 'CHECK FOR NEGATIVE VOLTAGES IF DELP > 0 GOTO OKSTATE DELP = 0 OKSTATE: VINVISCID = SQR((2 / 1.204) * (DELP) * (24884)) LOCATE 15, 5: PRINT USING "IMPLIES MAX INVISCID JET EXIT PLANE VELOCITY (In/S): ##1## "; VINVISCID 'IVIEASURE CHANNEL 1 HOTWIRE VOLTAGE, HWVl CHAN = 11 GGAIN = l DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, HWY] 1) IF DERR <> 0 THEN 100 BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF HWVl =HWV11 / 1000000 E1 = HWVl LOCATE 16, 5: PRINT USING "CHANNEL 1 HOTWIRE VOLTAGE (VOLTS)= ###.##";HWV1 'MEASURE CHANNEL 2 HOTWIRE VOLTAGE, HWV2 CHAN = 10 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, HWV21) IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF HWV2 = HWV2] / 1000000 E2 = HWV2 LOCATE 17, 5: PRINT USING "CHANNEL 2 HOTWIRE VOLTAGE (VOLTS)= ###.##"; HWV2; LOCATE 18, 3: PRINT "PRESS ANY KEY TO INITIATE HOTWIRE DATA ACQUISITION" 101 GOTO KEYCHECK KEYCHECK: LOOP WHILE CHECK$ = 1NKEY$ 'CHECK FOR KEYSTROKE OOPS: LOCATE 20, 5: INPUT "ENTER THE NUMBER OF VELOCITY SAMPLES To BE TAKEN"; NUMBER IF NUMBER < 2 GOTO OOPS 'READ HOTWIRE VOLTAGES 'INITIALIZE SUMS To ZERO FOR AVERAGES SUMI = 0 SUM2 = 0 SUM3 = 0 SUM4 = 0 HWV1(0) = 0 HWV2(0) = 0 dp(O) = 0 tcfl(0) = 0 FOR J = 1 To NUMBER 'MEASURE CHANNEL 1 HOTWIRE VOLTAGE ARRAY, HWVA1(J) CHAN = 11 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, HWVl 1) 102 IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF HWV1(J) -—- HWVII / 1000000 SUM1= SUM1+ HWV1(J) 'MEASURE CHANNEL 2 HOTWIRE VOLTAGE, HWV2 CHAN = 10 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, HWV21) IF DERR <> 0 THEN BEEP PRINT "ERROR"; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF HWV2(I) = HWV2] / 1000000 SUM2 = SUM2 + HWV2(J) 'MEASURE CHANNEL 2 HOTWIRE VOLTAGE, HWV2 CHAN = 12 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, dpl) 103 IF DERR <> 0 THEN BEEP PRINT "ERROR"; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF dp(J) -—— de / 1000000 SUM3 = SUM3 + dp(J) CHAN = 6 GGAIN = 1 DERR = KADRead%(DEVHANDLE, CHAN, GGAIN, :61) IF DERR <> 0 THEN BEEP PRINT "ERROR "; HEX$(DERR); " OCCURRED DURING 'KADRead'" STOP END IF 160(1) = 1C1 / 100 SUM4 = SUM4 + tCflU) NEXT 'CALCULATE THE AVERAGES HWlAVEL = SUMI /NUMBER HW2AVEL = SUM2 /NUMBER dpavg = SUM3 / NUMBER 104 h20 = dpavg * .1 tcavg = SUM4 / NUMBER 'CALCULATE THE STANDARD DEVIATIONS hwlst = 0 hw2st = 0 dpst = 0 tcst = 0 FOR J = 1 TO NUMBER hwlst = hwlst + ((HWlAVEL - HWV1(J)) A 2) A (.5) hw2st = hw25t + ((HW2AVEL - HWV2(J)) A 2) A (.5) dpst = dpst + ((dpavg - dp(J)) A 2) A (.5) tcst = tcst + ((tcavg - tcfl(J)) A 2) A (.5) NEXT hwlstdev = hwlst / (NUMBER - 1) hw2stdev = hw25t / (NUMBER - 1) dpstdev = dpst * .1 / (NUMBER - 1) tcstdev = tcst / (NUMBER - 1) CLS LOCATE 11, 5: PRINT USING "CHANNEL 1 HOTWIRE AVERAGE VOLTAGE (VOLTS)= ##.###"; HWlAVEL LOCATE 12, 5: PRINT USING "ST.DEV of Hot Wirel reading (VOLTS) = #.###"; hwl stdev 105 LOCATE 13, 5: PRINT USING "CHANNEL 2 HOTWIRE AVERAGE VOLTAGE (VOLTS)= ##.###"; HW2AVEL LOCATE l4, 5: PRINT USING "ST.DEV of Hot Wire2 reading (V OLTS) = #.###"; hw2stdev LOCATE 15, 5: PRINT USING "Pressure difference (in/I120): #.###"; h20 LOCATE l6, 5: PRINT USING "ST.DEV of Pressure reading (in/HZO) = #.###"; dpstdev LOCATE 17, 5: PRINT USING "Temperature of Flow = ##.#"; tcavg LOCATE 18, 5: PRINT USING "ST.DEV of Temperature = ##.##"; tcstdev 'WRITE THE DATA ARRAYS TO TWO FILES: HW1.DAT AND HW2.DAT 'FOR J = 1 TO NUMBER 'WRITE #1, J, HWV1(J) 'WRITE #2, J, HWV2(J) 'WRITE #3, J, dp(l) NEXT 'CLOSE THE DATA FILES 'CLOSE #1 'CLOSE #2 'CLOSE #3 LOCATE 24, 3: PRINT "Take another calibration point?" LOCATE 25, 3: INPUT " -1 to quit; I to set current position; 0 to move again; 2 TO TAKE DATA: ", DUMMY 106 CLS IF DUMMY = -1 GOTO FINISH IF DUMMY = 1 GOTO ZERO IF DUMMY = 2 GOTO DATALOOP GOTO THE.BEGIN: 107 APPENDIX 3. FLAT PLATE ISOTHERMAL VELOCITY PROFILES The velocity profiles taken of the flow below the two dimensional contraction are presented here for completeness. The profiles are not non-dimensionalized in the common format of the velocity divided by the inviscid velocity because at the time of the experiments, there was no pressure tap in the contraction (this has been added since). These results show the resolution of the stepper motor system combined with the hot wire probes. They also Show the flow development below the contraction in a limited fashion. Figure 34 shows all of the data obtained. Figure 35, Figure 36, and Figure 37 Show the velocity profiles at each specific X/W location. The affect of the number of data points can be seen in Figure 34, Figure 36 and Figure 37. The profiles taken with only 200 data points tend to appear jagged, an indication that the measurements at each point were not converged. The profiles with 2000 data points tend to appear smoother and also appear to define a mean value curve among the scatter shown in the 200 data point curves. The profiles at x/w = 1.59 show the inviscid core region of the wall jet as the flat region where V/V max = 1. The outer sheared layer and the inner boundary layer progress toward the center of the wall jet as the flow advances in the streamwise (x) direction. This can be seen as the slope becomes more less steep in the outer region and as the flat inviscid central core region disappears. 108 Compiled Profiles (Isothermal) 1.2 +xM=4.52 —x—xlw=1.59 +xlw=4.52 +xlw=10.95 +xIw-10.95 x —.—x1w==1.59 g —-— xlw =1o.95 (2000) 2 +xlw=4.52 (2000) > —I—xM=10.95(2000) 2.5 3 Figure 34. Compiled jet velocity profiles (w=2 cm). xIw=1.59 1 4 .4...“ 0.9 4» 0.8 4» 0.7 4 \ 5 0.6 i g 0.5 \ > 0.4 \ 0.3 \, 0.2 \‘ x 0.1 - ‘\ ‘.‘V’V‘Vv 7V..- Vita». .- o : -— -- — — -— —— — 4 o o 5 1 1 5 2 2.5 ylw Figure 35. W = 1.59 velocity profiles (w=2 cm). 109 xlw=4.52 1 0.9 ,7 l —1—2oo data points i 0.8 -9— 200 data mine 0 7 +200 data po'nts ‘ +2000 data po'nts § 0.6 4 g 0.5 l > 0.4 4 I 0.3 l 0.2 l 0.1 0 4 4 0 0 5 1 1.5 2 2 5 3 ylw Figure 36. W = 4.52 velocity profiles (w=2 cm). xlw=1 0 .9 5 1 l 0.9 ‘ . -o— 2000 data pohts 0.8 +200 data pohts —-—200 data pohts 0‘7 + 200 data pohts x 0.6 :-I—2000 data pohts a A ,, g 0.5 > 0.4 0.3 0.2 0.1 O .7 a 1 o o 5 1 1.5 2 2 5 3 ylw Figure 37. W = 10.95 velocity profiles (w=2 cm). 110 APPENDIX 4. HOT-WIRE CONSTRUCTION AND CALIBRATION The hot-wires used in this study were constructed of 5 micron diameter tungsten wire and are schematically shown in Figure 38. The hot-wires were fabricated in the Turbulent Shear Fluids Laboratory (TSFL) at Michigan State University. Probe body Broaches _[lmm . |._ 4T1.-- Figure 38. Schematic of hot-wire probe. The construction procedure, while simple in theory, was very diflicult to implement due to the small size of the tImgsten wire. The construction procedure is outlined below. 1. The broaches were soaked in flux for 30 seconds. The broaches were dipped slowly into and out of a solder pot to provide a uniform layer of solder. 2. The flux was cleaned off using a flux remover in an ultrasonic bath. 3. A 5 micron tungsten wire was strung across a steel “C” bracket and held in position by adhesive tape. 4. The span was placed in a device so it could be lowered into a copper oxide solution holder to be coated. The span was positioned so that the two regions to be plated were separated by approximately 1 mm (Figure 39). Ill Figure 39. Tungsten wire copper plating station. 6. Power was applied so that approximately 0.8 mA flowed through the device for 8 minutes. 7. The tungsten wire was then soldered across the broaches (Figure 38) using the special solder station to align them properly (Figure 40). Step seven is by far the most difficult and time consuming. A typical first attempt at manufacturing a hot-wire might take 3 hours. A high level of manual dexterity and practice are required to produce hot-wires with any spwd. It is suggested that the entire process could be greatly simplified if the need for human control of the soldering process was eliminated. 112 Figure 40. Precision solder station. Hot-wire calibration requires a well-characterized flow with a known velocity distribution to establish the relationship between the hot—wire output voltage and velocity. A hot-wire calibration system (Figure 41) fi'om the Turbulent Shear Fluids Laboratory (TSFL) was utilized to obtain the first cahbration data. This hot-wire cah'brator was designed to produce a uniform flow of known velocity (measured with a pressure transducer) with very low turbulence intensity. 113 Figure 41. TSFL hot-wire calibration flow contraction. The hot-wire was taken to the HTRL Flow Facility and mounted beneath the “coke bottle” contraction (Figure 42). This contraction has been well characterized in the TSFL to have a discharge coefiicient (Cd) of 0.98 to 1. Assuming that the fluid is incompressible and inviscid, the velocity along a streamline can be calculated from (Blevins [1984]) (22) 114 Figure 42. “Coke Bottle” contraction utilized for hot-wire calibration. Calculating the ( 1-(A2/A1)2)'1 term utilizing the venturi dimensions results in a value of 1.00196. Therefore, it incorporates little error to simplify Equation 22 to I7 = {fl} , (23) p which is the Bernoulli equation. Furthermore, if area A1 is considered to be a large hemisphere (control surface) above the nozzle, V12 quickly approaches zero. A pressure transducer was utilized to measure the difference in static pressure across the contraction, thus allowing the exit velocity to be calculated by applying Equation 23. The density of air was calculated from the ideal gas law. A typical hot-wire calibration performd in the TSFL contraction resulted in the calibration curve shown in Figure 43. The same hot-wire was then taken to the HTRL Flow facility and calibrated using the same pressure transducer. This resulted in a 115 calibration curve as shown in Figure 44. The data were curve-fit using TableCurve2D v3 software (AISN [1994]) and the relationship E=¢A+BV". (24) The coefficients and correlation factors were then used to compare the HTRL facility calibration curves and the calibrations obtained in the TSFL contraction. Typical cahbration curves for the TSFL calibrator and the HTRL facility are shown in Figure 43 and Figure 44 respectively. There was less than a 1% difference between the TSFL calibration coefficients and the HTRL facility coefficients. The conclusion was drawn that the HTRL facility could be used for future hot-wire calibration. A pitot tube was later used to verify that the flow velocity below the calibration nozzle agreed with the Bernoulli calculation. 116 Hot Wire Voltage 9l10197 Hot wire calibration in TSFL contraction (test 3) r2=0.99989089 DF Ad] 12:0.99985452 FIISIGEII=ODO23138116 Fstat=45820.757 3:8.3199168 b=2.56?3594 0:0.47403183 4.2 4 .1 4 3.9 f 3.8 / 3.7 / 3.6 / 3 .5 3 .4 2.5 5 7.5 10 12.5 Velocity (m/s) Figure 43. TSFL contraction hot-wire calibration curve. 117 15 9118437 Hot wire calibration in Ford Facility (test 1) rh099950373 DF Adi r2=0.99937966 FltStdErr=0.0075817315 Fstat=13091.133 0:9.2139295 1329585184 c=0.44175237 4 .5 / 4 .25 Hot Wire voltage 4 1’ / 3.75 // 3.5 / 3.25 0 5 10 1 5 20 25 Velocity (m/s) Figure 44. HTRL facility hot-wire calibration curve. Hot-wire calibrations drift with time due to several factors such as oxidation or stretching or damage of the ttmgsten filament due to impacts with particles. It is absolutely necessary to calibrate the hot-wire prior to its use. Post calibration is desirable as well if the hot-wire survived the experiment. The two cah'bration data sets define a drift that the hot-wire experienced over time. It is necessary to perform this calibration every time a hot-wire is used and after approximately 3 hours of operation (due to oxidation) in order to insure that accurate measurements can be taken. The data set fiom the first calibration is used to define the operating parameters and the post calibration data is strictly used to define the amount of deviation that has occurred relative to the operating 118 parameters used for the experiment. These two calibration data sets define the maximum accuracy of the velocity measurement possible with the hot-wire for that experiment. For example, Figure 44 shows the calibration curve for a hot-wire used in an experiment. The same hot-wire was re-calibrated after two and a halfhours with the resultant data shown in Table 10. These hot-wire voltage data and pressure (velocity) measurements from the post-cah'bration are then used with the curve-fit data from the pre- cahbration to define an error or drift quantity. Table 10. Post cahbration data Velocity Velocity (m/s, via Hot-wire predicted by % Error Equation 23) Po stcalibration Precalibration Output (V) curve (m/s) 1.89 3.5 2.01 6.4 3.09 3.63 3.22 4.1 4.09 3.7 4.03 1.4 4.76 3.76 4.81 1.1 6.04 3.85 6.16 2.0 9.56 4.026 9.49 0.7 11.38 4.11 11.42 0.4 13.85 4.19 13.5 2.6 14.44 4.218 14.29 1.1 16.63 4.28 16.13 3.1 17.37 4.313 17.17 1.1 18.17 4.345 18.22 0.3 19.26 4.371 19.11 0.8 19.81 4.39 19.78 0.2 21.07 4.4 20.13 4.6 21.82 4.44 21.61 0.9 119 The mean of the error is 2.7%. Therefore, a measured velocity of 10 m/S for this test can be reported as 10 m/s i 54 cm/s with a 95% confidence level. This error term does not incorporate variance that occurs during the experiment itself (due to a surging compressor) so discretion must be applied by the experimenter to insure accurate reporting of uncertainty. 120 APPENDIX 5. MAXIMIZING THE DYNAMIC RESPONSE OF THE INFRAMETRICS IR CAMERA AND DATA RECORDING SYSTEM Previous experiments by members of the Heat Transfer Laboratory have shown that the equipment used to record data from the Inframetrics IR camera efl‘ects the data. Following is a brief summary and recommendation from experiments run to optimize the dynamic range of the system. It was found that the Panasonic AG-2400 VCR effected the data stream fi‘om the IR camera. It would seem obvious that the conclusion should be to replace this piece of equipment. However, the Panasonic AG-2400 VCR increased the dynamic range. The For-A VTG-33 video timer and the Sony PVM-l343MD Trinitron monitor did not effect the data stream. The raw Signal fiom the Inframetrics camera has too much offset which results in an upward shift of the gray scale as a function of temperature output. This results in a reduction of the dynamic range capabilities of the system. Tests Show that as much as 40% of the dynamic range was being loss by the previous equipment configurations being utilized. The recommended equipment configuration for data recording and playback are as follows: 121 Data Recording Configuration IR , C omen ForA Timer ———lPanasonic VCR Monitor Data Playback Configuration IP Sony Plus I Parasonic VCR MW . z. Figure 45. Optimal equipment configurations for data acquisition and processing. On IP Plus Software: Set Brightness to 48 Set Contrast in the range of 66 to 70 Set Gamma Function to 1.0 (this is the default setting) 0n For-A Video Timer: Set 7552 switch to ON This arrangement resulted in obtaining 96.8% of the systems available dynamic response to temperature input. The results obtained and a comparison to the ideal system response are shown in Figure 46.. 122 Optima] equipment tests 20 degree IR range 250.0 4. * Test I Result /] y = 240.07x + 5. 1882 i 200“) " R2 = 0.9996 , q, I 3; 1500 4L [dad R8311” ; y=2flk l I? l i 23 100.0 1 l —-*—— mecca Test 1 l 50 0 mecca Test 2 l ——+— Idea! Respmse l / l 0.0 I ‘l T I o 0.2 0.4 0.6 0.8 1 Non Dimensional Temperature Figure 46. Results of optimized equipment configuration on gray scale - temperature relationship. 123 APPENDIX 6. VELOCITY AND TEMPERATURE CALCULATION PROGRAM FOR TWO HOT-WIRE METHOD The following is a program written in Mathematica° to calculate the velocity and temperature from the data obtained fi'om two hot-wires positioned in the flow field. The program is solves for velocity and temperature values in an iterative fashion. Clear[al ,a2,b1 ,b2,ts1 ,ts2,tinf,v1 ,v2,v,E1,E2,n1,n2,alpha,Rc1,Rc2,Rh1,Rh2,delta] (** Iterative Solution for hot wire measurements of velocity and temperature") (** Input Room Temperature at Calibration “) tcall=22.9; tca12=23.0; (** Input Parameters for specific wires and specific data points **) E1=3.454; E2=2.869; A1=7.1596309; B1=2.9114248; n1=0.4197877; Rc1=2.32; Rh1=3.94; A2=5.1325372; B2=1 .4904883; n2=0.45644473; 124 Rc2=1.99; Rh2=2.59; ts1=(Rh1/Rc1)*(273+tcall); ts2=(Rh2/R02)*(273+tca12); delta=.005; al=A1/(tsl-tcall); b1=Bl/(tsl-tcall); a2=A2/(ts2-tcal2); b2=B2/(ts2-tca12); vell [tinf__]=(((E1A2)/(ts1-tinf)-a1)/b1)A(1/n1); ve12[tinf_]=(((E2A2)/(ts2-tinf)-a2)/b2)A(1/n2); Plot[{ve11[tinf],ve12[tinf]},{tinf,15,35}] tinf=15; Do[{v1=(((E1A2)/(tsl-tinf)-al)/b1)A(1/n1), v2=(((E2A2)/(ts2-tinf)-32)/b2)A(l/n2), v=Abs[v2-v1], If[v