-IAV — _-—d7v ~ I'— <_,_.,__, A A ' 9qu“ ""I III ”II: VI, ' ...I ' ~ I IIII‘MI IIIII III‘II 0‘le I I II IIJJJ‘JJJJ' I I '. I I, ,III I _ III“ ‘WJ. lIJII I J EJJJJIJEII w .I JoJU IJ’III . I II Iv JIJ‘JIJ”.‘QI JJJ’J' IIIIJlII “fl” "6:: fi‘mph“ I JIJIIJJ JJIJIJI.JrJJ II .I 'I' J'JJII'.JJJJ IIII. JJJJJJJI ~——“n — —n A ' __—.-u-" P J JJIII. “J JJJJJJ JJ‘JJJJJJJJJJJJ II'JJ‘JJ'J I. JIJIJI JIJIIIII ‘JJJJJJJJJJ IIJIJJJIJJJ J JJJJIJJJI JJJ JJ JJJJI IJJIJ ‘J J JJJJJJ J JJ I I I I I I .II III .‘I III JJJJIJS JJJ. JIJIJI IIIJ IIIIIII I HJJIJJIIIJI JIII I IIIIII. I IIJI .:III N IJJJJJ JJJJIJI IIIIII JIJ'J J JJ JJ JJ‘IJIIJIII JJIJJIJIJJ IJJJJ IJIJIJJJIJJIJ I'II JIJI IIJJ II I ‘ JIIIII IJJJ J JJJ‘ I JJJIJJJJJJ'J IJJJJJJJ IJJ IJJ I "III? I IJJ‘JI' JIJ III III JIIII-IIIII III. III-u.» II . I “Mm JIIIJJJIJIIIIJI I. ".I ..I I MI‘I II .J JJ‘JJ I'IJJIJJI IJJJJIJJ'JJJJ JV‘ JJJJJ JJM“ -. I J I IM’I III JJJIIIJJ. JJJI JJJI JJ'JJ.‘ ‘JJJJ' JJ ~.,J.J"I'.I:JJJ JJILJJJJJ J I 'I " JJ QJ Jd.‘ PHIZ‘JJJJJ “I 0‘ H J, I II J ‘. I' .J“ fiJJIJJ‘zr NJ: I «3‘ JJ J'JJJ‘J' ‘JIIJJJ JJJIJJJ ' IIIJMJJ‘JI JJJJJJJI J , I J. J J J‘ J JJJJIJJJJ J L 'J ‘IJJIIJJJ-j’ J I Is I. I I. 1‘ . I I l I I III I “I‘lJJ ‘ 4J1” JJJJII J JJIJJJJJIJ III JJIJJJIJ I‘. JIJJ . :IJI'II JII [J I I I I IJIIIII‘;~” J I I I I 'V I ‘ I J :19] I I I I , . I I I JJ'IJ‘JJIJJ J JIIIJJIJIIIJJ:J‘IJ"J‘J~J JJIJI II I JIJJJJ JJI'I'I-JIJ‘IIIJ...II; JFJJJ II II J {I I.JJJI'¥?‘:I II I II I IIIII‘I-‘A‘II II .IIIJ ’JJIJJ'JJJ III III IIIIIIJII’PI I I; III ‘ IIIIII-III II. ‘ I ..I ' Ij'JIIII I ‘III': III JIJIII‘J IJIIIJJJJI'IIJIIIJIJIIJ' I IIIIIIIII-JII II'I‘ II IIIIIIIIII I'm” MI 1‘ I JJJIIJJ JIJIIIJJ JJJIJJJJIJJ IJJVP‘IJIquiy-lI-II JI'IJ-‘IJ‘ III JJJ'I'JJJ III III .5 I III J I JI'I In "J . JJJJ’ JJ "J I . J“ .JJ JIIIIII . I' .IJ.I II: JJJJ I. JIII' IJJ'JJ'IJ J I .I I! “J" II”; II iJIJIu-HII I JJ1I I M “II”; JJJJJJJJJIJJIJI JJJmlJJJJJIJIJIJJ J JJJJI JJJ J J I II JJJJ'I'I‘J' II‘._JI JJ'J JJJJJJ IJJJ JI JJIJJJJJJ IJIIIJ IJJJ JJIJJI I IIII JJJI JJJI I "'I II"I"I'J J‘J:IJI III (I I.JI|JI JJ'JIJJJIJIJJ ,IJI ‘IHIII JJIJ'J IIIIIIII IJJJJJJJ JJIJJ . {II J. JJJI | IJ‘JJIJJJ ‘IIIII' J‘JJJJIIII JJ J IIIJJ “I. JII I J I II IJJJJJJJJJJJJJJJ IJJ J I JJJJIJ JJ IIII J ‘JIIII J J JIJIJJIJI I II .JJ-IJ , I IIIJI IIIJ III‘III JJIJJJI' IIIIJJ' " ‘ ‘IIII IIIII‘ III III. .I ' -Mm-'t ’ V 1...) k ' ‘_". lllllll ”as i 644 +9 lllllllzllllllllll lllll 1006523 l l wchigan 5cm This is to certify that the thesis entitled Application of an automated stopped- flow system in clinical analysis presented by Wai Tak Law has been accepted towards fulfillment of the requirements for PhD Chemistry degree in fizz (4% Major professor Date lzr/llv/Z’ 0-7 639 u ‘ 4am.“- - it? ”Gila”; f." ' ,elz-‘Zifiilléfifig W .‘3 - 4 4, OVERDUE FINES: 25¢ per day per item RETURNING LIBRARY MATERIALS: \— Place in book return to remove charge from circulation records APPLICATION OF AN AUTOMATED STOPPED-FLOW SYSTEM IN CLINICAL ANALYSIS By Wai Tak Law A DISSERTATION Submitted to Michigan State University in partial fulfillment of the requirements for the degree of DOCTOR OF PHILOSOPHY Department of Chemistry 1978 ABSTRACT APPLICATION OF AN AUTOMATED STOPPED-FLOW SYSTEM IN CLINICAL ANALYSIS By Wai Tak Law Stopped-flow mixing with speetrophotometrie detec- tion has several desirable features that makes it a suit- able technique for fast clinical analysis. Mixing times are as low as a few ms, which enables analysis to be done in very short times by fast reaction-rate methods. A typical application is the reaction-rate determination of serum total protein by the biuret method. With the stopped-flow method reported here, the analysis time was 10 s for each sample, which is a lOO-fold decrease when compared with the normal equilibrium method. Preliminary kinetics studies of the biuret reaction were performed with a vidicon detector and the results provide a starting point for future mechanistic investigations. The short mixing time of stopped-flow systems is also advantageous for equilibrium-based applications. Serum albumin reacts very rapidly with bromcresol green. By taking the equilibrium measurements early (2-6 s after mixing), interferences from other slower reacting components Wai Tak Law are greatly reduced. 0 The automated stopped-flow system was also adapted for serum lipid analysis. Highly specific enzymatic methods of analysis for serum triglyceride and cholesterol were used, and the results obtained compared to other standard methods. Serum lipoprotein analyses were also performed. The complete serum lipid profile, including the increasingly popular serum high density lipoprotein (HDL) analysis, can be completed in about half an hour. Again the analysis time is reduced, in this case by at least lO-fold,in comparison to other prevalent methods. ACKNOWLEDGMENT "And we know that all things work together for good to them that love God . . . . ." Romans 8:26. As one goes through the disciplinary process of formal education, it is sometimes hard to remain convinced of this. These last five years have not been altogether smooth sail- ing. The days spent grinding out problem sets, the nights lost debugging programs, the hours spent brooding over me- chanical problems and experimental failures, not to mention all the despair, expasparation and frustration that accom- pany these. But through it all, this particular knowledge has been my mainstay, bouying me up between those scattered moments of eurekas, making it possible for me to culminate my schooling in this volume. Through all this, I am most thankful to my advisor, Dr. Stanley R. Crouch for his guidance. His willingness to give me the freedom to do what I want, but offering help when needed, built my confidence and independence in doing research. His sincere concern for my career and the pat— ience he showed me during the writing of this thesis are deeply appreciated. I would also like to thank the other members of my Ph.D. Committee: Dr. J. L. Dye, who served as my second 11 reader, Dr. J. Holland and Dr. A. Timnick, for their com- bined effort in seeing me through this project. I am also most obliged to Dr. Hans Lillevik for introducing me to the fascinating field of clinical chemistry. Then I wish to thank Dr. Roy Gall, who has been an excellent research partner all these years; always there to help when the stopped-flow system failed, and providing supportive comradeship when all else fails. Next I would like to acknowledge Michigan State Uni- versity and NSF for their financial support, Mr. Wai Kin Law and Ms. Elite Wong in preparing the manuscript, and Mrs. Peri-Anne Warstler for typing the final copy. Then, my most grateful thanks go to my parents; not only for their unswerving emotional and financial support, but most of all for all the years of their lives poured into mine, helping me become me. I also want to thank my brothers and sisters of the MSU Chinese Christian Fellowship for all their prayers and concern. And finally, I thank God, my Heavenly Father, for being my help and stay, and for reminding me time and again that all things do work together for good to them that love Him . . . 111 TABLE OF CONTENTS Chapter Page LIST OF TABLES. . . . . . . . . . . . . . . . . . .vii LIST OF FIGURES . . . . . . . . . . . . . . . . . . x I. INTRODUCTION. . . . . . . . . . . . . . . . . . 1 II. BACKGROUND . . . . . . . . . . . . . . . . . . A A. Clinical Analysis and Instrumentation . . . . . . . . . . . . . . A 1. Common Clinical Tests . . . . . . . . . 5 2. Automated Systems . . . . . . . . . . . 7 B. The Discrete Stopped-flow System and Kinetic Methods . . . . . . . . . . . . 12 l. The Automated Stopped-Flow system. 0 O O O O O O 0 O O O O O O O O 15 2. Reaction-Rate Method of Analysis. . . . . . . k . . . . . . . . 17 3. Differential Reaction- rate Methods. . . . . . . . . . . . . . 2A C. Application of the Stopped-Flow System in Clinical Analysis. . . . . . . . . . . . 29 1. Experimental Criteria and PPOblemS. o o o I o o o o o o o o o o o 29 2. Analyses Reported . . . . . . . . . . . 32 III. SERUM PROTEIN ANALYSIS. . . . . . . . . . . . 35 A. Introduction. . . . . . . . . . . . . . . . 35 l. The Biuret Reaction . . . . . . . . . . 36 2. Preliminary Observations. . . . . . . . 39 3. Protein Analysis Using Re- action-Rate Method. . . . . . . . . . . A2 B. Stopped-flow Method . . . . . . . . . . . . A2 iv Chapter Page 1. Reagents and Sample Preparation . . . . . . . . . . . . . .A2 2. Automated Stopped-flow System. . . . . . . . . . . . . . . . A3 3. Initial Rate Studies. . . . . . . . . AA A. Experiments with Human Control Lyophilized Sera. . . . . . . A8 5. Comparison with the Equilibrium MGLhOd. o o o o o o o o o o o o o o o 52 6. Choice of Protein Standards . . . . . 5A 7. Studies with a Vidicon Detector . . . 63 Co COHClUSiODS o o o o o o o o o o o o o o o 75 IV. SERUM ALBUMIN ANALYSIS . . . . . . . . . . . 77 A. Introduction. 0 o o o o o o o o o o c O o 77 l. Bromcresol Green Dye- binding Method. . . . . . . . . . . . 78 2. Proposed Solutions. . . . . . . . . . 79 B. StOpped-flow Method . . . . . . . . . . . 80 1. Reagent and Sample Preparation. . . . 80 2. Rate Studies I O O C O C O C C O O O O 81 3. Experiments with Human Control Lyophilized Sera. . . . . . . . . . . 83 C. Conclusions . . . . . . . . . . . . . . . 89 V. SERUM TRIGLYCERIDE ANALYSIS . . . . . . . . . 91 A. Methods of Analysis . . . . . . . . . . . 91 B. Stopped-flow Method . . . . . . . . . . . 95 l. Reagent Preparations. . . . ... . . . 96 2. Rate Studies 0 O O O O O O O O O O O O 96 3. Experiments with Human Serum. . . . . 101 C. Conclusions . . . . . . . . . . . . . . . 103 Chapter Page VI. SERUM CHOLESTEROL ANALYSIS . . . . . . . . . 10A A. Introduction. . . . . . . . . . . . . . . 10A B. The Stopped-flow Method . . . . . . . . . 107 1. 2. Reagent Preparation . . . . . . . . . 107 Rate Studies. . . . . . . . . . . . . 109 C. Conclusions . . . . . . . . . . . . . . . 116 VII. SERUM LIPOPROTEIN ANALYSIS. . . . . . . . . 119 A. Introduction. . . . . . . . . . . . . . . 119 1. 2. Lipoprotein Profile and Hyperlipoproteinemia. . . . . . . . . 119 Methods of Analysis . . . . . . . . . 122 Be StOppEd-flow MGthOd o o o o o o o o o o o 123 l. 2. 3. A. 5. General Scheme. . . . . . . . . . . . 123 Choice of Precipitation Agent . . . . 126 Separation of the Lipoprotein Fractions . . . . . . . . . . . . . . 129 The Ultrafiltration MethOdOOOOOOOOOOOOIOOOOOOOOOOOO O I .132 Experiment with Fresh Human Serum . . . . . . . . . . . . . . . . 136 C. Conclusions . . . . . . . . . . . . . . . 1A0 VIII. FUTURE DIRECTIONS. . . . . . . . . . . . . 1A3 APPENDIX A. APPENDIX B. APPENDIX C. REFERENCES. . . . . . . . . . . . . . . . . . . 1A7 . . . . . . . . . . . . . . . . . . . 176 . . . . . . . . . . . . . . . . . . 193 . . . . . . . . . . . . . . . . . . . 195 vi Table 2-1 2-A 3-1 3-2 3-3 3-5 3-6 3-7 LIST OF TABLES Routinely Performed Clinical Tests. Popular Models of Discrete Chemical Analyzers. . . . . . . . Important Properties of the Stopped-Flow System . . . . . . . Clinical Applications of the Stopped-flow System . . . . . . . Methods for the Determination of Total Serum Protein. . . . . . Results for Total Serum Protein Standards (At - 0.1 s). . . . . . Results for Total Serum Protein Standards (At - 10 s) . . . . . . Results of Determination of Total Human Serum Protein . . . . . . Comparison of the Reaction Rate Method with the Equilibrium Method. . . . . . . . . . . . . . Comparison of Rates of Reaction of Albumin from Different Sources Biuret Reaction with Different Albumin Compositions. . . . . . . vii Page 13 18 33 37 A9 51 53 56 69 61 Table Page 3-8 Biuret Reaction with Albumin from Different Sources. . . . . . . . . . 62 A-l Determination of Human Serum Albumin . . . . . . . . . . . . . . . . . 87 A-2 Serum Albumin Determination with Different Methods . . . . . . . . . . . . 88 5—1 Determination of Serum Tri- glyceride . . . . . . . . . . . . . . . . 100 5-2 Serum Triglyceride Determina- tion with Different Methods . . . . . . . 102 6-1 Cholesterol by Reaction-rate MGthOd. o o o o o o o o o o o o o o o o o 113 6-2 Cholesterol by Equilibrium MGthOd. o a o o o o o o o o o o o o o o o 115 7-1 Physical Properties of Serum Lipoproteins. . . . . . . . . . . . . . . 131 7-2 Comparison of the Centrifugation Method with the Ultrafiltration MethOdo o o o o o o o o o c o o o o o o o 135 7-3 Lipid Analysis on Three Human Sera. . . . . . . . . . . . . . . . . . . 138 7-A Lipoprotein Analysis on 3 Human Sera. . . . . . . . . . . . . . . . . . . 139 A-l Basic Moves of a Simplex. . . . . . . . . 151 viii Table Page Regression Analysis with the Simplex Method. . . . . . . . . . . . . . 165 Weighing by Variable Data Acquisition Rates . . . . . . . . . . . . 172 Effect of Length of Sampling Time on Error . . . . . . . . . . . . . . 17A ix Figure 2-1 3-1: 3-5 3-6 3-7 LIST OF FIGURES The automated stopped—flow system . . . . Classification of reaction-rate method. . . . . . . . . . . . . . . . . . The biuret reaction . . . . . . . . . . . Preliminary observations of the biuret reaction . Initial rate studies of the biuret reaction. (a) At = 1008; (b) At = 103. . Standard curves for two reaction rate methods used in the biuret reaction. . . . . . . . . . . . . . . . Standard curve for serum protein, At = 108. . . . . . . . . . . . . . Standard curve for serum protein using the equilibrium method. . . . . . . Comparison of rate of reaction between BSA and human serum protein . . . . . . . Time scan of the absorption spectra during the human protein-biuret re- action. . . . . . . . . . . . .'. . . . Apparent wavelength shift of the biuret reaction . . . . . . . . . . . . Page 16 22 38 A1 A5 147 50 55 57 6A 65 Figure Page 3—10 Absorbance versus time for the biuret reaction with human serum. (a) 752 nm; (b) 829 nm. . . . . . . . . . 67 2-11 Absorption spectrum of the copper- protein complex from the biuret reaction with human serum after about 25 minutes. . . . . . . . . . . . . 68 3-12 Absorption spectrum of the biuret reagent . . . . . . . . . . . . . . . . . 69 3-13 Rate studies of the biuret reaction. . . . . . . . . . . . . . . . . 70 3-1A Time scan of the bovine albumin- biuret reaction with a vidicon detector. . . . . . . . . . . . . . . . . 72 3-15 Time scan of the bovine globulin- biuret reaction with a vidicon detector. . . . . . . . . . . . . . . . . 73 3-16 Absorption spectrum of the copper- protein complex from the biuret re- action with bovine serum albumin after about 25 minutes. . . . . . . . . . 7A A-l Reaction of BCG with pure bovine albumin . . . . . . . . . . . . . . . . . 82 A-2 Reaction of BCG with serum albumin . . . . . . . . . . . . . . . . . 8A xi Figure Page A-3 Reaction of BCG with serum albumin of abnormally high concentration . . . . . . . . . .'. . . . 85 5-1 Enzymatic reaction of serum triglyceride. . . . . . . . . . . . . . . 97 5-2 Standard curve for enzymatic analysis of serum triglycerides . . . . . 99 6-1 Enzymatic reaction of serum cholesterol . . . . . . . . . . . . . . . 105 6-2 Enzymatic cholesterol reaction, At=lOOs . . . . . . . . . . . . . . . . . 110 6-3 Reaction rate of enzymatic cholesterol vs time (serum cholesterol). . . . . . . . . . . . . . . 111 6-A Reaction rate of enzymatic cholesterol vs. time (pure cholesterol dissolved in an aqueous solvent). . . . . . . . . . . . . 111 6-5 Standard curve for enzymatic cholesterol analysis. . . . . . . . . . . 11A 7-1 Characteristic lipoprotein fractions and electrophoretic patterns. . . . . . . . . . . . . . . . . 121 7-2 Proposed serum lipoprotein analysis procedure . . . . . . . . . . . . . . . . 125 xii Figure Page 7-3 The complete filtration assembly. . . . . 13A 7-A Electrophoretic pattern for serum sample. . . . . . . . . . . . . . . . . . 1A1 A-1 Basic moves of a simplex. . . . . . . . . 150 A-2 Procedure for the modified simplex . . . . .p. . . . . . . . . . . . 153 A-3 Procedure for the super- modified simplex. . . . . . . . . . . . . 15A A-A Safety factor in SMS. . . . . . . . . . . 160 A-5 Boundary violation in SMS . . . . . . . . 160 A-6 Effects of rate differences and noise on error. . . . . . . . . . . . . . 167 A-7 Effect of initial simplex size on efficiency . . . . . . . . . . . . . . 169 xiii I. INTRODUCTION With the advancement of instrumentation and fast analytical methods, the clinical sciences are becoming more and more chemically orientated. Massive screening chemical tests are being performed for individual patients nowadays, as evidenced by the popularity of the Techni- con continuous flow analyzers that can do six, twelve, or even more tests on each sample. Because of the constant demand for higher throughput and 'stat' analysis, together with the increasing use of enzymatic methods, reaction- rate methods of analysis are becoming more important than ever in clinical chemistry. It is not impossible that in the near future enough tests may be performed on each individual so that everyone would have his or her own normal ranges and no longer would it be necessary to com- pare test results with the population's average. The large volume of tests performed will certainly require a much faster throughput rate than the typical 120 samples/hour rate of most current instruments. The stopped-flow mixing system, although used pri- marily as a research tool until now, has the potential of being a very fast and versatile analyzer. It can mea- sure reaction half-lives from milliseconds to hours. Re- quired reagent volumes are as low as 100 pl, and the instru- mentation is easily automated. The only problem that has prevented wider applications of it in clinical chemistry 1 is the lack of research in examining the applications of this fast mixing system. In this thesis, the stopped-flow system was applied to routine analyses. Chapter 2 is a general discussion on some basic concepts about fast kinetics and the characteristics of several types of automated clinical chemistry analyzers. Then the automated stopped-flow system is described. Chapter 3 presents a reaction-rate analysis of the protein biuret reaction that cuts analysis time by more than 100-fold when compared to conventional methods. Chapter A describes an equilibrium method that allows serum albumin to be determined with bromcresol green within 2 8. Most of the interfering reactions are also reduced simultaneously through the use of the rapid equilibrium method. Then we move on to the important area of lipid analysis. Triglyceride analysis is described in Chapter 5, and cholesterol analysis in Chapter 6. These lipid analyses use slow enzymatic reactions so some means of treating the problem of the long analysis times had to be devised in order to utilize stopped-flow mixing ad- vantageously. Furthermore, the problems of formulating suitable aqueous standard solutions were also investi- gated. Chapter 7 concentrates on lipid profiling and analyses of lipoproteins. The analysis of high density lipoprotein has generated increasing interest recently, and its significance as an indicator of coronary heart disease has created a large demand for this test. The technique of getting a total lipid profile using serum triglyceride, cholesterol, and lipoprotein concentrations was investigated. We are able to report a rapid proced- ure that cuts the normal analysis time six to forty fold. Finally, in Appendix A, we report an application of the simplex technique in regression analysis and dif- ferential rate analysis. Although we have not applied the technique to actual clinical tests, it appears to have great potential in differential analysis of complex mixtures, especially those involving enzymes. A well documented super-modified version of the simplex program written in FORTRAN IV language is also included in this section. We conclude that the stopped-flow system has great potential in the field of clinical chemistry. Additional comments on the future trend of research in methodology may be found in the last chapter. II. BACKGROUND The first part of this chapter describes the commonly performed clinical tests and the most used clinical- chemical instruments. Next, a summary of kinetic methods of analysis is presented, followed by a brief descrip- tion of the automated stopped-flow system used in this research and its role in clinical chemistry. A. Clinical Analysis and Instrumentation The clinical field offers many exciting research opportunities for the analytical chemist who enjoys making meaningful measurements. While some may still regard quantitative chemical measurements merely as tools, many physicians and scientists are sharing with the analytical chemist a deep concern for accuracy, precision, and speed. Undoubtedly, the most recent trend is the increasing use of automated instruments to replace manual methods of analysis, and a wide variety of auto- mated analyzers and methods have been reported. But, the emphasis of this dissertation is on automated wet chemistry only, and instrumental techniques such as automated flame spectrometry and chromatography are not discussed. 1. Common Clinical Tests The number of different chemical tests performed in a clinical laboratory has risen from fewer than thirty in 1950 to more than one hundred in the past few years (23). The twelve most popular models of chemical analyzers provide 27 different routine tests that they can perform (See Table 2-1). Many factors determine what tests are to be performed routinely. First, they must serve as indicators of diseases, preferably with a one-to-one correspondence. For example, the use of the enzyme glucose oxidase for serum glucose analysis is a good test for diabetes mellitus. Second, the analysis time (both total work time and technician time) must be as short as possible. Tests such as serum Na+ and K+ using flame photometry require less than one minute and thus are performed routinely as stat tests. Third, the tests should be simple. Excellent reference methods such as the KJeldahl test for protein nitrogen involve complicated distillation and multiple-step reactions and, therefore, most laboratories prefer the simpler, but less precise blood urea nitrogen (BUN) test. Finally, low cost, availability, and sample size requirements are also important factors that govern the choice of routine methods. Recent research and development of new clinical chemical methods make increasing use of enzymes, which Routinely Performed Clinical Tests. Table 2-1. comm QmOmQHu Oauv uaumabuaa .ama Icmnuwa uuoamcmuu menu Hmuaoufiuon Amaze 0v mumawwuaou wHaBmm oau Immaa manmmoawap .Hmwucmsamm com 4mm a 330mm .0 m an cowlmN maoauumam oo ummu .aaam. wow voow .mummu uamumm luau om waucwsumm .vuwo Baum noun m nuwz nmaaaaam mum zuaga wum>uao manmmoamfiv vmmmxumamum BHMOwGwHU .n mummy uaoummmfiv pom nonmamn< m a: oqlma mamaaomam oo mmasvoa woumofivmc .Hmaamuma HwUHamao ca muwmu ucwummmav «NIH umuaaoo .q MXHH Admzn awe om .mdowuummu mnsu umwu .waouuan swan kn hamaomcauasaam m.mH.m Ha H.o no vmauomumm .mummu uaoummmwv xxumz mamawommm ooq OH on a: mo nowuumamm Hoax: .m mummu om you mamafiumam «Homcou ma0fiuowmu cont: Ha.» anoos com ca HmHHmumm usages» .N NH.HH.w some anoomlom mumou 00H Hmwucmavmm .mxumm uawwwmm unnswaouna .muouhama< kuaaono wumuuman mo mauve: unannom .Nim manna 1A make the stopped-flow method an attractive tool. First, it is a highly versatile system, for not only can it provide analytical data on reactions that occur in a matter of a few milliseconds, but it is also suitable for studying slow reactions. For example, by using kine- tic measurements one can use it to study the initial stage of reactions. ’By waiting for the reaction to go to completion, one can also use it to make equilibrium measurements. Therefore, unlike the continuous-flow systems that have a limited range of reaction half-lives the stopped-flow analyzer features rapid throughput of information and the study of reactions over a wide range of half-lives. Secondly, the system can be fully auto- mated. Using minicomputers or microprocessors to control the processes of solution preparation, data acquisition and data manipulation speeds data throughput, improves accuracy and increases precision. And finally, the small sample volume required (a few microliters) makes it attrac- tive for routine clinical measurements where sample size is often severely limited. This section describes briefly the stopped-flow system in our laboratory and some fundamentals of kinetic analysis. 15 l. The Automated Stopped-flow System Beckwith (A) first built the pneumatic driven stopped- flow system, and Notz (5) interfaced it with a PDP 8/e minicomputer. Figure 2-1 shows a diagram of the automated stopped- flow system. This system can be operated both manually and automatically. Operation begins by turning on the delivery valve (A), which connects the reagent and sample reservoirs to the delivery syringes, while closing off the remainder of the flow system. The syringes then draw in reagent and sample solutions, and the delivery valve then turns to connect the syringes to the mixing chamber and the observation cell. Air pressure is applied to the drive syringe piston, when the waste valve (B) opens and con- nects the flow system to the stopping syringe. The syringe drive piston rapidly strokes forward and forces the solutions through the mixing chamber, observation cell, and stopping syringe. The flow stops when the spring- 1oaded stopping syringe comes to rest against a mechani- cal stop. At the same time, an optical interrupter sends out a pulse which signals the end of flow and triggers the computer to begin gathering data. When the measure- ment is finished, the waste valve opens to the drain and l6 ‘ DRIVE AUTO ‘JR‘ snnmn , , q . MANUAL PDP We 7 . COMPUTER S 11 _ DIGITAL “P 08 , ‘-. ; Syringes , l 5 READOUI ! 3 DEVICE ' i Sam 1e ‘ * §Rea ent p ! MIXER 3 g l l f , e 9 = I near ‘ ossanvarrow 3 mm SOURCE 1 MONOCHROMAIOR CELL '—:,—’ PHI MANIPULA‘IION I l ' i l 3 h-—-.-- Thermostated W t to u e Stopping Syringe -— 7,...— To Trigger .\ OPTICAL INTERUPIER Figure 2-1. The Automated Stopped-Flow System. 17 the stopping syringe forces the old solution out the drain port (A). The valves then reset and run the next sample. Interactive FORTRAN programs are available to analyze the data. Results are stored on magnetic media, or the printer types out a copy of the results. Circulating water acts as the system's thermostat, keeping the tempera- ture of the observation cell slightly above that of the drive syringes and the mixing chamber, so that it will match the temperature of the solution after mixing (6). The design of the mixer and the flow rate ensure turbulent flow and efficient mixing. The rapid stopping of the solution completes the mixing and provides back pressure which prevents cavitation (5). Table 2-3 lists some other important information about the system. The dead time of the system (5:1 ms) dictates the practical limit on the fastest reaction which can be observed. However, for most clinically important reactions, the dead time imposes no serious problem. 2. Reaction-Rate Method of Analysis With the advancement of electronics and instrumen- tation, reaction-rate methods, previously difficult to do experimentally, are now becoming routine. Several books (36-38), review articles (39-A1), and papers (A2-AA) on the theory and application of these methods have been 18 Table 2-3. Important Properties of the Stopped-Flow System. Dead Time 511 ms Mixing Time N5 ms Stopping Time 3’1 ms Volume of Sampling Syringe 2.35 ml Volume of Each Push (6 pushes/fill) 0.39 ml Volume from Mixer Entrance to Exit of Observation Cell 0.16 ml Pathlength of Cell 1.9A 0.01 cm 19 published. The biggest advantage of the reaction-rate method compared to equilibrium methods is its greatly reduced experimental time. It is possible to measure the rate of change of the concentration of reactants or products immediately after mixing without waiting for the reaction to go to equilibrium. Since the measurement can be com- pleted before significant side reactions occur, and be- cause only relative rate measurements are made, many of the interfering factors are thus eliminated. Stopped- flow analyses often employ reaction-rate methods. By simply adjusting the experimental conditions, one can obtain first or pseudo-first order reaction conditions. For example, consider a simple reaction: kA A+R+P where kA is the rate constant, A is the substrate, R the reagent in excess, and P the product. The rate of dis- appearance of A can be expressed as d[A _ 71% " 'kAEA] (1) If Equation (1) is integrated from time t = 0 to time t = t, we have [A] = [AJOexp(-kAt) (2) 20 If Equation (2) is substituted into (1), 9§%J-= -kA[AlOexp(-kAt) (3) when t = 0, exp(-kAt) = 1 therefore, ‘3?’ = -kA[A30 <“) Equation (2) suggests that the initial concentration of the reagent [AJO can be obtained by measuring the concentration of A at any time t. [AJO can also be ob- tained by measuring the rate of change of A at time t, [Equation (3)]. But Equation (A) is the basis of the most commonly used rate method. By measuring the reac- tion rate at the very beginning so that the time is ap- proximately equal to zero, the exponential term approaches unity and thus the reaction becomes pseudo-zero order in time. This is known as the initial rate method, and requires that the measurement be made while the reaction curve (the exponential concentration Xi- time curve) is still linear. For 1% accuracy in the initial rate measurement, the measurement should be completed before the relative concentration changes by 5% for the fixed time method (31), and by 2% for the variable time 21 method (32). In the last few years, more clinical chemists have been using rate methods mainly due to the extensive use of enzymes (33). At the same time however, this wide- spread use of rate methods has also brought about am- biguity in the terminology. Pardue has suggested a comprehensive classification which has helped to clarify the terms somewhat (3A). Equilibrium methods were defined as those methods based upon reactions that go to completion, and rate methods as those based on measurements made while the reaction is approaching completion. The commonly used direct response rate methods can be classified as in Figure 2-2. The simplest is the one-point method wherein the signal is measured at some predetermined time after the reaction has been initiated. This method does not provide any information on the blank or the shape of the response curve. For the two-point method, there are two different approaches. The fixed- time approach measures the total signal change over a predetermined fixed time interval, while the variable- time approach measures the total time required for the signal to change a predetermined amount and relates it to the analyte concentration. The two-point method has an advantage over the one-point method in that it com- pensates for sample and reagent blank, but errors due to 22 fl all A Figure 2-2. "A 10 One-Point Method Simple. no information on blank or shape of curve Two-Point.Method Fixed-time or variable- time, simple, no information on curve shape Maggi-point Method A) Delta Method Suitable for 5-10 data points B) Regression Method Suitable for large number of data points Classification of reaction-rate method. 23 non-linearity of reaction curves caused by induction periods or substrate depletion still occur. When three or more data points are collected either in the fixed-time or the variable-time modes, it is de- fined as the multiple-point method. At least two dif- ferent approaches within this class can be identified. The "delta method" takes multiple absorbance data at fixed time intervals and computes values of A called "deltas". Delta values of the same magnitude are used to compute the substrate concentration. In the "regres- sion method", one collects many data points and then uses a mathematical algorithm, such as the linear least-squares method, to obtain the slopes from the linear region. An advantage of the regression method is its applicability to both linear and nonlinear responses. The first and second derivatives of A gs. t help to locate the linear region of the response curve, and overlapping groups of data will give many values of the slopes for averaging. It appears that with the help of computerized data acquisi- tion and analysis, the collection of large amounts of data is no longer a laborious task, and the regression multiple- point method, properly implemented with statistical analysis, is the preferred method. Reaction-rate methods can be very rapid and show a high degree of freedom from interferences, which are im- portant advantages in clinical analysis. The method is 2A also more specific than the equilibrium method so that separation steps can often be avoided. Differential analysis of complex mixtures can also be performed by taking advantage of the different reaction rates of two or more reactants with a common reagent (A9-52). But there are some rather complex problems in using reaction-rate methods. Rate methods usually require carefully controlled experimental conditions, such as temperature, pH, ionic strength, size and shape of the reaction vessels, etc. Minor changes in these condi- tions can change the reaction rate and thus the result. The half-life of the reaction must be greater than the mixing time of the instrument, and the precision also decreases with shorter observation time (A2). 0n the other hand, reactions with half-lives longer than a few hours are also not practical for routine analysis. For selected procedure, however, the overall advantages of rate methods can outweigh the disadvantages. 3. Differential Reaction-rate Methods Rarely does one find the sample for analysis in pure form, devoid of any other substances that might interfere in the determination of the sought after substance. So one must find some means to eliminate the interferences. Separation of the target sample component from interfering 25 substances is often too time consuming, or even impracti- cal. In some cases, by judiciously selecting the experi- mental conditions, one can alter the reaction rates of different components within the sample so that they are sufficiently different from each other to allow one or more of the closely related components in the mixture to be analysed without prior separation (AS-A8). Mark §t_a1. discussed the various methods for analysing mixtures whose reactants have large rate differences (A5), and Papa (36) and Mottola (A0) have evaluated the more commonly employed methods. In clinical analyses, we are mostly interested in first and pseudo-first order reactions with respect to total reactant concentration, or enzymatic activities. Several differential rate methods exist for the study of these reactions. The graphical extrapolation method (52) and the method of Roberts and Regan (5A) are very useful, but the method of proportional equation is of particular interest here (53). Generally this method requires less time than the other approaches, while it does not require prior knowledge of the total initial concentration of reactants. Most important of all, it is a mathematical method which is therefore easily adaptable for automa- tion (A0). This method is based on the principle of constant fractional life, which is applicable to most simple first 26 and pseudo-first order reactions. When a substance reacts with constant fractional life, its initial con- centration is directly proportional to that of the product at any given time. Consider the following first or pseudo- first order reaction: A + R -+ P (5) The concentration of the product at any given time t is [PJt = kA(l — e‘kAt)[A]O (6) or [Plt = GAEAJO (7) where GA = kA(1 - e'kAt) (8) The proportionality constant GA is dependent on the rate constant kA and the time t. As long as the factors that affect kA (such as temperature, pH, and ionic strength) do not change and the time t is fixed, GA remains constant. In Equation (7), the concentration of the product at any given time is therefore proportional 27 to the initial concentration of A. Very often an elec- trical signal St is observed instead of Pt' This signal is proportional to Pt by a factor n, and Equation (7) becomes m l or St = KAEAJO (10) and KA becomes the new proportionality constant. If there is a binary mixture (A + B) and the two different components react by first order kinetics to give the same product P, the concentration of the product at any time tl can be given by [Pjtl = GAlEAJO + GB1EBJO (11) where GB is the proportionality constant for substance B. At a different time t2, [P3t2 = GA2EAJO I GsafBjo (12) Thus, to determine the proportionality constants GAl and GBl experimentally, one can measure the concentration 28 of the product at time t1 for known amounts of pure A and pure B under the same experimental conditions as that of the mixture. [Pltl = GAllAlo (13> [P]t1 = GB1EBJO (1“) To find GA2 and GB2’ keep the experimental conditions constant and change the observation time to t2. Substi- tution of the four constants back into Equation(ll)znui (12)will allow the concentrations [A10 and [B]O to be obtained on an unknown sample. The method is applicable to mixtures that contain more than two substances (38). With the help of a computerized on-line system, 100 or more data points can be obtained during the course of reaction. One can set up 100 or more proportional equations and solve for the unknown concentrations [A30 and [B]O. Appendix A of this thesis describes the use of multiple proportional equations in some first and pseudo- first order reactions found in clinical analyses. 29 C. Application of the Stopped-flow System in Clinical Analysis Traditionally, only the few scientists who study the kinetics of rapid reactions use stopped-flow mixing systems. It was not until 1969 that Pausch and Margerum (A9), and Javier _p‘_1. (6A), first reported its use for analytical purposes. And it was only after Pardue gp'gl. had evaluated their discrete sampler stopped-flow mixer system in 1977 (1) that its use in clinical analysis ap- peared in a clinical journal. Two reasons account for this. First, instrument companies have been slow to realize the potential of this system as an analytical instrument. Therefore, very few commercial instruments are available. Also, the stopped-flow systems that are on the market are so inconvenient to use, they are in no way comparable to the "home-made" versions. Second, not enough clinical methods have been developed for recognition of the advantages of such a system and hence its adoption as a routine analytical tool. 1. Experimental Criteria and Problems An ideal clinical chemistry analyzer should be in- expensive, simple to operate, have a high sample through- put rate, and if possible, be portable. It should be easily automated, with no carry-over problem. It should 30 give precise and accurate results, and be capable of doing a wide variety of tests. The sample volume for analysis should be as small as possible, and finally, the reagents inexpensive. The advantages of the stopped-flow system in monitor- ing fast reactions, as discussed earlier, include high throughput rate, high precision and accuracy, easy auto- mation, good mixing and a large reaction time range. In addition, it is also suitable for equilibrium (end- point) analyses. In fact, Pardue 33 a1. (1) have just reported that attaching a sample loop to the system faci- litates further measurements for equilibrium methods. Most of the instruments currently used for kinetic analyses in clinical laboratories are designed for relatively slow reactions, usually requiring anywhere from a few seconds to a few minutes for a complete measure- ment. Thus they may not be adequate for such tasks as the.LHD isoenzymes determination by Everse _p{_1. (59), or the biuret protein reaction which is reported later. For the former, the first set of measurements must be taken within 0.2 s,while the measurement time for the latter is about 0.1 s. Pardue has already suggested that instruments with mixing and measurement capabilities that require approximately 0.01 5 would be highly applicable in clinical chemistry (1). To further reduce the sample size when using stopped- 31 flow systems, one can use new cells and mixer designs, or add zero dead-volume valves and fittings to the system. And, to cut reagent costs drastically, especially those needed for enzymatic reactions, one can employ reusable immobilized enzyme reaction loops such as those developed in our laboratory (2). The tremendous improvements in detection systems in recent years allow measurements with high sensitivity. Dual-channel and multichannel systems, such as the rapid scanning monochromator and vidicon have already been useful for fundamental studies of bio- chemical systems (67). For turbid samples, other non- optical detectors based on thermal or electrical prin- ciples can be used (68,67). Since no system is perfect, there still remain some problems in using the stopped-flow instrument described here. One of the major criticisms made of this system is the lack of sample preparation and initial separation, like that found in the autoanalyzer. But with the pos- sibility of adapting one of the recently reported auto- matic solution makers onto the stopped-flow analyzer, one of the most time consuming steps in the process would no longer be a problem (1,29). A second problem with using this system is that of carry-over, which exists because at present the observa- tion cell is not disposable. This makes repeated rinsings between samples necessary. It seems, however, that by 32 incorporating Pardue's proposed sample loops, this prob- lem too can be eliminated (l). A third disadvantage of this system is its inability to perform multiple tests in parallel due to its dis- crete mixing characteristics. One must remember, however, that a stopped-flow system is not designed for this, but is rather a 'stat' analyzer. And finally, sample and reagent are mixed in a 1:1 ratio in the stopped-flow system. Therefore dilutions must be made to obtain favorable experimental conditions. However, in some cases excessive dilution causes undesir- able problems, such as turbidity or slow reaction rate. Despite the above limitations, the stopped-flow system can be made applicable to many kinds of wet chemical analyses done in clinical laboratories, and as is shown throughout this dissertation, it has uniquely advantageous features when compared with other existing systems. 2. Analyses Reported Applications of the stopped-flow system in clinical analysis to date have been few, and they are all listed in Table 2-A. Most of these involve a reaction-rate method with adjustment of the experimental conditions so that the reactions occur under pseudo-first-order conditions. Enzymatic reactions are especially suitable 33 mm.:o OHpocHM COHpsHom oHoHo< CH «mzlmH mecoo dgmm mmzwwzmomH mag .HH H OHmemNcm .oaoooam +ocoo mzHHHpommmom o>Hpompmom .m H OHQEmm mo o>Hp E: 0mm pm mnpom tongumoolcoc .momLSOom Inm ochopzu mo mwch oHmeopo 5 oofiefl use Boom .ea mam to moaoaoo meson ooaoooa 53383 >2 .m m maowmsn oEom an mopomaoch .o.m can» oaaaooam whoa AHo\Ew H.0Imo.ov .pogdHn corp o>HuHmcom ucowmmp .o.m on» 29H; opoe moEHp OOHuom pcommon poasHp map ocHnEoo mazoq .o .m .2 Hommav m HmEpocnm mH Espom mH moonoEoo oSHp o>Hw ou COHpsHom tonne .OHmHooam ocHmeHm :H opmoanoEOpmeSp pom Ho\we m.HIm.o locomono anz cHopopo Eamon on» sopHMoOHo mpH>HpHmcom @000 no wosopw Hocond map mo COHpowom ICHHom .m AozmHpHmcom 30A meson ooHpooo :Hopono mo COHpomom poLSHm .m w cospoe ppmocmpm wHoEwm mo can no choEEoo pom: pcopcoo cowoppHc HopOp monummoz HnmoHonx .H .mom pamesoo COHpomom compo: .cHopopm Esaom Hmpoe no COHpmcHEmmpoQ on» non moonpoz .Hlm oHnme 38 7"7 A? :9 HzN-EC-Nflg-C-NHZ + eu'H” alkali”, Cu-Biuret Complex I u--4 Bi ret U Blue Amax(5AO nm) r"l? HZW'? :9-0H ++ alkaline Rl-CAC-NHPC-H + Cu -——————%~Cu-Biuret Complex L____:R2 Blue/Amax(5A0 nm) Protein Figure 3-1. The biuret reaction. 39 which gives the reaction its name. The biuret reaction only takes place between the cupric ion and compounds which have at least two NH2CO- groups, two NH2CH2- groups, or two NHZCS- groups. These groups must be directly joined to another similar group, or joined via a carbon or nitrogen atom. Therefore any dipeptides or amino acids present in the serum will not react and will not interfere with the protein analysis. The actual mechanism of the reaction is not known, but it is hypo- 2+ ion is coordinated to four to six thesized that each Cu nearby peptide linkages (8,72,85). The intensity of the color produced is proportional to the number of peptide bonds present, which is in turn proportional to the concen- tration of the protein. 2. Preliminary Observations The biuret method introduced by Kingsley in 1939 (72) is well known, and many modifications of the method have been published (73-75). In fact, quite often it is taught to students in clinical biochemistry courses. The standard procedure involves mixing blood serum or a protein standard solution with the biuret reagents. The color is allowed to develop for 30 minutes, and the absorbance at equilibrium is then measured not later than 60 minutes after mixing. No data points are taken during the 30 minutes waiting period. Because of the no long waiting period in the usual equilibrium biuret procedure, a preliminary investigation of the kinetics of the reaction was undertaken in order to determine if a reaction-rate method might be applicable. The standard biuret procedure was repeated, with absorbance values taken as a function of time during the 30 minute period before equilibrium. Typical absorbance-time curves are shown in Figure 3-2a. From this preliminary experiment two important observations were made. The first observation was that the reaction did not follow first or pseudo-first order kinetics, so that the log absorbance XE: time plot was a curve instead of a straight line, as shown in Figure 3-2b. The second one was even more interesting. Figure 3-2c shows that straight lines were obtained when the absorbance of each standard was divided by the first standard and plotted against time. This suggests that one may obtain an analytical curve long before the reaction reaches equilibrium. Thus, the 30 minute wait may be eliminated by using a reaction-rate method. Since the earliest time observable using manual mixing and a conventional spectrophotometer was 2 minutes, it was decided to study this reaction further with a stopped-flow mixing system. A1 4 .40— .30— 3 g .20- 4 TlME(min) (a) 01 O) I \ ABS RAND LOG(ABS no -A) 9 l \o 53 l l? l l l l 1 4L_ 5 '0 ngefigj 25 TIME(min) (b) (C) Figure 3-2. Preliminary observations of the biuret reaction. A2 3. Protein Analysis Using Reaction-Rate Method A survey of the literature showed that Hatcher and Anderson (83), then Koch, Johnson, and Chilcote (7A) have all reported various reaction-rate methods for determining total serum protein with the centrifugal analyzer. Ac- cording to Koch g3 a1., their method was reasonably pre- cise and rapid (90 s per run). But Savory 33,11. (75) reported that the method was relatively imprecise in their hands and was subject to errors in the analysis of sera from patients with dysproteinemias. Thus, we decided to try another kinetic approach, namely the stopped-flow mixing method, hoping to develop a more reliable, pre- cise and rapid procedure. B. Stopped-flow Method 1. Reagents and Sample Preparation The biuret reagent, which contains copper sulphate and tartrate in alkaline solution, was prepared according to a modified formulation of Cornell 32 a1. (8A). Biuret reagent. The working reagent contained cupric sulfate (2.A g L-l), sodium potassium tartrate (8 g L'l), _1) sodium hydroxide (A8 g L , and potassium iodide (l g L'l) Blank reagent. The blank contained sodium potassium tartrate (8 g L'l), sodium hydroxide (A8 g L'l), and A3 potassium iodide (l g L'l). Standards. Standards were dilutions of human serum protein standards (Dade Division, American Hospital Sup- ply Corp., Miami, FL 33152; Cat. No. B5158) in saline solution (NaCl, 9 g L-l). Quality Control Sera. Lyophilized control serum samples (Moni-trol I, Cat. No. B5103; Moni-trol II, Cat. No. B5113, Dade Division, American Hospital Supply Corp., Miami, FL 33152), were reconstituted according to the manufacturer's instructions and diluted 30 fold with saline solution before use. All reagents were filtered through a 5.0 pm dispos- able filter (Millipore Corp., Bedford, Mass., 01730) before use. The biuret reagent was stored in plastic bottles in the dark, and all protein solutions were freshly prepared from vacuum sealed prepackaged bottles stored at 2-8°C. 2. Automated Stopped-flow System Reaction-rate measurements were made on the computer controlled stopped-flow speetrophotometrie system des- cribed in Chapter 2. All data were acquired and processed by a PDP 8/e minicomputer (Digital Equipment Corp., May- nard, MA 0175A). Iteractive computer programs are written by Notz (PNSF3. PNF103)-and by Gall (RGMAIN) in our group. Notz's programs control the sequencing AA of the stopped-flow mixing system and the filling and rinsing of the drive syringes. They also perform the tasks of data acquisition and processing. Gall's program analyzes the data collected. It displays absorbance and rate on the CRT as a function of time. The slope of the absorbance-time curve as well as the maximum rate can also be calculated and displayed. The rates are obtained by 9 point first derivative, quadratic, Savitsky and Golay smooth on the absorbance- versus-time data. If the reaction is not pseudo-first order, the maximum rate is obtained. If the reaction is pseudo-first order, the initial portion of the rate curve is linear and the operator can identify the linear portion and obtain the average initial rate with the program. The stopped-flow system was thermostated at 25°C. Spectro- photometric measurements were made at 5A0 nm. 3. Initial Rate Studies Figures 3-3a-c show the absorbance-time curves for the formation of the biuret product from human serum protein standards. It is apparent that the biuret- protein reaction proceeds quite rapidly. The total ob- servation times for the 3 runs shown in Figure 3-3 were 1003 (Figure 3-3a), 108 (Figure 3-3b), and 0.18 (Figure 3-3c). As can be seen from Figure 3-3c and 3-3d, the A5 06-0 r- w u z d r— A” m <1 0 . J l 0 L00 TIME (sec) 1 IO" (a ) 0.20 - ms — '9’ q E o 0 “°” In (D <1 ombi:_f"".~ o o: s o 1 J A A I O LOO TIME (sec) I IO (0) (a) At = 1008; 040 v- 030 — assonsawce S O I OIOP 0.50 RATE (b) At A l l L TIME (sec) I I0" (b) LOO TIME (sec) xIO (d) Figure 3-3. Initial rate studies of the biuret reaction. = 2103 . A6 reaction rate reaches a maximum before 0.1 s. No other instrument could be used for studying the biuret reaction in so short a time scale. In fact, the desir- able feature of obtaining A0 for the sample-reagent mix- ture without running a separate blank, can be done by the stopped-flow system. Although Koch 33 £1. (7A) did the experiment under slightly different conditions (on a CentrifiChem analyzer), they took their first data point at 33 after the reaction began, and called that their blank. But as can be seen in Figure 3-3b, the reaction has proceeded to a substantial extent after 3 s. Fixed-time reaction-rate analysis of compounds is usually limited to those substances undergoing first- order or a pseudo-first order reaction (32,82). The biuret reaction, although not a simple first-order reaction, is suitable for fixed-time reaction-rate analysis, as shown in Figure 3-A. Here, a standard curve for human serum pro- tein using a two point fixed time method is shown. For the initial quantitative studies, we used two approaches: the two-point method, and the maximum rate method. In the two-point method we took data at time t0 and at 0.15, and used the difference in absorbance, AA, in the standard curve. Then we also measured the maximum rate and used it the with the other approach. Figure 3-A presents the resulting standard curves. As can be seen, the two-point method yields more precise results. A7 .050 LC O40 — i -* 0.8 .030 - — 0.6 MAXIMUM (I) RATE <1 .0I0 — 0.2 l l O 4 8 PROTEIN CONCENTRATION (G/ DL) Figure 3-A. Standard curves for two reaction—rate methods used in the biuret reaction. MAXIMUM RATE A8 The statistics of the calibration curve are summarized in Table.3-2. Note that the precision is only fair. For the three different series of standard solutions, the relative standard deviation was as high as 5.8%. The maximum rate method was even less precise, as can be seen in Figure 3-3d. The maximum rate does not last for a very long time period, and the relatively high noise level in the derivative curve causes imprecision. Since the biuret reaction is rapid, a longer measure- ment time, At, was used in subsequent trials. Longer At values give greater absorbance changes and thus better precision due to less instrumental noise and a smaller error in AA. A measurement time of 105 was chosen as a compromise between precision and analysis time. It was found that the precision improved greatly with a 10 8 measurement interval. Figure 3-5 shows that the standard curve is linear up to 8 g/dl serum protein concentration. The statistics of the plot are shown in Table 3-3. Note that the maximum relative standard deviation obtained was l.A%, which is equal or better than any equilibrium method. A. Experiments with Human Control Lyophilized Sera To test our method with actual serum samples, three 1yophilized sera (Moni—trol I, Patho-trol, and Moni-trol A9 Table 3-2. Results for Total Serum Protein Standards. (t - 0.1 sec) Protein Concentration Absorbance Relative Std. (gm/dL) (Average) Deviation % 8 0.0A5A 1.A 7 0.0381 2.6 6 0.0365 5.8 5 0.0329 5.1 A 0.02A6 2.2 3 0.0202 1.6 Slope = A.86 x 10-3 Intercept 6.22 x 10"3 A 1.57 x 10’3 Std. Error of Estimate 50 Me 0.30 ABSORBANCE o I“ o O.IO () l l l 1 J O LOO PROTEIN CONC.( G/DL) xIO'| Figure 3-5. Standard curve for serum protein, At = 10 s. 51 Table 3-3. Results for Total Serum Protein Standards. (t = 10 sec). Protein Concentration Absorbance Relative Std. (gm/dL) (Average) Deviation % .393 .3AA .308 ~235 .213 00000 Ol—‘OOH 0.151 NW-fimmflm Jr-lrl-‘WF-‘Nw I4 H 0.101 Slope = A.85 x 10'2 6.96 x 10‘3 A Intercept Std. Error of Estimate 9.57 x 10-3 52 II, DADE, Division American Hospital Supply 00., Miami, FL 33152) were used as controls. The manufacturer of these sera gives assigned values of total protein obtained by replicate analyses using well-established methods. A comparison of the results obtained here by the fixed- time rate method with those obtained by other methods is presented in Table 3-A. Note that the other reported rate method using the CentrifiChem analyzer gives rela- tively low results when compared to results from equilib- rium methods. The precision is also poorer, and is es- pecially noticeable with the Moni-trol I serum. As reported previously, the 0.1 s fixed-time method gave relatively poor precision, but we found the values still fell within the acceptable range provided by the manu- facturer. The 10 s fixed-time method gave good accuracy and precision. An abnormal serum (Patho-trol) was also analyzed using the 10 s fixed-time method. As shown in Table 3-A, the results we got were slightly higher than the value reported by the manufacturers, but still within the allowable range. 5. Comparison with the Equilibrium Method To further confirm the reliability of the results, the same control sera were analyzed using an equilibrium biuret method. A Beckman DB-GT spectrophotometer with a Table 3-A. Protein. 53 Results of Determination of Total Human Serum Moni-trol I l. 2. 3. A. Improved biuret CentrifiChem SMAC Mark X Moni-trol II 1. Improved biuret 2. CentrifiChem 3. SMAC A. Mark X Patho-trol 1. Improved biuret 2. Technicon AA II Repprted Values Mean (s/dL) 7.0:.06 6.5:.35 6.8:.10 7.3:.10 5.31.07 A.51.05 5.2i.08 5.61.10 5.3:.12 5.9i.05 Range 6.7-7.3 5.8-7.2 6.5-7.1 7.0-7.6 5.0-5.6 A.2-A.8 A.9-5.5 5.3-5.9 5.0-5.6 5.6-6.2 Observed Values At=0.ls At=lOs 6.791.15 6.79:.01 5035:085 5.1“1.0u 5.85:.08 5A 1 cm path length cuvet was used. The reagents and samples were mixed, and the color was allowed to develop for 30 minutes before the absorbances at 5A0 nm were measured. Dilutions were necessary for the more concentrated samples *to produce.a linear standard curve, which is shown in Figure 3-6. The results obtained were compared with the results from the rate method and are listed in Table 3-5. The equilibrium method gave practically the same results as the rate method. The differences found for the Moni-trol I and Moni-trol II samples were less than 1%, while the Patho-trol sample had about 7% difference. 6. Choice of Protein Standards Human serum protein was used as the standard in the stopped-flow biuret procedure. Bovine serum albumin is often the preferred standard material in equilibrium methods because it is much cheaper and readily available in pure crystalline form (81). However, Koch gp_gl. (7A) have reported that it reacts faster than human serum protein with the biuret reagent. We repeated their experi- ment with the stopped-flow analyzer and observed the same result (Figure 3-7). However, careful analysis of the data revealed one otherinsight. It was found that the rate of reaction of the human serum protein was about 0.65 times as rapid as the BSA. The human serum protein standard used contains 5‘g albumin and 3 g globulin per 55 (180- lu<160- Q) 2! <1 00 ' O: O 8 <1 m0'1 Figure 4-1. Reaction of BCG with pure bovine albumin. 83 2 s (Figure 4-2). The second slower phase of the reaction continued for about 30 min. Finally, the reaction of a serum whose albumin con- centration was beyond the linear range was also observed, and Figure “-3 shows the results of this experiment. From this figure it can be seen that the reaction approached equilibrium very rapidly, and there is also a rather evident breaking point, which demarks a slower reaction about 6 s after mixing. These rate studies indicate that there is indeed a second slower reaction in the overall serum-BCG reaction. This means that measurements must be made early in time if error is to be avoided. Prior to 1976, most procedures employed a reaction time of 5 to 30 min. Webster later tried to measure the "immediate reaction" within 30 s, but found that there was still a small positive differ- ence between his results and those obtained from electro- phoresis (100). Our results indicate that for the work- ing concentration range of serum albumin (1.0 g/dl-6.0 g/dl) an observation time of 2-5 s after mixing should be used to minimize interference from non-albumin materials. 3. Experiments with Human Control_Lyophilized Sera A standard curve was prepared with diluted human albumin standard solutions (7.9 g/dl, Sigma, St. Louis, MO). At first we prepared the ECG reagent as before, 0.5001 0.000 Figure “-2. ABSOR BANCE 'lUllLlllLUlilJlLLllllllLllLl_LlllIlLlJllLllllll 8A IIIIIIII'TFIIIIIIrlII‘IIIIIII'IIloTTFIjIIIIIIIIIIIIIIIIIWIFT‘1 O lllllllll 0.00 .000 0T1ME<_s-31Ec> X10 Reaction of 800 with serum albumin. 85 1.200 oooooooooooooooooooooooooooo ABSORBANCE 0.000 0.000 3.000 TIME(SEC) x10 '1 Figure “-3. Reaction of BCG with serum albumin of ab- normally high concentration. 86 but the Brij-35 caused minor mixing problems in the stopped- flow system. Thus, we decided to dilute all the solutions by a factor of two with saline solution. The sensitivity was un- affected and a linear standard curve up to an albumin concentration of 5.9 g/dl. was obtained. Table “-1 gives the relative standard deviation of the data and the standard error of estimate of the curve. We then followed the manufacturer's instructions and prepared 1:200 dilutions of three 1yophilized sera: Moni-trol I, Moni-trol II, and Patho-trol from Dade. A fixed-time equilibrium method was used for the determina- tions with a 2 8 measurement time. The measurement time of 2 s was chosen because our rate studies had shown that the BCG albumin reaction for the most dilute solution is completed in less than 2 s after mixing, while no sig- nificant amount of interference began until 6 s after mixing. The results of our analyses are compared with the values provided by the manufacturer in Table h-2, and it can be seen that our estimates of albumin concentrations within the normal range (3.5-5.9 g/dl) are lower than those obtained by other methods. For Moni-trol I, our results were 0.29 g/dl lower, and for the less concentrated Monitrol II, the difference was even greater (=0.52 g/dl). This is consistent with the fact that a decrease in serum albumin concentration is often accompanied by an increase 87 Table 9-1. Determination of Human Serum Albumin. Concentration A (g/loo ml) (At=2 sec) RSD 1.38 0.118 3.0% 2.06 ' o.1u7 2.0% 3.16 0.197 1.9% 9.10 0.230 2.9% 9.51 0.255 2.0% 5.90 0.312 1.2% Slope = 0.0928 Intercept =0.059 Std. Error of Est. = 3.78 x 10'3 88 Table 9-2. Serum Albumin Determination with Different Methods. Stopped-Flow Method Other Methodsa Mean RSD 1 2 3 9 5 Moni-trol I 9.11b 3.0% 9.9 9.7 9.9 9.9 --- Moni-trol II 2.58 1.3% 3.0 3.1 3.3 3.1 3.2 Pathotrol 9.19 2.7% --- --- 9.7 --- 9.1 a1 = DUPONT ACA 2 = CentrifiChem 3 = Technicon SMAC, or AAII 9 = Hycel Mark X 5 = Salt Fractionation, biuret method O‘ > H H concentrations expressed in mg/100 ml. 89 in globulin concentration, which further increases the magnitude of the slower reaction. In fact, all our results were lower than those from other automated BCG methods, which seems to indicate that our method is less susceptible to overestimations of albumin concentrations. Webster (98) reported that his results were consistently higher than those from electro- phoresis by 0.3 g/dl. Ferreria and Price said their analysis using bromcresol green gave results that were on the order of 0.5 g/dl higher than those obtained through immunoprecipitation (99). Thus based on these, one can safely assume that the manufacturer's reported values should be about 0.3-0.5 g/dL higher than the actual value, especially when their data are gathered from dis- crete and continuous-flow systems, some of which were measured 30 min after mixing. C. Conclusions We took advantage of the fast mixing feature of the stopped-flow system and were able to find conditions for determining serum albumin concentrations with bromcresol green, with the least amount of overestimation. Our rate studies suggest that a fixed-time equilibrium method with a measurement time of 2 3 gives the most precise results with least interference from other proteins in the serum. Studies with 1yophilized control sera show 90 that our method consistently gives lower values than the other automated techniques which use BCG. Electrophoresis experiments should now be done on the same samples whenever the necessary equipment is available. This would further validate our results from the stopped-flow method. V. SERUM TRIGLYCERIDE ANALYSIS It is known that serum triglyceride levels increase in patients with atherosclerosis (119) and diabetes meilitus (115). Thus the assay of serum triglyceride is an important clinical test for detecting these diseases. Serum triglyceride is also determined routinely in lipo- protein phenotyping, which is described in Chapter VII. In this chapter, the stopped-flow system is shown to be useful in the determination of serum triglyceride. By monitoring the initial rate of an enzymatic reaction, results are obtained rapidly, and an early absorbance reading A can be determined without the problems ex- 0 perienced in adapting the enzymatic procedure for the cen- trifugal fast analyzer (106). A. Methods of Analysis Many of the natural lipids are glyceryl esters of fatty acids, and triglycerides are the most common. Complete hydrolysis by strong acids, alkalis, or enzymes yields free glycerols and fatty acids as shown below for triolein. 91 92 O CH200(CH2)7CH=CH(CH2)7CH3 HZC-OH ? H20 I CH0C(CH2)7CH=CH(CH2)7CH3 :; HC-OH + 9 I CH20C(CH2)7CH=CH(CH2)7CH3 H2C-0H Triolein (Triglyceride) Glycerol 3 CH3(CH2)7CH=CH(CH2)7COOH Oleic Acid The triglyceride level in blood rises to its maximum within 9 to 6 hours after digestion of food. For reliable results, blood specimens should be drawn after the patient has fasted for 12 hours. Two colorimetric methods have been used extensively for serum triglyceride determination. Van Handel and Zilvermit (101) extracted the serum sample with chloro- formmethanol, removed the phospholipids, hydrolyzed the triglycerides in alkaline solutions, and then oxidized the glycerol to formaldehyde. The final product is pink, with a visible absorbance proportional to the original triglyceride concentration. The Hantzsch condensation method (102) uses similar extraction and hydrolysis steps, but the final product is the yellow diacetyl lutidine species. This same product is also the basis of 93 fluorometric procedures (109). These methods yield precise and accurate results, but they are cumbersome and time consuming because of the extraction and the chemical hydrblysis. 0n the other hand, hydrolysis of triglyceride by enzymes eliminates many of these complexities. Currently there are two such methods. Bucolo and David's lipase hydrolysis of triglyceride (105) is at present the preferred method. It involves the following sequence of reactions: 1. Triglyceride hydrolysis Lipase Triglycerides —————e.Glycerol + FFA 2. Glycerol reaction sequence GK Glycerol + ATP———’a-GP + ADP PK ADP + PEP-——>ATP + Pyruvate LDH Pyruvate + NADH-———>Lactate + NAD where ADP = Adenosine diphosphate, ATP = Adenosine triphosphate, FFA = Free fatty acids, GK Glycerol Kinase, a-GP c-Glycerol phosphate, 99 LDH = Lactate dehydrogenase, NAD = Nicotinamide - adenine dinucleotide, NADH= Nicotinamide-adenine dinucleotide, reduced, PEP - Phosphoenolpyruvate, and PK = Pyruvate Kinase. Lipase has been shown to hydrolyze preferentially the esters of the primary hydroxyl groups of glycerol (112). Thus, in order to complete the hydrolysis, a non- specific esterase (a-chymotrypsin) was added. When both enzymes are present, the hydrolysis is completed after 10 minutes at 30°C. The free glycerol then reacts as shown in the above equations resulting in the oxidation of NADH. The total decrease in absorbance at 390 nm is proportional to the concentratin of glycerol liberated from triglycerides and the free glycerol in the sample (free glycerol is usually subtracted as a blank). The method has been used on the centrifugal analyzer (106,107), and on the Gilford 3500 (110). Previous serum extraction is not needed. Thus, the analysis is much simpler than those techniques which require an extraction step. However, phosphatase contamination by the enzymes introduces a competing reaction that makes the final value appear higher than what it is, and a "reagent blank" run is necessary (105). Free glycerol present in the sample also requires a sample blank determination. Stinshoff gt a; (109) prescribed a correction factor 95 for free glycerol that follows a normal distribution:100 mg/l can be subtracted from each total triglyceride value instead of running a sample blank. A less empirical way to avoid running a blank is to use kinetic methods. Lehnus and Smith (110) reported using a Gilford 3500 system to measure the change in absorbance during a 19 second measuring time after a lag time of 30 seconds. The result was as good as the manual "end-point" method (110). Chong-Kit (106) and Tiffany g£_al. (107) used the centrifugal analyzer and made determinations by rate methods in less than 3 minutes. B. Stopped-flow Method Although the centrifugal analyzer can determine serum triglyceride rapidly, Chong-Kit reported that the inabi- lity to take an early absorbance reading AO was a principal problem in adapting the enzymatic procedure for the GeMSAEC (106). The mixing time for the GeMSAES is about 20 3, during this time, there is an appreciable amount of réaction, and the authors had to extrapolate back to zero time to correct for this. The stopped-flow system with its ability to obtain absorbance data very earily in time, should eliminate this problem. 96 l. Reagent Preparations 1. "Stat—Pack" Triglyceride Kit (Calbiochem, Post Office Box 12087, San Diego, CA 92112). Vial A was reconstituted with 80 ml of dis- tilled water, and 0.25 ml of undiluted serum or standard glycerol was added to this solu- tion. The contents were then mixed with Lipase in Vial C, and incubated at 30°C for 10 min. The Glycerol Kinase in Vial B was also reconstituted with 15.0 ml of distilled water. 2. Stock glycerol standard. Carefully weigh out 1.091 g of 99.8% pure glycerol (Drake Brothers, Menomonee Falls, WI 53051), and dilute to 1 liter. This is equivalent to 1000 mg triolein per liter. 3. Working standard solution. Dilute the stock glycerol standard with saline to make working standard solutions. 2. Rate Studies The enzymatic reaction was adjusted to follow pseudo- first order kinetics. All analyses were carried out at 30°C, and the reactions were monitored at 390 nm for 60 s. Figure 5-1 shows the absorbance-time curves. There is a lag phase of N15 3, followed by a linear decrease in 97 db LL u u b lllLlLllll llllllllll II‘TIIIIII TIMECSEC) llllllllll IIIIITIIII lllllllllll OITIIIIIIII lllllJllll rIrIIII 111111]! IIII’IIIIII. . 000 0 2. 000-”"°°" .moz X10 '2 Figure 5-2. Standard curve for enzymatic analysis of serum triglyceride. 100 Table 5—1. Determination of Serum Triglyceride. Concentration (mg/dL) Rate RSD 50 0.70 9.1% 100 0.15 2.9% 200 0.30 3.8% Slope = 1.53 x 10'3 Intercept = -0.005 Corr. Coeff. 0.9997 101 reaction times, and lag phases for the reactions are determined by the concentration of the coupling enzymes. For a general discussion of this, the article by Scopes (116) should be consulted. 3. Experiments with Human Serum The accuracy of the stopped-flow initial rate method was assessed by comparing the results obtained with re- ported values of a control 1yophilized human serum (Moni-trol I). This comparison is shown in Table 5-2. Our results appear to be comparable with those of the reported values, and since we did not run a separate serum blank sample, 10 mg/dL should be subtracted from the results to correct for the presence of free glycerol as suggested by Stinshoff gt El. (109). The 129 mg/dL triglyceride concentration is consistent with the average value obtained by the other methods. The imprecision of the triglyceride analysis as can be seen from Table 5-2, may also be dependent on the dynamic blank caused by phosphatase contamination of the lipase in the commercially prepared reagent, as reported by Tiffany e§_al. (107). We tested this with a blank reaction without glycerol, and found its contribution insignificant for the batch of reagent we used. 102 Table 5-2. Serum Triglyceride Determination with Dif- ferent Methods. Mean Concentration Method (mg/dL) S.D. DUPONT ACA 129 16.82 BIODYNAMIC 133 ° 3.61 GEMINI 130 2.80 SMAC 93 9.81 HYCEL MARK X 190 (150)* 2.80 Ave. = 129 Stopped-Flow 129 (139)* 9.95 *Values not corrected for serum blank (free glycerol), all other values are corrected for free glycerol. 103 C. Conclusions The enzymatic triglyceride reaction is relatively slow but has proved to be satisfactory with manual methods. Although our adaptation of it for the stopped-flow system did not improve the sample throughput rate, or decrease the sample volume requirements significantly, we were able to determine A0 at an early time period, which allowed more effective use of the fixed-time rate method. Furthermore, it was found that the stopped-flow method is at least as efficient and viable as other methods for the routine analysis of triglyceride. VI. SERUM CHLOESTEROL ANALYSIS A. Introduction The importance of cholesterol (and cholesterol esters) has long been recognized in clinical medicine. Serum cholesterol levels and their dietary intake have often been related to altherosclerosis, resultant strokes, cardiovascular diseases, etc. (8,117). New diagnostic approaches, such as hyperlipoprotein phenotyping, have considerably increased the interest in cholesterol de- terminations (118). A host of different methods to determine serum choles- terol fills the literature. These have been reviewed so that even the interferences and sources of variation to which these methods are prone have also been reported (119-121). In the last 90 years, most analysts have preferred the iron-surfuric acid reaction and the Lieber- mann-Burchard reaction as the methods of analysis (8,122, 129). However, the recent introduction of enzymes as reagents to determine cholesterol has focused attention on the limitations of the strong acid approach (129). The principle of the enzymatic procedure is presented in Figure 6—1. Cholesterol esters are first converted to free cholesterol by the enzyme cholesterol esterase. The free cholesterol is then oxidized by cholesterol oxidase, and the H202 produced is reacted with 109 105 PHENOL é-AMINOANTIPYRINE qumonsmms DYE Figure 6-1. Enzymatic reaction of serum cholesterol. 106 9—aminoantipyrine and phenol to form a quinoneimine dye that absorbs at 510 nm. The specificity of the enzymatic procedure is un— doubtedly an important reason for its popularity. The enzymatic procedure is much simpler to implement when compared with the other reference methods (122,125). Furthermore, results are as precise and accurate as those obtained with other methods. All these factors make it a good procedure to use for routine analysis. The typical sample analysis time varies from 12 min. to 5 min.,_depending on whether a CentrifiChem analyzer (128) or a DuPont ACA analyzer (129) is used. In this chapter results of serum cholesterol determina- tions with the stopped-flow system are presented. Two rather significant problems which must be solved in order to get satisfactory analytical results were discovered. They are the slow rate of the enzymatic reaction, which makes short analysis times difficult, and the difficulty in preparing suitable aqueous cholesterol standard solu- tions. We did not attempt to study the reported effect of interference by bilirubin (131) because the serum samples we worked with were all normal and contained insignificant bilirubin concentrations. 107 B. The Stopped-flongethod l. Reagent Prgparation Cholesterol is not readily soluble in water, and as a suitable aqueous primary standard for cholesterol is lacking, we are seriously handicapped in our efforts to make quantitative measurements of the steroid by non- extracting procedures. The organic solvents, such as ethanol or 2-propanol used for non-aqueous standard solu- tion preparations (8) inhibit enzymatic activity and are therefore unsuitable for our purposes (126). As for the commercial aqueous cholesterol standards, the "Presicet" standards produced by Boehringer Mannheim Corp. (Indianapolis, IN 96250) are the most popular. The kit contains cholesterol standard solutions of 50, 100, 150, 200, 300, and 900 mg/dl, combined with a solubilizer and a preservative. Due to the unique 1:1 reagent to sample volume requirement of the stopped-flow system, we had to dilute the cholesterol standards 50 fold with 0.2% phenol solution so that the reagent solu- tion contains all the buffers and enzymes and the sample solution contains the cholesterol sample. However, upon .dilution, the solutions turned turbid, and thus a new method to prepare aqueous standard solutions for stopped- flow use was sought. Solubilizer formulations that have been proposed 108 for aqueous standards are 25% Triton x-100, 12% Triton x-100 with unique mixing, and 15% Triton x-100 with 7% albumin (127). The 25% Triton x-100 preparation was found to be too viscous for efficient stopped-flow mixing, and several attempts to dissolve pure cholesterol in 15% Triton solution failed. We experimented with an alcoholic- aqueous mixture and prepared the standard solutions as follows: 2-propanol, reagent grade (Fisher Scientific Co., Fair lane, NJ 07910) Cholesterol (Sigma, St. Louis, MO 63178). Triton x-100 (Research Products International Co., Elk Grove Village, IL 60007). To prepare 25 m1 (500 mg/dl) cholesterol standard solution, dissolve 0.125 g cholesterol in 1.0 ml 2-pro- panol. Dilute to 25 ml with 15% Triton x-100, and warm in a water-bath at 60°C until the solution turns clear. Then prepare 50 to 900 mg/dl cholesterol standards by proper dilutions with 15% Triton x-lOO. This preparation procedure gave clear standard solutions in the normal concentration range, and the final 2-propanol content was small enough (N0.08%) that the enzymatic reaction was not inhibited. 109 2. Rate Studies We modified a commercial cholesterol Auto/STAT kit (Pierce Chemical Co., Rockford, IL 61105) working reagent preparation by carefully transferring 2 bottles (12.2 gm each) of enzyme reagent to a bottle containing 950 ml phenol reagent. After reconsitution, this solution contains the enzymes peroxidase, cholesterol oxidase, cholesterol esterase, 9-aminoantipyrine, phenol, a stabi- lizer and a buffer. The substrate reagent was prepared by diluting cholesterol standards or human serum samples 50-fold with the phenol reagent provided. The two solu- tions were mixed in the stopped-flow system at 30°C, and the absorbance was measured as a function of time at 510 nm. Initially the reaction was monitored for 100 s, as shown in Figure 6-2. There is a lag phase of about 20 s, and from approximately 35 to 55 s, the reaction becomes pseudo first-order. As can be seen in Figure 6-3, the rate is constant during this period. All the sera behaved similarly, and the rate was found to be proportional to the serum cholesterol concentration. However, the reaction with aqueous cholesterol standards was found to behave quite differently. It has been reported earlier that both alcohol and Triton x-100 retard the overall reaction rate (126,127), and we confirmed this observation in our studies. Figure 6-9 shows the rate of reaction with 110 — di- C d— ._ «- .1- db .h ‘- .1.- .I- I: 1:— 1- “F .0- ur- db oil- .1— in 111111] 1111111111111 IIIIIIIII L111111L111111111111' .0 [OLIJ L 11 IIIIIIII 0 IIIFIIIIIrrIIIIIIIrl .,0EBO 0.500 cu- cu- "F' tab -+ d ~— qu- .1.— b I. P 39 dr- db- -- db ‘- db qt— .1. ~0— -- ‘— II- dr— moz 53 9. o 8 p Normal clear a: "a? 7:7 "j Z, I") 7.7 :7/ creamy top. Type I ' c ear , ¢7 -.»//’ bottom <§Z§Wéé gzjr/ Type Ila ‘Zz' / clear .1209770094/ faintly Type 11b 7; 77 turbid / / 1'-/////7 ”7): turbid. 97,—], faint Type III #7 ¢7j creamy tOp /7/1" A //’/é "'6'? Type IV . EFT turbid 20350T7?3%66 2:] creamy top .. v/ A over Type V / 73/ turbid // 777 ~77 ‘1. . , / ,g 4/ . Figure 7-1. Characteristic lipoprotein fractions and electrophoretic patterns. 122 decades. Meanwhile there was relatively little interest in the role of the high density protein (HDL), which ordinarily carries about 20% of the total plasma choles- terol. In recent years, however, there has been an in- creased interest in HDL levels and their significance in the development of coronary heart diseases (139). Analysis of HDL will certainly replace that of LDL as an indicator of coronary heart diseases in the future. 2. Methods of Analysis Lipid profiles are performed to determine serum or plasma cholesterol and triglycerides, after which lipo- proteins are then measured by an analytical ultracentri- fuge or through electrophoresis (190—196). The former is more quantitative but takes 8 to 16 hours for each analysis (190). The most recently developed instrument of this kind is the Beckman lipoprotein profiling system, which uses a portable air driven ultracentrifuge with an oxygen sensitive electrode as detector (193). Cholesterol determinations in HDL, LDL, and VLDL can be made with 0.5 ml sample within 3 hours. The second method involves electrophoresis of lipo- proteins and is generally performed on paper or agarose gel. When done on paper it takes about 10 hours (191), but when agarose gel is used, the analysis time is reduced to 3 hours (192). Because lipoprotein electrophoresis 123 is difficult to control, particularly with regard to uniform sample application and irregular dye uptake by lipids, it is only used in pattern recognition. In fact, Iammarino (138) has suggested that the method be dis— continued as a routine procedure due to its high cost and because it offers no additional information to the data generated by lipid analysis. Helena Laboratories announced their new line of HDL electrophoresis instruments early in 1978 and claimed that their new method was both rapid and precise. They used a specific enzyme reagent instead of the conventional Oil Red 0 dye in developing the electrophoregrams, and thus were able to reduce the total analysis time to 1 hour. But, a careful analysis of their data showed that relative standard deviations as high as 25% were reported in their quantitative analyses (199). In this chapter, a stopped-flow method is reported that requires only about 1/2 hour to perform and can avoid the expense of using ultracentrifugation or electro- phoresis equipment. B. Stopped—flow Method 1. General Scheme In order to use the stopped-flow mixing technique for lipoprotein determinations, it was necessary to find 129 a method to separate the three different lipoprotein fractions without using either ultracentrifugation or electrophoresis. Thus, several procedures were investi- gated. Burstein and Scholnick (198,152) described several useful chemical precipitation methods fortflmeseparation of plasma lipoproteins. Fredrickson _£'a1. (156) used a combined chemical precipitation and preparative ultra- centrifugation method to separate the lipoproteins. Later they reported a modified method which totally eliminated the need for an ultracentrifuge (199). How- ever, their method cannot be used for patients with sub- stantial degrees of hyperglyceridemia (199). Wilson and Spiger (151,155) reported a dual precipitation method as an improvement of the above, but upon comparison with the ultracentrifuge method, it was also found to be inapprop- riate for serum with high level of triglycerides. Our proposed method is shown in Figure 7-2. The goal is to estimate the relative concentrations of VLDL, LDL, and HDL along with the values for total cholesterol and total triglycerides, such that the sample can be classi- fied into different phenotypes. Precise quantitation of the HDL fraction is important, especially in light of its increasing significance as a correlative index of coronary heart disease. The serum total cholesterol and total triglyceride concentrations are first determined. If I 125 chylomicro HDL LDL VLDL ns Determi Ctotal 1'18 Determi TGtotal Q 3 Yes [/4E;j:;;:>\g No \/ @ I I ppt. with ppt. agent Detn. TGnet ppt. with ppt. agent Detn. CHDL ppt. with ppt . agent CVLDL= TGnet/ S C LDL ‘ Ctotal'CHDL'CVLDL CVLDL= TGtotal/ 5 CLDL = Ctotal'CHDL'CVLDL Type I; III, V Normal, Type 11, IV Figure 7—2. Proposed serum lipoprotein analysis procedure. 126 the concentration of triglyceride is normal, precipitation agents are added to completely precipitate chylomicrons, LDL, and VLDL. The cholesterol level of the HDL fraction left in the supernatant is determined, and the choles- terol concentrations of VLDL and LDL can be estimated as suggested by Fredrickson (199). If the triglyceride level is high, one additional precipitation is performed (198), and the estimation is done by a modified equa- tion as shown in Figure 7-2. The proposed method has several advantages over other methods. First and most important, total serum cholesterol, HDL, and triglyceride concentrations are known quantitatively. This allows estimation of lipo- protein concentrations without separation by electro- phoresis or ultracentrifugation. Secondly, the total analysis time is short. Finally, the whole procedure can be easily automated. 2. Choice of Precipitation Agent Serum lipoproteins are known to form insoluble com- plexes with polyphosphates and detergents (152), with polycations (161,150), and with polyanions (153-160). The polyanions were first shown to give complete, im- mediate, and selective precipitates by Burstein (198, 152), and the Lipid Research Clinic Program of the NIH 127 adapted a polyanion method (heparin—Mn++) that is currently widely used in the United States. A mixture of LDL and VLDL can be prepared by precipitation with heparin and either MnC12 alone or MgCl2 plus sucrose. Other poly- anions may also be used either with dextran sulfate and MnClZ, or with sodium phosphotungstate and Mg012. It was found that the lower the protein:1ipid ratio, the easier is the formation of an insoluble lipoprotein- polyanion-metal ion complex, but the nature of the lipid moiety does not affect precipitation at all (198). The nature of the lipoprotein-polyanion-metal ion interation can be explained by assuming that the electrostatic interaction between the positive charges of the proteins and negative charges of the polyanion cause the poly- anions to combine with the protein moiety of the lipo- proteins to form soluble complexes. This is easily proved by agarose electrophoresis (163). These compounds are then precipitated by divalent cations. The heparin-Mn++ precipitation process has been shown to correlate well with the accurate but time consuming ultracentrifugation method (153,155,159,160). However, it has several undesirable problems. Not all heparin preparations are equally efficient, and thus the complete- ness of VLDL and LDL precipitation must be checked each time a different lot of heparin is used (158). Also the Mn++-heparin procedure requires that the sample be 128 placed in an ice-bath for 30 min. after the precipitating reagents are added. Sera with triglyceride concentrations of more than 900 mg/dL may require prior dilution before precipitation (153,169). If an enzymatic technique is ++ sometimes forms used to measure HDL cholesterol, the Mn a visible precipitate with the enzyme reagents and also contributes a variable and significant blank (157). The application of Mg++-dextran sulfate as the precipitation agent has recently been adapted for routine lipoprotein quantification (158). There has not been any further report on the reliability of the procedure, and therefore this method was not studied here. Lopes-Virella at 31. (157) described a routine method using Mg++-sodium phosphotungstate. Values thus obtained compared well with the ultracentrifugation method, and the method does not have theproblems experienced with the Mn++-heparin method described (157). Personal com- munication with a local clinical laboratory confirmed the reliability of the method (165). Pierce Chemical Co. (Box 117, Rockford, IL 61105) manufactured a HDL-choles- terol rapid STAT kit in 1977 that uses the phosphotung- state-MgCl2 method. This kit also has unique features that make it best suited for adaptation to the stopped- flow system. Therefore, we chose this as our precipita- tion agent. Details of the reagent have been presented in the cholesterol chapter. 129 3. Separation of the Lipoprotein Fractions The polyanion-metal precipitation method is very efficient in separating HDL from the other fractions. The HDL fraction remains in the supernatant while the chylomicron, VLDL, and LDL fractions are precipitated. Electrophoresis studies have shown that the precipitation is selective and complete (198). In order to further separate the remaining fractions quantitatively, ultra- centrifugation must be used. In fact, the Lipid Research Clinic program recommended using ultracentrifugation to separate VLDL from LDL and HDL before separating the HDL from the other fractions by precipitation (166). Friedewald gt_al. (199) described a simple method to estimate the concentration of VLDL and LDL. Their procedure is very helpful in the interpretation of lipo- protein profiles, and it is based on the following rela- tionships: CVLDL = TG/5 CLDL = Ctotal ' CHDL ' CVLDL where T0 = serum triglyceride concentration, and C = cholesterol concentration in the HDL, LDL, and VLDL fraction all units are expressed in mg/dL. 130 These equations are based on two observations: (1) the ratio of the mass of triglyceride to that of cholesterol in VLDL is apparently relatively constant and about 5:1 in normal subjects; and (2) when chylomicrons are not detected, most of the serum or plasma triglycerides are in the VLDL. The equations are fairly accurate when the TG level of the serum is less than 900 mg/dL, making it suitable for normal and type II patients (see Table 701). But the use of the ultracentrifuge remains necessary with the much rarer type I, III, and IV hyperlipoproteinemia. Dual-precipitation methods have been proposed. De- tergents such as sodium dodecyl sulfate were used to separate the VLDL fraction (159,155). However, in- complete aggregation of VLDL by sodium dodecyl sulfate and the long incubation time reported do not justify the adaptation of these methods in place of the ultracentri- fuge. A modified procedure, which was listed earlier in Figure 7-2, is proposed in this work. The total choles- terol and triglyceride levels are first determined. If no elevation of the serum triglyceride level is observed, the HDL and VLDL, LDL fractions are separated by the Mg++- phosphotungstate method using Friedewald's equations to calculate each concentration. If TG is larger than 350 mg/dL, an extra precipitation step is taken to determine the net concentration of TG present in VLDL before 131 ao>fia op xomn Hopmumoaono moooomoooo ouoomcoee o Hm.Hummo.H mm mm a om gem mHHoo on Hogopmmaono msocomooco opoomcope m mmo.Hnmoo.H mm m: HH am can po>HH on» :H woufimonp nose we ohoomoooe aloha moo.a-mm.o ma me em a qaq> Honoumoaozo can we mnmpofio pgoomcmpe cfiwfipo mm.o u n am m copofiEonno .d 0 Tu ..a q u. no .a O O O S I 3 . a e S I 1 9 u T. 8 d a I o D I cofipocsm coapompm AHE\mV :oHpHmOQEoo mfimogosooppooam mpfimcoo R mumsfixopaa< .mcfiopopaoafiq Edmom mo woaupmaopm Hoofimmnm .Huw magma 132 calculating the concentration in the different fractions. In our final lipoprotein profile, we will have total cholesterol concentration, total triglyceride concentra- tion, HDL cholesterol concentration, and an estimate of the concentrations of all 9 different fractions of lipo- proteins. This information combines the best of the old (Fredrickson's phenotyping) and the new (CHDL) indices for a complete profile. Finally we have to consider the physical method of separating the HDL in the supernatant from the VLDL and LDL in the precipitate. Traditionally, the sample was mixed and centrifuged in a table-top centrifuge (1500 x g, 9°C) for 30 minutes. But, since Burstein reported that the precipitation process is immediate (198), pre- cious time is lost in the centrifugation process. Warnick and Albers (160) used ultrafiltration in their attempt to clear up the HDL containing supernatant after centrifugation for 30 minutes. The flow rate was quite acceptable (1 mL/min). We explored the possibility of using ultrafiltration to replace centrifugation, thus cutting the total analysis time by 30 minutes or more. 9. The Ultrafiltration Method Fresh serum samples from fasting patients in Sparrow Hospital were analyzed. To each 1.0 mL serum sample, 0.1 mL phosphotungstate-Mg++ precipitating reagent (Pierce, 133 Rockford, IL) was added, making the final concentration of MgC12-6H20 1% (W/V) and phosphotungstic acid 0.9% (W/V). Three samples were then centrifuged at 1000 x g for 20 minutes. The clear supernatants were aspirated out, and the cholesterol contents of these HDL fractions obtained by the Pierce manual method using an HDL/Cholesterol Rapid STAT kit (product #99090,99002). Another three samples were processed as follows: a 0.22 um Millipore filter (Millipore Corp., Redford, Mass. 07130), 25 mm in diameter, was placed on the lower support of a Swinnex filter holder, and a silicone gasket was put on the filter. Then an AP 15 and an AP 20 glass depth prefilter (22 mm in diameter) were fitted into the gasket. The upper unit of the Swinnex holder was tight- ened over the lower support to form a complete filtration assembly. Figure 7-3 shows the filtration assembly. The 1.0 mL precipitated suspension is then forced through the filter assembly from a 10 mL Syringe, with a flow rate of approximately 1 mL/min. The clear supernatant was then collected for HDL cholesterol analysis. The results obtained are presented in Table 7—2. The difference in observed values between the two methods is compatible within experimental error. From these results, it seems reasonable to conclude that ultra- filtration is a separation technique which is at least comparable to centrifugation. The filtration process 139 *IOmI Syringe Silicone gasket APIS Swinnex_ ’W APZO Depth prefilter filter . . . holder 0.22pm Millipore filter Figure 7—3. The Complete Filtration Assembly. 135 Table 7-2. Comparison of the centrifugation method with the ultrafiltration method. Centrifugation Ultrafiltration HDL Conc.* RSD HDL Conc. RSD (mg/dL) (mg/dL) 55.6 3.6% 59.8 9.3% *Average of 3 runs, total serum cholesterol conc.=230 mg/dL. 136 takes 1 minute instead of 20 minutes, and the filtration assembly is disposable if necessary. Ultrafiltration has the additional benefit of eliminating possible turbid supernatants that are sometimes found with the centrifuga- tion method. Therefore, this separation technique was used in all further studies. 5. Experiment with Fresh Human Serum Lipoproteins are not stable complexes and therefore fresh serum samples are required in serum lipoprotein analysis. We obtained three fresh serum samples from Sparrow Hospital for analysis. The total triglyceride and cholesterol concentrations were determined as follows: Glycerol standards for serum triglyceride analysis were prepared as described in Chapter 5. A standard curve was prepared by using the initial rate method. All analyses were carried out at 30°C, and the reactions were monitored at 390 nm for 60 s. For each sample, 0.25 mL human serum was added to vial A of the "Stat- Pack" reagent kit and was reconstituted with 8.0 mL of distilled water. Details of the procedure can be found in Chapter 5. Cholesterol was analyzed with the equilibrium method. Aqueous cholesterol standards were prepared from pure cholesterol as described in Chapter 6. The Pierce cholesterol Auto/STAT kit was used for the working reagents. 137 Human serum samples were diluted 50-fold with the 2% phenol reagent provided. The two solutions were mixed in the stopped-flow system at 30°C, and the absorbance was measured at 510 nm 15 minutes after mixing. The results for the total cholesterol and total tri- glyceride determinations are listed in Table 7-3. Since all the triglyceride levels were found to be below 350 mg/dL, we proceeded to follow Scheme A of Figure 7-2 and added 0.1 mL phosphotungstate—Mg++ reagent to 1.0 mL of each serum sample. The separation of the precipitate was then performed by ultrafiltration, and the cholesterol content of the filtrates were determined. The results are again shown in Table 7-3. Using Fredrickson's equation, discussed earlier, con- centrations of cholesterol in the different fractions of lipoprotein were calculated, and they are listed in Table 7—9. The three samples we obtained all showed values that were within the normal range. If the serum were type II abnormal, the C concentration should be very high, LDL whereas if the samples were type IV, CVLDL would be elevated. We therefore conclude that all three serum samples are normal. Using the same samples, a cellulose acetate plate electrophoresis study was performed with a procedure developed in cooperation with Dr. Hans Lillevik of the Michigan State University Biochemistry 138 Table 7-3. Lipid Analysis on Three Human Sera. Serum #1 HDL = Chol TG = Serum #2 HDL = Chol = T0 = Serum #3 HDL = Chol TG RSD 98 (Norm = 90-75) 5% 215 (Norm = 150-250) 1.9% 81 (Norm = 30-150) 3.8% 62 (Norm = 90-75) 251 (Norm = 150-250) 115 (Norm = 30-150) 69 (Norm = 90-75) 190 (Norm = 150-250) 95 (Norm = 30-150) 139 Table 7-9. Lipoprotein Analysis on 3 Human Sera. quation CVLDL = TGtotal/S CLDL = Ctotal ’ CHDL ‘ CVLDL Regalia CVLDL CLDL CHDL Serum #1 16 155 98 Serum #2 23 166 62 Serum #3 9 122 69 Normal range (0-90) (62-185) (90-75) Type 11 range (0-78) (173-890) (18-82) Type IV range (6-356) (28-231) (15-79) 190 Department (See Appendix C for details of the procedure). The resultant electrophoretic pattern is shown in Figure 779. The normal lipoprotein pattern confirmed our stopped-flow results. HDL analysis has become increasingly important in the past two years. Unlike electrophoresis data, our results give quantitative concentrations of serum HDL. Since the normal range of HDL is not very large (90-75 mg/dL), a less precise method would create false positive or negative results. The stopped-flow method gives results with better than 9% RSD at this low concentration range, which is comparable to the currently available manual methodology. C. Conclusions Although we did not encounter any abnormal samples in our limited sampling, we have shown that the stopped- flow method is feasible for obtaining a total lipoprotein profile (total cholesterol, total triglyceride, HDL, and phenotype of serum) in about 30 minutes. Analysis time can easily be reduced to 15 minutes if the reaction- rate method of cholesterol analysis is used. At this moment, we have to use commercially prepared enzyme reagents, and the reagent cost for each complete analysis is almost $3.00. This high cost can be reduced by developing multiple—loop immobilized systems as described in the 191 I’m '73 Figure 7-9. DENSITY Electronhoretic pattern for serum sample. ORIGIN DISTANCE FROM THE ORIGIN 192 cholesterol chapter, because then the expensive enzymes would be reusable. The sample throughput rate would also be further increased with this arrangement. The whole process can be totally automated when the solution prepa- ration and the filtration procedures are brought under computer control. VIII. FUTURE DIRECTIONS The experiments performed in this thesis have il- lustrated several unique characteristics of the stopped- flow system in clinical applications. We found the system excellent for routine clinical analyses in many ways. The throughput rate is generally increased using the reaction-rate method. The fast mixing characteristic of the system allowed investigation of the initial rates of rapid reactions with ease. The same feature also enabled . us to study fast reactions, such as the bromcresol green- albumin reaction, and in the process devise a way to eliminate most of the interfering reactions. On the other hand, several undesirable problems were encountered. Several tests became uneconomical because of the necessity to rinse the stopped-flow system between samples with expensive reagents. We also met with tur- bidity and some mixing problems mainly due to the require- ment of using a 1:1 sample to reagent volume ratio. While the results presented demonstrated some of the analytical capabilities of the stopped-flow system, there are probably many more routine chemical analyses that have not been investigated by reaction-rate methods but are adaptable to the system. Pinnell and Northam reported a bromcresol purple (BCP) method for analyzing serum albumin. They claimed 193 199 that the dye is specific to albumin, thus eliminating the interfering problem to which BCG is susceptible (190). Furthermore, they reported that the BCP reaction with serum is an instantaneous one, making it very worthwhile for future investigations with the stopped-flow system. Enzymatic analysis of uric acid was done with a centri- fugal fast analyzer (189). The reaction is relatively fast, and the use of an early reading blank-corrected approach was found to be desirable(189). The uricase/ catalase/aldehyde dehydrogenase-coupled procedure is also ideally suited for adaptation to the stopped-flow system. The Jaffe reaction for serum creatinine appears to follow pseudo-first-order kinetics (192) and this routinely determined substrate can be easily analyzed by the reaction-rate method using the stopped-flow instru— ment. Another commonly determined substance is serum alcohol. Jung and Ferard described a specific and precise enzymatic reaction which reached equilibrium after only 9 min (193). No stopped-flow method for serum alcohol has been reported yet. One may also try to apply the simplex method des- cribed in Appendix A in differential rate analysis for mixtures. Pelizzetti 23 al, reported a kinetic determina- tion of adrenaline, L-dopa and their mixtures with a 195 stopped-flow system (191). The simplex technique should be superior to the single-point method of Lee and Kolthoff they were using. Another approach in research should be directed towards developing the immobilized enzyme loops as men- tioned in Chapters 6 and 7. Successful adaptation of these loops onto the stopped—flow system would mean an increase in throughput rate and a decrease in cost per analysis. Referring back to Table l in Chapter 2, one can see that we have only explored a small fraction of the clini- cal tests routinely performed. The largest unexplored area is the analysis of enzymes in serum. The stopped- flow system is especially suited for this kind of analysis because all procedures for serum enzyme analysis involve reaction-rate measurements; and the amount of enzyme present in serum can be measured only indirectly in terms of its activity (or rate of reaction) with its substrate. Commonly determined enzymes such as SGPT, SGOT, LDH, CPK, ACP and ALP all fit into this category. Another important area for further work is the de- termination of isoenzymes. Kaplan et_al. (59) were the first to apply the stopped-flow system in a differential determination of LDH isoenzymes. They pinpointed the exact source for abnormal levels of LDH enzyme and used the result as a sensitive test for myocardial infraction. 196 Their method was superior to conventional separation methods such as electrophoresis because analysis time was drastically reduced. Bostick and Mrochek ( 58) reported a procedure for creatine kinase MB isoenzyme using differential kinetics with a centrifugal analyzer. This should also be suitable for the stopped-flow system. Pardue had demonstrated that the stopped-flow system could have an attractive throughput rate even for conven- tional types of equilibrium analyses (1). Thus, the stopped-flow system seems to be applicable to virtually all types of wet chemical analyses performed routinely in clinical laboratories. And in addition, it has a capability for making fast measurements that are not pos- sible with any of the conventional instruments. We feel the stopped-flow system would make an excellent 'stat' analyzer in clinical chemistry. Further research would be centered around methodological development, complete automation of the system, and the decreasing of the dead volume in the system to minimize carry-over problems. Automated sample preparation, valve sequencing, and simple data processing can be brought under microprocessor control. Chemically inert and inexpensive high pressure liquid chromatography valves and fittings that have small dead volumes can be used in an improved design of the system. The continuous development in analytical methods should bring about wider acceptance of the system in the clinical chemistry field. APPENDICES APPENDIX A APPLICATION OF THE SIMPLEX METHOD IN DIFFERENTIAL KINETIC ANALYSIS A. Introduction A clinical chemist is often faced with the difficult task of determining the fastest way to handle several var- iables which are often known to interact. Spendley gt al. (171) introduced an idea in 1962 for tracking optimum operation conditions called the simplex optimization method. Essentially it searches for the optimum using empirical feedback strategy. Compared to other optimiza- tion methods, it requires fewer experiments for each move. It is also more compact, involves simple computations, and is particularly attractive for automation. Nelder and Mead (172) modified the original fixed- size simplex method. Later Long applied it to analytical chemistry (17“). Recently it has been applied in clinical chemistry (176). Several excellent reviews about the simplex optimization method have been published (167,169, 175,195), and at least one paper has indicated its pos- sibility in regression analysis (187). With the advent of high speed computers, linear least square analysis has been used in routine statistical analysis of data (179,180). However, more complex problems 197 198 such as non-linear or arbitrary functions sometimes require special constraints upon the data collections; or a user may face miserable failure (when an exasperated computer finally print out a message of defeat), or even disastrous failure (when the computer and user think they have found the answer) when improper procedures are employed. Matrix inversion is usually required in the curve fitting process and can lead to serious round-off errors in the calculation if handled improperly (185,186). A powerful general purpose curve fitting program called 'KINFIT' was designed by Dye and Nicely (182) and was very successful using CDC 3600 and 6500 computers as well as IBM 7099 and 360-75 computers. In the process of doing differential kinetics analysis in our laboratory, the need for a simple and computationally more compact program to be used with a 16K PDP 8/e minicomputer became evident. This need stimulated an investigation of the simplex method in curve fitting. This idea is explored in this chapter, and experiments with computer-simulated data are reported to demonstrate the feasibility of the idea. B. The Super-Modified Simplex 1. General Description The Nelder and Mead modified simplex has been used widely, but the procedure sometimes fails in stochastic 199 or noisy environments due to premature simplex contractions, adherence to false ridges, or reluctance to approach the boundary constraints (178). Denton gt al. (178) developed a super-modified simplex (SMS) procedure that increased the utility, reliability, and efficiency of the Nelder and Mead version. We further modified the SMS and applied it in least-square analysis as described later. The rules of the Nelder and Mead modified simplex are briefly reviewed below to facilitate the understanding of the SMS (169, 172). A simplex is a n-dimensional geometric figure containing n+1 vertices. For example, the initial figure of a two dimensional simplex is a triangle with vertices VBVNVW’ as shown in Figure A-l. These three vertices were measured and vertex VB was found to be the best response, Vw the worst response, and VN the next-to-worst response. ‘P is the centroid of the face remaining after the worst vertex has been eliminated (169). The optimum in the response surface can be found by one of several different moves as listed in Table A-1. To eliminate the worst response Vw, Table A-1 again provides us with several measures: it can be replaced by VR through reflection, by VE through expansion, or by VCR or VCW through contraction. A failed contraction occurs when the response at either VCR or VCw is worse than the response at VR or Vw, respectively. When this happens, a "Massive Contraction" move is then made, each leg of the simplex is contracted to one half 150 REFLECTION EXPANSION /\ \ L I I I I CONTRACTION Figure A-l. Basic moves of a simplex. 151 Table A-1. Basic Moves of a Simplex. VE = P + y(P - Vw) y = expansion coeff.=2 VCR: P'+ 8(P - Vw) B = contraction coeff.=0.5 <; O 2 ll “UI -e@-vm 152 its present length towards VB, forming a smaller new simplex VBVN'VW' or VBVN'VR" The responses are then determined and the optimization process continued. Boun- dary violations are handled by assigning a poor response to the vertex that violates the boundary and the new vertex is determined by a CW contraction. A flow chart of the modified simplex is shown in Figure A-2. Reflection is always the first move, followed by expansion or contrac- tion, while massive contraction is used only when the contraction has failed to give a better response than the original worst response. 2. Procedure for the Super—Modified Simplex The flow chart for the super-modified version is shown in Figure A-3. The initial vertex and the responses are determined as before. A reflection is made to determine the new response RVR. Now, instead of expanding or contract- ing with a fixed coefficient, a second order polynomial 2+Bx+C) is fitted through the responses at curve (y=Ax Vw, P, and VR' If the optimum sought for is a maximum, and the slope A is negative (i;§;, the curve is concave down), the maximum can be located within the interval by evaluation of the derivative of the curve. The response of this new vertex can then be determined, and after the safety and boundary checks, the optimization process can be continued. If the curve concaves up, no maximum can 153 @[mra Reflection No Yes No 7 7 Y No es NO RVE--—--N found? LJI Figure A-2. Procedure for the modified simplex. Iassive ontractio FE- 159 Reflection Flt Rw, RP'RVR through y-Ax2+Bx+C determine A.B Extend VW & VR by theoretical response Locate the mm‘xlo Pick direction C811 this VR' of expansion V Isafty R! (Expansion ) factor '® Va.-I boundary .Yes EXIT J Figure A-3. Procedure for the super-modified simplex. 155 be located. The curve is then expanded a percentage of (VWTE) and the extended interval boundary producing the highest predicted response is chosen as the new vertex location. Further expansion can be made by testing the slopes at these extended boundaries, deciding upon the direction of expansion, and expanding according to the magnitude of the slope. The SMS procedure thus allows the simplex to locate a new vertex more freely and move much faster towards the optimum by checking a larger response surface during each move by curve fitting. Although additional calculations are required, there is virtually no delay in experimental time since all calculations are performed by the mini- computer. A subroutine SMSMPX.FU was written following the algorithm described. It chains to other subroutines that input parameters in an interactive mode, print out results in several modes, and plot the simplex after each itera- tion as desired. A detailed description of the program, as well as the working program written in the FORTRAN IV language, is listed in Appendix B. 3. Design of the SMS Algorithm This section describes the basic factors considered while designing the algorithm of the SMS subroutine. 156 i). Fitting a second degree_polynomial curve - Instead of expanding or contracting with a fixed coefficient in each move, 3 points (VR,VW,F) are fitted to the equation y = Ax + Bx + C therefore we have three equations as follows: _ 2 O yl Axl + Bxl + Cx1 y2 = Ax22 + Bx2 + ng _ O y3 - Ax3 + Bx3 + 0x3 Solving for the unknowns A, B, and C by matrix methods, we have A = (y1X2+le3+V2X3-Y3X2-X3yl-Y2Xl)/D _ 2 2 2 2 2 2 B - (xly2+ylx3+x2y3-x3y2-y3x1 x2yl)/D c = y -Ax2-Bx l l l where 157 We can simplify the equations by letting x1 = -1, x2 = O, and x3 I l as the x coordinates for the 3 responses. The resulting equations are A = l/2(y1+y3-2y2) = 1/2(R+w-2C) B = 1/2(yl-y3) = l/2(R-W) where R - response at VR; w a response at Vw; C response at F. Finding A and B in this manner, we can proceed to find the shape of the curve as described in the next section. ii) Determination of minimum and maximum - The second derivative of the function Y = Ax2 + Bx + C was calculated to see if it is at a minimum or a maximum. If the second derivative is positive, the function curves upward and there is a minimum. If it is negative, the function curves downward indicating a maximum. And if the value is zero, it means that there is an inflection, which may be a maximum, or a minimum, or neither. iii) Expansion coefficient - The next step is to de- cide how much further we are going to expand or contract the vertex VR. Take the example of a maximization process. 158 If no maximum is found, the polynominal curve is then expanded further by a factor in order to investigate more portions of the response surface. The magnitude of this factor can be decided by looking at the slope of the new vertex at the extended boundary. Denton gt al. (12) suggested that the greater the slope (indicating rapid ap- proach to the optimum location), the smaller the required expansion coefficient; and conversely, the smaller the slope (indicating remoteness from the optimum location), the larger the required expansion coefficient. We found that there is a problem with this concept. It is difficult to define a "large slope" as such, because the value of a slope depends on the units one is using. An alternate way to attack this problem is to look at the ratio of the old and the new responses and decide upon the size of the expansion coefficient accordingly. In an optimization process for the maximum, if the ratio is large (indicating rapid approach to the optimum location), use a small expan- sion coefficient; but if the ratio is small or is close to l, we want to expand more to explore new grounds. The following equation was found to work quite well, x = new response/old response y = expansion coeff. = —2.75x + 6.75 159 iv) Safety factor - Due to the adaptation of the curve fitting method, the new vertex may occur at or very near the centroid. This will reduce the dimensionality of the process and may terminate the progress of the simplex. Therefore a small safety interval is placed between the centroid and the new vertex at the beginning of the search. We picked 0.2 out of 1.0 absolute scale to be this safety factor. As illustrated in Figure A-H, when the new vertex is located within this range, we set it equal to 0.2. Use of the safety interval is unnecessary when approaching the termination of the optimization search and can easily be eliminated near the optimum. v) Boundary violation - When VR is located outside of the boundary constraint, we simply set the point at the boundary which is intersected by the vector connecting VW and VB as the effective VR, we call this new vertex VR" The polynomial curve is then fitted through the responses Of VW’ F, and VR" and the search continues. This process is illustrated in Figure A-S. The method prevented the old problem of reluctance to approach the boundary constraint reported in the modified version of simplex (178). vi) Initializing a new simplex - Usually the first question that confronts a user of the simplex method is 160 I IV -1.o -o.2 L; R 0.2 .0 VW I'Pl VRO i I Figure A-h. Safety factor in SMS. Boundary I ,Constraint Figure A-5. Boundary violation in SMS. 161 how to design the initial simplex. Long (17h) described in detail the construction of initial simplexes of up to ten factors. Instead of selecting the step size and the initial vertex randomly, a systematic algorithm is used. The following is an example of a two dimensional simplex. 1) Choose the initial vertex (V1); 2) Choose step size (Si); 3) Calculate p1 and q1 with N = number of factors; p1 = (Si/(N*21/2))((N+l)1/2+N-l) q1 = (Si/(N'2l/2>)(