3. {‘13, 3.... ‘5: . . - . 3 .. Annual. .. 35“.... 0%” 3. .. .s c r «‘51 .t . .1“: Run» . ‘ . 2... Hi. . km 3; , J. . J. 23. . )1 .s... . human. “(an . r1“: 1...: . 1.53.".31... I .l D Q 1. .... .....: .muu(. lulu“? ’ up gggfiggnfig .9; $5”? 5%? loll-II 45am~§¥¥§vr gagafigfi . ... " '0‘??? I“. 3004 g / 3 (I? .':J' This is to certify that the thesis entitled REAL TIME FRUIT SPIRIT DISTILLATION ANALYSIS WITH HIGH SPEED GAS CHROMATOGRAPHY USING LIQUID PHASE SAMPLING presented by Robert Edward Boes Jr. has been accepted towards fulfillment of the requirements for the MS. degree in Chemistl ‘ ,./"\\ ‘ x. _ _- ; K I A. i‘ M 2 (w .t ’ ' ‘ 7 ~ I‘ . l . C— 4_ ”‘5 a k “\_ -‘f 1 ‘ 9—. ~ (H , ] Major Professor’s Signjmure I 5““ AA /\ \/ 2 ( v ff," Date MS U is an Affirmative Action/Equal Opportunity Institution LIBRARY Michigan State University PLACE IN RETURN Box to remove this checkout from your record. To AVOID FINES return on or before date due. MAY BE RECAILED with earlier due date if requested. DATE DUE DATE DUE DATE DUE 2/05 cJCTfic/omeoue.indd-p.1s REAL TIME FRUIT SPIRIT DISTILLATION ANALYSIS WITH HIGH SPEED GAS CHROMATOGRAPHY USING LIQUID PHASE SAMPLING By Robert Edward Boes Jr. A THESIS Submitted to Michigan State University in partial fulfillment of the requirements for the degree of MASTER OF SCIENCE Department of Chemistry 2005 Abstract REAL TIME FRUIT SPIRIT DISTILLATION ANALYSIS WITH HIGH SPEED GAS CHROMATOGRAPHY USING LIQUID PHASE SAMPLING By Robert Edward Boes Jr. Fruit brandy distillation is a growing industry. Several different spirit distillation techniques have been performed to improve both production and quality. Recent advances in high speed gas chromatography have made online analysis of distillation possible. Over the last couple years a few methods have surfaced that have provided a good starting point from which to continue with research. Two of these setups were thoroughly studied resulting in the development of a new high speed gas chromatography method. The new “liquid-phase sampler” method expanded upon the old setups, still relying heavily on prOper column selectivity and temperature programming for maximum compound separation. The new process introduced a peristaltic pump allowing for constant sample flow, a specially designed liquid-phase sampler for continued sampling, and the use of a GC auto injector for accurate sample injections. Synthetic brandy solutions were used to analyze the new setup’s ruggedness and reproducibility. Several distillations were performed examining changes in temperature, pressure, flow rate, and concentration as a function of distillation volume and time. The behavior of several compounds was analyzed in real time with the new method. The liquid-phase sampler setup proved to be cheaper, less complex, and a very efficient way of monitoring fruit brandy distillations with high speed gas chromatography. Table of Contents List of Tables ....................................................................................................... v. List of Figures ...................................................................................................... vi. 1. Introduction .................................................................................................... 1 1.1 Fruit Brandy Distillation ............................................................... 1 1.2 Structure of a Batch Still ............................................................... 2 1.3 Composition of Fruit Spirits ......................................................... 4 1.4 Quality Control and Analysis ........................................................ 7 1.5 Objective ....................................................................................... 8 1.6 Literature Cited ............................................................................. 9 2. High Speed Gas Chromatography ................................................................. 10 2.1 Background and Development ...................................................... 10 2.2 Speed and Separation .................................................................... 11 2.3 GC Columns .................................................................................. 13 2.4 Injector Considerations ................................................................. 14 2.5 Temperature Programming ........................................................... 15 2.6 Pressure Considerations ................................................................ 15 2.7 Carrier Gas F low Rate .................................................................. 16 2.8 Detectors ....................................................................................... 1 6 2.9 Literature Cited ............................................................................. 17 3. Discussion of High Speed GC Setup ............................................................. 18 3.1 Previous Setups and Results ......................................................... 18 3.2 Liquid-Phase Sampler Setup ......................................................... 20 3.3 Literature Cited ............................................................................. 26 4. Materials and Methods ................................................................................... 27 4.1 Spirits Production Process ........................................................... 27 4.1.1 Creation of Synthetic Solutions ............................... 27 4.1.2 Distillation ................................................................ 27 4.2 Pump Calibration ......................................................................... 30 4.3 Analysis ........................................................................................ 31 4.3.1 Gas Chromatograph Setup ....................................... 31 4.3.2 Top Tray - Online Analysis ..................................... 36 4.3.3 Post Distillation Analysis ......................................... 37 4.4 Literature Cited ............................................................................ 38 iii Modeling the Liquid-Phase Sampler .............................................................. 39 5.0 Ideal Reactors ............................................................................... 39 5.1 Residence Time Distribution of Ideal Reactors ........................... 41 5.2 Residence Time Distribution of Liquid-Phase Sampler .............. 43 5.3 Literature Cited ............................................................................ 50 Results: Liquid-Phase Sampler ...................................................................... 51 6.1 Still Reproducibility ..................................................................... 5 1 6.2 Method Optimization ................................................................... 57 6.3 Standardization ............................................................................ 57 6.4 Distillation Behavior .................................................................... 69 6.5 Literature Cited ............................................................................ 78 Summary and Conclusions ........................................................................... 79 Future Work .................................................................................................. 84 iv Table 4.1. Table 4.2. Table 4.3. Table 4.4. Table 4.5. Table 5.1. Table 6.1. List of Tables Congener types and volume percentages used in the liquid- phase sampler experiment .................................................................. 37 Pump calibration results for the two main settings used during the experiment ........................................................................ 41 Gas chromatograph temperature analysis parameters for the liquid-phase sampler setup ................................................................. 42 Gas chromatograph pressure analysis parameters for the liquid-phase sampler setup ................................................................. 43 Gas chromatograph auto injector analysis parameters for the liquid-phase sampler setup ........................................................... 44 Mean residence time results for the liquid-phase sampler. Two different flow-rates were used during the experiment ............... 56 Retention times for chromatograms taken using the liquid- phase sampler. Peaks were identified by comparing the retention times to standard retention time for each of the compounds. ........................................................................................ 68 Figure 1.1. Figure 1.2. Figure 3.1. Figure 3.2. Figure 3.3. Figure 3.4. Figure 4.1. Figure 5.1. List of Figures A 150 L Christian Carl Batch sun? ..................................... Milligrams of volatile congeners in distilled spirits per 100 milliliters of alcohol. Fruit spirits have higher congener concentrations than other distilled beverages”. .................... Schematic of the previously used headspace sampling setup connected to the Shimadzu GC-17A1. . ................................. A round plate made of plexiglas was created to replace the glass plate in the top tray of the still. A small hole was drilled in the center of the plate to allow for the transfer line to run through it. A similar plexiglas plate was also made for the bottom to allow for the sample to return to the still. . Schematic of the liquid-phase sampler. Liquid flows into the vial fi'om the bottom inlet filling up the vial. The liquid then exits through the top outlet to keep a constant flow running through the vial. ........................................................ Schematic of the liquid-phase sampler setup connected to the Simadzu GC-17A auto injector. Liquid sample travels from the top tray through a peristaltic pump and on to the GC. The unused sample returns to the still. .......................... Schematic of a 10L Arnold Holstein Still]. ........................... Two diagrams of ideal reactors; the plug flow reactor and the continuous stirred tank reactor. The liquid-phase sampler diagram is also shown in comparison. .................... vi ............ 3 ............ 5 ............ 19 ............ 21 ............ 23 ............ 24 ............ 29 ............ 4O Figure 5.2. Figure 5.3. Figure 5.4. Figure 5.5. Figure 6.1. Figure 6.2. Residence time distribution of a plug flow reactor, resulting from a step tracer input. The tracer output signal (F) is zero from the point of injection to the mean residence time (t). At this point the reactor is totally clear out and nothing but the tracer remains in the reactor (F=1)2. . .......................................... 42 Residence time distribution of an ideal continuous stirred tank reactor resulting fi'om a step tracer input. The tracer output signal increases in an exponential form until nothing but the tracer remains in the ractorz. .................................................. 44 The relationship between concentration of ethanol and the total volume exiting the liquid-phase sampler. Sample cuts of 0.25 ml were taken and analyzed by the GC. After about 5 ml of sample has been collected the concentration of ethanol is approximately zero. ........................................................... 45 The residence time distribution of the liquid-phase sampler plotted against the plug flow and continuous stirred tank reactor models at a mean residence time of 20.12 seconds. The liquid-phase sampler analysis is represented by the round plot. The CSTR analysis is represented by the square plot. The PFR analysis is represented by the diamond plot. The liquid—phase sampler appears to behave somewhere in between the two ideal models. ................................... 48 Pressure of the bottom pot on the still as a function of distillation volume. Pressure increases in a linear relationship with volume ................................................................... 52 Temperature as a function of distillation volume on the small still. The temperature levels off during the middle of the distillation with sharp increases both initially and vii Figure 6.3. Figure 6.4. Figure 6.5. Figure 6.6. Figure 6.7. Figure 6.8. Figure 6.9. towards the end portion of the run. This is characteristic of a third order polynomial relationship. ............................................... 53 Time of distillate cuts as a function of distillation volume. 75 milliliter cuts were taken and the time recorded. The data appears to increase in an exponential form. ............................... 55 Flow rate of the distillate as a function of distillation volume. Flow rate has the characteristics of a decreasing in a linear relationship with respect to distillation volume. .................. 56 Chromatogram for the distillation of a brandy-type mixture. The sample was collected fi'om the top tray of the still using the liquid phase sampler. The programs used were the same as in Tables 4.4 and 4.5. Peaks were identified by comparing the retention times from the analysis to standard retention times for each of the compounds. ........................................................................................ 58 The standard calibration curves for ethanol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. ........................ 61 The standard calibration curves for acetaldehyde. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. .................. 63 The standard calibration curves for ethyl acetate. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. ........................ 64 The standard calibration curves for methanol. The static measurement is represented by the diamond plot and the viii Figure 6.10. Figure 6.11. Figure 6.12. Figure 6.13. Figure 6.14. flow measurement is represented by the square plot. ........................ 65 The standard calibration curves for n-propanol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. ........................ 67 The standard calibration curves for isoamyl alcohol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. .................. 68 Ethanol concentration as a fimction of distillation time. Data was taken from the top tray with the liquid-phase sampler and also after distillation. Standard ethanol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. ............. 70 Acetaldehyde concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also after distillation. Standard acetaldehyde calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot ........................................................ 72 Ethyl acetate concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also afier distillation. Standard ethyl acetate calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond ix Figure 6.15. Figure 6.16. Figure 6.17. plot. ................................................................................................... 73 Methanol concentration as a function of distillation time. Data was taken from the top tray with the liquid-phase sampler and also after distillation. Standard methanol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. .................................................................................................... 74 N-propanol concentration as a function of distillation time. Data was taken from the top tray with the liquid-phase sampler and also after distillation. Standard n-propanol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. .................................................................................................... 75 Isoamyl alcohol concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also after distillation. Standard isoamyl alcohol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. ..................................................................................... 76 1. Introduction 1.1 Fruit Brandy Distillation Many challenges are encountered in producing high quality fruit spirits. First of all, fruit brandies contain hundreds to thousands of compounds, which greatly affect the taste and aroma characteristics of the distilled product]. Second of all, some of the compounds present in fruit brandy are regulated due to their toxic effects at high levels of exposurel‘u. Therefore, much research needs to be done on several aspects of fruit brandy production in order to ensure the quality of spirits for consumption4. Brandy is the product obtained from the distillation of wine or fermented fruit mash]. Distilling fruit spirits involves using one of two processes. First, there is distillation without rectification where vapors produced by boiling are condensed and do not return to the stills. This means that there is little or no reflux for the enrichment of the vapor in the still. This French or alambic style of brandy making requires multiple distillations to obtain high proof spirits]. Second, there is distillation with rectification (the use of trays and partial condensers) where part of the condensate returns to the still and comes into contact with the rising vapors. These points of contact are known as stages or trays and they help to make the vapor richer in the higher boiling component5 . It is known that the greater number of trays or stages used, the more separation can be achieved”. This German style of distillation is designed to capture the essences of the fruit and requires only a single pass through the still to obtain high proof spin'ts'. The German batch distillation process will be examined in this thesis. 1.2 Structure of a Batch Still The construction of the still is very important in controlling the various steps of a distillation. A 150 L Christian Carl still is shown in Figure 1.37. There are varying designs of these stills in terms of the number and placement of the trays, however, this figure is a good indication of what is typical for brandy stills. Although design opinions vary, it is agreed that the material used to make up the still must be copper. Copper is a very good heat conductor, shows good resistance to fruit acids, and has catalytic properties that contribute positively to the quality of the fruit spirits produced‘. The basic design includes a pot, a column with trays, and condensers. The pot contains the mash when the still is filled. In this figure, the pot is heated using a steam jacket connected to a steam inlet. Some fruit mashes are rather viscous and pose difficulties with uneven heating, foaming, and baking. Therefore, many stills may contain a stirrer to ensure not only that the mash is evenly heated, but also to reduce the occurrence of foaming and baking of mash to the surface of the pot. Trays can be opened or closed. The particular still contains four trays, one over the pot, and three in the column next to the still4. A still may also contain a catalytic converter, which is a vessel containing a high surface area copper packing used as both an extra tray in distillation and a means for removing cyanide containing compounds from the distillate7. There is a function that allows for the distillate to enter through the bottom of the catalytic converter, or allows for the catalytic converter to be bypassed". Stills contain a condensing column that condenses the distillate vapor into liquid when cooling water is passed through. There can also be partial condensers. The Steam Jacket Partial Condensers Total Condenser Copper Packing - - -Distillation Vapor Path —Cooling Water Path A Fermented Mash lnlet Figure 1.1 .........‘ I i i i Tray Control Levers Cooling Water Control To Drain A 150 L Christian Carl Batch Still7. “35,9.-._._._._._._._..1. -‘lll-‘U----V‘"'C’F ~‘\.~\.-W-sw--i-t-\-~-...-;-._---.__ displayed still contains a partial condenser above the tray over the pot and one above the three trays in the column adjacent to the pot. It is possible to choose to allow water to pass through these condensers, or to leave them off. There is a flow regulator to control the flow rate of cooling water through the condenser and partial condensers. By controlling the rate of cooling water that passes through the condensers, one can control the rate of the distillation4. 1.3 Composition of Fruit Spirits Distillation of fruit spirits is more complicated than just a simple ethanol/water distillation. Although the main components are ethanol and water, there are also hundreds of congeners (components other than ethanol and water) present in varying amounts. In fruit spirits production, the entire fruit is used, including the skins, the pits, the pectin and seeds. This results in various reactions in the fermentation and even distillation and storage that lead to the creation of many congeners“. In fact, fruit spirits contain a greater number of congeners compared to other types of beverages such as whiskey, vodka, or gin. This can be seen in Figure 1.3, which illustrates the typical components in various distilled beverages”. These congeners can either have a positive or negative effect on the taste and aroma of the resulting distillate. The distillation behavior of these congeners depends on factors such as boiling point and equilibrium relationship to ethanol and water. Conditions of the still and the environment are different each time a distillation is run and each batch of mash, even if it is from the same fermentation, tends to have slight differences. This leads to poor reproducibility of spirits from one distillation batch to the next4. Congener Amount (mgl100mL) I Methanol j 2000 « . El Higher alcohols 1800 . , I Carbonyl compounds 1600 - ‘ I Ethyl acetate 1400 . A ~ . DOther esters _j 1200 . I 1000 ~ 1 800 . . 600 I .. j j 400 I l 200 ' H I It i 0° 0% Q Q69” 0 3% Q 0040 r0960 09600 >9? “‘9 A c)“ \e 0" vi 0° 5 (0° 4‘ o 00 {9 08 Q 0060 Q80 9° \0 69 .5? (P 9 Figure 1.2 Milligrams of volatile congeners in distilled spirits per 100 milliliters of alcohol. Fruit spirits have higher congener concentrations than other distilled beverages”. There are various carbonyls, acids, and fusel alcohols present in distilled spirits. Most of these compounds are formed during the fermentation process. The main carbonyl compounds present include esters such as ethyl formate and ethyl acetate and ketones, such as acetone, aldehydes such as acetaldehyde and benzaldehyde. Fusel alcohols, which are alcohols with more than two carbons comprise the largest group of aroma compounds in fruit spiritsg. The most important fusel alcohol present is isoamyl alcohol. Other fusel alcohols include n-propanol, isobutyl alcohol, and amyl alcohol4. Methanol is a major issue in the production of distilled spirits. Although methanol is a desired compound for flavor and aroma, it is regulated for the health risks it poses. Methanol can cause damage to many human organs, such as liver, kidneys, and the nervous system. Methanol can also cause temporary or permanent blindness, or even death3 . Therefore, the United States Environmental Protection Agency has set limits on the amount of methanol that can be present in drinking water at 3.9 parts per million. Methanol in fruit spirits has been regulated in Europe from a range of alcohol depending on the variety of fruit spirit being produced. According to the Bureau of Alcohol, Tobacco, and F irearrns, the limit is 0.35%v/v for all fruit spirits, which translates to 700mg/100mL absolute alcoholz. The idea behind a distillation is to obtain the greatest amount of ethanol with optimal concentration of congeners to produce a product that is desirable in both taste and aroma. To accomplish this, the distillation is typically broken up into three fractional cuts; heads, hearts, and tails. The heads portion contains relatively higher concentrations of lower boiling point compounds and is generally discarded. The hearts fraction contains the optimal amount of ethanol and congeners and is therefore saved as the potable product. The tails contains relatively higher concentrations of higher boiling point compounds and is also discarded due to an undesirable aroma and taste]. 1.4 Quality Control and Analysis Adjustments may be made to control the quality of the distillate. First, appropriate fractionation is important. Separating the distillate into the heads, hearts, and tails is essential. The concentrations of congeners can be controlled during distillation through fractionation of the distillate through the use of trays, the catalytic converter, and partial condenser‘. It is important to know when it is appropriate to perform these various actions during a distillation in order to control the quality. This requires the sensory analysis of the distiller, in other words his or her sense of taste and smell]. This can be difficult as the senses can become overloaded after prolonged periods of exposure. Therefore an instrumental technique may be useful in monitoring the process of distillation“. Gas chromatography (GC) has been used for years to analyze the concentration of congeners present in the distillate]. It is a robust and reliable instrument with low limits of detection capable of detecting trace concentrations”). The drawback to conventional gas chromatography, however, is analysis time, which often ranges from 40 minutes to an hour for distilled spirits. GC has only been useful for analysis after the distillation‘. Advances in high speed gas chromatography (HSGC) have now made it possible to examine distillations using chromatographic methods as an online tool. HSGC can utilize faster flow rates, shorter columns, and sometimes special injection techniques to perform analyses that range from a few seconds to a few minutesl 1. With this technique, it is possible to collect a sample directly from a still and analyze it in a fast enough manner that adjustments could be made to control quality during a distillation4. 1.5 Objective The goal of this study was to examine methods of high speed gas chromatography and how they can be applied to online monitoring of a distillation. An efficient and effective sampling method for extracting sample from a brandy still and injecting it into a gas chromatograph was demonstrated. It was also important to examine the speed and efficiency of the actual chromatographic separation. Finally, the method was applied to an actual distillation in order to monitor compounds as they were volatilized“. 1.6 10. 11. Literature Cited Tanner, H. Brunner, H.R., Fruit Distillation Today 3'“ ed. , (English Translation) Heller Chemical Administration Society, Germany, 1982. Bureau of Alcohol, Tobacco, and Firearms. “Alcohol and Tobacco Newsletter (1) 1998.” Available at . Methanol, Material Safety and Data Sheet. Lindemann, D. “Real Time Monitoring of Fruit Spirit Distillation with High Speed Gas Chromatography”. MS. Thesis. McCabe, Warren, Smith, Julian, Harriott, Peter. Unit Operations of Chemical Engineering. McGraw, Hill, 2001. Wankat, P.C., Equilibrium Staged Separations. Prentice Hall, New Jersey, 1998. Claus, M. “An Investigation of the Relationship Between Tray Usage of the Still and Congener Compound Concentration in Distilled Fruit Spirits”. MS. Thesis. Piggot, J.R., Patterson, A. ed., Distilled Beverage Flavour: Recent Developments. Ellis Horwood Ltd., 1989. Suomalainen, H. and Lehtonen, M., “The Production of Aroma Compounds by Yeast.” J. Inst. Brew., 53, (1) 1972. Skoog, Douglas, Holler, James, Nieman, Timothy, Principles of Instrumental Analysis. Harcourt, Brace and Co., 1998. Sacks, R., Smith, H., Nowak, M., “High Speed Gas Chromatography.” Anal. Chem. 69 (1998) 29A. 2. High Speed Gas Chromatography 2.1 Background and Development The development of high speed gas chromatography began with the advent of open tubular capillary columns by Golay in 19571’2. Open tubular columns have greater permeability and require lower inlet pressures than packed columns. In addition, there is a less rapid loss in column efficiency with increasing velocity3. Since it was more practical to operate at higher flow rates, it was recognized that high speed gas chromatography was possible“. Open tubular columns gained slow acceptance due to competition with packed columns, which had been used for years. Packed columns are rugged and had high resolving power and high sample capacity. The first open tubular columns were made of rigid glass, which made them difficult to install and maintain. In addition, open tubular columns lacked good selectivity compared to packed columns since there were not as many available stationary phases. The advent of fused—silica columns along with other technological improvements made open tubular columns more flexible and viable for everyday usez. Unfortunately, once open tubular columns gained acceptance, the main goal in their application was not speed as chromatographers were primarily concerned with achieving high resolution. In fact, people often achieved much higher resolution than needed by using extremely long capillary columns. The reason for this was that there was a limited range of stationary phases, which resulted in low selectivity. So increasing the column length was the simplest way to bypass this problem. If a column is long enough, one can achieve adequate separation without having to be concerned with 10 stationary phases and temperature or pressure programming 2. Obviously, this leads to an increase in analysis time4. In addition there were technological deficiencies in the instrumentation early in the development of high-speed gas chromatography for which it took years to overcome. First of all, commercial instruments were unable to handle the high speeds. Detectors and other data collection systems were not fast enough to measure rapidly eluting compounds. Also, oven and pressure capabilities were not advanced enough to produce fast and precise programming rates. Additionally, the use of short columns needed for fast separations led to a loss in peak capacity. This shortcoming restricted the complexity of the sample that could be analyzed“ Fortunately, the needs of field portable GC devices and the need for chromato graphs with higher throughput led to greater developments in fast GC technology. Advances in injection systems have reduced the bandwidth of the initial injection to reduce band broadening. Faster detectors, such as flame ionization detectors and time of flight mass spectrometers, can keep up with the faster elution of compounds. In addition, gas chromatographs can be equipped with efficient ovens that are capable of achieving faster temperature program rates. They are also available with more sensitive pressure regulators. The growing need for faster separations combined with several technological advancements have led to the recent growth in the use of high speed gas chromatography“. 2.2 Speed and Separation Unfortunately, everything comes at a cost and high speed GC is no exception. The measures taken for decreasing analysis time ofien decrease sample capacity and 11 selectivity, and also lead to broadening of peaks. Therefore, a compromise between speed and separation must be found to meet the needs of the analysis at hand4. There are three major approaches that can be applied to achieve this compromise. Minimizing the resolution to a value that is just sufficient is one option5 . It was discussed before that peaks are often over separated by using extremely long columns. Instead, one could focus on the peaks or critical pairs that are important for the analysis and optimize just enough to separate those compounds. This is probably the simplest method for separation“. Another option is to maximize the selectivity of the chromatographic systems. Finding appropriate stationary phases, which are selective toward the compounds of interest has been done. It has been shown that more selectivity can mean less concern for column length and method development“. Finally, another approach involves speeding up analysis while maintaining the same resolutions. If the analysis time is too long even with the minimum accepted resolution, methods that speed up analysis without decreasing resolution can be applied. This is a more complicated route, as it often requires major equipment changes such as the use of narrow bore columns or new injectors4. There is really no universal method for high speed gas chromatography, so finding a method that works is a challenge. The application determines the approach needed to decrease analysis time. In choosing an approach, one needs to consider the amount of speed truly needed to accomplish the goal, the required ruggedness of the instrumentation needed, the complexity of the sample, sample phase, and method needed for sample extraction. Sometimes several methods may need to be used in combination in order to achieve the desired goal. Whatever the goal and method used, it is important 12 to look at every aspect of the GC instrumentation, from injection, to columns, to detectors, to temperature and pressure programming, and carrier gas flow4. 2.3 GC Columns Two simple methods for reducing the separation time are to shorten the length of the column and to decrease the stationary phase thickness3'5’6. Shortening the column obviously speeds up analysis by decreasing the distance the sample has to travel. Decreasing stationary phase thickness reduces the amount of time that the sample interacting with the stationary phase. These methods may speed up the analysis, however they also decrease the efficiency. One must be carefirl not to lose desired resolution in changing length or thickness4. Another common method for fast separation while maintaining column efficiency is the use of narrow bore columns. This method would fall under the category of maintaining constant resolution. A narrow bore column is one which has an inner diameter of 0.1 mm, and below, compared to a normal bore which has an inner diameter of 0.25mm and greater. Narrow bore columns make it possible to use shorter columns and faster flow rates while maintaining resolution4. The use of narrow bore columns is widespread, however it is not the universal method for performing HSGC. Narrow bore columns require the use of very small sample volumes often on the order of nanolitersl’3'6. Such small sample sizes are very difficult to create and injection repeatability is also difficult to. In addition, narrow bore columns require much higher pressures than some GC instruments can provide. Method 13 development is also difficult, especially in quality control7. However, it is becoming more convenient to use this method given recent technological advances“. It has also been shown by other research groups that fast gas chromatography can be performed without the use of narrow inner diameter columns. Good separation can be performed using ordinary inner diameter columns with regular injection techniques. In a great number of applications, the efficiency of a capillary column is much higher than necessary. It is believed that normal bore columns can then be shortened to meet the minimum separation requirements of medium complexity mixtures4. Loss of efficiency due to shortening columns can be compensated by carefiilly choosing selectivity in the stationary phaseé. One can attempt to find one specific column that meets the separation needs for the target compounds. However, selectivity can also be achieved using mixed stationary phases or series coupled ensembles of two capillary columnsg. These columns can be designed for specific use with the compounds of interest. Increasing the selectivity this way can allow for the use of a shorter column with a normal inner diameter. Using normal inner diameter columns helps to maintain sample capacity which is lost when using narrow inner diameter columns. Therefore, split ratios do not have to be extremely high or the GC instrument can be run in splitless mode. It can also help to avoid the need for more complicated instrumentation“. 2.4 Injector Considerations Most common injection systems are acceptable for high speed GC if applied properly. One must remember, however, that it is oflen useful to minimize the injection bandwidth whenever possibles. This will limit the amount of band broadening during 14 elution through the column and therefore, aid in separation. With a common splitless injection, sometimes decreasing the amount of sample injected is enough to achieve adequate separation4. Sometimes more sample is needed to decrease bandwidth, so split injection may be used. Many operating systems today are capable of achieving very high split ratios that produce injection bandwidths on the range of millisecondss. 2.5 Temperature Programming Temperature is a factor in controlling the speed and efficiency of a separation. For samples of medium to high complexity with a wide range of boiling points, a programmed analysis might be useful. In order to achieve high speed, it is helpful to increase the rate of temperature programming compared to a regular GC. Modern GC ovens can often produce programming rates of 50-100°C/min compared to the normally used 2-10 °C /min. The one drawback to temperature programming is the cooling and recovery time it takes to return the system back to initial conditions. This could be a disadvantage when fast repeated measurements are required. However, there are recently developed systems that not only have faster heating rates, but faster recovery times as well4. 2.6 Pressure Considerations Pressure is also an important factor in high speed gas chromatography. For example, the use of narrow bore columns requires high inlet pressures in order to operate3’5’6. Pressure programming has even been studied as an alternative to temperature programming in cases where the stationary phase or sample is thermally labile. With 15 pressure programming steeper gradients could be achieved and there is a more rapid return to initial conditionsg. The newest gas chromatographs can be equipped with electronic pressure control, which can achieve both extremely high pressures for narrow bore capillary use and tunable pressure for programming purposes”. 2.7 Carrier Gas Flow Rate Increasing the flow rate of the carrier gas is another way to speed up analysis time in high speed GC. This can be done until a loss of sufficient resolution is detected. Another option is to change the type of carrier gas usedz’s. One of the most widely used carrier gases is helium, has one of the highest rates of diffusion. However, hydrogen has an even faster diffusion rate. The drawback to hydrogen is its explosive nature. Despite this problem, many modern chromatographs can be equipped with safety shut-off features 5 that protect against hydrogen leaks and decrease the chance for dangerous explosions . This can make hydrogen gas a viable option4. 2.8 Detectors With faster elution of compounds from a high-speed GC, other instrumentation needs to be fast enough to handle the speed of separation. Detectors, for example, must have a fast enough response time to detect peaks with extremely narrow widths3'5'7. They also need to have low limits of detection, especially in cases where a very small sample size is used. Some examples of detectors with fast response times are flame ionization, thermal conductivity, and time-of-flight mass spectrometry“. l6 2.9 Literature Cited Sacks, R., Smith, H., Nowak, M., “High Speed Gas Chromatography.” Anal. Chem. 69 (1998) 29A. Klee, M., Blumberg, L.M., “Theoretical and Practical Aspects of Fast Gas Chromatography and Method Translation.” J. Chromatogr. Sci. 40 (2002) 234- 247. Ewels, B.A., Sacks R.D., “Electrically Heated Cold Trap Inlet System for High- Speed Gas Chromatography”, Anal. Chem. 57 (1985), 2774-2779. Lindemann, D. “Real Time Monitoring of Fruit Spirit Distillation with High Speed Gas Chromatography”. MS. Thesis. Korytar, P., Janssen, H., Matisova, E., Brinkman, U., “Practical Fast Gas Chromatography: Methods, Instrumentation and Applications.” Trends Anal. Chem. 21 (9-10) (2002) 558. Bicchi, C., Brunelly, C, “Conventional Inner Diameter Short Capillary Columns: An Approach to Speeding Up Gas Chromatographic Analysis of Medium Complexity Samples.”J. Chromatogr. A 931 (2001) 129-140 David, F., Gere, D.R., Scanlan, F., Sandra, P, “Instrumentation and Applications of Fast High-Resolution Capillary Gas Chromatography.” J. Chromatogr. A 842 (1999) 309-319. Coutant, C., Sacks, R., “Programmable Control of Column Selectivity for Temperature Programmed GC.” Anal. Chem. 72 (2000) 5450-5458. Peters, A., Sacks, R., “Inlet Pressure Programming for High-Speed Gas Chromatography.” J. Chromatogr. Sci. 30 (1992) 643-648. 17 3. Experiment: Discussion of the Liquid-Phase Sampler Setup 3.1 Previous Setups & Results A schematic of the previous headspace sampling setup connected to the still and GC is shown in Figure 3.1. This instrument consisted of an injection valve assembly, heated transfer line, and purge and trap adapter. Carrier gas supplied a constant flow to both the headspace sampler assembly and the GC. The headspace assembly was connected to the still through the sampling port just above the condenser. The sample traveled from the sampling port, to the sample loop in the injection valve, and through a heated transfer line. The purge and trap adapter connected this transfer line to the gas chromatograph where the sample traveled through the column and detector to be analyzed]. This setup required that work be done in the vapor/liquid phase. This created several problems. First, an inconsistent volume of sample was delivered to the CG. The pressure of the still was relied upon to extract a sample volume. With this setup it was very important that a fixed volume be injected into the chromatograph every time for precise analysis of compound retention times and peak areas. Even if a pump had been present, drawing a consistent vapor/liquid sample would still have had its challenges. Second, in order to keep the sample in the vapor phase a heated transfer line was needed. This line had to be kept at a high enough temperature to assure the gaseous phase was maintained. Pressure would also have to be controlled. While this was definitely able to be done it was not the most effective means of going about the experiment. Third, all the required items to make the process work properly made the entire setup very complex and bulky. This made the experiment very hard to move around from one place to 18 Actuator Heated Transfer Line Computer Valve III ..-\. III l.:'- “-1 I..\ III -'*- Temperature Controllers 10 L Holstein Still — - - -— Vapor Path from Still — - — Temperature Control - Heated Transfer Line Carrier Gas Path ----------- Temperature Control - . Valve Trigger Control Valve Figure 3.1 Schematic of the previously used headspace sampling setup connected to the Shimadzu GC-17Al. l9 another. These three main factors contributed in large part to the unsuccessfiilness of this setup. It was clear that a new, more efficient method would work much better then any of the current renditions. 3.2 Liquid-Phase Sampler Setup The new liquid-phase sampler setup involved replacing the current glass plate on the top tray of the column with a newly cut plexiglas plate. The plate measured 66mm in diameter with a depth of 5mm. This plate was needed so that a hole could easily be drilled through the material allowing for the fluid transfer line to pass into the top tray. See Figure 3.2. A small enough hole was drilled in the center of the plate (4mm in diameter) to create a good seal with the transfer line. Another identical plate was created to allow for the return sample to be placed back into the bottom of the still. The transfer line ran from the top tray of the still, through a peristaltic pump, into the liquid-phase sampler, and then back to the still. The line was made of platinum cured silcon tubing. The tubing was purchased from VWR Industries. It had an inner diameter of 0.79mm and an outer diameter of 3.99mm. The total length of the tubing was 122.3mm. This silcon tubing provided both flexibility and a wide temperature range on which to transfer fluids. A variable flow peristaltic pump was used to remove fluid from the top tray of the column and deposit it into the liquid-phase sampler to be analyzed by the GC. The peristaltic pump used for the experiment was purchased from VWR for $169.77. Its dimensions were 16.8cm L x 12cm H x 11.5cm D and weighed approximately 0.57kg 20 10 L Holstein Still Liquid Transfer Line Figure 3.2 Plexiglas Plate A round plate made of plexiglas was created to replace the glass plate in the top tray of the still. A small hole was drilled in the center of the plate to allow for the transfer line to run through it. A similar plexiglas plate was also made for the bottom to allow for sample to return to the still. 21 It consisted of 3 plastic rollers to reduce flow pulsation, prevent siphoning, and eliminate the need for a check valve. Depending on the tubing size it was able to produce flow- rates from 0.4 to 85.0 ml/min. It could support a temperature range of -62 to 260C. The pump provided a constant flow of sample through the liquid-phase sampler. The liquid-phase sampler is the key to the entire setup and is the main reason the experiment is successful. The liquid phase sampler’s design was based off of a regular GC auto injector vial purchased from National Scientific Company. The dimensions of the clear glass vial were 12mm x 32mm. It held approximately 2 milliliters in volume. The vial was modified by a glass shop to create two small holes in its side. This is seen in Figure 3.3. The first hole (the inlet) was made as close to the bottom of the vial as possible so that minimal sample residence time would result. The second hole (the outlet) was created about halfway up the vial. This was so that enough liquid would be present in the vial for the syringe to take up during analysis. Both an inlet and outlet were needed to make sure that a constant flow of liquid would continuously run through the vial. In order to deliver a consistent sample to the GC the AOC-20i Shimadzu Auto Injector was used. It consisted of a 10 micro-liter syringe, a sample holder vial rack, and a computer program to supply the parameters. The vial rack had to be modified to hold the liquid-phase sampler properly. A small piece was cut out of the side of the holder rack which allowed the liquid phase sampler to sit upright and provide sturdiness during the experiment. The auto injector provided the best method of sample injection into the GC. 22 Auto Injector Syringe Liquid GC vial \Xj/ Flow Outlet Liquid Flow Inlet Figure 3.3 i C <9 Schematic of the liquid-phase sampler. Liquid flows into the vial from the bottom inlet filling up the vial. The fluid then exits through the top outlet to keep a constant flow running through the vial. 23 Peristaltic Pump Computer \IIL l I |—_ - [:3 Shimadzu J : GC-l 7A __l Auto Injector E_. 10 L Holstein Still — -- — Liquid Transfer Path from Still to GC — — — Return Liquid Transfer Path from GC to Still Figure 3.4 Schematic of the liquid-phase sampler setup connected to the Shimadzu GC-17A auto injector. Liquid sample travels from top tray through a peristaltic pump and on to the GC. The unused sample returns to the still. The entire liquid-phase sampler setup can be seen in Figure 3:4. The liquid sample travels from the top tray of the column, to the peristaltic pump, and then through the liquid-phase sampler where it can be taken up by the GC auto-inj ector syringe. The unused sample travels from the liquid-phase sampler back to the still where it is redistilled. This whole process is a continuous cycle run throughout the experiment. GC analysis is conducted in the same way is if there were a normal static sample being examined. 25 3.3 Literature Cited 1. Lindemann, D. “Real Time Monitoring of Fruit Spirit Distillation with High Speed Gas Chromatography”. 2003 MS. Thesis. 26 4. Materials and Methods 4.1 Spirits Production Process 4.1.1 Creation of Synthetic Solutions Actual fermentations often take weeks to process, limiting the total number of distillations that can be performed. In order to run multiple batches within a given time period, standard solutions of mixtures were made. These solutions contained similar amounts of congeners present as in most fermentations. Solutions were made in 8 liter batches, enough to make a single distillation. As soon as the solution was made it was used for the distillation. This was to avoid any evaporation or possible compound formation. The standard solutions consisted mainly of ethanol and water. With hundreds of compounds present in fruit spirits it would be difficult to be completely accurate in analysis. Therefore only ethanol and the five main congeners were examined. The congeners used and their volume percentages are listed in Table 4.1. For the liquid-phase sampler experiment the congeners chosen were acetaldehyde, ethyl acetate, methanol, n- propanol, and isoamyl alcoholl. 4.1.2 Distillation Distillation was performed using a 10 L Arnold Holstein still, as shown in Figure 4.1. This still is mainly constructed out of copper and consists of a pot, a steam jacket, three bubble cap trays, a partial condenser, and a total condenser. The pot is where the solution to be distilled resides. It is heated by the steam jacket, which uses electrically heated steam. The bubble cap trays can be opened or closed so that one, two, or all three trays may be activated. Just above the pot is a partial condenser, which can be turned on 27 Compound Volume (mL) Percent v/v Solution Water 27507 91.69 Ethanol 2400 8.00 Acetaldehyde 1 5 0.05 Methanol 24 0.08 Ethyl Acetate 6 0.02 n-Propanol 24 0.08 Isoamyl Alcohol 24 0.08 Table 4.1 Congener types and volume percentages used in the liquid-phase sampler experiment'. 28 ‘ Distillation Vapor Path — Cooling Water In - - Cooling Water Out r. S wrava. nwr aumr kndw. COD aCOC che a l mfiflb wrath taou SPTB :fidfififififififififififijfififififififi Schematic of a 10L Arnold Holstein Stiu'. Figure 4.1 29 or off. The final component of the still is the total condenser, which is constantly cooled by running water in order to condense the vapor to form the resulting distillate. Cooling water is supplied to both condensers by a hose connected to cold tap water. A flow regulator located just behind the total condenser can control the flow of this water manually. This can either increase or decrease the rate of distillate exiting from the still. Distillation was performed first by adding the synthetic solution. Cooling water should fill the condensing column and the partial condenser if it is used. Rectification is relatively poor because of the small size of the still. Therefore, in order to obtain optimal rectification, the partial condenser was used, and all three trays were closed. The electric heater was turned on in order to heat the steam jacket. The distillation was then allowed to run until distillate appears. The distillate consists of mostly ethanol and water. Congeners exit the still at different times during the distillation depending on their boiling points. Therefore, depending on the point in the distillation, the distillate will contain different amounts of congeners and ethanol. Normally, the distillate would be broken up into the heads, hearts, and tails portions. However, to keep the analysis consistent, 75 milliter cuts were made throughout the distillation]. 4.2 Pump Calibration The peristaltic pump was not self calibrating and the flow rates depended upon the size of tubing used in the experiment. Two main settings were used during the experiment, High 6 and Low 6. The pump was calibrated to determine the flow rates at both these settings. For the Low 6 setting a 50mL graduated cylinder was used. The 30 cylinder was filled to the 50ml mark and the time was recorded. Six separate trials were done and the times averaged. The Low 6 setting was determined to have a flow rate of 3.73 mein. In the same way the High 6 setting was determined, only a lOOOmL graduated cylinder was used instead of the 100mL. The flow rate at this setting was found to be 9.77m1/min. Table 4.2 displays the overall results. 4.3 Analysis 4. 3.1 Gas Chromatograph Setup Online analyses were performed using a Shimadzu GC-17A gas chromatograph with flame ionization detectionl. The GC had both a tunable pressure and split injection analysis. All temperature, pressure, and GC parameters were entered through the computer’s analysis software then downloaded to the GC. Everything was directly controlled through the computer. For the liquid-phase sampler experiment, a 12.6 m Quadrex CW-007 Carbowax column was used. This column had an inner diameter of 0.25mm and a stationary phase thickness of 1pm. The temperatures of the injector and detector were 240°C and 255°C, respectively, and remained constant throughout the analysis]. The conditions of the column were set up for the experiment. The GC temperature and pressure analysis parameters are shown in Tables 4.3 and 4.4. The GC AOC-20i Shimadzu Auto Injector was used to inject a sample into the GC instead of manually injecting with a syringe. The auto-injector was controlled through the GC and computer software program. The auto injector parameters are shown in Table 4.5. 31 Low 6 Setting High 6 Setting Graduated Cylinder 50 ml 1000 ml Average Time 804 sec 6144 see Average Flow Rate 3.73 ml/min 9.77ml/min Table 4.2 Pump calibration results for the two main settings used during the experiment. 32 Injector Temperature 240°C Detector Temperature 255°C Max Temperature 399°C Equilibrium Time 0.3 minutes Column Tern erature Pro ram 400C held 0'4 minutes p g Ramp 70°C /min to 115°C Temperature Run Time 1.5 minutes Recovery (Cooling) Time 1.72 minutes Detector Analysis Time 2.0 minutes Total GC Run Time 3.22 minutes Table 4.3 Gas chromatography temperature analysis parameters for the liquid-phase sampler setup. 33 Injector Mode Split Gas Helium @ 600 kPa Column Pressure Program 157 kPa Isobaric GC Run Time 1.5 minutes Set P Pressure 100 kPa Set P Flow 3.0 ml/min Septum Purge 5.6 ml/min Oven Temperature 40 C Velocity 107 cm/sec Column Flow 5.8 ml/min Spilt Ratio 34:1 Split Flow 198 ml/min Table 4.4 Gas chromatography pressure analysis parameters for the liquid-phase sampler setup. 34 Dwell Time 0.3 seconds Number Rinses w/Sample 2 Number Rinses w/Solvent 2 Viscosity Composition Time 1 second Number Syringe Pumps 3 Injection Speed Fast Plunger Suction Speed Low Injection Type Traditional Syringe Injection Speed High Sample Volume Used 0.5 pl Needle Height From Vial Bottom 3.5 mm Table 4.5 Gas chromatography Auto Injector analysis parameters for the liquid-phase sampler setup. 4. 3.2 Top Tray — Online Analysis Acquisition with the liquid-phase sampler setup began by turning on the all three gas cylinders (helium, hydrogen, air) for the GC. The GC and auto injector were turned on. The temperature and pressure parameters were downloaded to the GC. Right before liquid started forming in the top tray of the still the pump was turned on. The pump was turned to its highest speed setting, in order to fill the line (prime it) with sample from the top tray. As soon as liquid started entering the liquid-phase sampler the pump was placed on the High 6 setting and the analysis program was started on the computer. The auto injector proceeded to operate and inject a sample from the liquid-phase sampler into the GC. Separation was best with a specific temperature program for this method of sampling. The temperature program began as soon as injection took place. The initial temperature of the oven was 40°C and this was held for 0.4 minutes. Next the temperature was ramped 70°C/minute to 115°C. The temperature program time was 1.5 minutes. In order to return the oven back to its initial conditions a cooling period was needed. The fastest way to achieve this was by opening the column door to the GC until the oven temperature had reached 40°C. The experiment was set up so that as soon as the oven reached 40°C a new sample was injected. This was adjusted by increasing the detector analysis time to 2.0 minutes, and by adjusting several of the auto injector parameters. The total GC run time was 3.22 minutes. Or in other words, every 3 minutes 13 seconds a new sample could be injected into the GC. 36 4.3.3 Post Distillation Analysis 75 milliliter cuts of distillate were taken from the still. As soon as distillate started appearing from the bottom of the condenser a stop watch was started to keep track of the times of the cuts. At each out several items were recorded. The time of each cut was recorded. The pressure of the still pot was recorded. The temperature above the total condenser was recorded. All these items were recorded throughout the experiment. Following distillation, samples of each 75 milliliter cut were placed into vials and then analyzed with the GC. The same temperature/pressure programs were used as that for the top tray - online analysis. Compounds were identified by their relative retention times during the runs. The retention times for the compound along with boiling points will be discussed in the results]. The concentrations of the six main congeners were measured using individual calibration curves. 37 4.4 Literature Cited 1. Lindemann, D. “Real Time Monitoring of Fruit Spirit Distillation with High Speed Gas Chromatography”. 2003 MS. Thesis. 38 5. Modeling the Liquid-Phase Sampler 5.1 Ideal Reactors The liquid—phase sampler was an important part of this experiment. All the data sampling and analysis was dependant upon the continuous flow of liquid through this vial. Understanding the fluid velocity distribution inside the vessel during the experiment would be very beneficial. This would allow us to relate the liquid that is being pumped into the vial with the sample that is actually taken up by the syringe and then injected into the GC. The behavior of the liquid—phase sampler can be predicted by modeling it after an ideal reactor. Two types of commonly used industrial reactors are the PF R and the CSTR. A PFR, or Plug-F low Reactor, is a tubular reactor. It typically consists of a long cylindrical pipe and is normally operated at steady state. The reactants are fed into one end and the products are withdrawn at the other end. The reactants are continually consumed as they flow through the reactor. Thus in modeling the tubular reactor, we assume that the concentration varies continuously in the axial direction through the reactor]. A CSTR, or Continuous-Stirred Tank Reactor, is normally run at steady state by feeding the reactants into the reactor and continuously withdrawing the products. The CSTR is generally modeled as having no spatial variations in concentration, temperature, or reaction rate throughout the vessel. Since the temperature and concentration are identical everywhere within the vessel, they are also the same at the point of discharge from the vessel]. The two different types of ideal reactors along with the liquid-phase sampler model used in the experiment can be seen in Figure 4.1. 39 Plug Flow Reactor Input >6 Output Mixed Flow CSTR 4., ‘\ W—a Output Input Liquid-Phase Sampler \/\/\/ Output <—--— § g b Input Figure 5.1 Two diagrams of ideal reactors; the plug flow reactor and the continuous stirred tank reactor. The liquid- phase sampler diagram is also shown in comparison. 40 5.2 Residence Time Distribution of Ideal Reactors The residence-time distribution of a reactor is a characteristic of the mixing that occurs in the chemical reactor. In an ideal plug-flow reactor, all the atoms of material leaving the reactor have been inside it for exactly the same amount of time. An atom of material entering a PFR will travel from the inlet to the outlet in a period of time equal to the mean residence time. For a steady-state, constant density system; the mean residence time (t) can be described as the total volume of the reactor (V) divided by the total flow-rate (0)1. The residence-time distribution of a PF R due to a step tracer input can be seen in Figure 4.2. A step tracer is one in which the entire contents of the inlet flow line are changed permanently. The tracer output signal (F) for the PFR is zero from the point of injection to the mean residence time (t). At this point the reactor is totally cleared out and nothing but the tracer remains in the reactor (F = 1). Any input tracer, or stimuli, injected at the inlet will exit in exactly the same manner after a lag time defined as t = Wu. The PFR is one class of reactors in which all the atoms have the same residence time'. In other reactor types, the various atoms spend different times inside the reactor; that is, there is a distribution of residence times of the material within the reactor. For an ideal CSTR the feed introduced at any given time becomes completely mixed with the material already in the reactor. Some of the atoms entering the CSTR leave it ahnost immediately, because material is being continuously withdrawn from the reactor. Other atoms remain in the reactor almost forever, because all the material is never removed from the reactor at one time. Many 41 l—d b------------------------------— F=1 {f ‘ 0.05 0.00 --= -— 0 100 200 300 400 500 600 700 800 900 1000 Cumulative Distillation Volume, ml -r __.. an __T___._. _ _..‘ _ 1-2%-- If...“ f__. Figure 6.1 Pressure of the bottom pot on the still as a function of distillation volume. Pressure increases in a linear relationship with volume. 52 Temperature, Degrees C l 75 l y= 115-07x3 -0.0002x2 + 0.0974x+48.586 l |= R2 = 0.9885 ‘ 70 ~ ‘ l l l 55 1 l 1 . 60 1 j l l 55 1 ~ 50 . —— —.— ———~ —- __. . ——-—. —— -. ——- . -——- —_~ -- .. 0 100 200 300 400 500 600 700 800 900 1000 Cumulative Distillation Volume, mi Figure 6.2 Temperature as a function of distillation volume on the small still. The temperature levels off during the middle of the distillation with sharp increases both initially and towards the end portion of the run. This is characteristic of a third order polynomial relationship. 53 temperature again starts to rise at a rapid rate, exceeding 75°C. This is characteristic of a third order polynomial function and the data fits this type of relationship. Another variable analyzed was time, or time of the cuts. 75 milliliter cuts of distillate volume were taken and the time of each cut was recorded. Figure 6.3 displays the time of each cut as a function of cumulative volume. As can be seen the data tends to rise in an exponential form. An exponential curve was fitted with the data. The R2 value was also shown indicating how well it matches up with the exponential curve. The curve doesn’t fit the data that great, but it does show a general trend in the data. The standard deviations increase as the amount of total volume collected increases. The flow rate of the still was also of interest. Even though the flow rate was set to a specific setting, the distillate volume did not exit the still at the same rate during the experiment. A number of factors such as pressure, temperature, and solution composition affect the flow-rate. The flow rate was calculated as a function of distillation volume and the plot is displayed in Figure 6.4. Not only does the flow rate decrease over time, but it has the characteristics of a decreasing in a linear relationship with respect to distillation volume, as indicated by the R2 value. It starts out initially at a flow rate of around 38ml/min. By the time the experiment had ended the flow rate had reduced to roughly 7ml/min, approximately five magnitudes less then when it started. The standard deviations of the measurements show no direct pattern. 54 Time of Cuts, See Figure 6.3 y = 151 .41e°-°°34" f R2 = 0.9522 l 100 200 300 400 500 600 700 800 900 1000 Cumulative Distillation Volume, mi Time of distillate cuts as a function of distillation volume. 75 milliliter cuts were taken and the time recorded. The date appears to increase in an exponential form. 55 Flowrate, milmin y = -0.0374x + 42.94 R2 = 0.9527 0 ‘i T_‘ i—l 7" TTTC _'_ ifl'“ T" _T_ _' 7 ‘TI_-—_ __ v" ‘ — " "_—_ _ T7 _ _ ' IV" _ ' 0 100 200 300 400 500 600 700 800 900 1000 Cumulative Distillation Volume, ml Figure 6.4 Flow rate of the distillate as a function of distillation volume. Flow rate has the characteristics of a decreasing in a linear relationship with respect to distillation volume. 56 6.2 Method Optimization Fast and efficient chromatographic readings could be obtained with the liquid- phase sampler setup. This was achieved with a shorter column and faster flow rate, along with the appropriate GC programming]. The GC temperature and pressure programs were shown previously in Tables 4.4 and 4.5. A temperature program was the best at achieving adequate separation. The elution time was approximately 1.5 minutes. Since a temperature program was used, time was needed to cool down the system to initial conditions]. This took approximately 1.72 minutes. The total run time of a sample acquisition was 3.22 minutes. The chromatograrn that resulted from this type of analysis is shown in Figure 6.2. As can be seen ethyl acetate and methanol come off extremely close to one another. Using isothermal conditions these two peaks would not be separated. Peaks were identified by comparing the retention times from the Chromatogram to standard retention times for each of the compounds'. A table with the retention times for the flow vial experiment is shown in Table 6.1. 6.3 Standardization All six main compounds needed to be standardized. A calibration curve for each compound was made. While producing the calibration curve for ethanol it was discovered that actually two calibration curves were needed for each compound. One set of curves was needed for post-distillation (static) analysis. This would be the normal way of analyzing the data. The other set of calibration curves was needed for analysis with the liquid-phase sampler. The movement of liquid through the vial did affect the 57 mV 1000 4 750 - 5 6 500 — 1 2 250 - 3 a. ii A I l T 0 0.5 1.0 1.5 2.0 Time, min 1) Acetaldehyde 2) Ethyl Acetate 3) Methanol 4) Ethanol 5) N-propanol 6) Isoamyl Alcohol Figure 6.5 Chromatogram for the distillation of a brandy-type mixture. The sample was collected from the top tray of the still using the liquid-phase sampler. The programs used were the same as in Tables 4.4 and 4.5. Peaks were identified by comparing the retention times from the analysis to standard retention times for each of the compounds. 58 Compound Retention Time (min) Boiling Points (°C) Acetaldehyde 0.355 :t 0.003 20.8 Ethyl Acetate 0.727 :t 0.005 77.0 Methanol 0.774 t 0.003 64.7 Ethanol 0.928 t 0.004 78.0 N-propanol 1.150 t 0.014 97.0 Isoamyl Alcohol 1.609 d: 0.009 132.0 Table 6.1 Retention times for the chromatograph taken using the liquid-phase sampler. Peaks were identified by comparing the retention times from the chromatograph to standard retention times for each of the compounds. 59 calibration data. Therefore a total of twelve calibration curves were needed to analyze both top tray data and post-distillation data. Ethanol is by far the main component of any alcoholic beverage and, therefore, one of the most important to examine. To develop the calibration curve for ethanol sixteen standards were made over the range of interest. Standards were created at 30%, 35%, 40%, 45%, 50%, 55%, 60%, 65%, 70%, 75%, 80%, 85%, 90%, 95%, and 100%. To produce the static calibration curve these standards were placed in 2 milliliter GC auto sampler vials and run with the GC using the parameters and temperature/pressure programs give in Tables 4.3, 4.4, and 4.5. To obtain the flow calibration curve these same standards were pumped through the liquid-phase sampler at 9.77ml/min (high 6 setting). This flow rate was chosen because it is the same flow rate at which the liquid- phase sampler experiment was conducted. Data was taken with the GC using the same software programs as before. Both the static and flow calibration plots can be seen in Figure 6.5. In both plots the peak area increases in a linear relationship with respect to the ethanol concentrations. The equations describing these linear trendlines are also supplied. The R2 values (20.99) and the standard deviations for each set of data, as seen on the graph, signify a good calibration. As can be seen the flow calibration data produced greater peak areas then the static calibration data at the same ethanol concentrations. This turned out to be the common theme in several of the compounds. Standardization with acetaldehyde proved to be quite challenging. Due to acetaldehyde’s low boiling point (approximately 208°C), its viscosity, and continuous evaporation in the hood it was very hard to produce volumes of accurate amounts. 60 Peak Area 9.0E+05 -. JI-rw~~- 8.5E+06 8.0E+06 i 7.5E+06 i 7.0E+05 i 6.5E+06 J 6.0E+06 -« 5.5E+06 j 5.01=.+05 l1 4.5E+06 4.0E+06 i—r —— 25 Figure 6.6 y(flow) = 47286x + 4E+06 R2 = 0.9938 y(static) = 44993x + 4E+06 R2 = 0.9953 oStatic Data l ‘ 1, [-———-——~—— --~ - - —v—-—-— ~ -—T——— -'—-—~ »-—.~— “m — 1~~——- - "v- ‘ - ‘~———— ‘ ~--- 35 45 55 65 75 85 95 Ethanol Concentration %VN The standard calibration curves for ethanol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. 61 - Flow Data . Syringing small samples of liquid acetaldehyde was just about impossible. In an attempt to solve this problem larger standard samples were created. The standards were also made up according to mass instead of volume. Using an analytical balance five standards were made over the range of interest. Volumetric samples, by mass, were made at 0.1%, 0.3%, 0.5%, 0.7%, and 1.0%. These were run in the same way as ethanol, with the two different methods (static and flow). The calibration curves for both can be seen in Figure 6.6. The R2 values are again really good, but the errors in measurement are worse. As the concentrations of acetaldehyde increased so did the standard deviations. Also it can be seen that the flow data is not necessarily above the static data as it was for ethanol. This is probably just an error in attempting to make good acetaldehyde standards based on the problems given above. Ethyl acetate and methanol were analyzed together. These two compounds have GC retention times very close to one another and thus their peak areas overlap considerably. It was decided that it would be best to standardize them together to model the synthetic solutions after actual brandy-type solutions. Standards were made at 0.1%, 0.3%, 0.5%, 0.6%, 0.7% for ethyl acetate and 0.3%, 0.4%, 0.5%, 0.7%, 1.0% for methanol. These standards were run with the GC using the static and flow methods. The calibration plots can be seen in Figures 6.7 and 6.8. As is expected the R2 values in general are not as good as the other compounds. Because of the peak overlap the software program must make certain approximations as to where the dividing line should be, and thus peak area approximations. These approximations obviously create error; and this can be seen in the calibration curves for both ethyl acetate and methanol. 62 Peak Area 4.0E+05 l l 3.0E+05 j l 2.0E+05 . l 1.0E+05 0.0E+00 i 0.0 Figure 6.7 y(flow) = 481795x + 30253 R2 = 0.9938 F ____._ ._ - 0 Static Data I Flow Data ___._ ._T_____ _—'I ___ ___ 0.4 0.6 0.8 Acetaldehyde Concentration %VN — ' I. 1.0 y(static) = 534590x + 13154 R2 = 0.9947 The standard calibration curves for acetaldehyde. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. 63 Peak Area 6.0E+05 i l i l 4.0E+05 j j 3.0E+05 a 2.0E+05 1 l 1.0E+05 j l 0.0E+00 r _— 0 Figure 6.8 y(static) = 824857x - 51033 I R2 = 0.9598 y(flow) = 636217x - 6991.7 R2 = 0.9844 j 0 Static Data j l - Flow Data j L7 -__. . . 7, l 0.1 0.2 0.3 0.4 0.5 0.6 0.7 Ethyl Acetate Concentration %VN The standard calibration curves for ethyl acetate. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. 64 l 0.8 Peak Area 4.0E+05 l l 3.01305 1. l l 1.0E+05 0.0E+00 4 ~._ 0.0 Figure 6.9 y(flow) = 426754x + 98218 R2 = 0.958 y(static) = 497046x + 47555 i R2 = 0.9974 , l l 1, LL -___ _, L l , oStatic Data j j j I Flow Data l 1 _WM , .-..-_ ”In- . _.___._ _.._ ‘..______..._ L. _. l 0.2 0.4 0.6 0.8 1.0 1.2 Methanol Concentration %VN The standard calibration curves for methanol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. 65 Five samples were made for N-Propanol. These standards were made at 0.1%, 0.3%, 0.5%, 0.7%, and 1.0%. The static and flow methods were used with the GC to analyze the data. The results are displayed in Figure 6.9. The R2 values and standard deviations are reasonable. The calibration curve for the flow method is again above that for the static. This is similar to the calibration data for ethanol and is expected. Isoamyl Alcohol was that last compound to be analyzed. Five samples were again prepared over the range of interest. Standards were made at 0.1%, 0.3%, 0.5%, 0.7%, and 1.0%. These were analyzed with the GC and the appropriate static and flow calibration curves were produced. The plots can be seen in Figure 6.10. R2 values and standard deviations are again good. The flow data resides above the static as in the previous compounds. 66 Peak Area 1.4E+05 - a v a L, v v a the i r f r r f __ —————~———~-—~ l 1.2E+06 l l y(flow) = 1E+06x + 100747 N j R2 = 0.999 1 1.0E+06 j l l l 8.0E+05 j j l l 6.0E+05 i y(static) = 978417x+ 92690 “ l R2 = 0.9949 4.0E+05 l ‘ i - l ‘ ‘ Static Data 2.05+05 i l ° i l 1 ;_ _ ___-___ m , 0.0900 Lua —-—e—— -——-—-—-~ +--—— _,______ —.——-———— -— ——..—t 0.0 0.2 0.4 0.5 0.8 1.0 1.2 N-Propanol Concentration %VN Figure 6.10 The standard calibration curves for n-propanol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. 67 Peak Area 1.6E-I-06 1— Jr--— --— —————— -—— ——- -A~—— fl..— -._-.# ___ -W _..___.-____- 1.4E+06 l l 1.2E+06 , l y(flow) = 1E+06x + 20555 t l R2=0.999 I 1.0E+06 - 8.0E+05 «‘ j y(static) = 1E+06x + 26751 60805 j R2=0.9952 4.0E+05 j __ I- __ --___ L I 20E 05 ‘ FioStaticData . . + '1 . l o ' I Flow Data I l a a- »— i 0.0E+00 - --——- —-—- . ___ . a- L- - _._ __ I 0.0 0.2 0.4 0.5 0.8 1.0 1.2 isoamyl Alcohol Concentration %VN Figure 6.11 The standard calibration curves for isoamyl alcohol. The static measurement is represented by the diamond plot and the flow measurement is represented by the square plot. 68 6.4 Distillation Behavior The behavior of all of the compounds during the distillation was studied using the liquid-phase sampler setup. Chromatography spectra were taken using this setup. The concentrations for each of the compounds as they materialized in the top tray were calculated by comparing the peak areas of the congeners to the peak areas of their respective calibration curves. These concentrations were then plotted to show percent volume of congener vs. distillation time. A post-distillation analysis was also performed on the cuts of distillate using the same GC programming as that for the top tray data. This was done in order to compare the behavior and amounts between the congeners present in the top tray and those present after distillation. The trends between the two sets of data were shown. The major component of the synthetic brandy solutions was ethanol. Ethanol measurements were taken during the distillation with the liquid-phase sampler and also on the 75 milliliter cuts alter the distillation. The concentrations of ethanol were found using the standard ethanol calibration curves. The behavior of ethanol for both methods during the distillation and how they relate to each other is shown in Figure 6.11. The post-distillation data shows ethanol starting out near 85%v/v ethanol. The trend slightly increases initially and maintains a concentration of about 90%v/v for the first 20 minutes of the distillation. The concentration then starts to drop off towards the end of the run, eventually reaching about 40%v/v afier 45 minutes. The top tray data shows ethanol starting out at about the same concentration as that of the post—distillation, 85%v/v. But instead of increasing initially, it proceeds to decline over time and continues to decrease 69 Ethanol Concentration %VN 120 -. —~-—- 110—1 100 _ Figure 6.12 Ethanol Behavior F—o— Post-Distillation Data I j 5 + TOP Tray - Online Data ; ‘ 1 0 1 5 20 25 30 35 40 45 50 Distillation Time, min Ethanol concentration as a function of distillation time. Data was taken from the top tray with the liquid-phase sampler and also after distillation. Standard ethanol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. 70 throughout the distillation. The data drops off rather sharply compared to the post- distillation data and is practically all but gone from the top tray after 30 minutes. The first major congener to emerge from the GC column is acetaldehyde. Figure 6.12 shows the behavior comparison between the top tray measurements and the post- distillation measurements. Both plots show a sharp decrease in concentration over time. Only one measurement could be obtained from the top tray due to the speed at which acetaldehyde passed through the system. Ethyl acetate was examined next. Its behavior can be seen in Figure 6.13. The top tray and post-distillation measurements again show similar trends. Both show a sharp decrease in concentration over time. After about 10 minutes ethyl acetate has blown through the top tray. After 15 minutes it can be seen to have left the still completely. The behavior of methanol is shown in Figure 6.14. The post-distillation plot shows methanol initially decreasing. Then towards the middle of the distillation it begins to show an increasing trend. The top tray data shows a pretty sharp decrease in methanol concentration for the first 5-10 minutes. After that though, it decreases in a more gradual fashion until the end of the distillation. Figure 6.15 shows the comparison of plots for n-propanol. Overall the plots for both the top tray and post-distillation tend to agree with each other. They both show a gradual decrease in concentration over time. However in the post-distillation data the concentration increases briefly at first before falling. Isoamyl alcohol came of the GC column last. Its behavior is seen in Figure 6.16. These two plots are very similar in comparison. Both show a slight increase in concentration and then a gradual decline over time. What is interesting about this set of 71 Acetaldehyde Concentration %VN 3.0 1 ——~ —— ”w —— _..-__.E LE ,v Acetaldehyde Behavior ‘w"—’ l l _._ Post-Distillation Data-raj j + Top Tray - Online Data j j 0.5 +— 0 5 Figure 6.13 1 0 1 5 20 25 30 35 40 45 50 Distillation Tme, min Acetaldehyde concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also after distillation. Standard acetaldehyde calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. 72 Ethyl Acetate Behavior 1.6‘1- ‘ww— WE‘LL A. l l 1.4 —" I _ g j 1' +Post—Distillation Data j 1.2 j +Top Tray-Online Data 1 l Ethyl Acetate Concentration %VN ‘0.2 +7 fl‘ " ‘1 1‘" r “ . ' -'*—~———‘r -- -- 5 10 15 20 25 30 35 40 45 50 Distillation Time, min ._1___ Figure 6.14 Ethyl acetate concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also after distillation. Standard ethyl acetate calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. 73 Methanol Concentration %VN Methanol Behavior 12 E_-L____._______ _# +. # #m* ,_.,A 1.0 0.8 1 k ._ // 0.6 7' 0.4 l i eo—Post-Distiilation Data i . I 0.2 l +Top Tray-Online Data l l l l 0.0 «l l l l 02 s--— «a —s 9*. _ . «H- , —. — ...—— —— ._.. ___ O 5 10 15 20 25 30 35 4O 45 50 Distillation Time, min Figure 6.15 Methanol concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also after distillation. Standard methanol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. 74 N-Propanoi Concentration %VN N-Propanol Behavior 2.0 —~ _._ — -._ _‘____ ___.._ ___ l 1.8 1' ,_ ., . * l ; +Post—Distillaiion Data 1.6 i ' ,_.______ *h __—‘..._—_+—.. .! _ 1r i —-— Top Tray - Online Data ! / ._ .. . _- __L, LL. 0 5 10 15 20 25 30 35 40 45 50 Distillation Time, min Figure 6.16 N-propanol concentration as a function of distillation time. Data was taken from the top tray with the liquid- phase sampler and also after distillation. Standard n- propanol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. 75 Isoamyl Alcohol Concentration %VN 2.7 ~- - ~* 2.4 2.1 - I 1.8 * 1.2 4 0.9 0.6 *1. 0.3 i 0.0 Figure 6.17 Isoamyl Alcohol Behavior If S;— Post-Distillation Data 1 ’ _.__ Top Tray - Online Data . Distillation Time, min Isoamyl alcohol concentration as a function of distillation time. Data was taken from the top tray with the liquid-phase sampler and also after distillation. Standard isoamyl alcohol calibration curves were used for each analysis. The online measurement is represented by the square plot and the post distillation measurement is represented by the diamond plot. 76 plots is that at the start the concentration from the top tray is higher then the post- distillation initial concentration. 77 6.6 Literature Cited 1. Lindemann, D. “Real Time Monitoring of Fruit Spirit Distillation with High Speed Gas Chromatography”. 2003 MS. Thesis. 78 7. Summary and Conclusions Previous problems with the headspace sampler setup lead to the development of the new liquid-phase sampler setup. This setup integrated many of the same ideas and techniques as before but allowed for a consistent sample transfer from the still to the GC. This created a successful high speed gas chromatography setup which was capable of analyzing distillations in real time. Understanding the flow process happening within the liquid-phase sampler was important. The relationship between the sample being extracted from the still and that being injected into the GC can be seen through liquid-phase sampler’s residence-time distribution. Its behavior with respect to a plug flow reactor and a continuous stirred tank reactor was observed. The residence-time distribution of the liquid-phase sampler resides somewhere in between the two ideal models. It does not follow the clear cut PFR distribution nor does it in any way adhere to the exponential form of the CSTR. The liquid-phase sampler appears to match up more accurately with the PF R model overall. In the beginning stages before the mean residence time and towards the end it is basically equivalent with the PFR. Around the mean residence time it deviates minutely from the PFR model, not following the clear cut ideal pattern. This is a result of the mixing happening within the vessel, reacting like a CSTR. The passage of liquid through the liquid-phase sampler appears to be uniform with a minimal dead zone where molecules might accumulate over time. Modeling the liquid-phase sampler after the PFR would be a decent approximation. While not absolute ideal, the data provides a good indication that the liquid-phase sampler is a good receptacle for GC analysis sampling. 79 The variable conditions of the IOL Arnold Holstein still were recorded to confirm its reproducibility. The pressure, temperature, time, and flow rate can be duplicated in a precise manner over several trials. A linear relationship is shown to exist for pressure and flow rate as a function of distillation volume. The cut times verses distillation volume expressed an exponential relationship. Temperature as a function of distillation volume exhibited a third order polynomial relationship. While this may suggest the still conditions are reliable, it does not necessarily imply a high quality control of distillation product. For this, compound composition and concentration were analyzed in detail. Calibration of the six main compounds was successful. To accurately represent the data it was determined that actually two separate calibrations were needed, one for flow sample and one for static sample. The standard solutions resulted in a good set of calibration curves for both the static and flow data. A repetitive trend was seen in the calibration. For ethanol, n-propanol, and isoamyl alcohol the flow set of data was clearly above that of the static data for a given concentration. For the other three compounds (acetaldehyde, ethyl acetate, and methanol) a similar occurrence was expected between the flow and static, but errors in program analysis of the peak area and preparing accurate standard solutions prohibited in showing this type of correlation. The reason for this type of relationship is unclear. It could be the result of the movement of liquid during analysis, creating slight temperature, pressure, and volume differences. It is plausible that a larger sample is being injected into the GC during the flow analysis then during a normal static injection. The volume of liquid in the liquid- phase sampler during flow analysis was not equivalent to the volume of liquid present in the GC vial during static analysis. This may have created some differences in injection 80 sample size. Also, the auto injector syringe may take up slightly more sample due to the flow of liquid through the liquid-phase sampler, thus forcing more solution into the syringe needle during injection. The larger sample size would create larger peak areas. Analyzing various standards at several different flow rates would provide a good indication about what is occurring here. The distillation behavior of ethanol, acetaldehyde, ethyl acetate, methanol, n- propanol, and isoamyl alcohol in the top tray was studied with the liquid-phase sampling method. In addition, the distillate was studied afierward under the same GC conditions for comparison. The amounts and the behavior determined from the top tray did not correspond directly to those of the distillate measured with the GC after distillation. Some general trends were observed. All of the top tray concentrations were shown to be less then those taken from the post-distillation. This was true from start to finish. In fact only once, in the case for isoamyl alcohol, did the concentration in the top tray exceed the post-distillation data. The top tray concentrations all decrease over time, except for isoamyl alcohol where there is a slight increase initially. This is not the case for the post- distillation concentrations of certain compounds. For acetaldehyde and ethyl acetate the decreasing plots are expected in both the top tray and after distillation. These two compounds reside in the still for a short period of time, due to their low boiling points, and come off rather rapidly. For the other compounds the fall and rise of concentrations observed post-distillation does not seem to directly correlate with those in the top tray. The top tray concentrations should give an indication of what to expect in the post-distillation data. The two sets of data should not be exactly similar, but should show some type of connection. Looking at the plots, there 81 may be a small indication due to the change in slope for the top tray data. It is hard to make any conclusions based on this though. In general, no specific relationship could be seen from these results. For the most part, all the top tray data simply decreased over time. Several reasons for this type of behavior are possible. First, the data observed could be the actual relationship that the top tray produces as the compounds pass through it. While this may be the most obvious explanation it does not seem likely. The top tray concentrations would be expected to show some type of relationship with the post- distillation data. Second, the calibration data may be faulty. Again this may be the case, and the reasons for this were described above. But overall the calibration results obtained do indicate a good standardization. Third, and probably the most likely, the Arnold Holstein still could be poorly constructed. The proportion of the column to the size of the pot indicates an engineering defect. Smaller stills in general do not produce the same quality of spirits as larger ones are capable of. The rate at which the distillation proceeded confirms this proposition. Attempts were made to slow down the speed of the still, changing pressure and distillate flow rate, but this seemed to have little effect. As a result, during distillation, all compounds were basically blown right through the column, with little or no rectification. This in turn would show a top tray concentration reduction for all compounds throughout the distillation. Data analysis on a larger, higher quality constructed still would be needed to confirm this hypothesis. In conclusion, the liquid-phase sampler proved to be a very feasible setup for high speed gas chromatography. This method achieved rapid sampling and adequate separation of all important compounds. The method produced chromatograms in a time much faster than a conventional GC so that measurements could be obtained every few 82 minutes during the distillation. Behavior of the compounds could be analyzed as they passed through the top tray of the column. More studies involving different distillation apparatuses and setup improvements are required to further the pursuit of monitoring fruit spirits in real time using high speed gas chromatography. 83 8. Future Work The liquid-phase sampler setup has been proven to work effectively. The next step would be to try out the setup on a larger, higher quality still. A properly engineered still would provide improvement in the results. Obtaining data from such a still would perhaps better show the unknown relationship between the top tray and the post- distillation data. Each of the individual compounds could then be studied more in depth. It would be beneficial to observe the other trays on the still to see how they compare with the top tray as well as with the post-distillation data. Observing the still pot while the distillation is occurring might also be profitable. Using these liquid-phase sampler results, manipulation of the still to achieve higher product quality should be examined. Actual fruit fermentations and distillations could be done and analyzed with the liquid-phase sampler setup. Synthetic solutions are great to do research with, but acquiring data from the actual fruit spirit distillations would be something of interest. Higher speed and resolution are always being strived for. As high speed gas chromatography becomes more common and available better techniques will surface, resulting in better performance. This perhaps will allow for the development of an even faster sampling rate then the 3.22 minutes now achieved while maintaining compound resolution. It might be adventitious to find a new technique that can separate more then just the six main compounds. This would probably require changing several GC parts, parameters, and programs. With hundreds of compounds present in any given fi'uit spirit, a way to examine these smaller concentrations as they move up through the trays would be of great use. 84 uijijjjjtjjjjjujijjju