THE RHEOLOGICAL PROPERTIES OF THE ANTERIOR CRUCIATE UGAMENT OF A DOG Thesis for the Degree of M. S. MICHIGAN STATE UNIVERSITY ROGER C. HAUT 1968 . . rm“. . 1.1V .rp-u— r w '- TTTTTTT TTTTTTTTTTTTT [l 3 1293 01070 543 6 Unwersxty 3'3 3 -———— 3 “’ Bun-Ema av r HUAG 8T SUNS' 300K BINUERY INC. - LIBRARY aluoans . ABSTRACT THE RHEOLOGICAL PROPERTIES OF THE ANTERIOR CRUCIATE LIGAMENT OF A DOG by Roger C. Haut The purpose of this research was to study the mechani— cal properties of the anterior cruciate ligament of canines. The changes in the shape and magnitude of the stress-strain curves for various strain rates are discussed in detail. The specimens were tested in tibia-anterior cruciate ligament- femur preparations which were subjected to environments of moistening and direct immersion in a saline solution. For these tests the specimens were constrained in the normal flexed positions of the hind limbs of canines. Information dealing with the per cent cumulative stress relaxation was obtained by allowing the specimens to relax at regular inter- vals during the loading. By testing each of the specimens twice, information regarding the possible damage, which might have occurred during the first test cycle below 10 per cent strain, was obtained. The stress-strain data fitted extremely well with an exponential relationship suggested by Y. C. Fung for the mesentery of the abdomen of a rabbit. The variation in Roger C. Haut environments appears to have an effect on the properties of the ligament. The results of the relaxation tests indi- cate that for high values of stress the per~cent cumulative relaxation is linearly proportional to the value of stress. It was observed that when 2 successive tests below 10 per cent strain were subjected to the same specimen, the second resulting test shows an extended initial toe region. How- ever, in the linear region of the curves the slope is almost identical with that of the first test. THE RHEOLOGICAL PROPERTIES OF THE ANTERIOR CRUCIATE LIGAMENT OF A DOG BY Roger C?‘Haut A THESIS Submitted to Michigan State University in partial fulfillment of the requirements for the degree of MASTER OF SCIENCE Department of Metallurgy, Mechanics, and Material Science 1968 ACKNOWLEDGMENTS The author wishes to express his appreciation and graditude to the following people for making this phase of his graduate work possible: To Dr. Robert W. Little, his major advisor, for his encouragement, friendship, and, most of all, for his valuable assistance in the preparation of this thesis. To Dr. Wade 0. Brinker for his constant concern and medical assistance for this thesis. To the National Institute of Health for the finan- cial support required for this study. To Dr. William Sharpe for his assistance in the de- sign and Operation of the inStruments needed for this study. To his mother for her continued concern and encour- agement for his academic education. To his wife's parents for their understanding and interest in his graduate work. To his wife, Judy, for her endless patience and encouragement throughout his graduate work. ii TABLE OF ACKNOWLEDGMENTS . . . . . . . LIST OF FIGURES . . . . . . . LIST OF TABLES . . . . . . . Section I. INTRODUCTION . . . . . II. SURVEY OF LITERATURE . III. EXPERIMENTAL METHODS . IV. PRESENTATION OF RESULTS V. DISCUSSION AND CONCLUSIONS BIBLIOGMPHY O O O O O O O 0 APPENDIX 0 O O O O O O O O 0 CONTENTS Page . . . . . . . . . . 11 . . . . . . . . . . iv . . . . . . . . . . vi . . . . . . . . . . l . . . . . . . . . . 9 . . . . . . . . . . 20 . . . . . . . . . . 30 . . . . . . . . . 53 . . . . . . . . . . 60 . . . . . . . . . . 64 iii Figure 1.1 3.1 3.2 3.3 4.11 4.12 4.13 4.14 LIST OF FIGURES Anterior Cruciate Ligament . . . . . . . . A Specimen in Fixtures . . . . . . . . . . A Ruptured Specimen . . . . . . . . . . . . Testing Setup . . . . . . . . . . . . . . . Measurement of Initial Length . . . . . . . Clip—Gage Calibration . . . . . . . . . . . Measurement of Area . . . . . . . . . . . . Stress-Strain Plot for Moistened Specimen . Stress-Strain Plot for Immersion at 72°F . Stress-Strain Plot for Immersion at 101°F . Strain-Strain Rate for Moistened Specimen . Strain-Strain Rate for Immersion at 72°F . Strain-Strain Rate for Immersion at 101°F . Apparent Young's Modulus-Strain Rate After Toe O O O O O O O O O O O O O O O O O Stress-Strain Plot for 5%/min. Strain Rate Stress-Strain Plot for 15%/min. Strain Rate Static Tests . . . . . . . . . . . . . . . Stress-Relaxation When Moistened . . . . . Stress-Relaxation in Saline at 72°F . . . . Stress-Relaxation in Saline at 101°F . . . Per Cent Relaxation-Strain Rate When M018 tened O O O O O O O O O O O O I O O O 0 iv Page 27 27 28 28 29 29 34 35 36 37 38 39 4O 41 42 43 44 45 46 47 Figure Page 4.15 Per Cent.Relaxation-Strain Rate When Immersed at 72°F . . . . . . . . . . . . . . 48 4.16 Stress After Relaxation—Strain for Immersion-at 72°F . . . . . . . . . . . . . . 49 4.17 Stress After Relaxation-Strain for Moistened Specimen- . . . . . . . . . . . . . 50 4.18 Stress After RelaXation-Strain for Immersion at 101°F . . . . . . . . . . . . . 51 4.19 Stress-Strain for First and Second Tests When Moistened . . . . . . . . . . . . 52 LIST OF TABLES Table Page 1 Values of a and C for Exponential Law . . . . 56 vi I . INTRODUCTION Most living organisms are divided into small units called cells. Robert Hooke, in 1665, first reported and named cells from an examination of cork. These vacuoles were not cells but rather depressions that were implanted in the matrix of intercellular material by the cells of the cork. In 1839, M. J. Schleiden, a botanist, and Theodor Schwann, a zoologist, formulated what is now called the Cell Theory. Their contribution was actually more of a synthesis of ideas and observations than a discovery. Two basic principles comprise the Cell Theory. 1. Principle of Cells--That all organisms are formed from cells. 2. Principle of Biogenesis--That cells are produced from other cells. These, coupled with the notion that sperms and eggs are also cells which unite in the formation of a new individual, constitute the basis of all life. When similar cells are joined together, either closely or loosely, with an intervening non-cellular mate- rial, the aggregate is called a tissue. In the animal king— dom there exists 4 primary types of tissues: epithelial, connective, muscle, and nervous tissue. The study of tissues is called histology. Epithelial tissue is the predominate type of tissue in most free internal and ex- ternal surfaces.‘ The cells of this type of tissue are closely joined, hence there is a minimum quantity of inter— cellular material. Epithelial tissue is further subdivided by its cellular shape. The epithelial surface which lines the mouth, esophagus, the outer surface of the cornea of the eye, and most of the female urogenital tract is called squamous epithelium because of its thin flat plate-like cells. Columnar, which are rectangular as the name sug- gests, constitute areas such as the pharynx, soft palate, and gland ducts. Cuboidal epithelium tissue is found in the kidneys, the retina of the eye, and on the inner sur- face of the lens of the eye. Muscle tissue, known for its contractibility and irritability, has a cellular structure in which the cells are elongated into muscle fibers. The contractibility of this particular type of tissue, as much as 33 per cent of its initial length, is theorized as being attributed to a dove-tailing effect of its fibers, myosin and actin. Nervous tissue is composed of nerve cells, or neurons, and some supporting cellular structures. Nerve impulses, according to the Membrane Theory, are conducted by a depolarization of charge along the tubular membrane of a neuron. These impulses are then transferred from one neuron to the next neuron via a synapse. At a synapse there is no physical contact of the neurons, therefore an electrical contact between neurons has been theorized. Connective tissue is by far the most predominate and diversified tisSue in the entire animal kingdom. With reference to the human body, connective tissue appears as bone, cartilage, ligaments, and tendons. Blood is also classified as a type of connective tissue. The distinguish- ing characteristic of this type of tissue is the presence of a large amount of intercellular material. This inter- cellular material may consist of collagenous or white fibers, elastic or yellow fibers, reticular fibers, and varying amounts of an amorphous ground substance. Elastic fibers are rather long, flat ribbons which contain a substance called elastin. Elastin is a very~low»modulus material. The large deformation which is seen in the posterior neck region of grazing animals results from the stretching of a large ligament composed ahmost entirely of elastin. This ligament is called ligamentum nuchae. The fiber giving the high/modulus effects to tendons and ligaments is col- Tagenous. The anterior cruciate ligament, which is the immediate concern at this time, is composed almost entirely of this type of fiber. These fibers appear as long wavy ribbons which are not branched. They appear to be the product of a long, flat, or star-shaped cell called a fibroblast. The production of collagenous fibers is ap- parently in the form of a secretion process in which an aggregate of extracellular material is secreted by many fibroblasts. The primary building blocks of these fibers are the collagen molecules: three chains of amino acids in certain sequences which coil into left-hand helices and intertwine to form a right-handed superhelix (9,17). The wavy pattern of a fiber will be considered later with reference to its effect on the stress—strain curve. Reti— cular fibers are thought to be immature collagen fibers which are highly branched. Connective tissue differs widely throughout the body. Loose connective tissue contains reticular, elastic, and collagenous fibers. The matrix of intercellular mate- rial has no orientation of its fibers, resulting in a low modulus material such as the mesentery of the abdomen. Dense connective tissue is essentially the same as loose connective except for a fewer number of cells and a larger percentage of collagen fibers. Regular connective tissue has an abundance of collagenous fibers, and it has a defi- nite arrangement of its fibers. For example, tendons have bundles of fibers which lie parallel to the central axis of the tendon with fibroblasts existing between the bundles. Ligaments are essentially the same as tendons, but they have a less regular arrangement of the collagenous and elastic fibers. These collagen bundles are presumably as long as the tendons or ligaments (18). The collagen bun- dles are surrounded by a woven mesh of loose connective tissue, the peritendineum internum, containing elastic fibers that tend to draw the bundle into a any formation in the relaxed condition. This structure, along with the pattern of the fiber itself, is significant to the stress- strain curves. Tendon and ligament fibers are continuous with the fibers of bones and at this point are called the fibers of Sharpey (24). Previously it was mentioned that ligaments and ten— dons had very little histological difference, but func- tionally they are very different. Ligaments serve as con- nectors between two bones, whereas tendons attach bones to muscles. Even though the point of attachment of ligaments and tendons appears to be continuous, a cellular transition can be noted in this region. In this region the fibrocytes undergo a transition to osteocytes (bone cells), and here they most resemble condrocytes (cells of cartilage) (24). In early investigations on tendons and ligaments using the intact bone-ligament-bone or bone-tendon preparations,most ruptures were encountered in this transition region. This failure was attributed to a weakness of the specimen in this transition region. There is reason to believe that failure might have been encountered in this region because of the high moments present when testing the Specimen in an unnatural position. This question is discussed further during the consideration of a proper testing procedure for the anterior cruciate ligament of canines. The stifle joint, located in the lower or hind limb, is the most complicated joint in the body. The stifle joint must transmit forces of a much larger magnitude than any other joint in the body, so that a high degree of sup- port must be present to stabilize this joint. To comply with the rigid requirements of stabilization as well as freedom to flex and extend under large forces, the stifle joint has the largest articulating surface in the body. The bones involved with this joint are the upper bone, the femur, which inserts into the pelvic girdle and the lower bone, the tibia, which articulates with the talus to form the ankle joint. Because of the support which is required in this joint, there are many muscles and tendons surround- ing this area. Collateral ligaments lie along the medial and lateral borders of the joint, while the quadriceps femoris muscle encases the knee cap which protects the anterior region of the joint. Posteriorly, the biceps femoris and the gastrocnemius muscles help support the joint. The entire joint is encircled with hyaline carti- lage which serves as a cushion and as a capsule for the encasement of synovial fluid. Synovial fluid is a light fluid which is present in the joint to reduce friction. In- terior to the joint are cruciate ligaments, so named because the two ligaments form a cross in the joint. The posterior cruciate ligament extends from the posterior intercondylar fossa of the tibia, a depression between the contact Med/'a/ half of / femur \\ ' / Pas/en'or lxg \ ' Anferior //g\ T‘ “r em.» Medial _ _ meniscus Lat memlscus- .1”; N ,r _‘ ' 'A. o" “‘- .. I S , ‘ ‘39“, ’ ' / I \1 Pare//o\ / q 0 ‘4‘ 3 3‘. Figure 1.1--Anterior Cruciate Ligament (Reproduced by courtesy of the J. B. Lippincott Co.. Phil.) surfaces, upward and forward to the lateral side of the medial condyle of the femur. The anterior cruciate ligament extends from the front of the intercondylar eminence of the tibia upward and backward to the medial side of the lateral condyle of the femur. Isolation of the anterior cruciate ligament without damage is possible due~to this central positioning in the joint. Injuries to the collateral and cruciate ligaments of the stifle joint are very common. During hyperextension, the anterior cruciate ligament and the collateral ligaments are under large tensile forces. The posterior cruciate ligament is apparently under a certain degree of tension at all times. From clinical studies of the stifle joint it is noted that when flexion with a subsequent medial rotation of the femur takes place, the anterior cruciate ligament and the collateral ligaments are under extreme tension. The most common type of injury is when the tibia is rotated laterally, as the joint is flexed. This type of injury is frequently encountered in football and skiing accidents (1). It has been suggested that similar movements may be the cause of the ruptures of the anterior cruciate ligaments in canines. II. SURVEY OF LITERATURE Many investigations have been conducted in an at- tempt to describe the mechanical properties of biological materials. More specifically, results dealing with regu- lar connective tissue such as ligaments and tendons were considered as relative background for the study of the mechanical properties of collagenous tissues. Correlation was extremely difficult and the results were sometimes contradictory because of the large number of various meth- ods of testing. For this reason a chronological survey of the results dealing with regular connective tissue, specifically tendons and ligaments, was chosen as the most appropriate description of earlier work. Some of the first bioengineers were Harvey, Helm- holtz, Poiseuille, Hooke, Galilei, Boyle, Euler, and Young. William Harvey was credited with the discovery of blood circulation in 1615. To von Helmholtz might go the title of "Father of Bioengineering." He discovered the focusing mechanism of the eye, and invented the Helmholtz resonator 'for the study of hearing. Other accomplishments were the invention of the phakoscope to view the retina, the ophthalmometer for measurement of eye dimensions, and the stereoscope with interpupillary distance adjustments for 10 stereo vision. Poiseuille invented the mercury manometer to measure-the blood pressure in the aorta of a dog while he was a medical student, and he discovered Poiseuille's law of viscous flow upon graduation. As previously men- tioned, Robert Hooke gave us the word "cell." Galileo Galilei discovered the constancy of the period of a pendu- lum, and used it to measure the pulse rate of humans. He invented the thermometer, and was also the first one to design a microscope in the modern sense. Robert Boyle studied the lung, and discussed the function of air-in water with respect to fish respiration. Leonhard Euler studied the propagation of waves in arteries. Thomas Young, who gave us Young's Modulus of Elasticity, was a physician in London. He worked on the wave theory of light while he was concerned with astigmatism in lenses and in color vision. One of the earliest investigations of ligaments was Annovazzi's work in 1928 (3). Although his work can- not be obtained, the literature states that he studied the properties of the anterior cruciate ligament of dogs. The only information known by most investigators is that he found that the mechanical properties of the ligaments changed drastically within the first hour after the death of the animal, but there is no indication as to how the specimen was preserved before or during the tests. In 1935, A. E. Cronkite conducted tests on almost every tendon in the human body (6). He fixed the tendons 11 in what was called "a normal laboratory manner“ for as long as 18 months. He reported that there was no significant difference in the fixed or non-fixed fresh tendons. Cron- kite reported that there was no one tendon which was stronger than another. He did note that tendons with a larger cross-sectional area seemed to have a little higher tensile strength, and suggested that this might be due to a greater density of collagen fibers in the larger area. Cronkite indicated that the tensile strengths of various tendons in the human body ranged from 2.8 x 108 dynes per square centimeter 1x) 2.1 x 109 dynes per square centi- meter’with variations in different cadavers being 6.0 x 108 dynes per square centimeter to 1.2 x 109 dynes per square centimeter. No correlation could be noted between the strengths of tendons and the age of the individuals. Also, tendons which normally are required to transmit large forces, such as flexors, tended to be no stronger than those which normally are required to transmit lower magnitudes of forces, such as extensors. In 1935, C. M. Gratz prepared 3 types of tendons and subjected them to loads under 500 kilograms per square centimeter, which he states as the maximum stress for which the specimens were still in the elastic range (10). The specimens were cut to test size and.moistened with Ringer's solution.until the test time. For the Achilles tendon he found a maximum tensile strength of 6.8 x 108 12 dynes per square centimeter with a Young's Modulus of 3.9 x 108 dynes per square centimeter. Similarly for the flexor digitorum longus, his results indicate a tensile strength of 7.3 x 108 dynes per square centimeter and a modulus of 3.4 x 108 dynes per square centimeter, and for the erector spinae tendon 6.7 x 108 dynes per square centi- meter, respectively. He noted that the Ringer's solution made no significant difference from the control group for 3 hours of testing. In 1951, R. H. Hardy conducted a series of tests in which ligamentum flavum and plantar calcaneo-navicular ligaments were subjected to various loads (12). Ligamentum flavum is a ligament which histologically is composed of elastic fibers whereas the plantar ligament is entirely collagenous. Hardy indicated that the plantar ligament displayed no deformation when a stress of approximately 70.4 kilograms per square centimeter was applied to it. On the other hand the ligamentum flavum strained approxi- mately 140 per cent under the same loading conditions. These results indicate that collagen fibers have an ex- tremely large modulus, whereas ligaments composed of elastic (or elastin) fibers display a very low modulus. In each of the above tests the specimens were fixed for 24 hours in a 10 per cent formol-saline solution. Since Hardy did not indicate his manner of testing the specimens, his work cannot be fully justified. 13 In 1954, J. W. Smith published a paper in which he examined the properties of anterior cruciate ligaments of rabbits (22). Histological tests show that anterior cruciate ligaments of rabbits are entirely collagenous. Smith re- ported that considerable histological changes took place within the first hour after the death of the animal. More precisely, he noted an increase in Young's modulus and in the elastic limit. Therefore, he performed all of his tests within 30 minutes after death. His results indicate that the rupture load of the anterior cruciate ligament is proportional to the cube of the body weight of the animal, although subsequent investigations have not been able to substantiate these results. Smith further noted that for a 5 pound rabbit the stress-strain curve was linear when loads below 15 pounds were applied to the specimen. With no reported specifications other than a uniform and rapid stress rate, Smith strained the ligament as much as 20 per cent in the elastic range. His testing apparatus suspended the cruciate ligament in a position such that the femur, tibia, and anterior cruciate ligament were in a line coin- ciding with that of the applied load. All the ruptures occurred at the tibial insertion of the ligament. In 1959, B. J. Rigby studied the mechanical proper— ties of rat~tail tendon (20). He found that the mechanical properties of rat tail tendons were reproducible within a strain range of 0-4 per cent. The tendons were rinsed in 14 normal saline (0.9 per cent NaCl), then refrigerated at 3°C in the same medium. A number of the tendons under study were frozen at -35°C. No significant differences in the mechanical behavior between the frozen or immersed speci- mens and the fresh specimens was noticed. Also, the mechan- ical properties did not show a temperature dependence within the 0-37°C range. As the strain rate was increased, the resulting stress-strain curves were identical except for a slight shift toward the stress axis. The average maximum slope of a stress-strain plot at a strain rate of 10 per cent per minute was 8.0 i 2.0 x 109 dynes per square centimeter. If the strain rate is held to approximately 1 per cent per minute, strains as large as 20 per cent can be obtained without any apparent breakdown of collagen fibers. In 1964, L. B. Walker performed a series of eXperi- ments in which he tested plantaris tendons in the foot of human cadavers (30). The tendons were dissected from hu- man cadavers which were embalmed and stored in a wetting solution of 5 per cent phenol. Tensile strengths of 7.3 x 108 dynes per square centimeter to 1.5 x 109 dynes per square centimeter were reported for tendons in the wetting solution. If the tendons were dried before testing 5 per cent higher strengths than the wetted specimen were re- corded. Walker indicated that there was no substantial re- laxation of the stress when the specimen-were held at a 15 specified strain level during each test. The stress-strain curves for 2 successive runs on the same specimen at the same strain rate for strain levels below 10 per cent dis- played no significant differences. Tests performed by J. D. VanBrocklin in 1965 on the extensor digitorum tendons of the human foot indicated no variations in the mechanical properties between the frozen or the fresh specimens (25). The data obtained shows, however, a marked increase in the elastic modulus as the rate of stress is increased. Furthermore, when the Specimens were loaded with a saw-tooth loading function, the energy dissipated for cycles of comparable amplitude remained essentially constant as the rate of stress appli— cation was increased. In 1966, E. H. Harris performed a series of tests on the tendons of the upper extremity of the human body and obtained results contradictory to those of Cronkite (14). Harris reported that the elastic modulus for tendons of the muscle flexor digitorum was significantly smaller than those of the extensor digitorum muscle. Clearly, the flexors in normal physiological activity are subjected to larger loads than the extensors. He reported that the av- erage Young's Modulus for embalmed flexor digitorum tendons was approximately 7.6 x 109 dynes per square centimeter. This indicates that if a certain tendon is required to withstand large loads it has the ability to deform much 16 more than those which are normally subjected to smaller loads. In 1967, M. Abrahams performed a series of tests relating the mechanical behavior of horse and human tendons to load, strain cycling, and strain rate (2). Human ten- dons were tested in Ringer's solution at 37°C within 36 hours after autOpsy, and the horse tendons were tested at 38°C in Ringer's solution within 48 hours after autopsy. All specimens were refrigerated until the time for testing. The tendons were placed in shelf-tightening spring loaded wedge-shaped jaws, and strain measurements were recorded by a 1 inch Instron extensometer attached to the specimen. The stress-strain curves described 3 distinct regions. The primary region was that of 0-5 per cent strain in which there was a considerable increase in length with only a slight increase in stress. In this region it is thought that the collagen fibers begin to straighten their wavy pattern. The secondary region was that of 1.5-3.0 per cent strain where the collagen fibers are thought to become fully oriented and begin to assume most of the load. In the final region, that of 3.0-5.0 per cent strain, it is thought that the entire response is due to the collagen fibers in pure tension. The stress-strain curve is a straight line in this region. Visible ruptures of the collagen fibers begin at 5.0 per cent strain level. In tests involving human Achilles tendons, human toe extensor 17 tendons, and horse extensor and flexor tendons, the maxi- mum stress level recorded was 352 kilograms per square centimeter for human tendons and 563 kilograms per square centimeter for horse tendons. Investigations in the cycling effects display complete recovery when the speci- mens were cycled to 2.0 per cent strain with a maximum stress of 113 kilograms per square centimeter. When the specimens were cycled to 3 per cent strain for 10 cycles, the specimens did not return to the same initial 1ength but acquired a 0.5 per cent "residual" strain. As the "residual" strain increased, the peak stress decreased from 228 kilograms per square centimeter to 187 kilograms per square centimeter. The last 10 cycles to a 4 per cent strain level indicated "residual" strains of l per cent with a maximum peak stress of 317 kilograms per square centimeter. These tests verify that when a specimen is strained beyond the 2.0-3.0 per cent strain level in cy- clic testing, a permanent "residual" strain will occur. The effect of strain rate on the stress-strain curve was to change its shape and magnitude. Tests were conducted on Achilles tendons in the range of l per cent strain per minute to 50 per cent strain per minute. Results indicated that as the strain rate was increased to reach a given strain level, the magnitude of the induced stress increased. This explains the ability of tendons to transmit large forces through bones and through joints in a very short time. 18 A. Viidik, in 1967, studied the properties of an- terior cruciate ligaments in rabbits (26). The specimens were kept in a saline moistened gauze for two hours before the tests, and then moistened periodically during the tests with saline solution. The general shape of the load- elongation curves was similar to that of Abraham's. How- ever, the second load-elongation curve exhibited a longer and more accentuated toe part than the corresponding one from the first loading. Also, the slope of the load- elongation curve was slightly higher for the unloading than for the loading cycle. The relative relaxation in per cent of initial load was a constant for any load level. The relaxation of load at a specified load level was largest for the first test and decreased in the following tests on the same specimen. The ligament experienced an asymptotic load relaxation value after an interval of ten minutes. Viidik also studied the effects of postmortal stor- age on the tensile strength characteristics of the anterior cruciate ligament of rabbits (28,29). He considered four types of storage: 1. (Direct immersion in a saline bath-at 20°C for 5 hours. 2. Refrigeration in a saline solution for 24 hours at 4°C. 3. Deep-freezing in a saline solution. . 4. Storage in 10 per cent formaldehyde. 19 The parameters he considered were: the gross shape of the load-elongation curve, its slope, its failure energy, fail- ure load, elongation at failure, and failure site on the bone-ligament-bone preparation. Viidik reported that none of the above parameters were comparable to that of the con- trol or fresh group. Viidik concluded his paper by sugges- ting that all tests on collagenous structures should be run immediately after death in an environment such as blood plasma or synovial fluid. Y. C. Fung has reported results dealing with the effects of strain rate on the mechanical behavior of the mesentery of the abdomen, which is a thin connective tissue membrane connecting the intestines to the abdominal wall (8). Fung noted that as much as one hundred per cent strain could be encountered before reaching the state at which nor- mal physiological observations are made. Fung also indicated that hysteresis loops did not depend significantly on the strain rate. The total amount of stress relaxation was lin- early proportional to the value of the load at which the re- laxation test had begun. Fung relates a term called the elastic tension, which is the stress value after the material has completely relaxed, to the slope of the elastic tension versus deformed length per initial length. Approximating this as a straight line, he integrated to get an exponential type of constitutive equation describing the mesentery of the abdomen. III. EXPERIMENTAL METHODS In order to study the mechanical properties of the anterior cruciate ligament of canines, a femur-anterior cruciate ligament-tibia preparation was chosen. Using this type of preparation the end effects caused by direct clamping to the ligament can be neglected. The fixtures which were used for these tests were designed such that the preparation could be flexed for the tests. This flexing of the prepara- tion was necessary so that the anterior cruciate ligament could be tested in a_natural position. The angle formed by the fixtures was set at approximately 140° for all of the tests. In the natural standing position, a dog flexes the stifle joint from approximately 135° to 140°.(Fig. 3.1).f Contradictions in the literature as to the effects of various testing environments suggested three different modes Which were considered: 1. Moistened slightly during the test with a saline solution. 2. Immersion in a saline bath at room temperature. 3. Immersion in a saline bath at the body temperature of 101°F. The saline solution used throughout all of the tests was Lock's solution. The solution was prepared in eighteen liter quantities according to the following formula: 20 21 l. 7.56 grams of KCl per 18 liters of water. 2. 2.70 grams of NaHCO3 per 18 liters of water. 3. 4.32 grams of CaCl2 per 18 liters of water. 4. 165.5 grams of NaCl per 18 liters of water. The water used in this formula was de-ionized with an Illco- Way Universal Ion-Exchanger. In order to maintain a prOper temperature bath, a "Chill-Chaser" electric immersion heater with a Type 316 stainless steel heating element was used. This particular type of instrument was capable of maintaining a temperature of 101°F to within i l°F throughout all of the tests. The specimens were tested within 6 hours after the death of the animals. Initial tests were started within 2 hours after death of the animals with individual tests requiring approximately 1 hour. While one specimen was being tested, the remaining specimens were kept at room temperature with the knee capsule left intact. Each spec- imen was subjected to the temperature bath for approximately 30 minutes before the beginning of each test. Due to its central location in the stifle joint, the anterior cruciate ligament was easily isolated. All interior cartilage and surrounding tendons and ligaments were removed from the joint area before testing,and the specimen was clamped to a table where the measurement of area was made. The cross-section of the anterior cruciate ligament was assumed to be elliptical. Since the anterior 22 cruciate ligament has a 90—degree twist as it passes from the femur to the tibia, it was necessary to adjust the clamps so that the ligament was as uniform as possible when the measurements of the major and minor diameters were being taken. A special set of micrometer probes were designed in order that easy access into the intact joint was possible. The probes were made from steel pins that were ground to a diameter of 1/32 of an inch and had an overall length of 1 inch. The probes were then attached .to a 2—3 inch micrometer (Fig. 3.6). The testing machine used for these tests was an Instron Model TT-CM metric with a 0-50 kilogram load cell inserted into the head. A cell range of 0-10 kilogram was sufficient for these tests (Fig. 3.3). As described earlier, the testing fixtures held the specimen in a flexed position. The fixtures were also designed such that either the femur or the tibia could be rotated to allow for better aligmment of the ligament during the tests. Attached to the stationary upper fixture was a system of coordinate axes. Using a pair of highly versatile calipers, 3 radii were drawn from the coordinate axes to the points of inser- tion of the anterior cruciate ligament on the tibia and on the femur (Fig. 3.4). 23 (019:0) ( x ',o,o) R3 r’jiL’I””/,///’[3(mgg) (0,0,2 ) In order to solve for the point P(x,y,z), a system of 3 simultaneous quadratic equations must be solved. x2 + (y - y')2 + 22 = R NM (x - x')2 + y2 + 22 II :0 UlN x2 + y2 + (z - z')2 = R paw Knowing R1, R2, R3, x', y', and z', the above system of equations can be solved for the points of insertion of the ligament into the tibia and the femur.* Therefore, after calculating for the tibial and femoral insertions, the *See Appendix. 24 length of the ligament was calculated using the following equation. L = {(xt - xf)2 + (yt - yf) + (zt - zf) t' yt, zt is the insertion point of the tibia, and xf, yf, zf is the insertion point of the femur. Where x The anterior cruciate ligament assumes approximately a vertical position in the test fixture. Due to this posi- tioning of the joint, the lateral displacement during the tests was minimal. Therefore the z-component of displace- ment was the only displacement that was considered for the calculation of the strain induced in the ligament. The new length of the ligament was calculated by a correspond- ing change in the z-component of the tibial insertion as follows: 1/2 2 2 L' = {(xt - xf) + (yt - yf) + (zt - zf + Az)2} The strain in the ligament was calculated by using the Lagrangian definition of strain. a = (L' — L)/L In order to eliminate false strains due to slipping of the femur and the tibia in the fixture, a clip gage was inserted into the joint (Fig. 3.1). The clip gage was con- structed from a strip of spring steel 0.01 inches thick 25 and 0.25 inches wide. Two SR-4 strain gages were attached to either side of the clip gage. The strain gages, Micro- Measurements Model EA-06-125AD-120, were bonded with East- man 910 cement and the lead wires were coated with W. T. Bean Gagekote #2. The strain gages were coated with Gage- kote #5, which is a waterproofing designed for direct im- mersion in oil or water. The output was recorded on a Sanborn Model 60-1300 Twin Viso Recorder with a Sanborn Model 64-500A Strain Gage Amplifier. To provide sufficient sensitivity and temperature compensation, the 2 SR-4 gages were attached to adjacent arms of the bridge network in the recorder. Slipping of the clip gage in the joint was avoided by the puncture of holes in the femoral and tibial intercondyles to receive the pointed tips of the clip gage. The calibration fixture used for calibration of the clip gage consisted of a Scherr Tumico dial gage graduated in 0.001 inch intervals attached to a sliding frame (Fig. 3.5). With this particular design for the clip gage, the cali- bration curve was linear throughout the range of interest. The procedure used for testing the anterior cruciate ligament was one of a steprloading pattern. Applying a constant strain rate, the load was allowed to reach a predetermined level and then there was a subse- quent stress relaxation at this particular strain level. This type of loading was applied until the strain level reached approximately 10 per cent where the loading was 26 reversed back to the point of a zero strain level. At each particular load level the specimen was reiaxed for 4 minutes at which time the amount of relaxation approached an asymptotic value. After the completion of the first test cycle, the specimen was subjected to a second identi- cal test. 27 Figure 3.2--A Ruptured Specimen 28 Figure 3.4--Measurement of Initial Length 29 Figure 3.5--Clip Gage Calibration Figure 3.6--Measurement of Area IV. PRESENTATION OF RESULTS The general shape of the stress-strain curves can be seen in Figures 4.1—3. The initial portion of the curves is called the toe region. In this region the magnitude of the strain increases much more rapidly than does the magni- tude of the stress. The next region, which is not so clearly visible, is called the transition region. This is the region in which the specimen begins to accumulate load quite rapidly. In the third and final region the curves become almost linear. As stated earlier in Section III, Abrahams described these three regions in terms of the response of the collagen fibers in the specimen (2). In the initial region the fibers have been said to organize along the axis of stress. In the tran- sition region the fibers have organized and begun to assume a tensile force. In the final region the response is due entirely to the collagen fibers in tension. The effect of strain rate on the shape and magnitude of the curves can also be seen in these figures. As the strain rate is increased, the toe region tends to decrease until at high strain rates the region is pushed to the bot- tom of the curves. This effect along with the increase in the apparent Young's Modulus, as seen in Fig. 4.7, tends to shift the curves toward the stress axis. Using Fig. 4.1 as 30 31 an example, for a change in strain rate from 2-40 per cent per minute the toe region decreases from approximately 6 per cent strain to a value of l.5_per cent strain- The apparent Young's Modulus increases from approximately 1.0 x 109 dynes per square centimeter to 2.0 x 109 dynes per square centimeter for the specimen which was moistened with saline solution throughout each test. Curves can be made from the graphs in Figures 4.1-3 which will relate the effects of strain rate for both the immersed and the moistened specimens at room temperature. Such curves are plotted in Figures 4.4-6. In these graphs the strain is plotted versus the strain rate for varying stress levels for each of the environments. It is clear that for an increasing strain rate, the stress required to induce a specified strain increases. Also, from these curves the approximate stress-strain curves for various strain rates can be plotted for each environment in order to compare them. By chosing a couple of representative examples, such as Fig- ures 4.8 and 4.9, the difference arising from these two en— vironments for shmilar strain rates can be determined. In these figures one can notice that the toe regions have essen- tially the same order of magnitude and the same shape for each of the environments. The major difference can be seen by observing the linear portion of the curves. This differ- ence stems from a variation in the slopes of the curves, as seen in Fig. 4.7, between the moistened and the immersed specimens. 32 By extending the curves in Figures 4.4-6 a zero strain rate or static test can be approxflmated, as given in Fig. 4.10. Here one can note that the slopes of the linear portions of the curves are just about identical. However, a difference in the toe regions can also be noticed. In Figures 4.11-l3 results dealing with the cumula- tive per cent relaxation of the initial stress versus the value of initial stress for varying strain rates for each of the environments are presented. In the region of 0-20 kilograms per square centimeter, the cumulative per cent relaxation changes very rapidly with initial stress. Above this level the curves become linear. There also appears to be a tendency for the curves to be shifted toward the stress axis as the strain rate is reduced, however, this result is not entirely consistent throughout all of the tests. The actual magnitudes of per cent relaxation fall between 10 and 30 per cent for all of the tests. If one sketches a plot of strain rate versus the cumulative per cent relaxa- tion for various stress values between the strain rates of 5 and 25 per cent per minute, Figures 4.14 and 4.15, the magnitudes of stress relaxation are almost identical for each of the environments. Outside of this range there is an insufficient number of data to determine the pattern. In the Figures 4.16-18 the value of stress after relaxation is plotted versus the strain for each of the strain rates. These curves follow the same shape as the 33 initial stress versus strain curves except for a decrease in each of the stress values. As previously mentioned, at each stress level the relaxation period lasted for 4 minutes, during which the stress reached an asymptotic value. The variation of results between the first and sec- ond tests on the same specimen are shown in Fig. 4.19. Three different strain rates for the moistened specimen at room temperature were used as representative examples of the existing trends apparent in all of the tests. As seen in this plot, the slopes of the linear portions of the curves are approximately identical for the two tests applied suc- cessively to the same specimen. The extended toe region in the second test was prevalent in all of the tests. 34 70 a 0 6 50 N .EU\.mM .o .mmmuum 20 8.0 6.0 e, % Strain, 4.1, Stress-Strain Plot for Moistened Specimen 4.0 2.0 Fig. 35 0 -8 0% 6 I n .1 a u 0 r as 4 0 .2 0 70- 60- 50- N 1 q 0 0 4 3 .EU\.mM .o .mmmuum 20' 10‘ Stress-Strain Plot for Immersion at 72°F 4.2 Fig. 36 2.0 f0 10. N o§\omVH s0 smmmHum Strain, % Er 4.3 Stress-Strain Plot for Immersion at 101°F Fig. 37 cofiwommm Uocmumfloz How uoam ovum camnumugflmupm v.v .wwm m .w .cflmuum o.v 0.00 Om. Os Om Gm .o_ - «.Eo\$x mmumem 0H ON om ow om 'eqeu urexqs ’3 'UTW/% moms. HM .GOHMHGEEH .HOM “GHQ 0.....mm GflflHflMICHMHDm a .0 .aamuum mé .mE ~.Eu.mox mmmmbw , o.m 0.x ooh Com com Dov oom O.N O.H o - L|F r p p L P p — o // O// Q / o x q .oH .ON x q , 8 3 .0m 10¢ Tom 0 X 0 X 4 q Axumv mxum Aumxunxumw mXUmw nxU_ ‘urw/g '3 'aqeg uterus moaoa um coamumssH now uoam 60mm camuumucamuum v.4 .mam m .w .aamuum 0.0 o.m 0.x. o.m o.m 03v on o.m o...” o p h h n o .OH // / / ’ wON 2 , E, dynes/cm. -8 Young's Modulus x 10 301 N O l 10- —’ ,0 /_—————-—"/’ 0 O o Immersion 72°F. / o 0 0 /00/0 40 Immersion lOl°F. O Moistened ’ 0 0 10 20 30 40 5'0 Strain Rate, é, %/Min. Fig. 4.7 Apparent Young's Modulus Versus Strain Rate After Toe 41 70. 60- 50- a 0 3 .mmmuum 20 . 10 10.0 Strain, % E: Strain Rate Stress—Strain Plot for 5%/Min. 4.8 Fig. 2 Stress, 0, Kg./cm. 42 70+ 60 ‘ O o 2‘" T \0 50 o <\ \0 ‘o K e 8e \ 40‘ o 30- o O 20 ‘ O o o 10‘ o 00 0 r . 4 r . 0 1.03 2-0 3.0 4.0 5.0 Strain e, % ' Fig. 4.9 Stress-Strain Plot for 15%/Min. Strain Rate 43 0 fi. 0 .l. 0 c To 0 fl. 6 0 r. 4 O 0 I. 2 a d d 0 0 0 o O 0 0 0 0 7 6 5 4 3 2 l .EO\.mM .b .mmoupm Strain, e, % Static Test 4.10 Fig. 44 30.0 70 60 503 N d o o 4 3 .E0\.mM .o .mmmuum 20 10 0 Relaxation, %0 Relaxation When Moistened 4.11 Stress vs. Fig. 45 7O 60 50 N a - 0 4. 3 .E0\.mM .p .mmmuum 20 10 50.0 Relaxation, %0 Relaxation in Saline at 72°F Stress VS . 4.12 Fig. 46 80 J 70 60 ' fi 0 5 N 0 4 .EO\.mM .o .mmmuum 30 20 10 50. 40 Relaxation, %0 Relaxation in Saline at 101°F 4.13 Stress vs. Fig. 47 omcmumfloz cons mumm cflmnum momnm> soflummemm ucmu mom va.v .mHm .cHE\w .m .mumm camuum ov om oh 0H 0 LEO x 0.0. 335 o 0.0m no“ 0.0m m“ 0\ .oa r0N .0m 0% ’uorqexeteu moms um cmmHmEEH coax oumm camuum mdmuo> cowumxmaom ucmo mom ma.e .mam .cflz\w .w..oumm.cflmuum om 0N 0H 0 h! D O O O O .Ei? 0.9 3,25 o - o T|\\\\\\\\\\\ o 0.0 m T \ 0.0m \ 0.0¢ 000 0000 00000 0 0.00 .OH .oN .om ’uoraexeteu 0% 2 Stress, quKg./cm. 49 Fig. 4.16 4'. 0 6T. 0 8'. 0 Strain, e, % Stress After Relaxation Versus Strain for Immersion at 72°F 2 Stress, 0R, Kg./cm. 50 7O ‘ 60 J 50 40 30 20 10 < 43 610 8.0 10.0 Strain, e, % Fig. 4.17 Stress After Relaxation Versus Strain for Moistened Specimen 51 .0. 1. o MJW o / O 0/0 0 0 8 IIIIIIIIIIo ////// N.n /0 0 / /o / o. O /O O .6 0/// /// ///// «.0 o o/ o Ill/Io/l/ ////d7/// 0 0 4M“ /0 v. 0/ /0 / /o W\ 0/0 / o /O / O /O m/.@Mv/O 0/0 /0 / O 0 IIIIIIIIb OOIIIIIIIIo Il/Io ////o o .L oon/\ at. QCNQkkm 00/0 07 0 O IIJWIToTIIIImHWI. ,/ . . TTT . . o T. O 0 0 0 0 0 0 m0 0 nu 8 7 6 5 4 3 2 1 .Eo\.mM .mb .mmmuum N Strain, e, % Stress After Relaxation Versus Strain for Immersion at 101°F 4.18 Fig. 52 O / O .///O / N\ / /0/ . / O / 0 O /0/// O/ O O [/0 I/O/.I[O/ / ’0 ——Firs’r ——Sec0nd 70¢ 60- 50 . TIT 0 0 4 3 20 « .E0\.mM .o .mmmuum 10 - 10. 8.0 Strain,e, % Stress Versus Strain for First and Second Tests When Moistened 4.19 Fig. V. DISCUSSION AND CONCLUSIONS The shape of the stress-strain curves follow the same pattern given most often in the literature. As seen in Figures 4.1-3, the shape of the stress-strain curves is not altered by various strain rates, except for high strain rates where the toe region appears at an early strain level. Unlike the range of 0-l.5 per cent strain for the toe regions of tendons given by Abrahams (2), the anterior cruciate ligament of dogs displays a toe region which varies with strain rate up to 6 per cent strain. Also, in these figures the increase in the apparent Young's Modulus with increasing strain rate is quite pronounced. For flow strain rates the modulus changes rapidly with strain rate, but for higher values of strain rate only a gradual increase can be seen. Results similar to these were described by VanBrocklin (25) and Abrahams (2). The average apparent Young's Modulus given by Frasher (11) for collagen fibers in blood vessels ranged from 3.0 x 107 to 1.0 x 109 dynes per square centimeter. Values of approxi- mately 2.0 x 109 dynes per square centimeter were obtained in these experiments on the anterior cruciate ligaments of dogs. 53 54 Y. C. Fung (12) described the elastic properties of the mesentery of the abdomen of rabbits by an exponen- tial function. He used the value of stress after a com- plete relaxation of the specimen at each stress level and plotted this value versus the value of strain. This rela- tionship describes the so-called "elastic" test condition corresponding to an infinitely slow application of load. By plotting the slope of the tension-deflection curve against the elastic tension, he was able to approximate the following linear relationship. =bT, arm >‘6 where T equals the elastic tension and 1 equals the exten- sion ratio (the deformed length divided by the relaxed length). An integration gives a u and m where c is an integration constant. This, however, does not give the required zero tension when the eXtension ratio is 1. Using the fact that the strain energy function W(ll,12,l3) of an isotrOpic compressible elastic body is expressible in terms of the strain invariants I1 and 12, the Lagrangian stress in simple elongation in given by 55 The zero factor must be of the form (A - £2] if the strain 1 energy has no singularity at the undeformed state. Insertion of this factor yields where D is a constant. If we describe the Lagrangian strain by then, 1 eae e + 28 + l £2 + 28 as e e + 26 + 1 By dividing the quantity in brackets and keeping terms to the third order, one gets 4 2 as c=C[e[l-e+§-e)]e where C and a are undetermined constants. This form of Fung's Law was plotted for various values of a and C in Figures 4.1-3. In Table l the constants which were used to draw these curves are given. It can be seen that Fung's Law very closely represents the 56 Table 1.-—Values of a and C for Exponential Law Strain Rate a C Environment %/Min. Kg./cm.2 0.0 20 147.7 1.7 18 167.6 Moistened 2.6 18 213.4 With saline 10.8 20 263.8 15.3 28 367.1 36.8 40 508.1 0.00 18 118.5 1.55 20 132.8 2$m$§§;°n 4.75 16 277.6 9.75 16 333.1 21.50 15 430.5 53.80 21 491.4 0.0 24 35.6 2.5 20 96.0 5.7 14 282.1 Immersion 9.7 1 1000.0 at 101°F 15.4 0 1600.0 16.2 0 2210.0 19.7 0 3070.0 57 results which were obtained in the tests. If one examines Table 1, it may be noted that the value of the constant a depends on the strain rate. In the tests for the moistened specimens the constant a increases with the strain rate, whereas for the immersed specimens the constant decreases with an increase in the strain rate. With the limited amount of data available at this time it is difficult to determine the exact nature of this dependence. Also in Table l the constants for the static test are given. These constants are used in the curves which are plotted in Figure 4.10. Again, Fung's Law represents the extrap- olated data very well. The constant C increases with an increase in strain rate and there is very little differ- ence between the moistened and immersed at 72°F tests. The difference between the first and second tests on the same specimen was noted (Fig. 4.19) to be an ex- tended toe region for the second test. Rigby first re- ported this result in his tests for the mechanical prop- erties of tendons (24). He "preconditioned" the specimen by straining it to a given value of strain before each test. He suggested that material other than collagen fibers was ruptured in this conditioning. After this procedure, he claimed that the tendons were elastic beyond the strain level of 4 per cent. If one considers the de- scription of the toe region given by Abrahams (2), it could be possible that the rupturing of these other 58 constituents might cause the toe region to become extended during the second test of a specimen. Furthermore, the fact that the lepes of the linear portions of the curves are identical for both of the tests shows that the actual collagen fiber response is not affected by the testing of the same specimen twice in the range considered. In Figures 4.11-13 the cumulative relaxation in per cent of initial stress versus the initial stress for varying strain rates is plotted. For lower strain rates it can be seen that the per cent relaxation changes rapidly with the value of stress. At higher values of stress, above 20 kilograms per square centimeter, the curves be- come linear. %0 = C0 relax or A0 = C02 , where Ac equals total amount of stress relaxation and 0 equals the initial stress. In the results from Figures 4.11-13 there is a trend for the curves to become shifted toward the stress axis as the strain rate is reduced, but- as stated previously this trend is not entirely consistent throughout all of the tests. Throughout this discussion of results there has been no reference to the results of the tests run at the elevated temperature of 101°F. If one looks through the results of these tests, the temperature of the saline bath 59 has rather drastic effects on the mechanical prOperties of the anterior cruciate ligaments of dogs. Due to the irregularities in the results of the tests conducted, a complete discussion is not possible at this time. In order to determine the actual effects of temperature on the mechanical prOperties, a series of varying temperature tests must be made to complete the investigation. The re- sults obtained for the single elevated temperature indicate that this study should definitely be made and that the ef- fect of using other bath fluids should be investigated. A further observation noticed by the author can be seen in Fig. 3.2. In this picture an apparent organized pattern of the fibers may be seen which took place while the specimen was being strained. Even though this was only an observation which was not dealt with in any de- tail, there is a possibility that a recrystalization might take place under tensile forces similar to that which has been seen to occur at the so-called "recrystalization temperature" of biological materials. BIBLIOGRAPHY 10. 11. 12. BIBLIOGRAPHY Abbott, L. C., "Injuries to the Knee Joint," Journal of Bone and Joint Surgery, 1944, pp. 503-517. Abrahams, M., "Mechanical Behavior of Tendon in Vitro, A Preliminary Report," Medical and Biological Engineer- ing, 1967, V01. 5, pp. 433-443. Annovazzi, G., "Osservazioni sulla elasticita dei legamenti," Arch. Sci. Biol. (Napoli), 1928, Vol. 11, Bloom, W., and D. W. Fawcett, A Textbook of Histology, 8. ed., W. B. Saunders Co., PhiIadeIphia, 1962. Burton, A. C., "Relation of Structure to Function of the Tissues 0f the Wall of Blood Vessels," Physiological Review, Vol. 34, 1954, p. 619. Carton, R. W., J. Dainauskas, and J. W. Clark, "Elastic Properties of Single Elastic Fibers," Journal of Applied Physiology, Vol. 17, 1962, p. 547. Contini, R., and R. Drillis, "Biomechanics," A lied Mechanics Reviews, Vol. 7, No. 2, February, 1955, pp. 49-52 0 Cronkite, A. E., "The Tensile Strength of Human Tendons," Anat. Rec., Vol. 64, 1936, pp. 173-186. Crouch, J. E., Functional Human Anatomy, Lea and Febiger, Philadelphia, 1965. DePalma, A. F., Diseases of the Knee, J. B. Lippincott Co., Philadelphia, 1954. Frasher, W. G., "What is Known About the Physiology of Larger Blood Vessels," Biomechanics, Ed. by Y. C. Fung, New York, 1966, pp. 1-19. Fung, Y. C., "Biomechanics," Applied Mechanics Reviews, V01. 21, N0. 1, January, 1968, pp. 1-20. 61 l3. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 62 Gratz, C. M., "Tensile Strength and Elasticity Tests on Human Fascia Lata," Journal of Bone and Joint Sur- gery, Vol. 13, 1931, pp.4334—340. Gratz, C. M., and S. N. Blackberg, "Engineering Methods in Medical Research," Mec. Engineering, Vol. 57, 1935, pp. 217-220. Gross, J., "Collagen," Sci. American, Vol. 204, N0. 5, 1961, pp. 120—130. Hardy, R. H., "Observations on the Structure and Pr0p- erties of the Plantar Calcaneonavicular Ligament in Man," Jr. of Anat., Vol. 85, 1951, pp. 135-139. Harris, E. H., B. R. Bass, and L. B. Walker, "Tensile Strength and Stress-Strain Relationships in Cadaveric Human Tendon," Anat. Rec., Vol. 148, 1964, pp. 289—295. Harris, E. H., L. B. Walker, and B. R. Bass, "Stress- Strain Studies in Cadaveric Human Tendon and an Anomaly in the Young's Modulus Thereof," Med. and Biol. Enging., Vol. 4, 1966, pp. 253-259. John, R. J., and V. Wright, "An Analytical Description of Joint Stiffness," Biorheology, Vol. 2, 1964, pp. 87- 95. Lindsay, W. K., and H. G. Thomson, "Digital Flexor Tendons: An Experimental Study Part 1," Br. Jr. of Plastic Surgery, Vol. 12, 1960, pp. 289-316. MacMaster, P. E., "Tendon and Muscle Ruptures," Journal of Bone and Joint Surgery, Vol. 15, 1933, pp. 705-722. Ramanathan, N., Collagen, Interscience Publishers, New York, 1962. Rich, A., and F. H. C. Crick, "The Molecular Structure of Collagen," Journal Mol. Biol., V01. 3, 1961, pp. 483-506. Rigby, B. J., N. Hairai, J. D. Spikes, and H. Eyring, "The Mechanical Properties of Rat Tail Tendon," Jour- nal of General Physiology, Vol. 43, 1959, pp. 265-283. Rigby, B. J., "Effect of Cyclic Extension on the Physi- cal Pr0perties of Tendon Collagen and its Possible Re- lation to Biological Aging of Collagen," Nature, Vol. 202, 1964, pp. B-2. 26. 27. 28. 29. 30. 31. 32. 33. 34. 35. 63 Smith, J. W., "The Elastic Properties of the Anterior Cruciate Ligament of the Rabbit," Journal of Anatomy, Vol. 88, 1954, pp. 369-380. Stucke, K., "The Elasticity of the Achilles Tendon in Loading Experiments," Langebeck Arch. Klin. Chir., Vol. 265, No. 5, 1950, pp. 5791599. Stucke, K., "Tendon Loads and Rupture in Animal Experi- ments," Chirurg., Vol. 22, 1951, p. 16. VanBrocklin, J. D., and D. G. Ellis, "A Study of the Mechanical Behavior of Toe Extensor Tendons Under Ap- plied Stress," Arch. of Physical Medicine and Rehabili- tation, Vol. 46, 1965, pp. 369-373. Viidik, A., "Biomechanics and Functional Adaptation of Tendons and Joint Ligaments," Studies on the Anatomy and Function of Bone and Joints, Ed. by F. G. Evans, Heidleberg, 1966, pp. 17-40. Viidik, A., "Experimental Evaluation of the Tensile Strength of Isolated Rabbit Tendons," Biomedical Engineering, Vol. 2, 1967, pp. 31-36. Viidik, A., L. Sandqvist, and M. Magi, "Influence of Postmortal Storage on Tensile Strength Characteristics and Histology of Rabbit Ligaments," Acta. Orthop. Scand., Suppl. 79, 1965. Viidik, A., and T. Lewin," Changes in Tensile Strength Characteristics and Histology of Rabbit Ligaments In- duced by Different Modes of Postmortal Storage," Acta. Orthop. Scand., Vol. 37, 1966, pp. 141-155. Walker, L. B., E. H. Harris, and J. V. Benedicts, "Stress-Strain Relationships in Human Plantaris Ten- dons: A Preliminary Study," Med. Elec. Biol. Enging., V01. 2, 1964, pp. 31-38. Wright, D. G., and D. C. Rennels, "A Study of the Elastic Properties of Plantar Fascia," Jr. Bone Joint Surg., Vol. 46-A, 1964, pp. 482-492. APPENDIX APPENDIX Ineorder to solve for the initial length of the anterior cruciate ligament before each test, 3 radii were drawn to the points of insertion of the ligament into the femur and the tibia. (0.y'.0) Pf(X.y.z) Cruciate ligament (X. 7010) X Z The point of intersection of the 3 spheres is the point Pf(x,y,z). The equations to be solved form a system of 3 simultaneous quadratic equations. (x - x')2 + y2 + 22 = R3 A-1 \ 2 ,2 2 2 x+(y—y)+z=R2 A-2 x2 + y2 + (z — z')2 = Ri A-3 65 66 From A-l, A-2 and A-3 one obtains -2xx' + x'2 + 2yy' - y'2 = R3 - R: , A-4 -2yy' + y'2 + 222' — 2'2 = R: - Ri , A-5 -222' + 2'2 + 2xx' - x'2 = Ri - R: . A-6 Using A-5 and A-6, yields _ l ,2 _ ,2 2 _ 2 _ z' _ y-ffr-{y z +Zzz'+Rl R2}—A2+T-z A7 _ l ,2 _ I2 , 2 _ 2 _ z' _ X - 5‘1- {X Z + ZZZ + R1 R3} - A1 + x-—'- Z , A 8 where _ l I,2 _ ,2 2 _- 2 A2 - 75,-1- {y Z + R1 R2} _ l , 2_ , 2 2__ 2 A]. - 23?.- {(X ) (Z ) + R1 R3} 0 Substituting these results into A-BV yields 2 2 z' z' 2 z' z' 2 , 2 _ 2 A]. + 2A1 i-r Z + (fr) + A2 + 2A2 if Z + (SIT) Z + (2-2 ) — R1 2 2 2 z' z' { z' 2 2'32 J} 2 2 ,+ ,2__ 2 Al + A2 + [A]. Err '1' A2 y-T) Z + (SET) 4" (yr) + Z - 22 Z - R1 67 Rearranging gives 2 2 A A I | {(Er) +T§r) +1}zz+2z'T;}-+§-2r- 112 + {21% + A: + 2'2 mi} = 0 Solving this quadratic equation for 2 produces the final result, .1. A1132 2A1A2 2 22 22 2'22'2 2 - Z'{l-}Tr,-y—r}i [Z' (x-fi-j' fir-1) -{A1+A2+Z' ‘R1}{(?)L+[§r)+n] z— ' 2 '2 c A-10 {(ET) + (gr) + 1} From equations A-7 and A-8 the remaining coordinates are obtained. Since this system of equations yields 2 solutions, it was necessary to determine which solution was applicable. This choice was easily made by determining the approximate coordinates during the tests. The point of insertion of the ligament into the tibia was calculated in a similar manner 0 436 "‘TLTT‘TTTTLTTTTTTTT'TTTTTTTTTTTTTT“